KIWI SOFT CUP WITH PALM PUMP MODEL NUMBER VFE-6000S, KIWI MUSHROOM CUP WITH PALM PUMP MODEL VFE-6000M
Applicant
Clinical Innovations, Inc.
Product Code
HDB · Obstetrics/Gynecology
Decision Date
Sep 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4340
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Use for fetal vacuum delivery assistance in conditions of: 1) dystocia, 2) uterine inertia, 3) maternal exhaustion (ineffective voluntary effort), or 4) maternal or fetal distress.
Device Story
Kiwi Fetal Vacuum Extractor is a manual obstetric device for assisted delivery. System comprises a vacuum cup for attachment to fetal head and a handle with an integrated palm-operated vacuum pump. Operator (physician) generates vacuum by pumping handle; device includes integral vacuum indicator and relief valve. Two cup styles: soft silicone (VFE-6000S) and Malmstrom design (VFE-6000M). Used in clinical labor and delivery settings to assist fetal descent during second stage of labor. Physician applies manual withdrawal force via handle. Device facilitates delivery in cases of dystocia, uterine inertia, maternal exhaustion, or fetal distress, potentially avoiding cesarean section.
Clinical Evidence
Safety and effectiveness demonstrated through laboratory and clinical testing. Biocompatibility testing confirmed material safety. Laboratory testing verified performance regarding vacuum measurement and device integrity. No specific clinical trial metrics (e.g., success rates, p-values) provided in the summary.
Technological Characteristics
Manual fetal vacuum extractor. Components: vacuum cup (soft silicone or Malmstrom design), handle with integrated palm-operated vacuum pump, vacuum indicator, and vacuum relief valve. Materials are biocompatible. Non-powered, manual operation. Standalone device.
Indications for Use
Indicated for fetal vacuum extraction in cases of prolonged second stage of labor (arrest of descent) with adequate fetopelvic relationships, non-severe presumed fetal jeopardy, or elective shortening of second stage for maternal/fetal conditions. Also indicated for trial of vacuum extraction in borderline fetopelvic relationships or mid-pelvic extraction when position/station are known. Contraindicated if no descent after 2 tractions, delivery not imminent after 4 tractions, or cup detaches twice.
Regulatory Classification
Identification
A fetal vacuum extractor is a device used to facilitate delivery. The device enables traction to be applied to the fetal head (in the birth canal) by means of a suction cup attached to the scalp and is powered by an external vacuum source. This generic type of device may include the cup, hosing, vacuum source, and vacuum control.
K021193 — FORTUNE SILICONE VACUUM SUCTION, MODEL #1300 · Fortune Medical Instrument Corp. · Feb 12, 2003
K013460 — SWIFT LOW PROFILE CUP (PRO CUP) · Swift Delivery Products · Jan 17, 2002
K041579 — MEDELA SINGLE USE SILC CUP, MODEL 077.0791 · Medela AG · Apr 21, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
SEP 22 1998
K981260
Kiwi Fetal Vacuum Extractor
Clinical Innovations, Inc.
## Section 6.0 510(k) Summarv
| Submitter: | Clinical Innovations, Inc. |
|------------|-------------------------------------------|
| Name: | Wm. Dean Wallace |
| Address: | 6477 So. Cottonwood St., Murray, UT 84107 |
| Telephone: | (801) 268-8200 |
| Fax: | (801) 266-7373 |
| Proprietary Names: | Kiwi Fetal Vacuum Extractors VFE-6000S and VFE-6000M |
|----------------------|------------------------------------------------------|
| Common/Usual Name: | Fetal Vacuum Extractor |
| Classification Name: | Extractor, Vacuum, Fetal |
The legally marketed devices to which equivalence is claimed are: Mityvac "M" style vacuum extractor (K890307). Mityvac Super "M" Style vacuum extractor (K943928), Mityvac Obstetric Vac Deliv. Kit w/ Univ. VAC REL (K934011), CMI Velvet Touch Vacuum Extractor (K942725). CMI Vacuum Pump #001C (K881967) and Swift Delivery Product #002 (K970170).
Description of the device: The Kiwi Complete Vacuum Delivery Systems are fetal vacuum extractors having a cup to attach to the fetal head, a handle with which to apply withdrawal force, and a vacuum pump built into the handle of the cup/handle combination. This palm pump includes a method for easily generating a vacuum by pumping the handle with the palm. A vacuum indicator and a vacuum relief valve are also integral to the palm pump. Two styles of cups are available; one of soft silicone (VFE-6000S) and the other of the Malmstrom design (VFE-6000M).
Intended use: Use for fetal vacuum delivery assistance in conditions of: 1) dystocia, 2) uterine inertia, 3) maternal exhaustion (ineffective voluntary effort), or 4) maternal or fetal distress.
The Kiwi Fetal Vacuum Extractors are substantially equivalent to the predicate devices because: they have the same intended uses, namely, use for fetal vacuum delivery assistance, and they have the same basic technological characteristics as predicate devices, namely, a hand pump generating a vacuum that attaches a cup to the fetal head and which has a handle for applying force to assist in the delivery of the fetus. They use the same or similar materials, all of which have been shown to be biocompatible and to function well in the intended application.
The safety and effectiveness are similar to existing devices as demonstrated in the laboratory and in clinical testing. Biocompatibility testing shows that the materials used in the Kiwi Fetal Vacuum Extractors are safe for this application. Effectiveness is the same as the predicate devices. The laboratory testing verified the performance in terms of vacuum measurement and device integrity.
Was. Dan Wallace
W. D. Wallace, M.D., Ph.D.
Wm. Dean Wallace, M.D., Ph.D.
4-2-98
Date
Page 8
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular pattern around the caduceus symbol. The image is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 2 1998
Wm. Dean Wallace, M.D., Ph.D. President CLINICAL INNOVATIONS 6477 S. Cottonwood Street Murray, UT 84107
Re: K981260
Kiwi Soft Cup with Palm Pump, Model Nos. VFE-6000M and VFE-6000S (Fetal Vacuum Extractors) Dated: July 8, 1998 Received: July 10, 1998 Regulatory Class: II 21 CFR 884.4340/Procode: 85 HDB
Dear Dr. Wallace:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionaliy, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely vours
Lillian Yin, Ph.D.
Director, Division of Repro Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
09/17/98
Kiwi Fetal Vacuum Extractor Clinical Innovations, Inc.
## 11.0 Indications For Use
Device Name: Kiwi Fetal Vacuum Extractor
K981260 510(k) Number:
Indications for use:
Standard vacuum extraction
Use for fetal vacuum extraction in conditions of 1) prolonged second stage of labor (arrest of descent) where fetopelvic relationships are adequate, 2) presumed fetal jeopardy which is not considered to be severe, or 3) elective shortening of the second stage for selected maternal or fetal conditions.
## Trial of vacuum extraction
Vacuum delivery should be regarded as a "trial" 1) if there is arrest of descent in the second stage and fetopelvic relationships are considered to be boderline, or 2) in a mid-pelvic extraction when position and station are known.
Vacuum extraction should be abandoned and birth completed by cesarean section 1) if no descent (progress) of the head occurs after 2 tractions. 2) if delivery is not achieved or imminent after 4 tractions, or 3) if the vacuum cup detaches ("pops-off") twice.
(PLEASE NO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devi 510(k) Number
Prescription Use
(Per 21 CFR 801.109) /
OR
Over-The-Counter Use (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.