← Product Code [GCJ](/productcode/GCJ) · K980787

# HERMES OPERATING ROOM CONTROL CENTER WITH THE WOM 20L INSUFFLATOR (K980787)

_Computer Motion, Inc. · GCJ · Jul 13, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K980787

## Device Facts

- **Applicant:** Computer Motion, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jul 13, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver and WOM 20L Insufflator. It can be used in laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy where a general laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic symph node dissection, laparoscopically assisted hysterectomy, laparoscopic & thorascopic anterior spinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting, where endoscopic visualization in indicated and examination of the evacuated cardiac chamber during performance of valve replacement. The users of the HERMES ORCC are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

## Device Story

HERMES OR Control Center is a robotic, computer-driven system for centralized control of surgical equipment; enables surgeon to adjust settings of Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver, and WOM 20L Insufflator via voice control. Eliminates need for manual interface manipulation or verbal communication with OR staff. Used in operating rooms by various surgical specialists. System simplifies equipment management, potentially reducing procedure time and improving surgical workflow efficiency.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Robotic computer-driven control system. Complies with IEC 601-1 (1988), IEC 601-1 Amendment 1 (1991), IEC 601-2-18 (1990), UL 2601-1, and CAN/CSA-C22.2 standards. Features voice-control interface for integration with specific endoscopic surgical equipment. Operates as a centralized control hub in the operating room environment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- HERMES Operating Room Control Center and Accessories ([K973700](/device/K973700.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUL 13 1998

K980787

## 16. Premarket Notification 510(k) Safety and Effectiveness Summary

## HERMES OR Control Center System 510(k) Summary

In accordance with 21 CFR section 807.92 Computer Motion is submitting the following safety and effectiveness summary.

1) Submitter Information

Computer Motion. Inc. 130-B Cremona Drive Goleta, CA 93117

Contact: David Thomas Prepared: February 26, 1998

- 2) Name of Device:
Proprietary Name: HERMES Operating Room Control Center and Accessories Common Name is HERMES Classification Name: Laproscope for Use in General and Plastic Surgery, Regulation Number 876.1500, Class II.

- 3) Substantially equivalent to HERMES Operating Room Control Center and Accessories K973700.
- 4) The HERMES Operating Room Control Center is a robotic computer-driven system whose basic function is offer the additional option for surgeon selection of attachment device parameter settings utilizing voice control.

The intent of the HERMES OR Control Center is to allow for simplified and more direct control of medical device settings by the physician, thereby eliminating the necessity of using the various interfaces existing on the Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver and WOM 20L Insufflator in the Operating setting, or relying upon verbal communications between the surgeon and other personnel in the operation room in order to adjust surgical equipment.

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver and WOM 20L Insufflator. It can be used in laparoscopy, nasopharyngoscopy, ear endoscopy, and sinuscopy where a general laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic surgeries are laparoscopic cholecystectomy, laparoscopic hernia repair, laparoscopic appendectomy, laparoscopic pelvic symph node dissection, laparoscopically assisted hysterectomy, laparoscopic & thorascopic anterior spinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting, where endoscopic visualization in indicated and examination of the evacuated cardiac chamber during

{1}------------------------------------------------

performance of valve replacement. The users of the HERMES ORCC are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists

- 5) The HERMES OR Control Center is designed and tested to the following Computer Motion and voluntary standards.
IEC 601-1 Second Edition 1988 International Standard for Medical Electrical Equipment IEC 601-1 Amendment 1 1991 International Standard for Medical Electrical Equipment IEC 601-2-18 First Edition 1990 International Standard for Medical Electrical Equipment UL 2601-1

EMC Directive European Union 89/336/EEC CAN/CSA-C22.2 NO. 601.1-M90 & NO. 601.2.18-92

HERMES System Functional Test Requirements

、イ

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 3 1998

Mr. Douglas P. Bueschel Director. Regulatory Affairs/Quality Assurance Computer Motion, Inc. 130 Cremona Drive, Suite B Goleta, California 93117

K980787 Re: Trade Name: Hermes Operating Room Control Center Regulatory Class: II Product Code: GCJ Dated: June 2, 1998 Received: June 3, 1998

## Dear Mr. Bueschel:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

## Page 2 - Mr. Douglas P. Bueschel

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

| 510(k) Number (if known): | K980787                  |
|---------------------------|--------------------------|
| Device Name:              | Hermes OR Control Center |

Indications For Use:

The HERMES OR Control Center is indicated for use with Stryker Endoscopy 882 Camera, Quantum 5000 Light Source, Stryker SE5 Shaver and WOM 20L Insufflator. It can be used in where nasopharyngoscopy, ear endoscopy, and sinuscopy a laparoscopy, general laparoscope/endoscope is indicated for use. A few examples of the more common endoscopic repair, laparoscopic cholecystectomy, laparoscopic hernia laparoscopic surgeries are ymph node dissection, laparoscopically assisted laparoscopic pelvic appendectomy, hysterectomy, laparoscopic & thorascopic anterior spinal fusion, decompression fixation, wedge resection, lung biopsy, pleural biopsy, dorsal sympathectomy, pleurodesis, internal mammary artery dissection for coronary artery bypass, coronary artery bypass grafting where endoscopic visualization is indicated and examination of the evacuated cardiac chamber during performance of valve replacement. The users of the HERMES ORCC are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |         |
|--------------------------------------------------------|---------|
| (Division Sign-Off)                                    |         |
| Division of General Restorative Device                 |         |
| 510(k) Number                                          | K980787 |

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K980787](https://fda.innolitics.com/device/K980787)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com