← Product Code [CAF](/productcode/CAF) · K981772

# HALOLITE AAD SYSTEM (K981772)

_Medic-Aid , Ltd. · CAF · Nov 3, 1998 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K981772

## Device Facts

- **Applicant:** Medic-Aid , Ltd.
- **Product Code:** [CAF](/productcode/CAF.md)
- **Decision Date:** Nov 3, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

A Nebulizer system designed to aerosolize commonly prescribed liquid medication ( except Pentamidine) for inhalation by the patient in the home care, nursing home or hospital environment.

## Device Story

HaloLite AAD is a portable, single-patient-use nebulizer system. It uses a pressure transducer to analyze patient breathing patterns and a microprocessor to control aerosol delivery. The device pulses aerosol exclusively during inspiration, matching pulse duration to inspiratory time. It continuously monitors breathing to adapt to changes throughout treatment. Once a pre-set dose is delivered, the system signals treatment completion. Used in homes, hospitals, and nursing homes by patients or clinicians. The device is battery-powered (9V), enhancing portability compared to bench-mounted systems. It eliminates the need for manual calibration or sensitivity adjustments required by earlier systems, simplifying operation for both patients and healthcare providers.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and design comparison.

## Technological Characteristics

Materials: Polypropylene (mouthpiece). Sensing: Pressure transducer for inhalation detection. Energy: 9V internal battery. Form factor: Hand-held portable unit. Connectivity: None. Sterilization: Provided clean, non-sterile. Software: Microprocessor-controlled, rule-based algorithm for breath-synchronized aerosol delivery.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- Vortran - AutoNeb ([K935693](/device/K935693.md))
- Medic-Aid - Ventstream ([K933535](/device/K933535.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Medic-Aid Limited Heath Place, Bognor Regis, West Sussex, UK PO22 9SL Telephone (01243) 840888 Fax (01243) 846100

3 1998
K98 1772

NOV

Image /page/0/Picture/2 description: The image shows the logo for Medic-Aid. The logo consists of three black triangles arranged to look like a mountain range. Below the mountain range is the name of the company, "MEDIC-AID", in black, sans-serif font. There is a registered trademark symbol to the right of the name.

#### Non-Confidential Summary of Safety and Effectiveness

page 1 of 4 May 18, 1998

Medic-Aid Ltd. Heath Place Bognor Regis, West Sussex PO22 9SL United Kingdom

| Tel - 011-44-1243-840888   | Fax - 011-44-1243-846100                                      |
|----------------------------|---------------------------------------------------------------|
| Official Contact:          | Ed Walters, Quality Manager                                   |
| Proprietary or Trade Name: | HaloLite ™ AAD                                                |
| Common/Usual Name:         | Nebulizer                                                     |
| Classification Name:       | Nebulizer, Direct Patient Interface                           |
| Predicate Devices:         | Vortran - AutoNeb K935693<br>Medic-Aid - Ventstream - K933535 |

Device Description: -- 1000

The Medic-Aid HaloLite AAD system is a single patient use, prescribed portable nebulizer designed to deliver a precise and reproducible dose of aerosolized liquid medication. The device consists of a hand-piece containing a pressure transducer, a microprocessor control system, and a medication chamber, based on the Medic-Aid -Ventstream Nebulizer, K933535, and mouthpiece.

The HaloLite system analyses the patient's breathing pattern to determine the aerosol pulse time. The system then pulses aerosol during inspiration only. Each pulse of aerosol is matched according to the inspiratory time. HaloLite continues to monitor the breathing pattern in order to adapt to changes in breathing throughout the treatment. When the preset dose has been delivered, the system indicates that the treatment is complete.

- Aerosolization of commonly prescribed liquid drug (except Pentamidine) 1. Intended use for inhalation by the patient.
05/17/98

Page 2-2

Image /page/0/Picture/13 description: The image contains two logos. The logo on the left is a circular seal with the text "QUALITY ASSURED FIRM" around the top half of the circle and "ISO 9001" in the center, with a checkmark below it and the letters "SGS" at the bottom. The logo on the right features a crown at the top, a large checkmark below it, and the text "NATIONAL ACCREDITATION OF CERTIFICATION BODIES" below the checkmark.

{1}------------------------------------------------

## Non-Confidential Summary of Safety and Effectiveness

page 2 of 4 May 18, 1998

- Single patient / multi-use for Home, Hospitals and Nursing Homes 2. Environment of Use -
- Patients requiring nebulized drug delivery via a 3. Patient Population mouthpiece.

# Comparison to Other Legally Marketed Predicate Devices

The following comparison table details the primary attributes of the intended device and legally marketed predicate devices. The most significant attributes have been listed.

| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>"The state of the state a status and a complete of the status and the may be the may be the may be the may be the may be the may be the may be the may be the may be the may b | Comprehensive of Children Company of Children Company of Children Company of Children Company of Children Comments of Children                             | Comments of the contraction of the comments of the consistence of the consistence of the comments of the comments of the comments of the comments of the comments of the comme                                                                                                                                                                                   | Carolin Call Carol                                                                                                                                                                                                                           |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| All Market States of Contract The Seat Child Control Concess of                                                                                                                                                                                                                                                                                                  | A STATE A CONTRACT A                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | Company of Children Comments of Children Comments of Children<br>Control Concession                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | September 1978 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 199                                                                                                                                                                                   |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  | A                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | Company of the program and the program and the comments of the<br>THE CONSUMICAL PRODUCTION COLLECTION COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR COLLECTION CONTRACTOR CONTRACTOR CONTRACTOR CONTR                                                                                                                 |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | and the contraction of the control of the control of the contribution of the comments of the comments of the comments of the comments of the comments of the comments of the c<br>was and the consideration of the control and the control of the contribution of the contribution of the contribution of the contribution of the contribution of the contributi |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | THE CALL PROPERTY CONTRACT COLLECTION COLLECTION OF COLLECTION OF CONSULTION                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | 2017                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            | 1000                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  | and the contract of the proportion of the proposed and the<br>A BREAD A STATE CONTRACT COLLECTION CONSULTER OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF |
| Comments of the control of                                                                                                                                                                                                                                                                                                                                       | The man a mail in Carliner Select Senate Senatif School Starts of Children<br>Comments of the control of the control of the contribution of the country of | and and the same of the comments of the comments of<br>Concession of the control of the control of                                                                                                                                                                                                                                                               | Company of the program and the may be                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                              |

| Use                                                    |     |     |     |
|--------------------------------------------------------|-----|-----|-----|
| Intended to nebulize drugs                             | Yes | Yes | Yes |
| Sidestream Jet System                                  | Yes | Yes | No  |
| Synchronized Delivery of<br>Nebulized drug             | Yes | No  | Yes |
| Drug Delivery on demand                                | Yes | No  | Yes |
| Used in hospitals, home care,<br>nursing home settings | Yes | Yes | Yes |
| Single Patient / Multi-use                             | Yes | Yes | N/A |

**Design**

| Flow Rates                  | 6 LPM                                      | 6-8 LPM                                    | 1.5-16LPM                                  |
|-----------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|
| Gas Source - Compressed Air | Yes                                        | Yes                                        | Yes                                        |
| Software Driven             | Microprocessor                             | No                                         | No                                         |
| Mode of Operation           | Turn on<br>During<br>Patient<br>Inhalation | Breath<br>Enhanced<br>Continuous<br>System | Turn on<br>During<br>Patient<br>Inhalation |

{2}------------------------------------------------

#### Non-Confidential Summary of Safety and Effectiveness

page 3 of 4

May 18, 1998

| Attribute                  | HaloLite AAD System                       | Medic-Aid Ventstream K933535 | Vortran AutoNeb K935693                   |
|----------------------------|-------------------------------------------|------------------------------|-------------------------------------------|
| Drug delivery triggered by | Patient Inhalation Airway Pressure Signal | None                         | Patient Inhalation Airway Pressure Signal |
| Used with Mouthpiece       | Yes                                       | Yes                          | Yes                                       |

Materials

| Materials in Contact with Patient Polypropylene | Mouthpiece | Mouthpiece | Mouthpiece |
|-------------------------------------------------|------------|------------|------------|
|-------------------------------------------------|------------|------------|------------|

Packaging

| Provided clean, non-sterile | Yes | Yes | Yes |
|-----------------------------|-----|-----|-----|
|-----------------------------|-----|-----|-----|

Performance Standards / Specifications

| None applicable under Section 514 | Yes | Yes | Yes |
|-----------------------------------|-----|-----|-----|
|-----------------------------------|-----|-----|-----|

### Differences Between Other Legally Marketed Predicate Devices

The differences between the intended device, HaloLite, and the predicates are -

- 1. The HaloLite system provides a hand-held, user-friendly device compared with the bench-mounted Vortran - AutoNeb which requires tubes to interconnect between the AutoNeb and a standard nebulizer.
- 2. The HaloLite is micro-processor controlled device. This permits the HaloLite device to operate across a wide range of patient ages and breathing patterns, without requiring any calibration or adjustment to sensitivity settings as required by the AutoNeb. The system can also adapt to any changes in the breathing pattern and adjust to the aerosol pulse time throughout the treatment. Each aerosol pulse is

{3}------------------------------------------------

#### Non-Confidential Summary of Safety and Effectiveness

page 4 of 4 May 18, 1998

summed to provide an accurate method of dose measurement so that a pre-set dose can be delivered. The HaloLite operates over a pulse range of 0.15 to 2.0 seconds compared to the Vortran AutoNeb 0.5 to 5.0 seconds. This range has been selected as being the most appropriate to match the patient's inhalation.

- 3. The Vortran AutoNeb can also be used to deliver a pre-set dose. This requires setting the controls for each patient's breathing pattern, and counting the number of pulses delivered to the patient. This is achieved automatically with the HaloLite, the press of a button initiates treatment and when the pre-set dose has been delivered the device signals that the treatment has been completed. This system provides ease of use for both the physician and the patient. The Vortran - AutoNeb is operated from 115 volt mains supply, and the HaloLite is powered by an internal replaceable 9 volt battery.
There is no other differences between the intended device and the predicate devices which would be significant to patient safety or effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

3 1998 NOV

Mr. Paul E. Dryden Medic-Aid, Ltd. c/o ProMedic, Inc. 6329 W. Waterview Court 46055-9501 McCordsville, IN

Re: K981772 HaloLite AAD System Regulatory Class: II (two) Product Code: 73 CAF Dated: September 8, 1998 Received: September 9, 1998

Dear Mr. Dryden:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. Paul E. Dryden

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,
Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

Pursuant to the Notice of February 6, 1996 regarding listing of Indications for Use on a separate sheet, the following is per that request.

| 510(k) Number: | K981772 (To be assigned)                                                                                                                                                                    |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:   | HaloLite AAD System                                                                                                                                                                         |
| Intended Use : | A Nebulizer system designed to aerosolize commonly prescribed liquid medication ( except Pentamidine) for inhalation by the patient in the home care, nursing home or hospital environment. |

Concurrence of CDRH, Office of Device Evaluation (ODE)

Water

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices

510(k) Number

Prescription Use
1
(Per CFR 801.109)
or

ﺔ ﺑ

Over-the-counter use

---

**Source:** [https://fda.innolitics.com/device/K981772](https://fda.innolitics.com/device/K981772)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
