ACCU-CHEK ADVANTAGE TEST STRIPS AND ACCU-CHEK ADVANTAGE H TEST STRIPS
Applicant
Boehringer Mannheim Corp.
Product Code
LFR · Clinical Chemistry
Decision Date
Sep 3, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Accu-Chek Advantage and Accu-Chek Advantage H Test Strips are to be used with the Accu-Chek® Advantage® and Accu-Chek® Complete™ Monitors. The Accu-Chek Advantage and Accu-Chek Complete monitors are designed for testing glucose in whole blood by persons with diabetes or by health care professionals in the home or in health care facilities. Professionals may use the test strips to test capillary, venous, arterial and neonate (including cord) blood samples; lay use is limited to capillary blood testing.
Device Story
Accu-Chek Advantage and Advantage H Test Strips function with Accu-Chek Advantage and Complete monitors to measure blood glucose. System uses whole blood samples; provides results reflecting laboratory plasma glucose values. Modification involves updated reference curve in code key to align with plasma-referenced values rather than traditional whole blood values. Used by patients with diabetes (home) or healthcare professionals (clinical settings). Provides quantitative glucose concentration to assist in diabetes management.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological comparison and modification of the reference curve to align with laboratory plasma glucose standards.
Technological Characteristics
In vitro diagnostic test strips for glucose measurement. Uses glucose dehydrogenase enzyme chemistry. System includes code key for calibration. Designed for use with Accu-Chek Advantage and Complete monitors.
Indications for Use
Indicated for blood glucose monitoring in patients with diabetes. Professionals may test capillary, venous, arterial, and neonate (including cord) blood. Lay users are restricted to capillary blood testing.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
K980731 — ACCU-CHEK COMFORT CURVE TEST STRIPS · Boehringer Mannheim Corp. · May 22, 1998
K971812 — PRECISION QID BLOOD GLUCOSE TEST STRIP, PRECISION G BLOOD GLUCOSE TEST STRIP · Medisense, Inc. · Aug 8, 1997
K091157 — GLUCO LAB AUTO CODING, MODEL IGM-0022 · Infopia Co, Ltd. · Nov 17, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K982089
## Accu-Chek Advantage and Accu-Chek Advantage H Test Strips
## 510(k) Summary
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) Submitter name, address, contact | Boehringer Mannheim Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 845-2000 |
| Contact Person: | Mike Flis |
| Date Prepared: | June 5, 1998 |
| 2) Device name | Proprietary name: Accu-Chek™ Advantage® and Accu-Chek™ Advantage H™ Test Strips<br>Common name: Blood glucose test system<br>Classification name: glucose dehydrogenase, glucose |
| 3) Predicate device | We claim substantial equivalence to the unmodified Accu-Chek Advantage Test Strip. |
| 4) Device Description | The Accu-Chek Advantage and Accu-Chek Advantage H Test Strips are to be used with the Accu-Chek® Advantage® and Accu-Chek® Complete™ Monitors. The test strips are designed for convenient, confident, and accurate testing of blood glucose in whole blood samples. |
Continued on next page
{1}------------------------------------------------
## 510(k) Summary, Continued
| 5) Intended use | The Accu-Chek Advantage and Accu-Chek Advantage H Test Strips are to be<br>used with the Accu-Chek® Advantage® and Accu-Chek® Complete™<br>Monitors. The Accu-Chek Advantage and Accu-Chek Complete monitors are<br>designed for testing glucose in whole blood by persons with diabetes or by<br>health care professionals in the home or in health care facilities.<br>Professionals may use the test strips to test capillary, venous, arterial and<br>neonate (including cord) blood samples; lay use is limited to capillary blood<br>testing. |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 6) Comparison<br>to predicate<br>device | The modified Boehringer Mannheim Accu-Chek Advantage and Accu-Chek<br>Advantage H Test Strips are substantially equivalent to other products in<br>commercial distribution intended for similar use. Most notably it is<br>substantially equivalent to the currently marketed Accu-Chek Advantage Test<br>Strip (k951887 and k954833). |
| | Blood glucose concentrations can be expressed two ways, as whole blood<br>glucose or as plasma (or serum) glucose. The concentration of glucose in<br>whole blood is lower (by about 8%) than the plasma glucose value for this<br>same blood sample because of the differences in distribution of water between<br>the two compartments (red blood cells and plasma). Test strips developed for<br>the measurement of whole blood samples were traditionally referenced to<br>provide a whole blood glucose value. However, many North American<br>laboratories have, for the last few decades, routinely measured and reported<br>only the plasma glucose values. Recent advancements in the management of<br>patients with diabetes have created a preference to have the results for blood<br>glucose test strips reflect the laboratory plasma value instead of the more<br>traditional whole blood value. For this reason Boehringer Mannheim has<br>modified the reference curve employed within the code key used with the<br>Accu-Chek Advantage and Accu-Chek Advantage H test strips to now<br>provide results which reflect the higher laboratory plasma glucose value. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP 3 1998
Mike Flis . Requlatory Affairs Specialist Boehringer Mannheim 9115 Hague Road P.O. Box 50457 46250-0457 Indianapolis, Indiana
K982089 Re : Accu-Chek® Advantage™ and Advantage H™ Test Strips Requlatory Class: II Product Code: LFR Dated: June 11, 1998 Received: June 15, 1998
Dear Mr. Flis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice; labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major ---regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰੇ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note -the regulation entification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure ... ...
{4}------------------------------------------------
510(k) Number (if known): Device Name: Accu-Chek™ Advantage® and Accu-Chek™ Advantage H™ Test Strips Indications for Use:
The Accu-Chek Advantage and Accu-Chek Advantage H Test Strips are to be used with the Accu-Chek® Advantage® and Accu-Chek® Complete™ Monitors. The Accu-Chek Advantage and Accu-Chek Complete monitors are designed for testing glucose in whole blood by persons with diabetes or by health care professionals in the home or in health care facilities.
Professionals may use the test strips to test capillary, venous, arterial and neonate (including cord) blood samples; lay use is limited to capillary blood testing.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K982084
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.