EXETER INTRAMEDULLARY BONE PLUG

K980843 · Howmedica, Inc. · JDI · Jun 2, 1998 · Orthopedic

Device Facts

Record IDK980843
Device NameEXETER INTRAMEDULLARY BONE PLUG
ApplicantHowmedica, Inc.
Product CodeJDI · Orthopedic
Decision DateJun 2, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3350
Device ClassClass 2

Intended Use

The Exeter II Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty. The Exeter Intramedullary Bone Plug is intended to be used with bone cement.

Device Story

Exeter Intramedullary Bone Plug is a molded acrylic, cone-shaped device; placed in femoral canal prior to femoral stem insertion. Function: restricts bone cement migration distally; enables cement pressurization. Design features: circumferential ridges for interference fit; slots for compression during insertion; threaded portion for insertion instrument. Used in surgical setting by orthopedic surgeons during total hip arthroplasty. Benefits: ensures optimal cement mantle formation and fixation of femoral prosthesis.

Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and operational similarity to legally marketed predicate devices.

Technological Characteristics

Molded acrylic device; cone-shaped with circumferential ridges and slots for compression; threaded interface for insertion tool; available in multiple diameters; non-electronic; non-implantable (temporary use during surgery).

Indications for Use

Indicated for patients undergoing total hip arthroplasty requiring bone cement for femoral stem fixation.

Regulatory Classification

Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 2 1998 JUN 510(k) Summarv K980843 Proprietary Name: Intramedullary Bone Plug, Exeter Hip System Common Name: Cement Restrictor Classification Name and Reference: 21 CFR 888.3350 Prosthesis, Hip, Semi-Constrained, Metal/Polymer Proposed Requlatory Class: Class II Device Product Code: LZN , 30I For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford. NJ 07070 Telephone: (201) 507-7875 Fax: (201) 507-6870 Date Summary Prepared: 3-3-98 The Exeter Intramedullary Bone Plug is a molded acrylic device that is designed to be placed in the femoral canal prior to the insertion of a femoral stem. The bone plug restricts the migration of bone cement down the femoral canal and permits cement pressurization within the canal. The Exeter Bone Plug described in this submission is the same design as the currently marketed device. The Bone Plug is a cone shaped device available in a range of diameters to accomodate various anatomical requirements. The distal portion of the plug is designed with circumferential ridges that create an interference fit between the plug and the femoral canal. Slots are cut into the ridge portion of the bone plug to allow the plug to compress slightly during insertion down the canal. The body of the plug contains a threaded portion for use with an insertion instrument. The substantial equivalence of the Exeter Bone Plug is based on an equivalence in intended use, design, materials, operational principles, and relative indications and contraindications to the Howmedica's polyethylene Exeter Bone Plug, (K933077) and Exeter II Distal Centralizer (K974054). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with flowing lines, representing movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is presented in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 1998 Ms. Margaret F. Crowe Group, Requlatory Affairs Manager Howmedica Inc. Pfizer Hospital Products Group 359 Veterans Boulevard Rutherford, New Jersey 07070-2584 Re: K980843 Exeter Intramedullary Bone Plug Trade Name: Requlatory Class: II Product Codes: JDI and LZN Dated: March 3, 1998 Received: March 4, 1998 Dear Ms. Crowe: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Ms. Margaret F. Crowe This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ia M. Witten, Ph.D., M.D. Cel Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use 510(k) Number (if known): K980843 Device Name: Exeter Intramedullary Bone Plug Indications for Use: The Exeter II Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty. The Exeter Intramedullary Bone Plug is intended to be used with bone cement. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96) (Division Sign-Off) Division of General Restorative Devices 510(k) Number. K980843