N-ASSAY GLU-UL

K980883 · Crestat Diagnostics, Inc. · CFR · Sep 3, 1998 · Clinical Chemistry

Device Facts

Record IDK980883
Device NameN-ASSAY GLU-UL
ApplicantCrestat Diagnostics, Inc.
Product CodeCFR · Clinical Chemistry
Decision DateSep 3, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Intended Use

The intended use for the N-ASSAY Glu-UL Reagent is for the quantitative determination of glucose in serum, plasma, urine, and cerebrospinal fluid in the diagnosis and treatment of diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. For in vitro diagnostic use only.

Device Story

N-ASSAY Glu-UL is an enzymatic, ready-to-use liquid reagent for clinical chemistry laboratories. It utilizes a hexokinase/glucose-6-phosphate dehydrogenase method to measure glucose concentrations in human serum, plasma, urine, and cerebrospinal fluid. Input samples are processed via a biochemical reaction: hexokinase phosphorylates D-glucose to glucose-6-phosphate (G-6-P) using ATP; G-6-PDH then oxidizes G-6-P to 6-phosphogluconate, reducing NADP to NADPH. The resulting NADPH absorbs light at 340 nm and 405 nm. The device output is a spectrophotometric absorbance measurement directly proportional to glucose concentration. Clinicians use these quantitative results to diagnose and manage diabetes and hypoglycemia. The reagent is intended for professional in vitro diagnostic use.

Clinical Evidence

No clinical trials; evidence based on bench-top comparative performance studies. The device was compared against predicate assays using serum, urine, and CSF samples. Results showed high correlation (r=0.99587 for serum; r=0.9989 for urine; r=0.9923 for CSF). Linearity established up to 1,000 mg/dl for serum; limit of quantitation 0.3 mg/dl for urine/CSF; minimum detectable level 1 mg/dl.

Technological Characteristics

Enzymatic hexokinase/glucose-6-phosphate dehydrogenase liquid reagent. Principle: Spectrophotometric detection of NADPH at 340 nm and 405 nm. Ready-to-use liquid form. No specific materials of construction or software algorithms described.

Indications for Use

Indicated for quantitative glucose determination in serum, plasma, urine, and cerebrospinal fluid to aid in diagnosis and treatment of diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 3 1998 # CRESTAT DIAGNOSTICS, INC. K980883 25549 Adams Avenue Murrieta, CA 92562 #### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. The assigned 510(k) number is: March 2, 1998 Date: Submitted by: Colin Getty KAMIYA BIOMEDICAL COMPANY 910 Industry Drive, Seattle WA 98188 TEL: 206-575-8068; FAX: 206-575-8094 Crestat Diagnostics, Inc. For: 25549 Adams Avenue Murrieta. CA 92562 N-ASSAY Glu-UL (Glucose Assay Reagent) Product: Blood qlucose determinations are one of the most frequently performed clinical chemistry laboratory procedures and are commonly used as an aid in the diagnosis and treatment of diabetes. Early methods were based on the reducing properties of glucose: however, these methods are susceptible to positive errors caused by nonglucose-reducing substances present in normal blood. Chemical methods using otoluidine offered improved specificity, but other sugars not normally present in serum can also react with o-toluidine and be a potential source of error. Enzymatic methods using purified enzymes which act on glucose provided greater specificity. The N-ASSAY Glu-UL reagent is based on an enzymatic hexokinase/glucose-6phosphate dehydrogenase method, shows good correlation with similar glucose reagents, practically no interference by coexistent substances, high sensitivity with good reproducibility, wide assay range, and is a convenient ready-to-use liquid type reagent. In this method, serum D-glucose is phosphorylated by hexokinase (HK) in the presence of adenosine triphosphate (ATP) to produce glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PDH) specifically oxidizes G-6-P to 6-Phosphogluconate with the concurrent reduction of nicotinamide adenine dinucleotide phosphate (NADP) to nicotinamide adenine dinucleotide phosphate reduced (NADPH). The NADPH produced absorbs light at 340 nm (main) and 405 nm (sub) and can be detected spectrophotometrically. The increase in absorbance measured at 340 nm (main) and 405 (sub), due to the formation of the NADPH, is directly proportional to the glucose concentration in the sample. {1}------------------------------------------------ ## CRESTAT DIAGNOSTICS, INC. ### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS (Continued) The safety and effectiveness of the liquid Crestat N-ASSAY Glu-UL Reagent is demonstrated by its substantial equivalence to the Medical Analysis Systems Glucose liguid reagent (K853464) which is also based on an enzymatic hexokinase/glucose-6phosphate dehydrogenase method. Both test systems are intended to quantitatively measure glucose in human serum. In comparison studies against the predicate assay, a correlation coefficient of 0.99587 and a regression equation y = 0.8861 x + 4.7012 was obtained with serum samples. Precision studies indicate acceptable values can be obtained on a day to day basis. The minimum detectable level of this method is 1 mg/dl. The N-ASSAY Glu UL reagent is linear to 1,000 mg/dl. For urine and CSF samples, the liquid Crestat N-ASSAY Glu-UL Reagent was compared to the Boehringer Mannheim Glucose assay (K812303) which is also based on an enzymatic hexokinase method. Using urine samples, a correlation coefficient of 0.9989 and a regression equation y = 1.038 x + 1.528 was obtained. Using CSF samples, a correlation coefficient of 0.9923 and a regression equation of y = 1.080 x + -1.233 was obtained. The limit of quantitation is 0.3 mg/dl. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the HHS logo, which consists of a stylized caduceus with three snakes intertwined around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the logo. The seal is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 3 1998 Crestat Diagnostics, Inc. Colin Getty C/O KAMIYA Biomedical Company 910 Industry Drive Seattle, Washington, 98188 K980883 Re : N-ASSAY Glu-UL Requlatory Class: II Product Code: CFR June 30, 1998 Dated: Received: July 2, 1998 Dear Mr. Getty: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure ... ... {4}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 12980883 510(k) Number (if known): __ N-ASSAY Glu-UL (Glucose Assay Reagent) Device Name: #### Indications For Use: The intended use for the N-ASSAY Glu-UL Reagent is for the quantitative determination of glucose in serum, plasma, urine, and cerebrospinal fluid in the diagnosis and treatment of diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. For in vitro diagnostic use only. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cun (Division Sign-Off) Division of Clinical Laborato 510(k) Number . Prescription Use (Per 21 CFR 801.109) OR 5 Over-The-Counter Use Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...