K980811 · Quest Intl., Inc. · LKP · May 12, 1998 · Immunology
Device Facts
Record ID
K980811
Device Name
SERAQUEST ANTI-SM/RNP
Applicant
Quest Intl., Inc.
Product Code
LKP · Immunology
Decision Date
May 12, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
For the qualitative and semi-quantitative detection of human antibodies to Sm/RNP nuclear antigen human serum by enzyme immunoassay. For use as an aid in the diagnosis of systemic rheumatic disease, particularly mixed connective tissue disease. For manual use, or for use with the HyPrep System Plus. For In Vitro Diagnostic Use Only.
Device Story
SeraQuest Anti-Sm/RNP is a solid-phase enzyme immunoassay (EIA) for detecting IgG antibodies against Sm/RNP nuclear antigen in human serum. Input: human serum samples. Principle: samples incubated in microwells coated with Sm/RNP antigen; immobilized antibodies detected via enzyme-labeled goat anti-human IgG conjugate; substrate added to produce yellow end-product. Output: photometric reading at 405 nm converted to Index values. Used in clinical laboratories; manual operation or semi-automated via HyPrep System Plus. Healthcare providers use results as an aid in diagnosing systemic rheumatic disease, specifically mixed connective tissue disease. Benefits: provides standardized immunological testing for specific autoantibodies.
Clinical Evidence
Clinical study compared SeraQuest Anti-Sm/RNP against the Shield Diastat Anti-Sm/RNP predicate using 252 serum specimens. Results showed 100% relative sensitivity (95% CI: 99.9-100%) and 90.4% relative specificity (95% CI: 86.4-94.4%), with an overall agreement of 95.4% (95% CI: 88.6-95.3%).
Technological Characteristics
Solid-phase enzyme immunoassay (EIA) using plastic microwells coated with calf thymus-derived Sm/RNP antigen. Employs goat anti-human IgG conjugated with alkaline phosphatase. Detection via p-Nitrophenyl phosphate substrate and photometric readout at 405 nm. Manual or semi-automated operation via HyPrep System Plus.
Indications for Use
Indicated for the qualitative and semi-quantitative detection of IgG antibodies to Sm/RNP in human serum as an aid in the diagnosis of systemic rheumatic disease, particularly mixed connective tissue disease (MCTD). For in vitro diagnostic use.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
Shield Diastat Anti-Sm/RNP test (Shield Diagnostics)
Submission Summary (Full Text)
{0}------------------------------------------------
K980811
APPENDIX 3 (revised 4/24/98)
Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 124
MAY 1 2 1998
510(k) SUMMARY
| Applicant: | Quest International, Inc.<br>1938 N.E. 148th Terrace<br>North Miami, FL 33181 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------|
| Registration No. | 1061839 |
| Contact Person: | Robert A. Cort, V.P., Quality Assurance |
| Telephone: | (305) 948-8788 |
| Telefax: | (305) 948-4876 |
| Manufacturing Site: | Same as above |
| Device: | SeraQuest® Anti-Sm/RNP |
| Device Name: | Antinuclear antibody immunological test system (21CFR § 866.5100)<br>Device Classification: Class II (performance standards) |
#### >scription:
The SeraQuest Anti-Sm/RNP test is a solid-phase enzyme immunoassay (EIA), which is performed in microwells, at room temperature, in three thirty minute incubations. It has been developed to detect IgG antibodies which are directed against Sm/RNP nuclear antigen (Anti-Sm/RNP), in human serum.
The Calibrators in the SeraQuest Anti-Sm/RNP test set have been assigned Index values based on an in-house standard. Test results are reported as Index values.
### Principle:
Diluted samples are incubated in wells coated with Sm/RNP antigen. Antibodies against Sm/RNP (if present) are immobilized in the wells. Residual sample is eliminated by washing and conjugate (enzyme-labeled antibodies to human IgG) is added and incubated. If IgG antibodies to Sm/RNP are present, the conjugate will be immobilized in the wells. Residual conjugate is eliminated by washing and draining, and the enzyme substrate is added and incubated. In the presence of the enzyme, the substrate is converted to a yellow end-product which is read photometrically at 405 nm.
### Intended Use:
For the qualitative and semi-quantitative detection of human antibodies to Sm/RNP nuclear antigen human serum by enzyme immunoassay. For use as an aid in the diagnosis of systemic rheumatic
{1}------------------------------------------------
APPENDIX 3 (revised 4/24/98)
sease, particularly mixed connective tissue disease. For manual use, or for use with the HyPrep رو. ystem Plus. For In Vitro Diagnostic Use Only.
## Predicate Device:
----
The SeraQuest Anti-Sm/RNP test is substantially equivalent in intended use and performance, to the Shield Diastat Anti-Sm/RNP test, Shield Diagnostics, Dundee, U.K.
#### Summary of technological characteristics:
| Characteristic | SeraQuest Anti-Sm/RNP | Shield Diastat Anti-Sm/RNP |
|-----------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|
| Description: | Enzyme Immunoassay | Enzyme Immunoassay |
| Intended Use: | The detection of<br>antibodies against Sm/RNP<br>in human serum. | The detection of<br>antibodies against Sm/RNP<br>in human serum. |
| Antigen Source: | Calf Thymus | Calf Thymus |
| Solid Phase: | Plastic Microwell | Plastic Microwell |
| Number of Incubation Periods: | Three | Three |
| Sample Dilution: | 1:50 | 1:500 |
| Sample Incubation<br>Duration: | 30 minutes | 60 minutes |
| Incubation Temperature: | Room temperature | Room temperature |
| Enzyme-labeled Conjugate: | | |
| Antibody | Goat anti-human IgG | Goat anti-human IgG |
| Enzyme | Alkaline phosphatase | Alkaline phosphatase |
| Conjugate Volume: | 100 μl | 100 μl |
| Conjugate Incubation<br>Duration: | 30 minutes | 30 minutes |
| Substrate: | p-Nitrophenyl<br>phosphate | phenolthalein<br>monophosphate |
| Substrate Volume: | 100 μl | 100 μl |
| ubstrate Incubation | 30 minutes | 30 minutes |
| Stop Reagent: | 0.5 M Trisodium<br>phosphate | Sodium<br>Hydroxide |
| Stop Reagent Volume: | 100 µl | 100 µl |
| Readout: | Spectrophotometric<br>405 nm | Spectrophotometric<br>540-565 nm |
{2}------------------------------------------------
APPENDIX 3 (revised 4/24/98)
Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 126
Summary of Clinical Testing:
Of the 252 specimens tested, 40 were positive, and 189 were negative in both the SeraQuest and Shield tests (please see Table C-3). Of the remaining specimens, 23 specimens which were negative by the Shield test, 20 were positive and 3 equivocal by the SeraQuest test. Please see Table 1 below.
#### TABLE 1.
Sm/Anti-RNP SHIELD ASSAYS AND RESULTS OF SeraQuest Anti-Sm/RNP ASSAYS OF 252 SERUM SPECIMENS.
| SeraQuest Anti-Sm/RNP | | | | | |
|-----------------------|----------|-----------|----------|-------------------------------------------------------------|------------------------------|
| Shield<br>Anti-Sm/RNP | Positive | Equivocal | Negative | % | 95 % CIv |
| Positive | 4 0 | 0 | 0 | Relative sensitivity<br>100 | 99.9 to 100 |
| Negative | 2 0 | 3 | 189 | Relative specificity*<br>Overall agreement*<br>90.4<br>95.4 | 86.4 to 94.4<br>88.6 to 95.3 |
· Excluding equivocal results.
* Calculated by the normal method.
Reference: Gardner, M.J. and Altman, D.G., Confidence Intervals Rather Than Hypothesis Testing. Brit. Med. J., 292: 746-750, 1986.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads facing to the right. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY | 2 1998
Robert A. Cort Vice President, Ouality Assurance QUEST INTERNATIONAL, INC. 1938 N.E. 148th Terrace North Miami, FL 33181
Re: K980811 Trade Name: SeraQuest® Anti-Sm/RNP Regulatory Class: II Product Code: LKP Dated: April 28, 1998 Received: April 28, 1998
Dear Mr. Cort:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597,or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Quest International, Inc., 1938 N.E. 148th Terrace, N. Miami, FL 33181 Page No. 154
APPENDIX 8
Page _ 1 of _ し
510(k) Number (if known): K 980 & //
Device Name: SeraQuest Anti-Sm/RNP
Indications For Use:
- 1 . For in vitro diagnostic use only.
- 2. For the qualitative and semi-quantitative detection of IgG antibodies to Sm/RNP in human serum by enzyme immunoassay.
- For use as an aid in the diagnosis of systemic rheumatic disease, particularly 3. mixed connective tissue disease (MCTD).
- For manual use, or for use with the HyPrep System Plus semi-automated fluid 4. handler.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices | |
| 510(k) Number | L990811 |
| Prescription Use | <div style="text-align:center;">✓</div> |
|----------------------|-----------------------------------------|
| (Per 21 CFR 801.109) | |
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.