CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720, 721 AND 722
K980670 · Cryomedical Sciences, Inc. · GEH · Jul 17, 1998 · General, Plastic Surgery
Device Facts
Record ID
K980670
Device Name
CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720, 721 AND 722
Applicant
Cryomedical Sciences, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Jul 17, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CMS Blizzard 700 Series Models 710, 711, 712, 720, 721and 722, depending on the cryogen selected, are intended to be used: Liquid Nitrogen: - as a cryosuraical tool for destruction of unwanted tissue in the fields . of dermatology, general surgery, neurosurgery, thoracic surgery, urology, proctology, oncology, gynecology and ENT; - to ablate prostatic tissue; . - for the ablation of prostatic tissue in cases of prostate cancer and . benian prostatic hyperplasia; - for ablation of cancerous or malignant tissue; . - for ablation of benign tumors; . - for palliative intervention; . - for ablation or freezing of skin cancers and other cutaneous . disorders; - for the ablation of malignant neoplasia or benign dysplasia of the . female genitalia; - for ablation of leukoplakia of mouth, angiomas, sebaceous . hyperplasia, basal cell tumors of the eyelid and canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, and fissures, peri-anal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemanaiomas, recurrent cancerous lesions; - for the destruction of warts or lesions: ● - for the palliation of tumors of the oral cavity, rectum, and skin;. . - for ablation of arrhythmic cardiac tissue; ● - . for the ablation of benign or malignant growths of the anus and rectum; - for the ablation of hemorrhoids; ● Prescription Use (Per 21 CFR 801.109) Nitrous Oxide: - as a tool in the destruction of unwanted tissue in the fields of . dermatology, gynecology, general surgery, and urology; Carbon Dioxide: - as a tool in the destruction of unwanted tissue in the fields of . dermatology, gynecology, and general surgery; Argon and Krypton: - as a cryosurgical tool in general surgery, dermatology, neurology, . thoracic surgery, ENT, gynecology, oncology, proctology, and urology for the ablation of tissue including liver metastases, skin lesions, warts and prostate tissue; Prescription Use (Per 21 CFR 801.109)
Device Story
Portable cryosurgical unit; utilizes liquid nitrogen, nitrous oxide, carbon dioxide, argon, or krypton; cryogen forced under pressure through interchangeable cryoprobes; probe placed at target site to freeze/destroy unwanted tissue; operated by physicians in clinical settings; available in wheeled pole, tabletop, or suitcase configurations; single or dual cryoport capacity; controlled via console switch or foot pedal; provides targeted tissue ablation for various surgical specialties; benefits patients through minimally invasive destruction of lesions, tumors, and diseased tissue.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Cryosurgical unit; supports liquid nitrogen, CO2, nitrous oxide, argon, krypton; interchangeable cryoprobes; single or dual cryoport configurations; console or foot pedal control; portable (wheeled, tabletop, or suitcase form factors); external cryogen source or packet; mechanical operation.
Indications for Use
Indicated for destruction of unwanted tissue in dermatology, general surgery, neurosurgery, thoracic surgery, urology, proctology, oncology, gynecology, and ENT. Specific applications include ablation of prostatic tissue (cancer/BPH), skin cancers, benign tumors, malignant neoplasia, benign dysplasia, leukoplakia, angiomas, warts, hemorrhoids, and arrhythmic cardiac tissue. Prescription use only.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
CMS Cryo-lite™ Series
CMS AccuProbe® 600 Series
Cabot Medical Cryomedics MT700
Erbe Erbokryo CA
Submission Summary (Full Text)
{1}------------------------------------------------
K980670
# 510(k) SUMMARY
February 20, 1998
### Submitted by:
Cryomedical Sciences, Inc. 1300 Piccard Drive Suite L105 Rockville, Maryland 20850 (301) 417-7070 Fax (310) 417-7077
#### Contact:
Richard J. Reinhart, Ph.D. President and CEO or Susan Hayes Requiatory Affairs
Proprietary name: CMS Blizzard 700 Series Models 710, 711, 712, 720, 721 and 722
Common name: Cryosurgical units and accessories
Cryosurgical units with Liquid Nitrogen, Nitrous Oxide, Carbon Classification: Dioxide. Class II, under CFR § 878.4350. Under CFR § 882.4250 a cryogenic surgical device also is classified in Class II.
The CMS Blizzard 700 Series Models 710, 711, 712, 720, 721 and 722 are simple, portable cryosurgical devices. All models feature physician-selected cryogen choices as well as a selection of cryoprobes of various sizes and shapes. Models 710, 711 and 712 are single cryoport devices while models 720, 721 and 722 have a two cryoport capacity. Although all models operate in the exact same way, their casings differ to allow desired positioning. The 710 and 720 are wheeled pole type configuration; the 711 and the 721 are tabletop configuration; while the 712 and the 722 are suitcase or attaché type configuration.
The CMS Blizzard 700 Series Models 710, 711, 712, 720, 721 and 722 are similar in form and function to CMS's own Cryo-lite™ System, and its AccuProbe® 600 Series, as well as Cabot Medical's Cryomedics MT700 and the Erbe Erbokryo CA. For comparisons, please see below.
Page 1 of 5
{2}------------------------------------------------
The devices are used to destroy unwanted tissue by the application of cold to the selected site. The cryogen is forced through the cryoprobe under pressure. The cryoprobe having been placed in the appropriate position, then becomes cold and freezes the tissue.
# COMPARISON OF FEATURES BETWEEN THE CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720, 721 AND 722 AND THE CMS CRYO-LITE™ SERIES .
| FEATURES | Cryo-liteTM Series<br>(Tab G) | Blizzard 700 Series<br>Models 710, 711, 712,<br>720, 721and 722 |
|-------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Cryogens | Choice of liquid nitrogen,<br>carbon dioxide, nitrous oxide<br>and argon which may be<br>used sequentially | Choice of liquid nitrogen,<br>carbon dioxide, nitrous oxide<br>argon or krypton which may<br>be used sequentially |
| Number of<br>Cryoprobes | One cryoprobe | Models 710, 711 and 712<br>have a one cryoprobe<br>capacity while models 720,<br>721 and 722 have a two<br>cryoprobe capacity |
| Cryogen Containers | Cryogen packet and/or<br>transfer line to larger cryogen<br>source | Cryogen packet and/or<br>transfer line to external<br>cryogen source |
| Size | Portable and Hand-held | Portable,- Table-top or<br>wheeled stand |
| Cryoprobes Types | Various, interchangeable | Various, interchangeable |
Page 2 of 5
{3}------------------------------------------------
# COMPARISON OF FEATURES BETWEEN THE CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720, 721 AND722 AND THE CMS ACCUPROBE® 600 SERIES
| FEATURES | AccuProbe® Series<br>(Tab. H) | Blizzard 700 Series<br>Models 710, 711, 712,<br>720, 721and 722 |
|-------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Cryogens | Liquid nitrogen | Choice of liquid nitrogen, carbon dioxide, nitrous oxide argon or krypton which may be used sequentially |
| Number of<br>Cryoprobes | Depending on the model, from one to eight cryoprobe ports | Models 710, 711 and 712 have a one cryoprobe capacity while models 720, 721 and 722 have a two cryoprobe capacity |
| Cryogen Containers | Internal cryogen container | Cryogen packet and/or transfer line to external cryogen source |
| Size | Portable | Portable, - Table-top or wheeled stand |
| Cryoprobes Types | Various, interchangeable | Various, interchangeable |
.
Page 3 of 5
{4}------------------------------------------------
# COMPARISON OF FEATURES BETWEEN THE CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720 ,721 AND 722 AND THE CABOT MEDICAL CYROMEDICS MT700
| FEATURES | Cryomedics MT700<br>(Tab I) | Blizzard 700 Series<br>Models 710, 711, 712,<br>720, 721 and 722 |
|-------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Cryogens | Choice of nitrous oxide and<br>carbon dioxide | Choice of liquid nitrogen,<br>carbon dioxide, nitrous<br>oxide, argon, or krypton |
| Number of<br>Cryoprobes | One probe | Models 710, 711 and 712<br>have a one cryoprobe<br>capacity while models 720,<br>721 and 722 have a two<br>cryoprobe capacity |
| Cryogen Containers | Commercial Cryogen<br>Container with Transfer Line<br>to Probe | Cryogen packet and/or<br>Transfer Line to external<br>cryogen source |
| Size | Portable - Stand with wheels<br>and hand-held probe | Portable,- Table-top or<br>wheeled stand with hand-<br>held cryoprobe |
| Cryoprobes Types | Various, interchangeable | Various, interchangeable |
:
Page 4 of 5
{5}------------------------------------------------
# COMPARISON OF FEATURES BETWEEN THE CMS BLIZZARD 700 SERIES MODELS 710, 711, 712, 720 , 721 AND 722 AND THE ERBE ERBOKRYO CA
.
| FEATURES | Erbokryo CA<br>(Tab J) | Blizzard 700 Series<br>Models 710, 711, 712,<br>720, 721 and 722 |
|-------------------------|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Cryogens | Choice of nitrous oxide and<br>carbon dioxide | Choice of liquid nitrogen,<br>carbon dioxide, nitrous<br>oxide, argon, or krypton |
| Number of<br>Cryoprobes | One probe | Models 710, 711 and 712<br>have a one cryoprobe<br>capacity while models 720,<br>721 and 722 have a two<br>cryoprobe capacity |
| Cryogen Containers | Commercial cryogen<br>container with transfer line to<br>probe | Cryogen packet and/or<br>transfer line to larger cryogen<br>source |
| Size | Portable - Stand with wheels<br>and hand-held Probe | Portable - Table-top or<br>wheeled stand with and-<br>held cryoprobe |
| Cryoprobes Types | Various, Interchangeable | Various, Interchangeable |
| Control | Foot switch | Console unit switch or foot<br>pedal |
.
・
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The eagle is composed of three thick, curved lines that suggest the shape of a bird in flight. The text is in all caps and is evenly spaced around the circle.
FEB 2 1 2008
Food and Druq Administration 9200 Corporate Boulevard Rockville MD 20850
Cryomedical Sciences Inc. c/o Mr. Richard J. Reinhart, Ph.D. President and CEO 1300 Piccard Drive, Suite L105 Rockville, MD 20850
Re: K980670 Trade Name: CMS Blizzard 700 Series Models 710, 711, 712, 720, 721, 722 Regulatory Class: II (two) Product Code: OCL, GEH Dated: June 2, 1998 Received: June 2, 1998
Dear Dr. Reinhart
This letter corrects our substantially equivalent letter of July 17 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Richard J. Reinhart, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
ell. Mall.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
The CMS Blizzard 700 Series Models 710, 711, 712, 720, 721and 722, depending on the cryogen selected, are intended to be used:
#### Liquid Nitrogen:
- as a cryosuraical tool for destruction of unwanted tissue in the fields . of dermatology, general surgery, neurosurgery, thoracic surgery, urology, proctology, oncology, gynecology and ENT;
- to ablate prostatic tissue; .
- for the ablation of prostatic tissue in cases of prostate cancer and . benian prostatic hyperplasia;
- for ablation of cancerous or malignant tissue; .
- for ablation of benign tumors; .
- for palliative intervention; .
- for ablation or freezing of skin cancers and other cutaneous . disorders;
- for the ablation of malignant neoplasia or benign dysplasia of the . female genitalia;
- for ablation of leukoplakia of mouth, angiomas, sebaceous . hyperplasia, basal cell tumors of the eyelid and canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, and fissures, peri-anal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemanaiomas, recurrent cancerous lesions;
- for the destruction of warts or lesions: ●
- for the palliation of tumors of the oral cavity, rectum, and skin;. .
- for ablation of arrhythmic cardiac tissue; ●
- . for the ablation of benign or malignant growths of the anus and rectum;
- for the ablation of hemorrhoids; ●
**Prescription Use**
**(Per 21 CFR 801.109)**
pc
Division Sian-Off) of General Restorative Device 510(k) Number
{9}------------------------------------------------
# Indications for Use (continued)
### Nitrous Oxide:
- as a tool in the destruction of unwanted tissue in the fields of . dermatology, gynecology, general surgery, and urology;
### Carbon Dioxide:
- as a tool in the destruction of unwanted tissue in the fields of . dermatology, gynecology, and general surgery;
### Argon and Krypton:
- as a cryosurgical tool in general surgery, dermatology, neurology, . thoracic surgery, ENT, gynecology, oncology, proctology, and urology for the ablation of tissue including liver metastases, skin lesions, warts and prostate tissue;
**Prescription Use**
(Per 21 CFR 801.109)
cale
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number ***_***__ 196067
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.