K981626 · Siemens Medical Solutions USA, Inc. · IYO · May 27, 1998 · Radiology
Device Facts
Record ID
K981626
Device Name
SONOLINE ELEGRA DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYO · Radiology
Decision Date
May 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The SONOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of 3-D Imaging will not add new indications for use to the Elegra. It will allow the user to display regions of interest in 3-dimensional format, either by showing views in three orthagonal planes or allowing surface rendering of objects. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
The SONOLINE Elegra is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays it in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler, and Continuous (CW) Doppler. This submission adds 3D imaging capabilities, allowing users to visualize regions of interest through three orthogonal planes or surface rendering. The system is operated by clinicians in clinical settings. It provides anatomical measurements and analysis packages to assist in clinical diagnosis. The 3D reconstruction technique is functionally equivalent to the previously cleared SieScape panoramic imaging feature. The device benefits patients by providing enhanced visualization of anatomical structures for diagnostic purposes.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Mobile, software-controlled diagnostic ultrasound system. Supports B-Mode, M-Mode, Color, PW Doppler, and CW Doppler. Includes 3D reconstruction and surface rendering. Complies with UL 2601, CSA 22.2 No. 601-1, AIUM/NEMA 1992 (thermal/mechanical indices), and 93/42/EEC. Connectivity includes on-screen display for indices.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in general radiology, abdominal, intraoperative, small parts, transcranial, OB/GYN, pelvic, neonatal/adult cephalic, urology, vascular, and musculoskeletal applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
SieScape panoramic display for SONOLINE Elegra Advanced (K961833)
Submission Summary (Full Text)
{0}------------------------------------------------
### MAY 2 7 1998 510(K) SUMMARY
## Addition of 3 D Imaging to SONOLINE Elegra Diagnostic Ultrasound system
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary.
#### Submitted By: 1.
Siemens Medical Systems, Inc., Ultrasound Group 22010 S.E. 51st Street Issaquah, WA 98027-7002
## Contact Person: Steve Hesler Manager of Regulatory Affairs (425) 557-1629
## Date Prepared: March 31, 1998
- 2. Proprietary Name: SONOLINE Elegra Advanced SONOLINE Elegra
Common/ Usual Name: Diagnostic Ultrasound System with Accessories
## Classification Name: Ultrasonic Pulsed Doppler Imaging System (Product Code 90 IYN, 21 CFR 892.1550)
#### 3. Predicate Device:
K945072, 11/21/95, cleared as the Q4000, marketed as the SONOLINE Elegra Advanced
K950517, 4/5/96, cleared as Q64XX, marketed as the SONOLINE Elegra K961833, 10/29/96, SieScape panoramic display for SONOLINE Elegra Advanced
#### 4. Device Description:
The SONOLINE Elegra is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire ultrasound data and display it in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, or in a combination of modes, on a CRT display.
The SONOLINE ® Elegra, has been designed to meet the following product safety standards:
- UL 2601, Safety Requirements for Medical Equipment .
- CSA 22.2 No. 601-1, Safety Requirements for Medical Equipment .
- Standard for Real Time Display of Themal and Mechanical Indices on . Diagnostic Ultrasound Equipment, AlUM/NEMA, 1992.
- 93/42/EEC Medical Devices Directive
EN60601 = (IEC 601-1-1 + IEC 601-1-2), Safety and EMC Requirements for Medical Equipment
{1}------------------------------------------------
#### 5. Intended Uses:
The SONOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of 3-D Imaging will not add new indications for use to the Elegra. It will allow the user to display regions of interest in 3-dimensional format, either by showing views in three orthagonal planes or allowing surface rendering of objects.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
#### Technological Comparison to Predicate Device: 6.
SONOLINE Elegra is a previously cleared device. The purpose of this submission is to receive clearance for the addition of 3 dimensional imaging to the already-cleared svstem. This real-time reconstruction technique is equivalent to the SieScape panaoramic imaging feature already cleared by 510(k) (see K961833, 10/29/96).
## End of 510(k) Summary
t with 5
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three birds in flight, arranged in a row.
# MAY 2 7 1998
Siemens Medical Systems. Inc. Ultrasound Group c/o Joseph P. Murnane Underwriters Laboratories, Inc. 1285 Walt Whitman Road Melville, New York 11747-3081 Rc:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
K981626 Sonoline® Elegra Diagnostic Ultrasound System Dated: May 1, 1998 Received: May 7, 1998 Regulatory class: II Procodes: 90 IYN, 21 CFR 892.1550 90 IYO, 21 CFR 892.1560 90 ITX, 21 CFR 892. 1570
Dear Mr. Murnane:
We have reviewed your Section 510%) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Festeral Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions agginst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devise in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electrotic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmam.html".
Sincerely yours,
William Yu
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
510 (k) Number (if
known) :
SONOLINE Elegra with SieScape and 3 D Imaging Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | P | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | P | P | P | P | P | P | | BMDC (P) | |
| Pediatric | | P | P | P | P | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | P | P | P | P | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Cardiac | | P | P | P | P | P | P | | BMDC (P) | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | P | P | P | | BMDC (P) | |
| Transvaginal | | P | P | P | P | P | P | | BMDC (P) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David G. Seymore
(Division Sign-Off)
(Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{4}------------------------------------------------
510 (k) Number (if
known) :
2.0 CW probe Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
## Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|---------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | P | | | | | |
| Intraoperative<br>(Specify) | | | | | P | | | | | |
| Pediatric | | | | | P | | | | | |
| Small Organ<br>(Specify) | | | | | P | | | | | |
| Neonatal Cephalic | | | | | P | | | | | |
| Adult Cephalic | | | | | P | | | | | |
| Cardiac | | | | | P | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | P | | | | | |
| Transvaginal | | | | | P | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | P | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
under Appendix E
Additional Comments:___
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Severson
(Division Sign-Off)
(Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K481626
{5}------------------------------------------------
510 (k) Number (if
known) :
## 2.5PL20 Phased Array Transducer Device Name :
Intended Use:
2.5F E20 Friased Aray andy or fluid flow analysis of the human body as follows:
## Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
|----------------------------------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | E | E | E | E | | E | | BMDC (E) | | |
| Abdominal | | E | E | E | E | | E | | BMDC (E) | | |
| Intraoperative<br>(Specify) | | | | | | | | | | | |
| Pediatric | | E | E | E | E | E | E | | BMDC (E) | | |
| Small Organ<br>(Specify) | | | | | | | | | | | |
| Neonatal Cephalic | | E | E | E | E | E | E | | BMDC (E) | | |
| Adult Cephalic | | E | E | E | E | E | E | | BMDC (E) | | |
| Cardiac | | E | E | E | E | E | E | | BMDC (E) | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | E | E | E | E | E | | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added<br>under Appendix E<br>Additional Comments: | | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Segerson
Division Sign Off
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number _ K981626
{6}------------------------------------------------
## 510 (k) Number (if
known) :
### 3.5PL28 Phased Array Transducer Device Name :
Intended Use:
5.5F E20 Thused Airay Truid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | P | P | P | P | | BMDC (P) | |
| Fetal | | P | P | P | P | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | P | | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added | | | | | | | | | | |
| under Appendix E | | | | | | | | | | |
| Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Jarid a. Jesson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
981626 510(k) Number
{7}------------------------------------------------
510 (k) Number (if
known) :
## 3.5C40 Curved Array Transducer Device Name :
Intended Use:
3.5040 Cul ved Array Transade of the data visis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
|----------------------------------------------------------------------------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | |
| Intraoperative (Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | | E | E | | BMDC (E) | |
| Laparoscopic | | E | E | E | | E | E | | BMDC (E) | |
| Musculo-skeletal Conventional | | E | E | E | | E | E | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lyons
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K487422
{8}------------------------------------------------
510 (k) Number (if known) : Device Name :
# 5.0HDPL40 Linear Array Transducer
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. Bergman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{9}------------------------------------------------
510 (k) Number (if known) :
### 5.0C50 Curved Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication: P = previously cleared by FDA: E = added | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Hanson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{10}------------------------------------------------
510 (k) Number (if
known) :
### 6.5EC10 Curved Array Transducer Device Name :
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | E | E | E | | E | E | | BMDC (E) | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | E | E | E | | E | E | | BMDC (E) | |
| Transvaginal | | E | E | E | | E | E | | BMDC (E) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added<br>under Appendix F | | | | | | | | | | |
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seigman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{11}------------------------------------------------
510 (k) Number (if
known) :
### 6.5EV13 Curved Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | P | P | P | | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC (P) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | BMDC (P) | |
| Transvaginal | | P | P | P | | P | P | | BMDC (P) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymon
(Division Sign-Off) (Porision of Reproductive, Abdominal, ENT,
and Radiological Devices and Radiological Devices
510(k) Number K981626
{12}------------------------------------------------
510 (k) Number (if known) : Device Name :
# 7.5C30 Curved Array Transducer
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Application | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | |
| Intraoperative | | | | | | | | | | |
| (Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | |
| Small Organ | | E | E | E | | E | E | | BMDC (E) | |
| (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | | E | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | | E | E | | BMDC (E) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added | | | | | | | | | | |
under Appendix E Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Thail G. Seyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number_K 78/626
(
{13}------------------------------------------------
510 (k) Number (if
known) :
7.5L40 Linear Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|---------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | P | P | P | | P | P | | BMDC (F) | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | P | P | P | | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC (P) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication. D = previously cleared by FDA. E = added. | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Hind G. Hyam
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number_K481626
{14}------------------------------------------------
510 (k) Number (if
known) :
### 7.5PL13 Phased Array Transducer Device Name :
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | E | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | E | E | E | | BMDC (E) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | E | E | E | | BMDC (E) | |
| Small Organ<br>(Specify) | | E | E | E | E | E | E | | BMDC (E) | |
| Neonatal Cephalic | | E | E | E | E | E | E | | BMDC (E) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | E | E | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | E | E | E | | BMDC (E) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Thind le. Beynam
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.