TIMX LOW BACK SYSTEM TITANIUM MX ROD BASED LOW BACK SYSTEM
Applicant
Acromed Corp.
Product Code
MNH · Orthopedic
Decision Date
Jun 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When labeled for pedicle screw fixation, the TiMX Low Back System is intended for use in Grade 3 and 4 spondylolisthesis at L5-S1, when affixed to the lumbosacral spine, utilizing autologous bone graft, and intended to be removed after solid fusion is attained. Levels of attachment for this indication range from L3 to the sacrum. When labeled for non-pedicle screw posterior fixation, the TiMX Low Back System is intended for hook, wire and/or sacral/iliac screw fixation from the thoracic spine to the ilium/sacrum. Properly used, the posterior TiMX Low Back System will provide temporary stabilization as an adjunct to spinal bone grafting process. Specific indications are: 1. Idiopathic scoliosis. 2. Neuromuscular scoliosis/ kyphoscoliosis with associated paralysis or spasticity. 3. Scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele. 4. Spinal fractures (acute reduction or late deformity). 5. Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies). 6. Spondylolisthesis. 7. Neoplastic disease. 8. Previous Failed Fusion. Levels of attachment for these indications range from T1 to the sacrum. This system is intended for single use.
Device Story
TiMX Low Back System; rod-based spinal fixation construct. Components: pedicle/sacral screws, washers, slotted/cross connectors, Twister connectors. Material: implant-grade titanium alloy (ASTM F-136). Used in OR by surgeons for posterior spinal stabilization. Screws/connectors provide screw-to-connector-to-rod union; V-Groove Hollow Ground (VHG) design ensures secure rod fixation; Twister connector allows intraoperative sagittal alignment in 7-degree increments. System provides temporary stabilization as adjunct to spinal bone grafting; intended for removal after solid fusion in spondylolisthesis cases. Benefits: mechanical stability for spinal fusion; reduced risk of hardware failure via improved torque/bending performance compared to standard titanium VSP/Isola systems.
Clinical Evidence
Bench testing only. Static and fatigue testing performed on titanium constructs. Results: TiMX pedicle screws showed improved torque/bending performance and 30% higher endurance limit vs. standard titanium VSP screws. TiMX Low Back System constructs demonstrated equal or superior mechanical properties (compression, bending, torsion) compared to Isola system constructs. Twister connector constructs showed 250N endurance limit, 755.8N static bending strength, and 38.4N torsional strength.
Technological Characteristics
Implant-grade titanium alloy (ASTM F-136). Rod-based spinal fixation system. Components: pedicle/sacral screws, washers, slotted/cross connectors, Twister connectors. V-Groove Hollow Ground (VHG) locking mechanism. Mechanical fixation via screws, hooks, and wires. Single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring spinal stabilization via pedicle screw fixation for Grade 3-4 spondylolisthesis at L5-S1 (L3-sacrum attachment) or non-pedicle posterior fixation (hook, wire, sacral/iliac screw) for scoliosis, kyphoscoliosis, spinal fractures, degenerative disc disease, neoplastic disease, or failed fusion (T1-sacrum attachment). Contraindicated for pedicle screw use outside of L5-S1 Grade 3-4 spondylolisthesis.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
· Jun. 9. 1998 3:02PM 0. 20. 20. 8 F. 8/13
pl
JUN 3 0 1998
# TiMX Low Back System K98/7/4
510(k) SUMMARY
of the rod-based version of the TiMX Low Back System
COMPANY:
AcroMed Corporation 3303 Carnegie Avenue Cleveland, OH 44115
TRADENAME:
TiMX Low Back System
CLASSIFICATION:
For Pedicle Screw Use: Unclassified- Spondylolisthesis Spinal Fixation Device System
For Non-Pedicle Screw Use: Spinal Interlaminal Fixation Orthosis
DESCRIPTION:
. - -
All components of the TiMX Low Back System are made from implant grade titanium alloy conforming to ASTM F-136 specifications. TiMX pedicle and sacral screws, TiMX washers, TiMX slotted and cross connectors and TiMX Twister connectors are used in conjunction with the Isola System titanium components to comprise a rod based spinal fixation system.
## TiMX Pedicle Screws
TiMX Pedicle Screws are made from implant grade titanium alloy conforming to ASTM F-136 specifications. The TiMX Pedicle Screw is a variation of the existing titanium alloy pedicle screw previously cleared for the Isola Spinal System under K952236 and is the same screw that was found substantially equivalent for use with plates in K981274.
The machine thread portion of the TiMX pedicle screws are available in two thread lengths: Standard (30.0 mm) and No Cut The No Cut +5 machine thread requires no +5 (17.0 mm). cutting. Both the Standard and No Cut +5 thread length TiMX
8
---
{1}------------------------------------------------
## DESCRIPTION:
t l
(Continued)
#### TiMX Pedicle Screws
Pedicle screws are available in four cancellous diameters: 5.50 mm, 6.25 mm, 7.00 mm, and 7.75 mm. The larger size pedicle ' ; screws, 7.00 mm and 7.75 mm, may also be used in the sacrum. " The cancellous portion of the Standard thread length is available in seven lengths that range from 25 mm to 55 mm in five millimeter The cancellous portion of the No Cut +5 thread increments. length is also available in seven lengths that range from 30 mm to 60 mm in five millimeter increments.
#### TiMX Sacral Screws
TiMX Sacral Screws are made from implant grade titanium alloy conforming to ASTM F-136 specifications. The TiMX Sacral Screw is a variation of the existing titanium alloy sacral screw previously cleared for the Isola Spinal System under K952236 and is the same screw that was found substantially equivalent for use with plates in K981274.
The TiMX Sacral Screw is designed with a larger diameter, 8.5 mm, for placement into the sacrum. TiMX Sacral Screws are available in the No Cut +5 (17.0 mm) machine thread length only. The No Cut +5 machine thread requires no cutting. The cancellous portion is available in four lengths that range from 35 mm to 50 mm in five millimeter increments.
## TiMX Washers
. –
TiMX Washers are manufactured from implant grade titanium alloy conforming to ASTM F-136 specifications. The TiMX Washer design is the same as previously cleared for the TiMX Low Back System under K981274. Washers are available in two styles: flat and tapered. Flat washers come in three sizes, 3 mm, 4.50mm and 5.0 mm. All edges of the washers are rounded. All washers have a chamfered inner hole for placement on the machine threaded portion of the TiMX screws. The tapered washer comes in one size with two different shaped inner holes: one round and the other oblong.
9
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{2}------------------------------------------------
Jun. 9.1998 - 3:03PM
P 10/13
# DESCRIPTION: Continued
...
# TiMX Rod Based Slotted and Cross Connectors
TiMX Connectors are manufactured from implant grade titanium alloy conforming to ASTM F-136 specifications. The Modular Cross Connector utilizes J-hooks which can be positioned g anywhere along the construct to provide rod to rod connection. " The TiMX Connector designs are a variation of titanium alloy Isola Slotted Connectors and the Modular Cross Connector and J-Hooks previously cleared for the Isola Spinal System under K952236.
The TiMX slotted connector assembly has two parts: a body and a Slotted connectors provide screw-to-connector-to-rod set screw. union. Slotted and offset designs are available in 14 inch diameters to accommodate the rod size. The slotted portion of each connector provides surgical latitude for screw placement. The machine threaded portion of the connector is locked to the screw with the nut. The rod is locked to the connector with the hexlobe The TiMX slotted connectors are offered in five set screw. designs: straight and extended (each with a 90° angle at the rod locking end), angled (with a 45° angle), offset left, and offset right.
# TiMX Twister Connector
,
TiMX Twister Connector is manufactured from titanium alloy conforming to ASTM F-136 specifications. The TiMX Twister Connector is a variation of the titanium alloy Isola Twister Connector previously cleared for the Isola Spinal System under K965046.
The Twister Connector, like other slotted connectors, provides a stable, strong and durable screw-to-connector-to -rod union. To provide secure fixation to the rod, a patented V-Groove Hollow Ground (VHG) design is used in all connectors. The two piece Twister connector design utilizes the attributes of the one piece connectors. It consists of a slotted transverse member with splines which mate with the splines of the V-Groove body. Together these two pieces create the slotted connector assembly. The spline connection or joint is the medium by which the screw/connector interface can be manipulated and secured. This two piece design allows for intraoperative sagittal alignment in 7º increments.
10
{3}------------------------------------------------
3:03PM Jun. 9. 1998 -
:
E. 11.13 10. 2645
# MATERIAL:
The implant components of the TiMX Low Back System are all manufactured from implant grade titanium alloy conforming to ASTM F-136 specifications.
### INDICATIONS:
| |
When labeled for pedicle screw fixation, the TiMX Low Back System is intended for use in Grade 3 and 4 spondylolisthesis at L5-S1, when affixed to the lumbosacral spine, utilizing autologous bone graft, and intended to be removed after solid fusion is attained. Levels of attachment for this indication range from L3 to the sacrum.
When labeled for non-pedicle screw posterior fixation, the TiMX Low Back System is intended for hook, wire and/or sacral/iliac screw fixation from the thoracic spine to the ilium/sacrum. Properly used, the posterior TiMX Low Back System will provide temporary stabilization as an adjunct to spinal bone grafting process. Specific indications are:
- 1. Idiopathic scoliosis.
- 2. Neuromuscular scoliosis/ kyphoscoliosis with associated paralysis or spasticity.
- 3. Scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele.
- 4. Spinal fractures (acute reduction or late deformity).
- 5. Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies).
- 6. Spondylolisthesis.
- 7. Neoplastic disease.
- 8. Previous Failed Fusion.
Levels of attachment for these indications range from T1 to the sacrum. This system is intended for single use.
{4}------------------------------------------------
. Jun. 9. 1998 3: 3 = 3 % 8. 12/13
12
# PERFORMANCE DATA:
Statistical analysis of the results of static and fatigue component testing of the TiMX Pedicle Screw showed significant improvement of the TiMX design in torque plus bending with the VSP plates, better torque to failure performance of the hexlobe feature and 30% 14 improvement in endurance limit of the TiMX design compared to standard titanium VSP pedicle screws.
A full battery of testing was conducted on titanium constructs which included static tests in compression bending and torsion, and fatigue tests in compression bending. Fatigue test results have indicated that the TiMX Low Back System has a significantly higher endurance limit than the standard titanium Isola system. Static and dynamic compressive bending tests have shown that the TiMX Low Back System constructs have equal or superior mechanical properties when compared to standard titanium Isola constructs. Static torsion testing showed a lower torsional stiffness and similar torsional strengths for the TiMX Low Back System compared to the titanium standard Isola system.
Additional full battery of testing was undertaken to characterize the TiMX Low Back System with TiMX Twister constructs. Fatigue test results have indicated that the TiMX Low Back System with four Twister connectors had an endurance limit of 250N. The static bending strength of the constructs was 755.8N with a bending stiffness of 23.1N/mm. The torsional strength was 38.4N with a stiffness of 5.5Nm/deg.
{5}------------------------------------------------
3:04PM Jun. 9. 1998
P. 13/13 10. 2649
। ਤੇ
# SUBSTANTIAL EQUIVALENCE:
1 1
The TiMX Low Back System is a variation of and substantially equivalent to AcroMed's titanium alloy Isola Spinal Fixation System as previously cleared under K952236 and to the Harrington System, manufactured by Zimmer beginning in the 1960's.
The TiMX Twister Connector is a component of the rod based is version of the TiMX Low Back System and is a variation of the existing titanium alloy Isola Twister Connector previously cleared under K965046.
TiMX Low Back System Pedicle and Sacral screws are The variations of the existing titanium alloy pedicle and sacral screws previously cleared for the Isola Spinal System under K952236 and pre the same screws that were found substantially equivalent for use with plates in K981274.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is black and white, and the text is in a simple, sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 3-0 1998
Ms. Mary Lewis Product Approval Partner AcroMed Corporation 3303 Carnegie Avenue Cleveland, Ohio 44115
Re: K981714 TiMX Plate Based Low Back System (rod-based version) Regulatory Class: II Product Codes: KWP and MNH Dated: May 14, 1998 Received: May 15, 1998
# Dear Ms. Lewis:
We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal, Food, Drug, and Cosmetic Act (Act). This decision is based on your device system being found equivalent only to similar device systems labeled and intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. You may, therefore, market your device system subject to the general controls provisions of the Act and the limitations identified below.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment for indications other than severe spondylolisthesis, as described above, would cause the device system to be adulterated under 501(f)(1) of the Act.
This device system, when intended for pedicular screw fixation/attachment to the spine for indications other than severe spondylolisthesis, as described above, is a class III device under Section 513(f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
{7}------------------------------------------------
# Page 2 - Ms. Mary Lewis
- All labeling for this device, including the package label, must state that there are 1. labeling limitations. The package insert must prominently state that the device system using pedicle screws is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
- You may not label or in anyway promote this device system for pedicular, screw 2. fixation/attachment to the cervical, thoracic or lumbar vertebral column for intended uses other than severe spondylolisthesis, as described above. The package insert must include the following statements:
#### WARNINGS:
- When used as a pedicle screw system, this device system is intended only for . grade 3 or 4 spondylolisthesis at the fifth lumbar - first sacral (L5-S1) vertebral joint.
- The screws of this device system are not intended for insertion into the . pedicles to facilitate spinal fusions above the L5-S1 vertebral joint.
- Benefit of spinal fusions utilizing any pedicle screw fixation system has not . been adequately established in patients with stable spines.
- Potential risks identified with the use of this device system, which may . require additional surgery, include:
- device component fracture, loss of fixation, non-union, fracture of the vertebra, neurological injury, and vascular or visceral injury.
See Warnings, Precautions, and Potential Adverse Events sections of the package insert for a complete list of potential risks.
- 3. Any pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described by item 1, for this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment for intended uses other than severe spondylolisthesis, as described above, must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation.
- 4. Any previous warning statements identified as part of previous 510(k) clearances or required by OC/Labeling and Promotion which stated that a component/system was
{8}------------------------------------------------
Page 3 - Ms. Mary Lewis
not approved for screw fixation into the pedicles of the spine must be replaced by the warnings of items 1 and 2 above.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
FDA advises that the use of your device system with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or those of other manufacturers, may also be required. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K981714 (Under Review)
Device Name: AcroMed TiMX Low Back System ( ( nd - pass
### Indications for Use:
When labeled for pedicle screw fixation, the TiMX Low Back System is intended for use in Grade 3 and 4 spondylolisthesis at L5-S1, when affixed to the lumbosacral spine, utilizing autologous bone graft, and intended to be removed after solid fusion is attained. Levels of attachment for this indication range from L3 to the sacrum.
When labeled for non-pedicle screw posterior fixation, the TiMX Low Back System is intended for hook, wire and/or sacral/iliac screw fixation from the thoracic spine to the ilium/sacrum. Properly used, the posterior TiMX Low Back System will provide temporary stabilization as an adjunct to spinal bone grafting process. Specific indications are:
- 1. Idiopathic scoliosis.
- 2. Neuromuscular scoliosis/ kyphoscoliosis with associated paralysis or spasticity.
- 3. Scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele.
- 4. Spinal fractures (acute reduction or late deformity).
- 5. Degenerative Disc Disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies).
- 6. Spondylolisthesis.
- 7. Neoplastic disease.
- 8. Previous Failed Fusion.
Levels of attachment for these indications range from T1 to the sacrum. This system is intended for single use.
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | | | |
|----------------------------------------|---------|----|--------------------------|
| Division of General Restorative Device | | | |
| 510(k) Number | K981714 | | |
| Prescription Use | X | OR | Over-The-Counter Use |
| (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.