INNOVATIVE TECHNOLOGIES TRANSPARENT FILM AND INTELLIGENT FILM WOUND DRESSING

K981753 · Innovative Technologies , Ltd. · MGP · Jun 30, 1998 · SU

Device Facts

Record IDK981753
Device NameINNOVATIVE TECHNOLOGIES TRANSPARENT FILM AND INTELLIGENT FILM WOUND DRESSING
ApplicantInnovative Technologies , Ltd.
Product CodeMGP · SU
Decision DateJun 30, 1998
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Innovative Technologies Transparent & Intelligent Film Wound Dressings, are conformable, sterile wound dressings intended to provide a moist environment ideally suited for wound management. In particular for partial thickness wounds. Innovative Technologies Transparent Film dressings act as a semi occlusive barrier to exogenous moisture and bacteria whilst still allowing permeability to moisture vapour and oxygen. The film transmits water, but retains other exudate components creating the ideal environment for wound healing and extending the life of the dressing. The Innovative Technologies Intelligent wound dressings utilize an intelligent film to enhance product performance. The semi occlusive dressing acts as a barrier to exogenous moisture and bacteria whilst still allowing permeability to moisture vapour and oxygen. The film transmits water, but retains other exudate components creating the ideal environment for wound healing and extending the life of the dressing. Innovative Technologies Transparent Films & Intelligent Films can be used for partial thickness wounds. They can be used for pressure sores. superficial wounds such as minor cuts. Iacerations and abrasions, minor scalds and burns (18 & 2" degree), donor sites, dermal lesions, trauma wounds and post operative surgical wounds.

Device Story

Sterile, conformable, semi-occlusive film wound dressings; provide moist wound environment; act as barrier to exogenous moisture and bacteria; permeable to moisture vapor and oxygen; transmit water while retaining exudate components. Used for partial thickness wounds, pressure sores, minor burns/scalds, surgical wounds, and as secondary fixation. Applied by patients (OTC) or healthcare professionals. Benefits include wound protection and maintenance of healing environment. Discontinue if infection suspected.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Semi-occlusive film wound dressing; sterile; conformable; moisture vapor and oxygen permeable; water-transmitting; exudate-retaining barrier.

Indications for Use

Indicated for partial thickness wounds, pressure sores, superficial wounds (abrasions, lacerations, cuts), minor scalds and burns (1st and 2nd degree), donor sites, dermal lesions, trauma wounds, and post-operative surgical wounds. Also indicated for IV site dressing and secondary fixation for other wound care products (alginates, gels, foams) used for diabetic ulcers.

Submission Summary (Full Text)

{0}------------------------------------------------ ## Attachment 2 K981753 JUN 3 0 1998 ## Summary of Safety and Efficacy Summary of Innovative Technologies Transparent and Intelligent Film Wound Dressings Manufacturer: Regulatory Affairs Contact: Telephone: Date Summary Prepared: Device Trade Name: Common or Usual Name: Classification: Description:G Intended Use: Innovative Technologies, Limited Road Three, Winsford Industrial Estate Cheshire CW7 3PD, United Kingdom Christopher Oakes, Manager 44 1606 863 500 May 13, 1998 Transparent Film & Intelligent Film Wound Dressing. Film Wound Dressings Wound Dressings, currently unclassified by FDA. Innovative Technologies Transparent & Intelligent Film Wound Dressings, are conformable, sterile wound dressings intended to provide a moist environment ideally suited for wound management. In particular for partial thickness wounds. Innovative Technologies Transparent Film dressings act as a semi occlusive barrier to exogenous moisture and bacteria whilst still allowing permeability to moisture vapour and oxygen. The film transmits water, but retains other exudate components creating the ideal environment for wound healing and extending the life of the dressing. The Innovative Technologies Intelligent wound dressings utilize an intelligent film to enhance product performance. The semi occlusive dressing acts as a barrier to exogenous moisture and bacteria whilst still allowing permeability to moisture vapour and oxygen. The film transmits water, but retains other exudate components creating the ideal environment for wound healing and extending the life of the dressing. Innovative Technologies Transparent Films & Intelligent Films can be used for partial thickness wounds. They can be used for pressure sores. superficial wounds such as minor cuts. Iacerations and abrasions, minor scalds and burns (18 & 2" degree), donor sites, dermal lesions, trauma wounds and post operative surgical wounds. Discontinue use if any infection of the wound is suspected and seek guidance from a health care professional. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The eagle is depicted in a simple, graphic style, with bold lines and a sense of movement. The text is in a sans-serif font and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 3 0 1998 Mr. Christopher Oakes Innovative Technologies, Limited Road Three Winsford Industrial Estate Winsford, Cheshire United Kingdom CW7 3PD > Re: K981753 Trade Name: Innovative Technologies' Transparent and Intelligent Film Wound Dressing Regulatory Class: Unclassified Dated: May 13, 1998 Received: May 18, 1998 Dear Mr. Oakes: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations: - 1. This device may not be labeled for use on third degree burns. - 2. This device may not be labeled as having any accelerating effect on the rate of wound healing or epithelization. - 3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin. - 4. This device may not be labeled as a treatment or a cure for any type of wound. The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. {2}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html. Sincerely yours, Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page lof I K981753. 510(k) Number (if known): _ Device name: Innovative Technologies: Transparent Film & Intelligent Film Dressings Indications For Use: Innovative Technologies' Transparent Film and Intelligent Film Dressings are indicated for use on wounds including: - . Minor scalds and burns - Superficial Wounds such as Abrasions, Laccrations, and cuts t This wound dressing may be also used under the care of a health care professional for such wound as, Pressure sores, post operative surgical wounds, donor sites, trauma wounds and dermal Icsions. And for such uses as a IV site and us a secondary fixation device for other wound care products such as alginates, gels and foams used for diabctic ulccrs. (PLUASE DO NOT WRITE RELOW THIS LINE. CONTINUE ON ANDTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | | |----------------------|--| | (Per 21 CFR 801.109) | | | | OR Over The Counter Use <span style="text-decoration: overline;">✓</span> | |--|---------------------------------------------------------------------------| | | (Optional Format 1-2-96) | | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of General Restorative Devices | | | 510(k) Number | K981753 | P - 02
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...