← Product Code [IYO](/productcode/IYO) · K981404

# FUKUDA DENSHI ULTRASOUND SCANNERS MODELS UF-3500 AND UF-4500 (K981404)

_Fukuda Denshi USA, Inc. · IYO · May 20, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K981404

## Device Facts

- **Applicant:** Fukuda Denshi USA, Inc.
- **Product Code:** [IYO](/productcode/IYO.md)
- **Decision Date:** May 20, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

As originally submitted, these devices are intended to be used for applications in fetal, abdominal, intraoperative (defined as the abdominal region and periphery), pediatric, small organ (defined as thyroid, breast and testes), cardiac (UF-4500 only), transrectal, transvaginal (UF-4500 only),, and peripheral vessel scanning. This submission adds conventional and superfical musculoskeletal imaging to both devices. Both the UF-3500 and the UF-4500 are prescription devices intended to by used by or on the order of a physician or similarly qualified health care professional. Both are intended to be used in a doctor's office and all hospital environments; ER ICU, CCU, OR, etc. These devices are intended to be use on any patient; neonate, pediatric, or adult; where the placement and positioning of the transducer does not interfere with or complicate the treatment of the patient. These devices are not intended for home use.

## Device Story

General-purpose ultrasound systems (UF-3500, UF-4500) utilize linear probes to perform B-Mode imaging. Devices capture acoustic echo data to generate real-time anatomical images of musculoskeletal structures. Operated by physicians or qualified healthcare professionals in clinical settings (doctor's office, hospital, ER, ICU, OR). Output displayed on-screen for clinician review to identify musculoskeletal soft tissues. No hardware or software modifications were made to support the new musculoskeletal imaging indication; acoustic output remains within established Track 1 limits. Benefits include accurate and reproducible visualization of musculoskeletal anatomy to aid clinical assessment.

## Clinical Evidence

Bench testing and clinical evaluation. Skilled sonographers performed imaging using UF-3500 and UF-4500. Images were reviewed by a radiologist to render findings and clinical impressions. Study demonstrated anatomical identification of musculoskeletal soft tissues and confirmed capability for accurate, reproducible imaging.

## Technological Characteristics

General-purpose ultrasound systems using microprocessors. Linear probes (FUT-L104, FUT-L106) used for B-Mode musculoskeletal imaging. Acoustic output complies with Track 1 limits (ISPTA3, MI). No changes to hardware, software, or patient-contact materials. Standalone operation.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- ATL model Level 10 HDI Ultrasound System ([K961459](/device/K961459.md))
- GE Model LOGIQ 500 ([K970901](/device/K970901.md))

## Reference Devices

- Fukuda Denshi UF-3500 ([K955543](/device/K955543.md))
- Fukuda Denshi UF-4500 ([K922208](/device/K922208.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 2 0 1998

# 510(k) Summary

# New Intended use ... Musculoskeletal Imaging

Fukuda Denshi General Purpose Ultrasound Scanners UF-3500 (K955543) and UF-4500 (K922208)

# Submitter:

FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St., Bldg C Redmond WA 98052 425/881-7737 Tel: Fax: 425/869-2018

## Contact Person:

David J. Geraghty Regulatory Affairs Manager FUKUDA DENSHI AMERICA CORP. 17725 NE 65th St., Bldg C Redmond WA 98052 425/881-7737 Tel: 425/869-2018 Fax:

### Date Prepared:

March 26, 1998

# Device Name:

### New Intended Use:

Conventional and Superficial Musculoskeletal Imaging for the

- Fukuda Denshi FF sonic model UF-3500 General . Purpose Ultrasound Scanner (K955543)and the
- Fukuda Denshi FF sonic model UF-4500 General . Purpose Ultrasound Scanner (K922208)

### Common Name:

Conventional and Superficial Musculoskeletal Imaging

### Classification Name:

Ultrasonic Pulsed Echo Imaging System (§892.1560) Diagnostic Ultrasonic Transducer (§892.1560)

# Legally Marketed Device:

- ATL model Level 10 HDI Ultrasound System (K961459) ●
- . GE Model LOGIQ 500 (k970901

{1}------------------------------------------------

# Description:

The Fukuda Denshi model UF-4500 (K022208) and the model UF-3500 (K955543) are general-purpose ultrasound systems that have been previously found to be substantially equivalent under the Food and Drug Administration's 510(k) process. These devices have not been changed or modified in any way. All software and hardware, all portions of the device that control the acoustic power out are unchanged.

The addition of both conventional and musculoskeletal imaging as intended uses require no changes to the operating instructions.

# Intended Use:

As originally submitted, these devices are intended to be used for applications in fetal, abdominal, intraoperative (defined as the abdominal region and periphery), pediatric, small organ (defined as thyroid, breast and testes), cardiac (UF-4500 only), transrectal, transvaginal (UF-4500 only),, and peripheral vessel scanning. This submission adds conventional and superfical musculoskeletal imaging to both devices.

Both the UF-3500 and the UF-4500 are prescription devices intended to by used by or on the order of a physician or similarly qualified health care professional. Both are intended to be used in a doctor's office and all hospital environments; ER ICU, CCU, OR, etc. These devices are intended to be use on any patient; neonate, pediatric, or adult; where the placement and positioning of the transducer does not interfere with or complicate the treatment of the patient. These devices are not intended for home use.

{2}------------------------------------------------

# Technological Characteristics

The UF-3500 and UF-4500 incorporate microprocessors in a manner similar to the predicate devices. No changes to the hardware or software that control the device or the acoustic output were required to support the claims for musculoskeletal imaging. Conventional and superficial musculoskeletal imaging is possible with linear probes in B-Mode. Patient contact materials have not changed

Both devices' derated acoustic output limits remain below the maximums established for Track 1 devices as listed in the "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" issued September 30, 1997, and restated below:

| Use                    | ISPTA3<br>(Mw/CM²) | ISPTA3<br>(Mw/CM²) | MI   |
|------------------------|--------------------|--------------------|------|
| Peripheral Vessel      | 720                | 190                | 1.9  |
| Cardiac                | 430                | 190                | 1.9  |
| Fetal Imaging & Other* | 94                 | 190                | 1.9  |
| Ophthalmic             | 17                 | 28                 | 0.23 |

# Testing:

Skilled sonographers using both the UF-3500 and the UF-4500 obtained images. All imaged were reviewed by a radiologist who rendered findings and clinical impression.

# Conclusion:

Based on the anatomical identification of musculoskeletal soft tissues and the diagnostic images obtained from the UF-3500 and the UF-4500, these devises have exhibited the capability of accurately and reproducibly imaging the musculoskeletal structure. Fukuda Denshi has demonstrated through this study that the device is safe and effective in obtaining conventional and superficial musculoskeletal images and is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized eagle or bird in flight, with three curved lines forming the body and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 0 1998

David J. Geraghty Regulatory Affairs Manager Fukuda Denshi America Corp. 17725 NE 65th Street Redmond, WA 98052

Re: K981404 Fukuda Denshi Ultrasound Scanners Models UF-3500 and UF-4500 Dated: March 23, 1998 Received: April 17, 1998 Regulatory class: II 21 CFR 892.1560/Procode: 90 IYO 21 CFR 892.1570/Procode: 90 ITX

Dear Mr. Geraghty:

We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Fukuda Denshi Ultrasound Scanners Models UF-3500 and UF-4500, as described in your premarket notification:

Transducer Model Number

FUT-L104 (5.0 MHz, 60mm) FUT-L106 (7.5 MHz, 40mm)

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you. ... may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - David J. Geraghty

Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA's position on this use.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro-diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Lillian Yin, Ph.D.

Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{5}------------------------------------------------

## Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-3500 System

# Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

| Clinical Application             | A | B | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic                       |   |   |     |                  |                      |                              |                       |                    |
| Fetal                            |   | P |     |                  |                      |                              |                       |                    |
| Abdominal                        |   | P |     |                  |                      |                              |                       |                    |
| Intraoperative (Specify)         |   | P |     |                  |                      |                              |                       |                    |
| Intraoperative                   |   |   |     |                  |                      |                              |                       |                    |
| Neurological                     |   |   |     |                  |                      |                              |                       |                    |
| Pediatric                        |   | P |     |                  |                      |                              |                       |                    |
| Small Organ (Specify)            |   | P |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |   |   |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |   |   |     |                  |                      |                              |                       |                    |
| Cardiac                          |   |   |     |                  |                      |                              |                       |                    |
| Transesophageal                  |   |   |     |                  |                      |                              |                       |                    |
| Transrectal                      |   | P |     |                  |                      |                              |                       |                    |
| Transvaginal                     |   |   |     |                  |                      |                              |                       |                    |
| Intravascular                    |   |   |     |                  |                      |                              |                       |                    |
| Peripheral Vessel                |   | P |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |   |   |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |   | N |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |   | N |     |                  |                      |                              |                       |                    |
| Other (specify)                  |   |   |     |                  |                      |                              |                       |                    |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Intraoperative is defined as the abdominal region and the periphery Other Indication or Modes:

Small Organ is defined as thyroid, breast, and testes

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vieinity of the shoulder, elbow, wrist, knee, and/or ankle.

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Bergman

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number K981404

Prescription Use (Per 21 CRF 801.109)

Page: 1 of 6

{6}------------------------------------------------

### Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-3500 w/ FUT-L104 Probe

## Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

| Clinical Application     | A | B | Mode of Operation |               |                   |                        |                    |                 |  |
|--------------------------|---|---|-------------------|---------------|-------------------|------------------------|--------------------|-----------------|--|
|                          |   |   | CWD               | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |  |
| Ophthalmic               |   |   |                   |               |                   |                        |                    |                 |  |
| Fetal                    |   |   |                   |               |                   |                        |                    |                 |  |
| Abdominal                |   | P |                   |               |                   |                        |                    |                 |  |
| Intraoperative (Specify) |   |   |                   |               |                   |                        |                    |                 |  |
| Intraoperative           |   |   |                   |               |                   |                        |                    |                 |  |
| Neurological             |   |   |                   |               |                   |                        |                    |                 |  |
| Pediatric                |   | P |                   |               |                   |                        |                    |                 |  |
| Small Organ (Specify)    |   | P |                   |               |                   |                        |                    |                 |  |
| Neonatal Cephalic        |   |   |                   |               |                   |                        |                    |                 |  |
| Adult Cephalic           |   |   |                   |               |                   |                        |                    |                 |  |
| Cardiac                  |   |   |                   |               |                   |                        |                    |                 |  |
| Transesophageal          |   |   |                   |               |                   |                        |                    |                 |  |
| Transrectal              |   |   |                   |               |                   |                        |                    |                 |  |
| Transvaginal             |   |   |                   |               |                   |                        |                    |                 |  |
| Intravascular            |   |   |                   |               |                   |                        |                    |                 |  |
| Peripheral Vessel        |   | P |                   |               |                   |                        |                    |                 |  |
| Laparoscopic             |   |   |                   |               |                   |                        |                    |                 |  |
| Musculo-skeletal         |   | N |                   |               |                   |                        |                    |                 |  |
| Conventional             |   |   |                   |               |                   |                        |                    |                 |  |
| Musculo-skeletal         |   | N |                   |               |                   |                        |                    |                 |  |
| Superficial              |   |   |                   |               |                   |                        |                    |                 |  |
| Other (specify)          |   |   |                   |               |                   |                        |                    |                 |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Other Indication or Modes: Small Organ is defined as thyroid, breast, and testes

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vicinity of the shoulder, elbow, wrist, knee, and/or ankle.

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT and Radiological Devic

510(k) Number

Prescription Use (Per 21 CRF 801.109)

Page: 2 of 6

{7}------------------------------------------------

### Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-3500 w/ FUT-L106 Probe

# Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

|                                  | Mode of Operation |   |     |                  |                      |                              |                       |                    |  |  |
|----------------------------------|-------------------|---|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|--|
| Clinical Application             | A                 | B | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |  |  |
| Ophthalmic                       |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Fetal                            |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Abdominal                        |                   | P |     |                  |                      |                              |                       |                    |  |  |
| Intraoperative (Specify)         |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Intraoperative                   |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Neurological                     |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Pediatric                        |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Small Organ (Specify)            |                   | P |     |                  |                      |                              |                       |                    |  |  |
| Neonatal Cephalic                |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Adult Cephalic                   |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Cardiac                          |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Transesophageal                  |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Transrectal                      |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Transvaginal                     |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Intravascular                    |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Peripheral Vessel                |                   | P |     |                  |                      |                              |                       |                    |  |  |
| Laparoscopic                     |                   |   |     |                  |                      |                              |                       |                    |  |  |
| Musculo-skeletal<br>Conventional |                   | N |     |                  |                      |                              |                       |                    |  |  |
| Musculo-skeletal<br>Superficial  |                   | N |     |                  |                      |                              |                       |                    |  |  |
| Other (specify)                  |                   |   |     | ・・、              |                      |                              |                       |                    |  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Small Organ is defined as thyroid, breast, and testes Other Indication or Modes:

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vicinity of the shoulder, elbow, wrist, knee, and/or ankle.

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Higgins
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Device

510(k) Number

Prescription Use (Per 21 CRF 801.109)

Page: 3 of 6

{8}------------------------------------------------

### Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-4500 System

Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

|                               | Mode of Operation |   |     |               |                   |                        |                    |                 |
|-------------------------------|-------------------|---|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application          | A                 | B | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| Ophthalmic                    |                   |   |     |               |                   |                        |                    |                 |
| Fetal                         |                   | P |     |               |                   |                        |                    |                 |
| Abdominal                     |                   | P |     |               |                   |                        |                    |                 |
| Intraoperative (Specify)      |                   | P |     |               |                   |                        |                    |                 |
| Intraoperative                |                   |   |     |               |                   |                        |                    |                 |
| Neurological                  |                   |   |     |               |                   |                        |                    |                 |
| Pediatric                     |                   | P |     |               |                   |                        |                    |                 |
| Small Organ (Specify)         |                   | P |     |               |                   |                        |                    |                 |
| Neonatal Cephalic             |                   |   |     |               |                   |                        |                    |                 |
| Adult Cephalic                |                   |   |     |               |                   |                        |                    |                 |
| Cardiac                       |                   | P |     |               |                   |                        |                    |                 |
| Transesophageal               |                   |   |     |               |                   |                        |                    |                 |
| Transrectal                   |                   | P |     |               |                   |                        |                    |                 |
| Transvaginal                  |                   | P |     |               |                   |                        |                    |                 |
| Intravascular                 |                   |   |     |               |                   |                        |                    |                 |
| Peripheral Vessel             |                   | P |     |               |                   |                        |                    |                 |
| Laparoscopic                  |                   |   |     |               |                   |                        |                    |                 |
| Musculo-skeletal Conventional |                   | N |     |               |                   |                        |                    |                 |
| Musculo-skeletal Superficial  |                   | N |     |               |                   |                        |                    |                 |
| Other (specify)               |                   |   |     |               |                   |                        |                    |                 |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Intraoperative is defined as the abdominal region and the periphery Other Indication or Modes:

Small Organ is defined as thyroid, breast, and testes

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vicinity of the shoulder, elbow, wrist, knee, and/or ankle.

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Timith le. Sefrom

---

(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devic

510(k) Number

Prescription Use (Per 21 CRF 801.109)

Page: 4 of 6

{9}------------------------------------------------

## Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-4500 w/ FUT-L104 Probe

# Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

| Clinical Application             | A | B | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic                       |   |   |     |                  |                      |                              |                       |                    |
| Fetal                            |   | P |     |                  |                      |                              |                       |                    |
| Abdominal                        |   | P |     |                  |                      |                              |                       |                    |
| Intraoperative (Specify)         |   |   |     |                  |                      |                              |                       |                    |
| Intraoperative                   |   |   |     |                  |                      |                              |                       |                    |
| Neurological                     |   |   |     |                  |                      |                              |                       |                    |
| Pediatric                        |   | P |     |                  |                      |                              |                       |                    |
| Small Organ (Specify)            |   | P |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |   |   |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |   |   |     |                  |                      |                              |                       |                    |
| Cardiac                          |   |   |     |                  |                      |                              |                       |                    |
| Transesophageal                  |   |   |     |                  |                      |                              |                       |                    |
| Transrectal                      |   |   |     |                  |                      |                              |                       |                    |
| Transvaginal                     |   |   |     |                  |                      |                              |                       |                    |
| Intravascular                    |   |   |     |                  |                      |                              |                       |                    |
| Peripheral Vessel                |   | P |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |   |   |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |   | N |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Superficial  |   | N |     |                  |                      |                              |                       |                    |
| Other (specify)                  |   |   |     |                  |                      |                              |                       |                    |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Small Organ is defined as thyroid, breast, and testes Other Indication or Modes:

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vicinity of the shoulder, elbow, wrist, knee, and/or ankle. ----------------------------------------------------------------------------------------------------------------------------------

> (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David G. Flynn

(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devic

510(k) Number

Prescription Use (Per 21 CRF 801.109)

Page: 5 of 6

{10}------------------------------------------------

### Fill out one form for each ultrasound system or transducer

510(k) Number (if known):

Device Name:

Fukuda Denshi model UF-4500 w/ FUT-L106 Probe

### Indications for Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows

|                                  |   | Mode of Operation |     |                  |                      |                              |                       |                    |  |
|----------------------------------|---|-------------------|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application             | A | B                 | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |  |
| Ophthalmic                       |   |                   |     |                  |                      |                              |                       |                    |  |
| Fetal                            |   | P                 |     |                  |                      |                              |                       |                    |  |
| Abdominal                        |   | P                 |     |                  |                      |                              |                       |                    |  |
| Intraoperative (Specify)         |   |                   |     |                  |                      |                              |                       |                    |  |
| Intraoperative                   |   |                   |     |                  |                      |                              |                       |                    |  |
| Neurological                     |   |                   |     |                  |                      |                              |                       |                    |  |
| Pediatric                        |   |                   |     |                  |                      |                              |                       |                    |  |
| Small Organ (Specify)            |   | P                 |     |                  |                      |                              |                       |                    |  |
| Neonatal Cephalic                |   |                   |     |                  |                      |                              |                       |                    |  |
| Adult Cephalic                   |   |                   |     |                  |                      |                              |                       |                    |  |
| Cardiac                          |   |                   |     |                  |                      |                              |                       |                    |  |
| Transesophageal                  |   |                   |     |                  |                      |                              |                       |                    |  |
| Transrectal                      |   |                   |     |                  |                      |                              |                       |                    |  |
| Transvaginal                     |   |                   |     |                  |                      |                              |                       |                    |  |
| Intravascular                    |   |                   |     |                  |                      |                              |                       |                    |  |
| Peripheral Vessel                |   | P                 |     |                  |                      |                              |                       |                    |  |
| Laparoscopic                     |   |                   |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Conventional |   | N                 |     |                  |                      |                              |                       |                    |  |
| Musculo-skeletal<br>Superficial  |   | N                 |     |                  |                      |                              |                       |                    |  |
| Other (specify)                  |   |                   |     |                  |                      |                              |                       |                    |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Other Indication or Modes: Small Organ is defined as thyroid, breast, and testes

Musculo-skeletal Superficial is defined as muscle, tendon, ligament, bursa, and other soft tissue in the vicinity of the shoulder, elbow, wrist, knee, and/or ankle.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

David U. Hlyson

---

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT and Radiological Device

510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K981404](https://fda.innolitics.com/device/K981404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).