The intended function and use is of the Invacare CPAP is to treat Obstructive Sleep Apnea (OSA) in the home, in adults by delivering continuous positive airway pressure.
Device Story
Electromechanical CPAP device; treats Obstructive Sleep Apnea (OSA) in home setting. Inputs: room air via inlet filter. Operation: blower motor pressurizes air; microprocessor controls motor speed to maintain prescribed pressure (3-18 cmH2O). Output: continuous positive airway pressure delivered via standard nasal mask/tubing. Clinician sets pressure and programs therapy mode (ramp-up) via hidden interface; monitors compliance via hour/compliance meters. Patient uses push-button toggle for standby/therapy modes. LED display shows current pressure. Benefits: maintains airway patency during sleep.
Clinical Evidence
Bench testing only. Device tested for electrical, mechanical, and environmental performance requirements per FDA's November 1993 'Reviewer Guidance for Premarket Notification Submissions'. Device met all performance criteria.
Technological Characteristics
Electromechanical CPAP; blower motor with microprocessor-based control system. Pressure range: 3-18 cmH2O (max static 27 cmH2O). Interface: 5-digit LED display, LED indicators, push-button controls. Connectivity: none. Power: AC. Materials: standard medical-grade components for respiratory air path. Dimensions: standard home-use form factor.
Indications for Use
Indicated for treatment of Obstructive Sleep Apnea (OSA) in adults via continuous positive airway pressure. Contraindicated in patients with respiratory failure/insufficiency, pneumothorax, pneumomediastinum, bullous lung disease, untreated cardiac failure/hypotension, massive epistaxis, pneumoencephalus, cranionasopharyngeal fistula, acute sinusitis, otitis media, or perforated ear drum.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{1}------------------------------------------------
K98 2242
JAN 27 1999
#### 510(k) SUMMARY
#### Invacare Corporation's Revised CPAP 10/9/98
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.
Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036-2125 Phone: (440) 329-6000 Facsimile: (440) 365-4558
Edward A. Kroll Contact Person: Director, TQM and Regulatory Affairs
June 24, 1998 Date Prepared:
#### Name of Device and Name/Address of Sponsor
Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036-2125 Phone: (440) 329-6000 Facsimile: (440) 365-4558
#### Common or Usual Name
СРАР
#### Classification Name
Ventilator, Noncontinuous
#### Predicate Devices
The Invacare CPAP is substantially equivalent to the Resmed Sullivan V CPAP System (K930656, 3/15/94) and the Respironics Solo CPAP System (K961626, 7/18/96).
#### Intended Use
The intended function and use is of the Invacare CPAP is to treat Obstructive Sleep Apnea (OSA) in the home, in adults by delivering continuous positive airway pressure.
{2}------------------------------------------------
#### Contraindications
The Invacare nasal CPAP system is for the treatment of Obstructive Sleep Apnea (OSA) in adults.
#### CONTRAINDICATIONS
Nasal CPAP therapy should be used under the supervision of a licensed physician. The following conditions contraindicate the use of Nasal CPAP for some patients:
- 1. Existing respiratory failure of insufficiency resulting in potential risk of rebreathing.
- 2. Pneumothorax or pneumomediastinum.
- 3. Bullous lung disease.
- 4. Untreated cardiac failure or hypotension.
- 5. Massive epistaxis or previous history of massive epistaxis (nose bleed).
- 6. Pneumoencephalus, recent trauma or surgery that may have produced cranionasopharyngeal fistula.
- 7. Acute sinusitis, otitis media or perforated ear drum.
The clinician should assess on a case by case basis the relative risks and benefits of nasal CPAP therapy in such a subject. When assessing the relative risks and benefits, the clinician should understand that the Invacare CPAP system can deliver pressures up to 18 cmH2O. Also in the unlikely event of certain fault conditions, a maximum static pressure of 27cmH2O is possible. If such a pressure could present risk to particular patients, then this device MUST NOT be used.
#### Technological Characteristics and Substantial Equivalence
#### A. Device Description
The Invacare CPAP is an electromechanical device that produces Continuous Positive Airway Pressure. Its' intended function and use is to treat Obstructive Sleep Apnea (OSA) in the home, by delivering continuous positive airway pressure to patients with OSA.
It consists basically of a blower motor and a microprocessor based control system. Air enters the device through an inlet filter which traps dust and airborne particles. The filtered air is then directed to the blower motor where it is pressurized, and then delivered to an outlet nozzle located on the front of the unit. The nozzle accepts standard, smooth, flexible, 22mm diameter tubing.
Pressurized air is delivered to the patient through a standard nasal mask with an exhaust port. The microprocessor controls the pressure delivered by controlling the speed of the motor. Pressure is adjustable from 3 to 18 cm H+O, and is preset by the Clinician during set up in the home. The amount of pressure selected in based upon the Physician's prescription.
{3}------------------------------------------------
Pressurized air is delivered to the patient through a standard nasal mask with an exhaust port. The microprocessor controls the pressure delivered by controlling the speed of the motor. Pressure is adjustable from 3 to 18 cm H2O, and is preset by the Clinician during set up in the home. The amount of pressure selected in based upon the Physician's prescription.
The CPAP controls and displays include a 5 digit numerical LED display (each digit made of 7 segments) three additional standard LED indicators, a large control button, and two sections: the patient controls and displays, and the therapist's setup controls and displays. In normal operating condition, only the patient interface is available. The therapist must open the top cover of the device to reveal the additional two small buttons and operate the set-up controls.
The patient interface consists of the "Invacare" control button, the 5-digit numerical display, and three LED indicators. The meaning of the symbols associated with the three LED indicators are. "standby," "delay," and "pressure." The five digit numerical display, displays the pressure being provided by the unit. The "Invacare" control button is a push button toggle switch which places the unit in normal, therapy (if desired), or standby modes. In therapy mode, airway pressure starts at a rate that is less than the prescribed pressure, and gradually increases to full prescribed pressure. This mode is used when patients find full prescribed pressure to be uncomfortable during initial use, and prefer a gradual increase. In standby mode, the unit is on, but, there is no power to the blower motor.
As stated above, the clinical set up interface is located under a cover which is fastened to the outer housing. Removing the cover exposes the clinical interface controls which include the patient interface items described above and two additional buttons. The two additional buttons are labeled with symbols indicating "up" and "down." The symbols under the associated with the three LED indicators under the cover are, "time," "delay," and "pressure." The up and down buttons are used by the clinician to set the prescribed pressure and program the therapy mode if desired. Additionally, the clinician can also determine patient compliance with the prescribed use of the device, by viewing the hour and compliance meters. The hours and compliance meters monitor the hours of device use and the pressure delivered.
#### A. Substantial Equivalence
The Invacare CPAP covered by this submission is substantially equivalent to other legally marketed CPAP devices intended for home use. Specifically, the Invacare CPAP is substantially equivalent to the Resmed Corporations' Model "Sullivan" CPAP device (K930656, March 15, 1994), and Respironics Model "Solo" CPAP system (K961626, 7/18/96). The Invacare CPAP has the same intended use, princible of operation and technological characteristics as the two previously cleared devices.
{4}------------------------------------------------
The Invacare CPAP and its predicate devices are all electromechanical devices that produce Continuous Positive Airway Pressure. All have the same intended function and use which is to treat Obstructive Sleep Apnea (OSA), by delivering continuous positive airway pressure to patients with OSA. They consist of the same basic materials, are of similar construction, and the same technological characteristics.
Each of these devices consist basically of a blower motor and microprocessor control system. Technology is similar in that all function by taking in room air through an inlet filter, directing it to a blower motor where it is pressurized, and channeling it to an outlet nozzle, for delivery to the patient through a standard mask and tubing arrangement. Performance characteristics for operating pressure, pressure adjustment, power requirements and maintenance are also similar. Finally all of these devices vary operating pressure by controlling motor speed.
### Performance Data
The Invacare CPAP was tested in accordance with the electrical, mechanical and environmental performance requirements for home use respiratory devices set forth in the Anesthesiology and Respiratory Devices Branch's November 1993 document entitled "Reviewer Guidance for Premarket Notification Submissions" In all instances the Invacare CPAP met the required performance criteria and functioned as intended.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 27 1999
Mr. Edward A. Kroll Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, OH 44036-2125
Re: K982242 Invacare Nasal CPAP System Regulatory Class: II (two) Product Code: 73 BZD Dated: January 14, 1999 Received: January 19, 1999
Dear Mr. Kroll:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Mr. Edward A. Kroll
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 510(k) Number (if known): TBD
Device Name: Invacare CPAP
Indications For Use: To treat Obstructive Sleep Apnea (OSA) in adults by delivering continuous positive airway pressure.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) (Division Sign Off)
Division of Cardiovascular, Respiratory, Prescription Use \ Division of Orical Devices OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ .Per 21 CFR 801.109) 510(k) Number _
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.