K982058 · Encompass Group, LLC · FYA · Jan 14, 1999 · General, Plastic Surgery
Device Facts
Record ID
K982058
Device Name
ROTECNO FR (FLUID RESISTANT) SURGICAL GOWN
Applicant
Encompass Group, LLC
Product Code
FYA · General, Plastic Surgery
Decision Date
Jan 14, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Intended Use
Intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Device Story
Rotecno® FR Surgical Gown is a reusable protective garment for operating room personnel. Fabricated from 99% polyester and 1% carbon, the material is chemically treated to provide fluid resistance. Designed for durability, the gown maintains protective performance through up to 150 laundry and sterilization cycles. Used in surgical settings to prevent cross-contamination between patients and staff by blocking transfer of microorganisms, body fluids, and particulates.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included breaking strength, tear resistance, air permeability, water repellency, impact penetration, hydrostatic head, flammability, and aqueous bacterial barrier testing. Toxicity testing was also performed and found acceptable.
Technological Characteristics
Reusable surgical gown; 99% polyester, 1% carbon chemically treated fabric. Designed for up to 150 laundry/sterilization cycles. Provides fluid resistance and bacterial barrier protection.
Indications for Use
Indicated for operating room personnel during surgical procedures to provide a barrier against microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
K241214 — ProPel SG3 Surgical Gown · Standard Textile Co., Inc. · Sep 10, 2024
K032631 — INNERBLOC LR SURGICAL GOWNS BASED ON GORE LR FABRIC (REUSABLE 100X) · Lac Mac, Ltd. · Feb 5, 2004
K012268 — TRIMAX · Standard Textile Co., Inc. · Dec 3, 2001
K962666 — BEMISS JASON-STERILE,NON-STERILE,REINFORCED STERILE,AND REINFORCED NON-STERILE SURGICAL GOWN · Bemiss-Jason Corp., Triad Medical Div. · Feb 27, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
11. Al
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
:
| Manufacturer: | Dowling Textile Company<br>615 Macon Road<br>McDonough, Georgia 30253<br>(770) 957-3981 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Contact: | Tim Cummins |
| Common Name: | Reusable Surgical Gown |
| Trade Name: | Rotecno® FR Surgical Gown |
| Classification Number: | 79FYA |
| Class, Regulation: | II per 21 CFR §878.4040 |
| Predicate Device: | SafeCare™ Reusable Surgical Gown |
| Device Description: | A reusable surgical gown fabricated from 99% polyester, 1% carbon, chemically treated fabric and designed to provide fluid protection through up to 150 laundry/sterilization cycles. |
| Intended Use: | "Intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material." |
| Substantial Equivalence: | The Rotecno® FR Surgical Gown is substantially equivalent to the SafeCare™ Gown in material, design, form, fit, and function. |
| Toxicity Testing Summary: | The material used in the fabrication of the Rotecno® FR Surgical Gown has been found acceptable for its intended use. |
| Performance Testing Summary: | The Rotecno® FR Surgical Gown has been tested for breaking strength, tear resistance, air permeability, water repellency, impact penetration, hydrostatic head, flammability, and aqueous bacterial barrier and found acceptable for its intended use. |
K982058 510 (k) Summary XVI.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the agency's commitment to health, human services, and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 4 1999
Mr. Tim Cummins RA/QA Manager Dowling Textile Company P.O. Box 598 615 Macon Road 30253 McDonough, Georgia
Re : K982058 Rotecno® FR (Fluid Resistant) Surgical Gown Trade Name: Requlatory Class: II Product Code: FYA December 14, 1998 Dated: December 21, 1998 Received:
Dear Mr. Cummins:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{2}------------------------------------------------
Page 2 - Mr. Cummins
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Diredtor Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K982058
Rotecno® FR Surgical Gown, Non-sterile and Reusable Device Name:
Indications For Use:
"Intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material."
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Neurre 1. Miller for Chin S. Tim, Ph D
(Division Sign-Off) Division of Dental. Infection Control, and General Hospital Devi 510(k) Number
Prescription Use_
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.