SPACELABS MEDICAL 90519 ANESTHESIA DELIVERY SYSTEM
K981530 · Spacelabs Medical, Inc. · BSZ · Nov 20, 1998 · Anesthesiology
Device Facts
Record ID
K981530
Device Name
SPACELABS MEDICAL 90519 ANESTHESIA DELIVERY SYSTEM
Applicant
Spacelabs Medical, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Nov 20, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Spacelabs Medical 90519 Anesthesia Delivery System provides intermittent or continuous gas inhalation for adults and children (neonatal and pediatric). It allows the administration of operator selected gas mixtures of oxygen, nitrous oxide and air with any of the anesthetic agents: Halothane, Isoflurane, Enflurane, or Sevoflurane. It provides safe and accurate gas flows to maintain patient respiration during anesthesia, and incorporates a ventilator, an oxygen monitor, and a respiratory monitor. The ventilator provides the necessary power, as air or oxygen, to generate volumes and pressures in the ventilating system to ventilate a patient connected to the anesthesia machine. It is recommended for use only by trained physicians, in the operating room or similar surgical environments.
Device Story
Anesthesia delivery system for hospital operating rooms; operated by trained physicians. Inputs: oxygen, nitrous oxide, air, and anesthetic agents. System uses 'bag in bottle' technology to control ventilation cycles; valveless patient circuitry provides low-resistance pathway for spontaneous breathing. Modes: manual (hand bag or spontaneous) or mechanical (time-cycled, volume-controlled, pressure-limited). Features: 'sigh' switch for CO2 expiration; 'hold' switch for temporary ventilation cessation during imaging. Outputs: gas delivery, monitored respiratory parameters (resistance, minute volume), alarms, and error messages. Benefits: safe, accurate gas delivery and ventilation support for diverse patient populations.
Clinical Evidence
Bench testing only. Safety and performance testing conducted to ensure compliance with functional requirements and industry standards (UL2601-1, C222.2 No. 601-1, IEC 601-1, IEC-1-2, ASTM, EN, ISO).
Technological Characteristics
Bag-in-bottle ventilation; time-cycled, volume-controlled, pressure-limited. Valveless patient circuitry. Gas inputs: O2, N2O, air. Compatible with Halothane, Isoflurane, Enflurane, Sevoflurane. Integrated O2 and respiratory monitors. Standards: UL2601-1, C222.2 No. 601-1, IEC 601-1, IEC-1-2, ASTM, EN, ISO.
Indications for Use
Indicated for intermittent or continuous gas inhalation in adult, pediatric, and neonatal patients. Used for administration of oxygen, nitrous oxide, and air with anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane) to maintain respiration during anesthesia in surgical environments.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Spacelabs medical. The logo is in black and white, with the word "Spacelabs" in a bold, sans-serif font. The word "medical" is below "Spacelabs" in a smaller font. A curved line is above the word "Spacelabs". The date "NOV 20 1998" is in the upper left corner of the image.
15220 N.E. 40th Stree PO Box 97013 Redmond, Washington 98073-9713 425-882-3700
1981530 510(k) SUMMARY
## Spacelabs Medical 90519 Anesthesia Delivery System
| 1. | Submitter's Name/ Contact Person: Company | Russ Garrison Director of Regulatory Affairs Spacelabs Medical Inc. 15220 N.E. 40th Street Redmond, WA 98073 |
|----|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Telephone Facsimile | (425) 882-3913 (425) 867-3550 |
| 2. | Name of Device | Spacelabs Medical 90519 Anesthesia Delivery System |
| | Classification | Gas-Machine, Anesthesia 73BSZ; 868.5160 Class II |
| 3. | Predicate Device | We consider the Spacelabs Medical 90519 Anesthesia Delivery System to be substantially equivalent to currently-marketed devices with the identical intended uses. Specifically, the Spacelabs Medical 90519 Anesthesia Delivery System is substantially equivalent to the Falcon marketed by Medical and Industrial Equipment Limited (MIE) (510[k] reference K971030) for the administration of general anesthetics, continually or intermittently, and to maintain a patient's ventilation. |
| 4. | Device Description | The Spacelabs Medical 90519 Anesthesia Delivery System is designed to deliver anesthetic gases in a controlled hospital environment. |
The System incorporates a valveless patient circuitry design which offers very low resistance during spontaneous breathing since it allows for the ebb and flow of natural breathing Oxygen, nitrous oxygen, and air are supplied to the gas distribution system for mixing with the selected anesthetic agent and delivery to the patient through the ventilation system. Another circuit separately provides for the cycling of patient expiration gases.
{1}------------------------------------------------
gases.
The operator may select from two operating modes options: in manual mode the patient can be ventilated manually via a hand bag or the patient can breathe spontaneously; alternately, the controls can be set so that the patient is continually ventilated by the device.
The control settings provide ease of use for both pediatric and adult patient populations in that the System may be set by the anesthesiologist for the appropriate tidal volume range based on patient needs. . The System uses standard, commercially-available anesthesia system accessories appropriate to meet the needs of the clinician and patient.
The delivery of gases is time cycled, volume controlled, pressure limited, and monitored for alarm states and error messages. Additional user-selectable convenience features include a hold switch to permit the temporary cessation of mechanical (continual) ventilation to facilitate the taking of imagery (e.g. x-rays) and a "sigh" switch to provide extra volume above the set tidal volume at a defined interval to allow for the more complete expiration of carbon dioxide.
- The Spacelabs Medical 90519 Anesthesia Delivery 5. Intended Use System provides intermittent or continuous gas inhalation for adults and children (neonatal and pediatric). It allows the administration of operator selected gas mixtures of oxygen, nitrous oxide and air with any of the anesthetic agents: Halothane, Isoflurane, Enflurane, or Sevoflurane. It provides safe and accurate gas flows to maintain patient respiration during anesthesia, and incorporates a ventilator, an oxygen monitor, and a respiratory monitor. The ventilator provides the necessary power, as air or oxygen, to generate volumes and pressures in the ventilating system to ventilate a patient connected to the anesthesia machine. It is recommended for use only by trained physicians, in the operating room or similar surgical environments.
- The intended use, design, materials, accessories, energy 6. Comparison of Technological source, and principles of operation are similar to the Characteristics Falcon Anaesthesia System. Both systems utilize a "bag in bottle" technology to provide fresh and anesthetic gases and to control the ventilation cycle. Both systems offer the convenience for use in both pediatric and adult patients by user adjustment of the tidal volume range
7
{2}------------------------------------------------
control to the appropriate mode setting. Both units are time cycled, volume controlled, and pressure limited at user-selectable settings, generating alarms and error messages as appropriate.
The Spacelabs Medical 90519 Anesthesia Delivery System differs from the Falcon in offering a valveless patient circuitry to provide a low resistance pathway for the delivery of fresh gas to the patient. The 90519 System additionally features the capabilities for user selection of a "sigh" capability for more complete expiration of CO2 gas, a "hold" feature to temporarily maintain a breath to reduce patient motion during a photographic (e.g. x-ray) session, and the automated calculation and display of monitored parameters such as resistance and minute volume for user convenience.
- The Spacelabs Medical 90519 Anesthesia Delivery Standards 7. System is designed to meet the general safety requirements for medical equipment, including the requirements of UL2601-1, C222.2 No. 601-1, IEC 601-1 and IEC-1-2. The 90519 also meets the requirements of industry standards for anesthesia delivery systems, including ASTM, EN, and ISO Standards. The device will be CE marked as a Class IIB device to the Medical Devices Directive.
- The Spacelabs Medical 90519 Anesthesia Delivery 8. Testing System has been subject to extensive safety and performance testing, to ensure that the device meets all its functional requirements and performance of specifications. Safety and standards testing has been, or will be, performed by third party agencies to ensure the System complies to applicable industry and safety standards.
In conclusion, the Spacelabs Medical 90519 Anesthesia Delivery System is as safe and effective as its predicate device and raises no new issues.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 20 1998
Ms. Nancy Gertlar Spacelabs Medical, Inc. 15220 N.E. 40th Street P.O. Box 97013 Redmond, WA 98073-9713
Re: K981530 Spacelabs Medical 90519 Anesthesia Delivery System Requlatory Class: II (two) Product Code: 73 BSZ Dated: August 27, 1998 Received: August 28, 1998
Dear Ms. Gertlar:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Nancy Gertlar
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
sincerely yours,
Thomas J. Callahan
Thomas J. Čallahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
Page 1 of 1
K981530 510(k) Number (if known): Not Known (New Submission)
Device Name: Spacelabs Medical 90519 Anesthesia Delivery System
Indications for Use:
The Spacelabs Medical 90519 Anesthesia Delivery System provides intermittent or continuous gas inhalation for adults and children (neonatal and pediatric). It allows the administration of operator selected gas mixtures of oxygen, nitrous oxide and air with any of the anesthetic agents: Halothane, Isoflurane, Enflurane, or Sevoflurane, It provides safe and accurate gas flows to maintain patient respiration during anesthesia, and incorporates a ventilator, an oxygen monitor, and a respiratory monitor. The ventilator provides the necessary power, as air or oxygen, to generate volumes and pressures in the ventilating system to ventilate a patient connected to the anesthesia machine. It is recommended for use only by trained physicians, in the operating room or similar surgical environments.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
| Prescription Use<br>(Per 21 CFR 801.109)<br>(Optional Format 1-2-96) | <span style="font-size: 2em;">✓</span> |
|----------------------------------------------------------------------|----------------------------------------|
|----------------------------------------------------------------------|----------------------------------------|
OR
| Over-The-Counter Use | |
|----------------------|--|
|----------------------|--|
ukame
6/19/98
le modos
mukame
"1
12-17-98
Division Sign-Off) (Division Sign Off, On-)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number ***_***
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.