← Product Code [CAF](/productcode/CAF) · K981944

# BREATHERITE (K981944)

_Ventlab Corp. · CAF · Aug 24, 1998 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K981944

## Device Facts

- **Applicant:** Ventlab Corp.
- **Product Code:** [CAF](/productcode/CAF.md)
- **Decision Date:** Aug 24, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

VENTLAB Corporation's BreatheRite MDI Spacer is designed to work in conjunction with a metered dose inhaler (MDI) to provide a dispersion chamber that will aid the sufferer in inhaling a suitable asthmatic medication properly. The BreatheRite MDI Spacer consists of a chamber that secures the metered dose inhalers mouthpiece with an elastomeric flange adapter at one end, a chamber connected to a one-way valve adapter, which opens on inhalation, and an exhalation adapter that allows the air to exhaust to the atmosphere both of which are housed within the mouthpiece. Federal (USA) law restricts this device to sale by or on the order of a physician.

## Device Story

BreatheRite MDI Spacer functions as a dispersion chamber for metered dose inhalers (MDI); device attaches to MDI mouthpiece via elastomeric flange adapter. Operation involves one-way valve opening upon patient inhalation to deliver medication; exhalation adapter allows air exhaust to atmosphere. Used in clinical or home settings under physician prescription; assists patients in coordinating inhalation with MDI actuation; improves medication delivery efficiency for asthma sufferers.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Dispersion chamber; elastomeric flange adapter; one-way valve; exhalation adapter; mouthpiece housing. Non-powered, mechanical device.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 24 1998

Mr. E.J. Smith Ventlab Corporation c/o Smith Associates P.O. Box 4341 Crofton, MD 21114-4341

Re: K981944 BreatheRite Regulatory Class: II (two) Product Code: 73 CAF Undated Dated: Received: June 3, 1998

Dear Mr. Smith:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{1}------------------------------------------------

Page 2 - Mr. E.J. Smith

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahon

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## 510(k) Number (if known):

Device Name: BreatheRite MDI Spacer

Classification Panel: 73 CAF

Indications for Use:

VENTLAB Corporation's BreatheRite MDI Spacer is designed to work in conjunction with a metered dose inhaler (MDI) to provide a dispersion chamber that will aid the sufferer in inhaling a suitable asthmatic medication properly. The BreatheRite MDI Spacer consists of a chamber that secures the metered dose inhalers mouthpiece with an elastomeric flange adapter at one end, a chamber connected to a one-way valve adapter, which opens on inhalation, and an exhalation adapter that allows the air to exhaust to the atmosphere both of which are housed within the mouthpiece.

Federal (USA) law restricts this device to sale by or on the order of a physician.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use | <span style="text-decoration: overline;">✓</span> | or | Over-the-Counter Use <span style="text-decoration: overline;"></span> |
|------------------|---------------------------------------------------|----|-----------------------------------------------------------------------|
|------------------|---------------------------------------------------|----|-----------------------------------------------------------------------|

|               | <span style="text-decoration: overline;">Lank W. Wedos 8-18-99</span> |
|---------------|-----------------------------------------------------------------------|
|               | (Division Sign-Off)                                                   |
|               | Division of Cardiovascular, Respiratory, and Neurological Devices     |
| 510(k) Number | <span style="text-decoration: overline;"></span>                      |

---

**Source:** [https://fda.innolitics.com/device/K981944](https://fda.innolitics.com/device/K981944)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
