K981841 · Respiraid , Ltd. · FRF · Oct 13, 1998 · General Hospital
Device Facts
Record ID
K981841
Device Name
BREATHE EASY, MODELS AD AND CD
Applicant
Respiraid , Ltd.
Product Code
FRF · General Hospital
Decision Date
Oct 13, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BREATHE EASY device is a medical recirculating air cleaner designed to remove airborne particles and allergens, such as: dust, smoke, pollen, mold spores, animal hair and dander, dust mites, and harmful fibers, that may lead to allergic reactions.
Device Story
BREATHE EASY is an adjustable, personal medical recirculating air cleaner for treating air in a localized area. System comprises a housing unit with air inlet/outlet, blower, and filter to remove airborne contaminants. Model AD includes integrated heater and humidifier to adjust ambient air to preselected respiratory comfort levels. Device mounts via adjustable support arm to furniture, walls, or wheeled structures. It guides treated air outflow to create a localized 'treated air envelope' around the user's respiratory openings. Used in home or clinical settings to reduce exposure to allergens; helps manage allergic reactions by providing filtered, conditioned air directly to the user.
Clinical Evidence
No clinical data provided; bench testing only. Device complies with IEC 601-1 safety standards.
Technological Characteristics
Medical recirculating air cleaner; includes blower, filter, housing, and adjustable support arm. Model AD includes heater and humidifier. Complies with IEC 601-1. No specific materials or ASTM standards cited.
Indications for Use
Indicated for individuals suffering from allergic reactions to airborne particles and allergens including dust, smoke, pollen, mold spores, animal hair, dander, dust mites, and harmful fibers.
Regulatory Classification
Identification
A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.
{1}------------------------------------------------
P 1/2
OCT 1 3 1998
510(k) Summary
# RespirAid Ltd. BREATHE EASY
# 510(k) Number K981841
# Applicant's Name:
RespirAid Ltd. POB 12, Katzrin 12900, Israel
#### Contact Person:
Shoshana Friedman Push-med Ltd. 117 Ahuzah St. Ra'ananna 43373, Israel Telephone: 972-9-771-8130 Fax: 972-9-771-8131
### Date Prepared:
May 24, 1998
## Trade Name:
BREATHE EASY (Models AD and CD)
# Classification Name:
Medical Recirculating Air Cleaner
## Classification:
The FDA has classified medical recirculating air cleaners as class II devices (product code 80 FRF, regulation no. 880,5045) and it is reviewed by General Hospital Devices.
{2}------------------------------------------------
K981841
p2/2
#### Predicate Device:
- · Enviracaire EV-1 (Enviracaire, Co.) K872359
#### Indication for Use:
The BREATHE EASY device is a medical recirculating air cleaner designed to remove airborne particles and allergens, such as: dust, smoke, pollen, mold spores, animal hair and dander, dust mites, and harmful fibers, that may lead to allergic reactions.
#### Device Description:
The BREATHE EASY is an adjustable, personal system for treating air in a specified area of a room. The BREATHE EASY device contains an air treatment system, including a housing unit with an air inlet and a treated air outlet, a blower and a filter, for removing contaminants from the air flowing along the flow path.
The BREATHE EASY Model AD device contains also an air filtering system with a heater and humidifier for treating ambient air so as to bring it to a preselected respiratory comfort level. The BREATHE EASY device provides a method for guiding the treated air outflow so as to obtain a flow distribution in close proximity to the head of a user, thereby forming a treated air envelope surrounding his respiratory openings.
The BREATHE EASY device is supported on an adjustable support arm, adjustably mounted onto a household furniture (e.g., bed), wall, or wheeled support structure.
#### Performance Standards:
No performance standards have been established for such devices under Sections 514 of the Federal Food, Drug, and Cosmetic Act, However, the BREATHE EASY complies with the IEC 601-1.
#### Substantial Equivalence:
The BREATHE EASY device is substantially equivalent to Enviracaire EV-1 cleared under K872359 in respect to intended use, technological characteristics, performance, and labeling.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or other bird, with three parallel lines forming the body and wings. The bird is facing to the right.
. .
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 13 1998
RespirAid Limited C/O Ms. Shoshana Friedman Push-med Limitd 117 Ahuzah St. Ra'ananna 43373 ISRAEL
Re: K981841 Trade Name: BREATHE EASY, Models AD and CD Regulatory Class: II Product Code: FRF Dated: Auqust 24, 1998 Received: September 4, 1998
Dear Ms. Friedman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration .................................................................................................................................................................
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਕਿ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{4}------------------------------------------------
Page 2 - Ms. Friedman
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address
"http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Contro. and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
.......
{5}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K981841
Device Name:
### BREATHE EASY
Indications for Use:
The BREATHE EASY device is a medical recirculating air cleaner designed to remove airborne particles and allergens, such as: dust, smoke, pollen, mold spores, animal hair and dander, dust mites, and harmful fibers, that may lead to allergic reactions.
Qlin S. him
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
of Reptaductive, Abdominal, Ear, Nose and Rhroat, and Radiological Devices Division
510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over the Counter Use X
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.