← Product Code [DWF](/productcode/DWF) · K981601

# CALIFORNIA MEDICAL LABORATORIES INC. MALLEABLE VENT CATHETER (K981601)

_Sorin Group Italia S.R.L. · DWF · Jul 28, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K981601

## Device Facts

- **Applicant:** Sorin Group Italia S.R.L.
- **Product Code:** [DWF](/productcode/DWF.md)
- **Decision Date:** Jul 28, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Malleable Vent Catheter is intended for use during cardiopulmonary bypass Surgery to vent/drain the left ventricle.

## Device Story

Malleable Vent Catheter used during cardiopulmonary bypass surgery to vent/drain left ventricle. Device consists of soft PVC tubing with radius tip and vent holes along distal tip; dual-lumen design includes one large drainage lumen and one smaller lumen encapsulating stainless steel wire. Encapsulated wire allows for manual preshaping and structural support during placement. Proximal end features rigid PVC connector and depth markings. Operated by surgical staff in OR setting. Device facilitates venting/drainage of left ventricle, assisting in surgical management of patient during bypass.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use.

## Technological Characteristics

Materials: soft polyvinyl chloride (PVC) tubing, stainless steel wire. Design: dual-lumen catheter, radius tip, vent holes 3.5 cm along distal tip, rigid PVC connector, depth markings. Function: malleable wire allows preshaping and support during insertion. Non-powered, mechanical device.

## Regulatory Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

## Predicate Devices

- Research Medical, Inc. Left Ventricular Vent Catheter
- DLP, Inc. Left Ventricular Malleable Vent Catheter

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K981601

# 111 28 1998

## Summary of Safety and Effectiveness

#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE

The California Medical Laboratories, Inc. devices are substantially equivalent to the DLP and Research Medical predicate device. The California Medical Laboratories, Inc. devices have substantially equivalent intended uses as the predicate device. The California Medical Laboratories, Inc. devices have technologic characteristics which are substantially equivalent to the DLP and Research Medical predicate device.

## COMPANY AND CONTACT PERSON

California Medical Laboratories Inc. 2681 Kelvin Avenue Irvine. California 92614

Mehmet Bicakci President

#### DEVICE NAME

California Medical Laboratories Inc. Malleable Vent Catheter

## NAME OF PREDICATE OR LEGALLY MARKETED DEVICE

The claim of substantial equivalence is based upon the following devices:

- Research Medical, Inc. Left Ventricular Vent Catheter .
- . DLP, Inc. Left Ventricular Malleable Vent Catheter

## DESCRIPTION OF DEVICE

The Malleable Vent Catheter is designed to vent/drain the left ventricular during coronary artery bypass surgery.

It consists of soft polyviny] chloride (PVC) tubing with a radius tip and vent holes 3.5 cm along the distal tip. The soft PVC tubing consists of two lumens; one large lumen for venting/drainage and one smaller lumen which encapsulates the stainless steel wire. The encapuslated stainless steel wire enables preshaping of the catheter prior to insertion and support during catheter placement. The proximal end of the catheter contains a rigid PVC connector and depth markings along the farge lumen.

#### STATEMENT OF INTENDED USE

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

The Malleable Vent Catheter is intended for use during cardiopulmonary bypass Surgery to vent/drain the left ventricle.

# STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES

The predicate devices have the same intended use as stated above.

# STATEMENT OF COMPARISON OF TECHNOLOGIC CHARACTERISTICS BETWEEN DEVICE AND PREDICATE DEVICE

The California Medical Laboratories, Inc devices have technologic characteristics which are substantially equivalent to the predicate device.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird-like figure, depicted with flowing lines to suggest movement or flight. The overall design is simple and symbolic, representing the department's role in health and human services.

JUL 2 8 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Mehmet Bicakci President California Medical Laboratories, Inc. 2681 Kelvin Avenue Irvine, CA 92614

Re : K981601 Malleable Vent Catheter Regulatory Class: II (Two) Product Code: DWF Dated: Mav 1, 1998 Received: May 4, 1998

Dear Mr. Bicakci:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

## Page 2 - Mr. Mehmet Bicakci

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Also, please note the regulation Compliance at (301) 594-4639. entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours
Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): K

Device Name: California Medical Laboratories Inc. Malleable Vent Catheter.

Indications For Use: The Malleable Vent Catheter is indicated for use during cardiopulmonary bypass surgery to vent/drain the left ventricle.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kira Kuiperle

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number_k 98 160 /

X Prescription Use Per 21 CFR 801.109

OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K981601](https://fda.innolitics.com/device/K981601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
