K981680 · Dayton Water Systems · FIP · Dec 31, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K981680
Device Name
ULTRAPURE WATER TREATMENT SYSTEM FOR HEMODIALYSIS
Applicant
Dayton Water Systems
Product Code
FIP · Gastroenterology, Urology
Decision Date
Dec 31, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device. Federal law restricts this device to sale by or on the order of a physician.
Device Story
Water purification system for hemodialysis; removes organic/inorganic substances and microbial contaminants from feed water. System components include reverse osmosis (RO) unit, deionizer (DI) exchange tanks, activated carbon filters, water softeners, cartridge filters, and temperature blending valves. Optional components: multimedia filters, calcite filters, UV disinfection, iodinators, and booster pumps. Central control panel monitors water levels, DI quality, and RO production; triggers remote audible/visual alarms at nurses' station for malfunctions (low pressure, low quality, low tank level). Repressurization pumps deliver treated water to dialysis loop. Used in clinical settings; operated by facility staff/technicians under physician oversight. Benefits patient by ensuring dialysate water meets AAMI quality standards, reducing risk of contamination during hemodialysis.
Clinical Evidence
Bench testing only. Performance data demonstrated that water samples collected from the return to the storage tank 48 hours after system start-up met AAMI Hemodialysis Water Quality Standards.
Technological Characteristics
Modular water purification system; components include RO membranes, DI exchange tanks, activated carbon media, water softeners, and submicron post-filters. Sensing/monitoring: pressure gauges, flow meters, temperature-compensated water quality monitors, and level sensors. Energy: electrical power for pumps, control panel, and UV disinfection. Connectivity: remote audible/visual alarm interface. Materials: standard industrial water treatment components. System design is configurable based on feed water quality and AAMI standards.
Indications for Use
Indicated for use in hemodialysis treatment to purify water used to dilute dialysate concentrate.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Better Water Inc. Water Treatment System for Hemodialysis (K920186)
Submission Summary (Full Text)
{0}------------------------------------------------
DEC 3 1 1998
Image /page/0/Picture/2 description: The image shows handwritten text on a white background. The text includes "P.3/9" at the top, followed by "K981680" in the middle. Below that, the text "p. 1/7" is written. The handwriting appears to be somewhat rough, with the numbers and letters not perfectly formed.
# Section 5 Summary of Safety & Effectiveness
Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis
> submitted by R.D. Baker Enterprises dba Dayton Water Systems 1288 McCook Ave. Davton, OH 45404 937-461-5225 fax 937-461-8838
Contact: Cindy Helton February 2, 1998 updated: December 29, 1998
Common or Classification Name: Water Purification System for Hemodialysis
Proprietary Name: Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis
## Claiming substantial equivalence to:
Better Water Inc. Water Treatment System for Hemodialysis (K920186)
Intended Use: Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device. Federal law restricts this device to sale by or on the order of a physician.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows handwritten text on a white background. The text includes "P.4/9", "K981680", and "p.2/7". The numbers appear to be written with a thick marker or pen, and the handwriting is somewhat uneven.
#### Device Description
The Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis is a complete water purification system and consists of a reverse osmosis unit (RO) and/or deionizer exchange tanks (DI) to demineralize the water. Our recommendation to use RO and DI in conjunction or individually is based upon feed water quality and the demand of water to be used. Pretreatment includes activated carbon filtration media to remove organics, namely chlorine and chloramines. In order to optimize the performance of the RO system, the following items are utilized: 1) a water softener to remove scale-forming minerals: 2) cartridge filters to remove suspended solids: 3) a temperature blending valve to produce the desired water temperature. In order to compensate for fluctuations in demand, an air-tight storage tank with vent filter, level control, and internal spray array is utilized to store RO product water until needed. Repressurization pumps are employed to provide adequate flow and pressure to points of use.
Whenever a storage tank or deionizer exchange tanks are used in a water purification system for hemodialysis, submicron membrane post filtration is used to remove any suspended solids and bacteria. Also, a temperaturecompensated quality-indicating device is installed on all systems utilizing deionizer exchange tanks. For additional safety, a remote audible/visual alarm is installed at the nurses' station to indicate any problems that may occur in the water treatment room.
When feed water quality demands or when requested by the facility physician, optional devices can include the following components: 1) cartridge filtration for the removal of silt or feed water sediment: 2) automatic backwashing multimedia filter for the removal of suspended solids: 3) automatic backwashing carbon dioxide reduction filter (calcite filter) for the reduction/removal of CO2; 4) ultraviolet disinfection to reduce bacteria either in the influent water supply or on the product supply. 5) iodinator for the introduction of an iodine residual to the storage tank to inhibit bacteria growth (NOTE: DI tanks must be used to remove the residual iodine from the product water before use in hemodialysis treatment); and 6) a booster purnp to increase influent water pressure sufficient to provide adequate flow and pressure to the water treatment system.
{2}------------------------------------------------
Water treatment components are recommended based on the demand of water required, a feed water analysis, and the ability to produce water which meets AAMI standards. Dayton Water Systems will always recommend a water purification system which will exceed Hemodialysis Water Quality Standards as set forth by AAMI (Fig. 5-1).
# Summary of Technological Characteristics
Dayton Water Systems designs, recommends and assembles the components of the system. Dayton Water Systems is not the manufacturer of several of the components and is not the manufacturer of component parts which are assembled by Dayton Water Systems.
Although, Dayton Water Systems recommends the water treatment system and components, the dialysis facility's physician has ultimate authority of the equipment installed for water purification. The recommended system (Fig. 5-2) consists of the following components. A complete system and the intended purposes of each components are described below. Our design is of equivalent to that of the predicate device.
Hot and cold filtered water supplies are mixed together using a temperature blending valve to the desired 77°F. Sufficient water pressure must be supplied in order to operate the water treatment equipment effectively. If sufficient pressure is not available, a booster pump is installed to supply the required flow and pressure of feed water. If silt, fine sediment, or low pH conditions are present in the feed water supply, optional filtration is used. A calcite filter can be used to neutralize low pH due to high levels of dissolved carbon dioxide. A .35 micron filter cartridge can be used to remove the fine sediment and silt. Water is then directed to the water softener where calcium and magnesium ions are exchanged for non-scaling sodium ions. Next, the water flows through dual exchange carbon filters to remove the chlorine, chloramines and other organics from the water supply. The carbon filters are sized to provide adequate empty bed contact time of 3-5 minutes for chlorine and 6-10 minutes for chloramines.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows handwritten text on a white background. The text includes the numbers "981680" and the letters "K" and "p". There are also page numbers "P.6/9" and "p. 4/7" written in the image.
The "pre-treated" water then enters into the reverse osmosis membrane where dissolved minerals are rejected from the product stream. The product water, permeate, enters into a storage tank. The storage tank is complete with a submicron vent filter to prevent airborne bacteria from entering into the storage tank. An iodinator can be installed as an option on the return water line to the storage tank to add a residual of iodine to inibit bacteria growth. Stored product water is then repressurized by the repressurization pump for distribution through the dialysis loop. After the repressurization pump and before the dialysis loop, mixed-bed deionizer exchange tanks are installed to remove dissolved minerals and dissolved gases from the water to be used to dilute dialysate. In rare circumstances when requested by the physician in charge of the facility, an ultraviolet disinfection unit is used to kill bacteria. Submicron membrane post filters are used to remove bacteria and pyrogens from entering into the dialysis loop. Submicron membrane post filters must be used in conjunction with storage tanks, UV and DI. Fibercore ultrafilters are used as post filtration. However, the WET filter may be used as a primary filter to the Fibercore ultrafilter to remove larger particles down to .2 micron.
# Identification of Components
5 Micron Prefilters (Hot and Cold) - are used to filter particulate and suspended solids from the incoming water supply. Any suspended solids in the water supply could cause a malfunction of any of the water treatment components.
Temperature Blending Valve - is used to blend hot and cold water to achieve a temperature of 77%. Tempered water increases the efficiency of the reverse osmosis unit.
Booster Pump - (optional) is used to increase the influent water pressure. The reverse osmosis unit, as well as the other pretreatment equipment, requires sufficient feed water pressure to operate.
Calcite Filter - (optional) is used to neutralize the pH of the influent water source due to dissolved carbon dioxide in the water supply. Excessive CO2 in the influent water supply will adversely effect the pH of the product water of the reverse osmosis unit. CO2 will also accelerate the exhaustion of the deionizer exchange tanks.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows handwritten text and numbers. The text at the top reads 'P.7/9'. Below that, there is a sequence of characters and numbers: 'K9816880'. At the bottom, there is another line that reads 'p 5/7'.
Multi-media Filter - (optional) is used to remove suspended solids from the feed water supply. Excessive suspended solids can cause malfunction of pretreatment components, i.e. media plugging, moving parts binding, etc.
.35 Micron Filter - (optional) is used to remove very small particulate and suspended solids, such as silt, that are not removed by the 5 micron prefilters. Suspended solids will cause the reverse osmosis membrane to "plug up", thus reducing production of treated water.
Water Softener - is used to remove calcium and magnesium bicarbonate from the influent water supply. Calcium and magnesium bicarbonate will cause the reverse osmosis membrane to scale, thus reducing production of treated water.
Exchange Carbon Filters - are used to remove organics, namely chlorine and chloramines, from the influent water supply. Chlorine will attack and destroy the reverse osmosis membrane, thus reducing the quality of the treated water. Also. AAMI recommends that chlorine and chloramines be removed from the treated water supply.
Reverse Osmosis Unit - rejects inorganics, dissolved minerals, suspended solids and microbiological contaminants from the product water stream. In order to meet AAMI standards for dissolved inorganics, a reverse osmosis unit is utilized.
Storage Tank - stores reverse osmosis product water and acts as a integral part of the dialysis water loop. The reverse osmosis unit produces water at a steady rate which is sometimes slower than demand. The storage tank will allow for fluctuations in demand. The storage tank is equipped with level sensors to turn on the reverse osmosis unit when water is required, and to turn off when the tank is full. A vent filter removes airborne bacteria while allowing the tank to breathe. A spray array is utilized so that the interior of the tank is continually rinsed.
Repressurization/Recirculation Pump - serves two purposes. First, it is used to repressurize the water in the storage tank and delivers the water to the dialysis loop and the machines. Second, it is used to continuously recirculate the water in the dialysis loop piping at the recommended velocity of 5 feet per second, to inhibit bacteria growth.
{5}------------------------------------------------
Central Control Panel - acts as the integral central processing unit which monitors the level of the water the storage tank, includes the quality monitor for the DI tanks, and controls the production of RO permeate. The central control panel also controls the operation of the repressurization/recirculation purnp. The central control panel interacts with the remote audible/visual alarm to alarm at the nurses' station when one of the following alarm conditions arise: 1) low water level in the holding tank; 2) low quality water in the product loop; 3) RO malfunction; 4) low feed water pressure; and 5) other alarm/alert levels as desired by the facility.
Remote Audible/Visual Alarm - is mounted at the nurses station and activates when an alamn/alert condition occurs in the water treatment room.
Deionizer Exchange Tanks - are used to remove inorganic water contaminants and dissolved gases from the water supply. Using water quality monitoring devices, this method assures treated water will meet AAMI recommended standards for inorganic removal. When utilizing dejonizer excliange tanks, submicron membrane filtration must be used.
Ultraviolet Disinfection Unit - is used to kill bacteria at up to 99.9% efficiency utilizing 254 nm ultraviolet waves. Ultrafiltration or submicron membrane filtration is recommended to inhibit endotoxins from entering into the process stream.
Ultra Filters & Submicron Post Filters - are used to remove any suspended solids, "free floating" bacteria, and inhibit pyrogens from entering the water used in dialysis. Post filters must be used in conjunction with storage tanks, DI and UV.
Iodinator - is used to add a residual of iodine to the holding tank. The iodinator is installed as a split-stream so that water returning from the loop to the holding tank has a residual of iodine. The iodine residual inhibits bacteria growth in the storage tank where water flow velocity is not 5 feet per second. The iodine is removed from the treated water by the deionizer exchange tanks before distribution to the loop. An iodinator can be utilized only when deionizer exchange tanks are used.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows handwritten text on a white background. The text includes "P.9/9" at the top, followed by "K981680" in the middle, and "p.717" at the bottom. The handwriting appears to be clear and legible.
Monitoring Devices - are used to assure the proper performance of the water trèatment equipment. Pressure gauges, flow meters, temperaturecompensated water quality monitors, test ports, temperature gauge, water meter, etc. are installed at specific points throughout the system to measure each component's performance.
## Performance Data
Results of clinical testing were performed on a water purification system and are attached as Fig. 5-3. This test sample was pulled from the test port which is on the return to the storage tank approximately 48 hours after start-up. This testing indicates that after 48 hours water in the dialysis loop still meets the AAMI standards. Thus, as with the predicate device, this water purification system successfully meets performance standards.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center is an abstract symbol that resembles an eagle or bird with three wing-like shapes. The symbol is black, and the text around the perimeter is also in black.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 31 1998
Ms. Cindy Helton Quality Assurance Manager Dayton Water Systems 430 Leo Street Dayton, OH 45404
Re: K981680
Ultrapure Water Treatment System for Hemodialysis Dated: November 5, 1998 Received: November 6, 1998 Regulatory Class: II 21CFR 876.5665/Procode: 78 FIP
Dear Ms. Helton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
sitt
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Section 1 Device Name
The classification device name of this system is Water Purification System for Hemodialysis. The proprietary device name is Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis.
### Statement of Intended Use
Dayton Water Systems' UltraPure Water Treatment System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.
Prescription
vision of Reproductive, Abdominal, ENT d Radiological Device 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.