K980831 · Philips Medical Systems (Cleveland), Inc. · LNH · Aug 27, 1998 · Radiology
Device Facts
Record ID
K980831
Device Name
IMAGE POST-PROCESSING TECHNIQUES
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
LNH · Radiology
Decision Date
Aug 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
ProPak Techniques The ProPak Package for Picker MR systems or workstations provide supplemental information regarding contrast changes over time for those images extracted from MR temporal datasets. The indications for use for the MR system remain unchanged. Apparent Diffusion Coefficient Mapping The Picker Diffusion-Weighted MR Imaging Package has been designed to image the diffusive mobility of water or other proton-containing molecules. One important clinical application is to visualize the apparent loss of mobility of water molecules in brain tissue affected by acute stroke. Areas of decreased diffusion, as is observed in acute cerebral infarcts, appear as areas of higher image intensity. Post-processing using ADC mapping produces parametric images with further contrast manipulation. Diffusion weighted MR pulse sequences are more accurate than conventional MRI pulse sequences in identifying the occurrence of acute stroke during the first 24 hours after onset of symptoms.
Device Story
Device consists of software post-processing techniques for existing, reconstructed MR images; includes ProPak for temporal datasets and ADC mapping for diffusion-weighted images. Operates on Picker MR systems or workstations; used by clinicians to analyze contrast changes or water molecule mobility. Inputs: MR images, single/multi-slice datasets. Processing: pixel-by-pixel or region-of-interest analysis; generates parametric images, time-intensity plots, or tabular data. Output: visual parametric maps and quantitative data. Assists clinicians in evaluating perfusion and identifying acute cerebral infarcts; provides supplemental diagnostic information without changing primary MR system indications.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Software-based post-processing for MR images. Features include semi-automated analysis, color parametric imaging, and time-intensity plotting. Operates on existing reconstructed MR datasets. No specific hardware materials or energy sources; software-only enhancement package.
Indications for Use
Indicated for patients undergoing MR imaging, specifically for visualizing contrast changes over time in temporal datasets (ProPak) and imaging diffusive mobility of water molecules in brain tissue to identify acute stroke (ADC Mapping).
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
AUG 27 1998
# K 980831
# SUMMARY OF SAFETY AND EFFECTIVENESS (As required by 21 CFR 807.92)
#### General Information 1.
| Classification: | Class II<br>Magnetic Resonance (MR) Diagnostic Device |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Magnetic Resonance (MR) Device Option |
| Proprietary Name: | Image Post-Processing Techniques |
| Establishment Registration: | Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
| Performance Standards: | No applicable performance standards have been<br>issued under section 514 of the Food, Drug and<br>Cosmetic Act. |
#### Intended Uses 2.
)
The intended use and indications for use for each of the types of Post-Processing Techniques described in this submission are described below.
### ProPak Techniques
The ProPak Package for Picker MR systems or workstations provide supplemental information regarding contrast changes over time for those images extracted from MR temporal datasets. The indications for use for the MR system remain unchanged.
## Apparent Diffusion Coefficient (ADC) Mapping
This submission adds the following sentence to the indications for use statement for Picker's Diffusion-Weighted MR Imaging Package.
Post-processing using ADC mapping produces parametric images with further contrast manipulation.
{1}------------------------------------------------
#### Device Description 3.
There are two basic types of post-processing techniques included in this submission. The first type is known as ProPak and is a set of techniques for processing temporally resolved image data sets and performing general perfusion analysis. The second type of technique is called ADC mapping and is for processing data from diffusion-weighted imaging sequences. All of the techniques included in this submission are for processing existing images that have already been reconstructed.
#### Safety and Effectiveness 4.
The Post-Processing Techniques described in this submission are substantially equivalent in technological characteristics and intended use to the GE Functool Option, the Philips Quantitative Analysis Package and the Picker Diffusion-Weighted Imaging Package. The following table has been compiled in order to demonstrate this substantial equivalence.
| Parameter | Image Post-Processing<br>Techniques | Predicate Devices<br>GE FuncTool Option (K960265),<br>Philips Quantitative Anal. (K971965),<br>Diffusion Imaging (K974530), |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compatibility | Same. | Available on Independent Workstations<br>or the MR System Operator's Console<br>(See K960265) |
| Inputs | • MR images<br>• Single or multi-slice datasets<br>with equally spaced time<br>intervals (ProPak)<br>• Diffusion-weighted images<br>(ADC Mapping) | • CT and MR images<br>• Single or multi-slice datasets with<br>equally spaced time intervals<br>(See K960265) |
| Features | • Semi-automated<br>• Color parametric images<br>(optional)<br>• Analysis on a pixel-by-pixel<br>basis or region of interest<br>• Time intensity information in<br>plot or tabular form | • Automated<br>• Color parametric images<br>• Overlay of parametric images on to<br>anatomical reference images<br>• Analysis on a pixel-by-pixel basis or<br>region of interest<br>• Time intensity information in plot or<br>tabular form<br>(See K960265 and K971965) |
# SUBSTANTIAL EQUIVALENCE CHART
PICKER INTERNATIONAL, INC.
(POSTPRO)
3/3/98
{2}------------------------------------------------
| Parameter | Image Post-Processing<br>Techniques | Predicate Devices<br>GE FuncTool Option (K960265),<br>Philips Quantitative Anal. (K971965),<br>Diffusion Imaging (K974530), |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Filming and Archiving | Same. | Images can be filmed, stored in image<br>database or archived with rest of patient<br>exam. (See K960265) |
| Anatomy of Interest | Same. | Brain and Body Imaging<br>(See K960265) |
| Processing Algorithms | • Area Mapping<br>• Time to Peak Mapping<br>• ADC Mapping | • Negative Enhancement Integral<br>• Mean Time to Enhance<br>(See K960265) |
| Indications for Use | ProPak Techniques<br>The ProPak Package for Picker<br>MR systems or workstations<br>provide supplemental<br>information regarding contrast<br>changes over time for those<br>images extracted from MR<br>temporal datasets. The<br>indications for use for the MR<br>system remain unchanged. | FuncTools (K960265)<br>The FuncTool option to the Advantages<br>Windows workstation is a software<br>module that provides supplemental<br>information to those images extracted<br>from CT and MR temporal datasets.<br>Quantitative Analysis (K971965)<br>The CT/MR Quantitative Analysis<br>Package is intended for use where<br>visualization and analysis of CT and MR<br>dynamic studies, showing changes in<br>contrast over time, are useful or<br>necessary. |
ﺰ
/
, ジ
(POSTPRO)
{3}------------------------------------------------
| Parameter | Image Post-Processing<br>Techniques | Predicate Devices<br>GE FuncTool Option (K960265),<br>Philips Quantitative Anal. (K971965),<br>Diffusion Imaging (K974530), |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(cont.) | ADC Mapping<br>Same as Diffusion Imaging with<br>the addition of the following<br>sentence: Post-processing using<br>ADC mapping produces<br>parametric images with further<br>contrast manipulation. | Diffusion Imaging (K974530)<br>The Picker Diffusion-Weighted MR<br>Imaging Package has been designed to<br>image the diffusive mobility of water or<br>other proton-containing molecules. One<br>important clinical application is to<br>visualize the apparent loss of mobility of<br>water molecules in brain tissue affected<br>by acute stroke. Areas of decreased<br>diffusion, as is observed in acute<br>cerebral infarcts, appear as areas of<br>higher image intensity. Diffusion<br>weighted MR pulse sequences are more<br>accurate than conventional MRI pulse<br>sequences in identifying the occurrence<br>of acute stroke during the first 24 hours<br>after onset of symptoms. |
PICKER INTERNATIONAL, INC.
)
(POSTPRO)
3/3/98
I - 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a series of flowing lines, possibly representing people or services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 27 1998
Elaine K. Keeler, Ph.D. Manager, Clinical Science Picker International, Inc. 595 Miner Road Cleveland, OH 44143
K980831 Re:
Image Post-Processing Techniques (ProPak and Apparent Diffusion Coefficient Mapping) Dated: June 8, 1998 Received: June 9, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH
### Dear Mr. Keeler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdch/dsmaldsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K980831 510(k) Number (if known): _
Image Post-Processing Techniques Device Name:
### Indications for Use:
The intended use and indications for use for each of the types of Post-Processing Techniques described in this submission are described below.
### ProPak Techniques
The ProPak Package for Picker MR systems or workstations provide supplemental information regarding contrast changes over time for those images extracted from MR temporal datasets. The indications for use for the MR system remain unchanged.
# Apparent Diffusion Coefficient Mapping
The Picker Diffusion-Weighted MR Imaging Package has been designed to image the diffusive mobility of water or other proton-containing molecules. One important clinical application is to visualize the apparent loss of mobility of water molecules in brain tissue affected by acute stroke. Areas of decreased diffusion, as is observed in acute cerebral infarcts, appear as areas of higher image intensity. Post-processing using ADC mapping produces parametric images with further contrast manipulation.
Diffusion weighted MR pulse sequences are more accurate than conventional MRI pulse sequences in identifying the occurrence of acute stroke during the first 24 hours after onset of symptoms.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K980831
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.