POLYVUE 43 (OCUFILCON A) SPHERICAL, TORIC AND ASPHERICAL MULTIFOCAL HYDROPHILIC CONTACT LENSES FOR DAILY WEAR
K982110 · Optech, Inc. · LPL · Nov 6, 1998 · Ophthalmic
Device Facts
Record ID
K982110
Device Name
POLYVUE 43 (OCUFILCON A) SPHERICAL, TORIC AND ASPHERICAL MULTIFOCAL HYDROPHILIC CONTACT LENSES FOR DAILY WEAR
Applicant
Optech, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Nov 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lens (clear and visibility tinted) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia), presbyopia and astigmatism in aphakic and non-aphakic persons with non-diseased eyes. Eyecare practitioners may prescribe the lens for daily wear in a Frequent Replacement Program. The lenses may be disinfected using heat, chemical or hydrogen peroxide disinfection systems.
Device Story
PolyVue 43™ are hydrophilic soft contact lenses manufactured from ocufilcon A (HEMA and methacrylic acid crosslinked with EGDMA). Lenses are hemispherical shells designed to fit over the corneal surface to provide refractive correction for myopia, hyperopia, presbyopia, and astigmatism. Features include spherical, aspherical, and toric designs with beveled edges to aid centration and comfort; multifocal designs provide additional add power for near vision and block spherical aberration. Lenses are prescribed by eyecare practitioners for daily wear, including frequent replacement programs. Patients use standard lens care regimens (heat, chemical, or hydrogen peroxide disinfection) to maintain the lenses. The device functions by conforming to the corneal shape and providing corrective refractive power corresponding to the eye's needs.
Clinical Evidence
Bench testing only. Studies included cytotoxicity, systemic injection, eye irritation, compatibility, preservative uptake/release, and leachability of color additives/monomers. Results confirmed material is non-toxic, compatible with standard care, and leachables are within acceptable limits. Manufacturer also verified ability to produce multifocal lenses across various specifications.
Technological Characteristics
Material: ocufilcon A (HEMA/methacrylic acid crosslinked with EGDMA). Water content: 43%. Refractive index: 1.427. Dk (35°C): 11.340 x 10^-11. Modulus: 36.50 g/mm². Form factor: Hemispherical shell, lathe cut from molded buttons. Diameter: 14.0-15.0mm. Base curve: 8.0-9.3mm. Power range: -20.00 to +20.00 D. Multifocal add power: up to 3.25 D. Sterilization: Not specified (standard manufacturing).
Indications for Use
Indicated for daily wear for correction of myopia, hyperopia, presbyopia, and astigmatism in aphakic and non-aphakic persons with non-diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Specialty UltraVision's Specialty T (ocufilcon A) Toric Hydrophilic Contact lens (K963764)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 6 1998
# 510 (K) Summary
## SUBMITTER:
#### Submitted on behalf of:
| Company Name: | Optech, Inc. |
|---------------|----------------------------------------------------|
| Address: | 6341 South Troy Circle<br>Englewood, CO 80111-6415 |
Phone! Fox:
303 708-1390 303 708-1392
#### CONTACT PERSON: Jarnes A. Brooks
#### DATE SUMMARY PREPARED: June 10, 1998
7 : 0 : 0 : 0000000
TRADE NAME: The PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lenses for Daily Wear (clear and - i ------------------------------------------------------------------------------------------------------------visibility tinted)
### COMMON NAME: contact lens
### SUBSTANTIALLY EQUIVALENT TO: -- ------
The PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophlic Contact Lenses for Dally Wear (clear and visibility tinted) is equivalent to Specially UltraVision's Specialty T (ocufilcon A) Toric Hydrophlic Contact lens for daily wear as currently marketed in the U.S.
The PolyVue 43™ (ocurilcon A) Spherical, Toric and Aspherical Multifocal l lydrophille Contact Lenses for Dally Wear (clear and visibility tinted) are substantlally equivalent to Specialty UltraVision's Specialty T (ocufilcon A) Toric lens. PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lens (clear and visibility tinted) is the same lens " material that received marketing clearance pursuant to K963764 for Specialty UltraVision, Inc. This lens conforms to and Is substantially equivalent to spherical, aspherical, presbyopic and astigmatic lens designs that are currently marketed in the United States. The intended use and target population are substantlally equivalent.
This lens is in Group 'J lonic, low water content polymers as established by the FDA and located in the Guldance Document for Dally Wear Contact Lenses,
{1}------------------------------------------------
Revised Edition May 1994. The physical, optical and chemical propertles of the PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lenses for Daily Wear (clear and visibility tinted) and the Specialty T (ocufilcon A) Soft (Hydrophilic) Toric Contact Lens for Dally Wear are substantially equivalent.
| PARAMETER | | PolyVue 43™ | Specialty T Toric | |
|---------------------|--|--------------------------------------------------|-----------------------------------|--|
| material | | ocufilcon A | ocufilcon A | |
| indication for use | | myopia, hyperopia, presbyopia<br>and astigmatism | myopia, hyperopia and astigmatism | |
| water content | | 43% | 44.1% | |
| light transmittance | | 98.513 | 98.493 | |
| Dk (35°C) | | $11.340 X 10^{-11}$ | $11.267 X 10^{-11}$ | |
| modulus | | 36.50 g/mm² | 35.96 g/mm² | |
| tensile strength | | 34.68 g/mm² | 33.98 g/mm² | |
| elongation @ break | | 169.26 % | 165.27% | |
| toughness | | 31.9668 g/mm² | 31.8488 g/mm² | |
| color | | clear and visibility tinted | clear | |
| refractive index | | 1.427 | 1.427 | |
| Powers | | -20.00 TO +20.00 d | -20.00 to +20.00 D | |
## SIMILARITIES and DIFFERENCES:
#### DESCRIPTION of the DEVICE:
Soft contact lenses are hemispherical shells manufactured of polymerized material of HEMA and methacrylic acid crosslinked with EGDMA, which yield the appearance of lenses, which are designed to fit over the corneal surface of the eyc. These lenses are designed with varying base curves which conform to the shape of the radius of the cornea and center over the apex of the cornea to provide corrective refi action of functional conditions of the eye including myopla (nearsightedness), hyperopia (farsightedness), presbyopia and astigmatism (multiple foci). Each lens provides corrective power, which is to correspond to the refractive power of the eye to which It Is being treated. Each lens is designed with a base curve on the internal side of the lens and an optical zone in the center of the lens which is generally of a diameter greater than 6mm. Spherical and aspheric curves as well as beveled edge configurations are built into the lens for alding in lens centration and comfort as well as providing additional add power for near and blocking spherical aberration.
#### PERFORMANCE
Optech, Inc. has carried out a complete performance evaluation of the PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact
{2}------------------------------------------------
Lenses for Daily Wear (clear and visibility tinted). This assessment addressed the following issues:
- Cytotoxicity Tast 1.
- SystemIc Injection I est 2.
- 3. Eye Irritation Test
- 4. Compatibility tasting
- ഗ് Preservative Uptake and Release
- Leachability of color additives and residual monomers 6.
These studies demonstrate that the material Is non toxic, compatible with standard lens care regimens and that the residual monomers and color additive leachability is within acceptable limits.
Concerning compatibility testing, the recommended lens care products (cleaning, rinsing and disinfection) have been approved for use with lenses of the same lens group. Therefore, no additional compatibility testing is Included.
with regard to preservalive uptake and release studies, since the subject contact lens has the same surface electric charge as the predicate device (same material being used) no additional studies need be conducted.
We have also included in this submission the results of 10 multifocal lenses, manufactured to variely of prescribed specifications to verify ablity of the manufacturer to make these lenses.
## INDICATIONS FOR USE:
Device Name: PolyVue 43™ (ocufiicon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lous (clear and visibility tinted)
The Polyvue 43m (ocutiicon A) Spherical, Toric and Aspherical Multifocal Hydrophlic Contact Lens (clear and visibility tinted) Is Indicated for dally wear for the correction of refractive ametropia (myopla and hyperopia), presbyopia and astigmatism In aphakic and non-aphakic persons with non-diseased eyes.
Lyecare practitioners may prescribe the lens for dally wear in a Frequent Replacement Program. The lenses may be disinfected using heat, chemical or hydrogen peroxide disinfection systems.
{3}------------------------------------------------
## PARAMETERS AVAILABLE:
## PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophlic Contact Lens (clear and visibility tinted)
・
### Spherical and Aspherical Multifocal
Powers: Center Thickness: Diameter: Bose Curve: Cylinder Powers: Axis:
+20.00 to -20.00D 0.03 το 0.40mm 14.0, 14.5mm 8.3 to 8.6mm N/A N/A
Toric +20.00 to -20.00D 0.101 to 0.520mm 14.0, 14.5, 15.0mm 8.0 to 9.3mm -0.50 to -16.00 in 0.25 D steps 10 to 1800 in 10 sleps
#### For the Multifocal Design
Add Powers:
Continuous adds to 3.25 D
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
1998 6 NOV
James A. Brooks 6341 S. Troy Circle Unit E Englewood, CO 80111-6415
Re: K982110/S1
Trade Name: Poly Vue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lenses for Daily Wear (clear and visibility tinted, lathe cut from molded buttons) Regulatory Class: II Product Code: 86LPL Dated: August 25, 1998 Received: August 27, 1998
Dear Mr. Brooks:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. James A. Brooks
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS STATEMENT
#### Device Name: PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophilic Contact Lenses for Dally Wear (clear and visibility tinted)
The PolyVue 43™ (ocufilcon A) Spherical, Toric and Aspherical Multifocal Hydrophlilic Contact Lens (clear and visibility tinted) is indicated for daily wear for the correctlon of refractive ametropia (myopia and hyperopia), presbyopla and astigmatism in aphakic and non-aphakic persons with non-diseased eyes.
Eyecare practitioners may prescribe the lens for daily wear In a Frequent Replacement Program. The lenses may be disinfected using heat, chemical or hydrogen peroxide disinfection systems.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | <u>Ming-chuen Shut</u><br>(Division Sign-Off)<br>Division of Ophthalmic Devices |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use | <div style="display:inline-block; vertical-align:middle;"> <img height="5" src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAUAAAAFCAYAAACNbyblAAAAHElEQVQI12P4//8/w+r8+KXEy8vLgYmJicEA4v8AAQ1OBN52kHQAAAAASUVORK5CYII=" width="5"/> </div> |
| or Over-The-Counter Use | 510(k) Number <u>K982110</u> |
(Optional Format 1-2-9G)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.