← Product Code [GEX](/productcode/GEX) · K982113

# UMECO OPTIMA 694 RUBY LASER SYSTEM (K982113)

_Umeco, Inc. · GEX · Sep 30, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K982113

## Device Facts

- **Applicant:** Umeco, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 30, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Umeco OPTIMA 694 system is intended for the use in the removal of blue/black tattoos and benign pigmented lesions in dermatology and plastic surgery.

## Device Story

The Umeco OPTIMA 694 is a Ruby laser system used in dermatology and plastic surgery for tattoo and pigmented lesion removal. It consists of an optical cavity with a Ruby laser rod, flashlamps, and a red helium-neon aiming beam. The system delivers light at 694 nm with a 20-nanosecond pulsewidth via an articulated arm and focusing handpiece in non-contact mode. It features an internal closed-loop water-air heat exchanger for thermal control. The device is operated by clinicians. The laser energy is absorbed by tattoo ink or melanin, allowing for precise tissue action. The system includes an electronic power supply and interface circuitry.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and performance specifications compared to predicate devices.

## Technological Characteristics

Ruby laser system; 694 nm wavelength; 20 ns pulsewidth; 2.2 Joules max output; articulated arm delivery; focusing handpiece; internal closed-loop water-air heat exchanger; Class I aiming beam; non-contact operation.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Derma-Lase DLR-1
- Spectrum RD 1200

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

982113

ದ್ದಾರೆ ರಾಜ

# 510k summary: Umeco OPTIMA 694 Ruby Laser System

{1}------------------------------------------------

## 510k summary

Submitter's name: lain D. Miller, Ph.D. Medical Laser Solutions LLC 65 School Street Charlestown, MA 02129 on behalf of : Umeco Corporation 501/5F Chungang Apt. Plant # 278-51 Sungsoo 2-ga-1-dong Sungdong-ku Scoul, Korea 133-120 Dovice Trade Name: Umeco OPTIMA 694 Ruby Laser System Common Name: Dermatology Laser Classification Name: Laser Surgical Instrument for use in General and Plastic Surgery and Dermatology

Device Class:

· ··· class II

{2}------------------------------------------------

#### Summary of Safety and Effectiveness Data Appendix E .
Callery of the final of the final of the finalisting of the finalismes of the finalisting of the finalisting of the finalismes of the finalisting of the formation of the fo

## General Information and Description

The Umeco OPTIMA 694 system is based on an optical cavity containing a Ruby laser rod. which is activated by means of the use of flashlamps. After the cavity, a red helium neon aiming beam is reflected onto a coaxial beam path using a beamsplitter assembly. The combined therapuetic and aiming beams are guided down an articulated arm delivery system to a focusing handpiece. The laser is used in non-contact mode.

The System is capable of emitting up to 2.2 Joules of light at 694 nm, with a nomimal pulsewidth of 20 nanoseconds. The laser is intended to be used for intended for the removal of blue/black tattoos and benign pigmented lesions in dermatology and plastic surgery.

The Ruby laser system is designed with 3 major sub-systems:

- a) An articulated arm delivery system terminated in a focusing handpiece.
- b) An electronic power supply and interface circuitry.
- c} An optical chamber containing laser rod and laser cavity optics.

No accessories are available for use with the OPTIMA 694 ruby laser system.

## Summary of Substantial Equivalence

Umeco believes that its OPTIMA 694 Ruby laser is substantially equivalent to the Derma-Lase DLR-1 and Spectrum RD 1200 Ruby laser systems.

The Derma-Lase DLR-1 and Spectrum RD 1200 Ruby laser systems are both intended for the removal of blue/black tattoos and benign pigmented lesions. They therefore have the same Intended Use as the its OPTIMA 694 Ruby laser system.

All three lasers emit at a wavelength of 694nm, a wavelength which is highly absorbed in the ink comprising blue/black tattoos and in the naturally occurring melaning comprising benign pigmented. All three lasers are therefore absorbed in an equally small volume of tissue, allowing for very precise action on the target tissue components.

All three lasers have a nominal pulsewidth in the range 20-30 nanoseconds and therefore realize similar peak power with minimal thermal conduction during the pulse.

All three lasers offer a range of available spotsizes with diameter in the range 4 - 8 mm.

{3}------------------------------------------------

All three lasers utilize an articulated arm delivery system for maximum flexibility and reliability. Both lasers utilize class I aiming beams which pose no hazard to the user.

.

All three systems utilize an internal closed loop water-air hear exchanger circuit for optimal thermal control of the laser cavity.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines representing hair or water. The overall design is simple and symbolic, representing the department's focus on health and human well-being.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 3 0 1998

Umeco Incorporated c/o Iain D. Miller, Ph.D. President Medical Laser Solutions LLC 65 School Street Charlestown, MA 02129

Re: K982113 Trade Name: Umeco OPTIMA 694 Ruby Laser System Regulatory Class: II Product Code: GEX August 25, 1998 Dated: Received: August 31, 1998

Dear Dr. Miller:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ದ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Iain D. Miller, Ph.D.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

C. Celia M. Whitten, Ph.D., M.D.

Celia M. Witten, Ph.D., M.D. Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### Appendix l

510(k) Number (if known): K98 2113

Device Name: Umeco OPTIMA 694

Indications for Use:

.

The Umeco OPTIMA 694 system is intended for the use in the removal of blue/black tattoos and benign pigmented lesions in dermatology and plastic surgery.

(PEEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                             |
|--------------------------------------------------------|-----------------------------|
| (Division Sign-Off)                                    |                             |
| Division of General Restorative Devices                |                             |
| 510(k) Number                                          | K982113                     |
| Prescription Use<br>(Per 21 CFR 801.109)               | OR                          |
|                                                        | Over-the-counter use ______ |

---

**Source:** [https://fda.innolitics.com/device/K982113](https://fda.innolitics.com/device/K982113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).