← Product Code [LNH](/productcode/LNH) · K981190

# HIGH IMPEDANCE CARDIAC GATING CABLE (K981190)

_GE Medical Systems · LNH · Jun 30, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K981190

## Device Facts

- **Applicant:** GE Medical Systems
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Jun 30, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The GE High Impedance Cardiac Gating Cable is designed for MR applications where cardiac gating is desired to reduce motion and flow artifacts. The cable may be used in the presence of surface coils.

## Device Story

High impedance cardiac gating cable; four-lead ECG patient cable. Used in MR scanner environments to facilitate cardiac gating; reduces motion and flow artifacts. High distributed impedance limits RF currents induced by MR scanner. Operated by clinical staff in MR imaging suites. Output provides ECG signal for scanner synchronization; enables improved image quality by minimizing motion-related artifacts. Benefits patient through reduced need for repeat scans due to motion.

## Clinical Evidence

Bench testing and volunteer testing completed. Results demonstrated no significant cable heating and successful cardiac gating performance during MR imaging.

## Technological Characteristics

Four-lead ECG patient cable; high distributed impedance design to mitigate RF-induced currents. Designed for MR compatibility.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Invivo Cables #9240A
- GE Flat Film Leads #E8811JA

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters "GE" intertwined within a circular border. A portion of the logo is cut off on the left side, giving it an incomplete appearance.

K981/190

## GE Medical Systems

P.O. Box 414, W-709 Milwaukee, WI 53201 USA

# SUMMARY OF SAFETY AND EFFECTIVENESS

- o This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
- o Identification of Submitter Larry A. Kroger, Ph.D., 414-544-3894, March 25, 1998
- o Identification of the Product High Impedance Cardiac Gating Cables

Manufactured by:

GE Medical Systems 3200 N. Grandview Blvd. Waukesha, WI 53188

## o Marketed Devices

The High Impedance Cardiac Gating Cable is substantially equivalent to the currently marketed Invivo Cables #9240A and the GE Flat Film Leads #E8811JA.

### o Device Description

The high impedance cardiac gating cable is a four lead ECG patient cable used for gating the MR scanner. The cable has a high distributed impedance to limit any RF currents induced into the cable by the MR scanner.

## o Indications for Use

The GE high impedance cardiac gating cable is designed for MR applications where cardiac gating is required to reduce motion and flow artifacts. The cable may be used in the presence of surface coils.

#### 0 Comparison with Predicate

The GE high impedance cardiac gating cable is comparable to the Invivo cable in mechanical flexibility and with the GE Thin Film cable in regard to impedance.

#### o Summary of Studies

In-house as well as volunteer testing was completed. There was no significant heating of the cables and the cables gated the MR System properly.

#### o Conclusions

It is the opinion of GE that the High Impedance Cardiac Cables are substantially equivalent to the presently marketed cables. These cables do not include any new indications for use, nor does use of this cable result in any new potential hazards.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Image /page/1/Picture/3 description: The image shows a date stamp that reads "JUN 3 0 1998". The text is in a bold, sans-serif font. The date is likely stamped on a document or piece of paper. The image is in black and white.

Larry A. Kroger, Ph.D.

Senior Regulatory Affairs Programs Manager General Electric Systems P.O. Box 414. W-827 Milwaukee, Wisconsin 53201

Re:

K981190 High Impedance Cardiac Gating Cable Dated: March 25, 1998 Received: April 2, 1998 Regulator Class: II 21 CFR 892.1000/Procode: 90 LNH

Dear Dr. Kroger:

We have reviewed your Section 510/k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrl/dsmaldsmam.html".

Sincerely yours,

Lillian Yin, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: High Impedance Cardiac Gating Cable

Indications For Use:

The GE High Impedance Cardiac Gating Cable is designed for MR applications where cardiac gating is desired to reduce motion and flow artifacts. The cable may be used in the presence of surface coils.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Division Sign-Off                                                  |         |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |         |
| 510(k) Number                                                      | K981190 |

| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|---------------------------------------|----|----------------------|
|---------------------------------------|----|----------------------|

---

**Source:** [https://fda.innolitics.com/device/K981190](https://fda.innolitics.com/device/K981190)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com