K981536 · Hellen Professional Services · MVE · Jun 29, 1998 · Immunology
Device Facts
Record ID
K981536
Device Name
QUANTECH MYOGLOBIN ASSAY
Applicant
Hellen Professional Services
Product Code
MVE · Immunology
Decision Date
Jun 29, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5680
Device Class
Class 2
Indications for Use
The Quantech Myoglobin Assay is intended to be used as an aid in diagnosing myocardial infarction in patients exhibiting chest pain. It is intended to be used in conjunction with EKG, and physician examination, as well as possibly other biochemical blood tests to rule in or out non-cardiac damaging origin of the chest pain.
Device Story
In vitro diagnostic assay for quantitative measurement of human myoglobin in serum; aids diagnosis of myocardial infarction in patients with chest pain. Principle: two-site sandwich immunoassay using monoclonal anti-myoglobin antibody immobilized on gold surface; utilizes surface plasmon resonance (SPR) to measure surface mass increase proportional to myoglobin concentration. Instrument uses microprocessor for control, data acquisition, and reduction. Used in clinical laboratory settings by trained personnel. Results interpreted by physicians alongside EKG and clinical exam to rule in/out cardiac damage.
Clinical Evidence
Bench testing included dilution linearity (107% recovery), accuracy (108% recovery), analytical sensitivity (6 ng/mL), and precision (intra-assay CV 9.5-11.9%; inter-assay CV 6.2-10.3%). No hook effect observed up to 10,000 ng/mL. Clinical evaluation included normal range determination (14-70 ng/mL) and patient sample correlation against a fluorogenic ELISA (r=0.970, slope=1.16).
Indicated for patients exhibiting chest pain to aid in the diagnosis of myocardial infarction. Used in conjunction with EKG, physician examination, and other biochemical blood tests.
Regulatory Classification
Identification
A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.
{0}------------------------------------------------
K981536
**510(k) SUMMARY**
Revision 02
### Company Information
Quantech Ltd. 1419 Energy Park Drive St. Paul, MN, (612) 647-6370 Thomas Witty, Ph.D. - Vice President, Research and Development
### Contact Information
Robin J. Hellen, M.S. Hellen Professional Services (818) 709-5646
### Product Name
Classification Name: Trade Name: Common Name:
Myoglobin Immunological Test System, Class II Quantech Myoglobin Assay Mvoalobin Test Kit
### · Substantial Equivalence
The Quantech Myoglobin Assay is substantially equivalent to the OPUS® Myoglobir assay marketed by Behring Diagnostics Inc. since 1991.
### Intended Use
The Quantech Myoglobin Assay is intended to be used as an aid in diagnosing myocardial infarction in patients exhibiting chest pain. It is intended to be used in conjunction with EKG, and physician examination, as well as possibly other biochemical blood tests to rule in or out origin of the chest pain.
### Device Description
The Quantech Myoglobin Assay is based on the principle of two site, or sandwich immunoassay in combination with SPR surface mass measurement. Each test module contains a solid phase anti-myoglobin monoclonal antibody immobilized onto a gold surface. An anti-myoglobin enzyme conjugate solution and wash solution are used to enhance the specific detection of myoglobin.
The Quantech assay utilizes myoglobin-specific antibody to capture the myoglobin in the This is followed by a quantitation of the surface mass increase using surface sample. plasmon resonance (SPR), to measure the myoglobin in serum.
{1}------------------------------------------------
## Comparison of Technological Characteristics
The Quantech Myoglobin Assay is similar to the Opus® Myoglobin assay (K915850) as follows. Both assays are in vitro immunological assays for the quantitative measurement of human myoglobin. Additionally, both assays use antibody to myoglobin coated on a solid support, and both instruments utilize a microprocessor for instrument control, data acquisition, and data reduction.
### Summary of Non-Clinical Performance Data
Dilution Linearity/Parallelism - The parallelism study was conducted to evaluate the linearity of the Quantech MYOGLOBIN Assay. Serum samples were separately spiked with myoglobin and serially diluted with corresponding unspiked serum. The average percent of expected was 107% with individuals varying from 101-115%.
Recovery - Accuracy of the Quantech MYOGLOBIN Assay was calculated from test results as the percentage of added analyte, corrected for endogenous analyte, recovered by the assay. After correcting for endogenous myoglobin content, the average recovery was 108%.
Analytical Sensitivity - Multiple duplicates of zero samples (blank) were assayed to determine the minimum quantity of myoglobin detectable by the Quantech Assay. The average SPR signal shift plus two standard deviations (2 S.D.) was calculated and translated into a dose. The calculated analytical sensitivity of the Quantech myoglobin assay is 6 ng/mL.
Precision - INTRAASSAY variation was calculated by replicate evaluation of three levels of sera in one day. The mean mvoolobin concentrations (with % C.V.) were 61 (11.9%), 131 (9.7%), and 378 (9.5%) ng/mL for the low, medium and high pools, respectively.
The INTERASSAY precision was determined by evaluating three pools (different from INTRAASSAY pools) in multiplicate on different days. The mean myoglobin concentrations (with % C.V.) were 51 (7.1%), 110 (6.2%), and 219 (10.3%) ng/mL for the low, medium and high pools, respectively.
Interfering Substances - Physiological interference was evaluated by spiking a serum pool of myoglobin with bilirubin, triglycerides and hemoglobin at levels ten times the highest expected physiological concentration. The percent recovery of myoglobin was determined to be acceptable in all three solutions and no interference was noted by the endogenous in the Quantech Myoglobin assay.
Hook Effect - Samples well beyond the standard curve range were assayed. No high dose hook effect was observed. Therefore, the Quantech Myoglobin Assay does not give erroneously low results for grossly elevated samples up to at least 10,000 ng/ml.
{2}------------------------------------------------
# Summary of Clinical Performance Data
Normal Range - Testing of apparently healthy individuals demonstrated a normal range consistent with published literature (male 19-92 ng/mL and female 12-76 ng/mL). The overall normal range observed with the Quantech MYOGLOBIN assay was 14-70 ng/mL with 16-70 ng/mL for males and 17-55 ng/mL observed for females.
Patient Sample Correlation - Results from human samples with values distributed throughout the quantitative range of the Quantech MYOGLOBIN Assay, were compared with those obtained with a commercially available method (fluorogenic ELISA). The correlation coefficient was 0.970 (slope = 1.16, y-intercept = 7.94 ng/mL).
# Conclusions Drawn From Performance Tests
The Quantech Myoglobin Assay provides results which are internally accurate, unaffected by ordinary variation of sample matrix and equivalent to the results obtained using the approved device in a valid laboratory setting.
Additionally, both clinically-based studies (normal range, patient correlation) demonstrated essential equivalence between the two devices as measured by their correlation and the degree to which assay results are linearly related to one another over a broad range of values. Likewise, the normal range evaluation provided empirical evidence that the log of the assay value is statistically similar for both devices, and in agreement with published data.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, with its wings spread.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 2 9 1000
Quantech, Ltd. C/O Robin J. Hellen, M.S. Hellen Professional Services 9418 Lasaine Avenue Northridge, California 91325
K981536 Re : Quantech Myoglobin Assay Regulatory Class: II Product Code: DRD April 28, 1998 Dated: Received: April 29, 1998
Dear Ms. Hellen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "dsmo@fdadr.cdrh.fda.gov".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DUANTECH MYOGLOBIN ASSAY emarket Notification
PART I - 510(k) Information
### Statement for Indications for Use 111.
510(k) Number (if known): K981536
Device Name: Quantech Myoglobin Assay
Indications for Use:
The Quantech Myoglobin Assay is intended to be used as an aid in
diagnosing myocardial infarction in patients exhibiting chest pain. It is
intended to be used in conjunction with EKG, and physician examination,
as well as possibly other biochemical blood tests to rule in or out non-
cardiac damaging origin of the chest pain.
Concurrence of the CDRH, Office of Device Evaluat=ion (ODE)
Prescription Use: ✓
OR
Over the Counter Use: **__**
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K981536
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.