← Product Code [JAQ](/productcode/JAQ) · K980601

# VARISOURCE WRIGHT VAGINAL CUFF APPLICATOR FOR VARIAN VARISOURCE REMOTE HIGH DOSE RATE AFTERIOADER (K980601)

_Varian Assoc., Inc. · JAQ · Oct 21, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K980601

## Device Facts

- **Applicant:** Varian Assoc., Inc.
- **Product Code:** [JAQ](/productcode/JAQ.md)
- **Decision Date:** Oct 21, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5700
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Varian VariSource™ Remote High Dose Rate Afterloader [system, including applicators and accessories] is a device intended to be used by properly trained and licensed medical personnel to provide radiation brachytherapy. The VariSource Wright Vaginal Cuff applicator which is the subject of this 510(k) is a component of the VariSource system.

## Device Story

The VariSource Wright Vaginal Cuff Applicator is a component of the VariSource Remote High Dose Rate (HDR) Afterloader system. It functions as a delivery interface for radiation brachytherapy. The device consists of a vaginal cylinder (polysulfone shell with solid water filling) and four stainless steel irradiation tubes arranged in a dome-shaped configuration. It is used in clinical settings by trained medical personnel to deliver high-activity radionuclide sources to internal body locations via remote control. The applicator facilitates precise source positioning, allowing clinicians to administer targeted radiation therapy for cancer treatment. By enabling remote afterloading, the system reduces radiation exposure to medical staff compared to manual methods.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Vaginal cylinder: 3.4 cm x 12.0 cm, polysulfone shell with solid water filling. Irradiation tubes: 4 stainless steel tubes in dome configuration. System: Remote HDR afterloader component. Connectivity: Coupling catheter fittings for remote source delivery.

## Regulatory Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

## Predicate Devices

- Mick Radio-Nuclear Instruments, Inc., Wang Front Loading Applicator ([K890485](/device/K890485.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 | 1998

Attachment 3 Page 1 of 2

varian@ oncology systems

## 510(k) SUMMARY AS REQUIRED BY 21 CFR 807.92

K980601

- 1. Submitter: Varian Oncology Systems 3045 Hanover Street Palo Alto, CA 94304
Linda S. Nash. Manager Contact: Regulatory Compliance & Radiation Safety Phone (650) 424-6990 FAX (650) 424-4830 linda.nash@os.varian.com

Prepared: February 12, 1998 July 23, 1998 Revised:

- VariSource Wright Vaginal Cuff Applicator for Varian 2. Device Name: VariSource™ Remote High Dose Rate Afterloader.
- 3. Predicate Device: Mick Radio-Nuclear Instruments, Inc., Wang Front Loading Applicator, K890485.
- Applicators for the Varian VariSource Remote High Dose Rate 4. Description: Afterloader are a part of a remote controlled radionuclide applicator system, including an electromechanical device to enable an operator to apply, by remote control, a radionuclide source of high activity at various internal or surface body locations for radiation brachytherapy. The shape and materials of the applicator determine where it will be utilized for treatment.
- The Varian VariSource Remote High Dose Rate 5. Intended Use: Afterloader [system, including applicators and accessories] is a device intended to be used by properly trained and licensed medical personnel to provide radiation brachytherapy The VariSource Wright Vaginal Cuff Applicator which is the subject of this 510(k) is a component of the VariSource system.
- 6. Technological See attached comparison chart. Characteristics

Image /page/0/Picture/13 description: The image shows the logo for "Partners for Life". The word "Partners" is stacked on top of the word "For", which is stacked on top of the word "Life". The word "Life" is written in a cursive font.

{1}------------------------------------------------

## Comparison to Predicate Device

| # | Feature                       | MRNI Wang Front<br>Loading Applicator,<br>K89045                                  | VariSource Wright<br>Vaginal Cuff Applicator<br>K980601                                               |
|---|-------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 1 | Afterloading Method           | Manual                                                                            | Remote HDR                                                                                            |
| 2 | Coupling Catheter<br>Fittings | No                                                                                | Yes                                                                                                   |
| 3 | Vaginal Cylinder              |                                                                                   |                                                                                                       |
|   | Diameter and Length           | 3.0 cm X 7.0cm                                                                    | 3.4 cm X 12.0 cm                                                                                      |
|   | Material                      | Polysulfone                                                                       | Solid Water inside<br>Polysulfone Shell                                                               |
| 4 | Irradiation Tubes             | 2                                                                                 | 4                                                                                                     |
|   | Material                      | Stainless steel                                                                   | Stainless Steel                                                                                       |
|   | Configuration                 | 1 central and 1 lateral<br>with source positioned<br>perpendicular to<br>central. | 4 central at 0, 90, 180,<br>and 270 degrees. Ends<br>come together to form<br>dome (eggbeater) shape. |
|   |                               |                                                                                   |                                                                                                       |

{2}------------------------------------------------

Image /page/2/Picture/16 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. The eagle is facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 1 1998

Linda Nash Regulatory Compliance and Radiation Safety Manager Varian Associates, Inc. 3045 Hanover Street Palo Alto. CA 94304

Re:

K980601 VariSource Wright Cuff Applicator for VariSource HDR Afterloader Dated: June 24, 1998 Received: August 4, 1998 Regulatory class: II 21 CFR 892.5700/Procode: 90 JAQ

Dear Ms. Nash:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 80) and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrb/dsmadsmam.html".

Sincerely yours,

Kilian Yin

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## varian@

## STATEMENT of INDICATIONS for USE*

I state in my capacity as Manager, Regulatory Compliance and Safety, of Varian Oncology Systems that the Product which is the subject of this premarket notification, is intended to be used for the following:

> The Varian VariSource™ Remote High Dose Rate Afterloader [system, including applicators and accessories] is a device intended to be used by properly trained and licensed medical personnel to provide radiation brachytherapy. The VariSource Wright Vaginal Cuff applicator which is the subject of this 510(k) is a component of the VariSource system.

Charles Howell

Charles H. Will, Manager Regulatory Compliance & Safety

February 12, 1998

Date

*Suggested language and format to meet the requirements of section 513(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and 21 CFR sections 801.4 and 809.92(a)(5).

K98060/

510(k) Number

William Yin

Division Sign-off Office of Device Evaluation

> Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

---

**Source:** [https://fda.innolitics.com/device/K980601](https://fda.innolitics.com/device/K980601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
