← Product Code LZU · K980437

# SILICONE PUNCTAL PLUG (K980437)

_Oasis Medical, Inc. · LZU · Apr 30, 1998 · OP · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K980437

## Device Facts

- **Applicant:** Oasis Medical, Inc.
- **Product Code:** LZU
- **Decision Date:** Apr 30, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** OP
- **Attributes:** Therapeutic

## Intended Use

The OASIS® Medical Silicone Punctal Plug is intended for use in patients experiencing dry eye symptoms such as redness, burning, reflex tearing, itching or foreign body sensations which can be relieved by blockage of the punctum Silicone Punctal Plug may be used in the treatment of dry eye syndrome and the dry eye components of various ocular surface diseases such as corneal ulcers, conjunctivitis, pyterygium, blepharitis, keratitis, red lid margins, recurrent chalazions, recurrent corneal erosion, filamentry keratitis and other external eye diseases. Other patients that may benefit are: cataract patients with arthritis, patients medicated for hypertension, contact wearers of any age, seasonal allergy sufferers, patients who live in dry climates, or spend extended periods in air conditioning, and any others who suffer from dry eye irritation. It is also reasonable that eye drops of many kinds would be more effective if retained on the surface of the eye, rather than drained into the sinus.

## Device Story

Silicone Punctal Plug is a medical-grade silicone device designed for insertion into the punctal opening to block tear drainage through the canaliculus. It is supplied premounted on an inserter. The device is used by clinicians to treat dry eye syndrome and associated ocular surface diseases by retaining tears and eye drops on the ocular surface. It provides relief from symptoms like redness, burning, and foreign body sensation. The device is available in five sizes (Micro, Mini, Petite, Small, Medium) to accommodate anatomical variations.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Medical grade silicone; five sizes ranging from 0.4 to 0.8 gauge; head diameters 0.6-0.9 mm; nose diameters 0.6-1.4 mm; lengths 1.1-2.0 mm. Mechanical insertion via provided inserter tool.

## Predicate Devices

- OASIS Medical Silicone Punctal Plug ([K945906](/device/K945906.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

980431/

APR 30 1998

February 2, 1998

# 510(k) SUMMARY

Submitted by:

James Christensen OASIS Medical, Inc. 514 South Vermont Avenue Glendora, CA 91741 (626) 914-2891 (Phone) (626) 914-2285 (Fax)

Device Name:

Common Name: Punctal Plug Proprietary Name: Silicone Punctal Plug

### Device Classification:

Lacrimal system plugs have not been officially classified, but have been historically been regulated through the 510(k) process.

#### Indication

The Silicone Punctal Plug is intended for use in patients experiencing dry eye symptoms such as redness, burning, reflex tearing, itching or foreign body sensations which can be relieved by blockage of the punctum

Silicone Punctal Plug may be used in the treatment of dry eve syndrome and the dry eye components of various ocular surface diseases such as corneal ulcers, conjunctivitis, pyterygium, blepharitis, keratitis, red lid margins, recurrent chalazions, recurrent corneal erosion, filamentry keratitis and other external eye diseases.

Other patients that may benefit are: cataract patients with arthritis, patients medicated for hypertension, contact wearers of any age, seasonal allergy sufferers, patients who live in dry climates, or spend extended periods in air conditioning, and any others who suffer from dry eye irritation. It is also reasonable that eye drops of many kinds would be more effective if retained on the surface of the eye, rather than drained into the sinus.

{1}------------------------------------------------

K980437

## Description:

The OASIS Medical Silicone Punctal Plug is designed to be inserted into the punctal opening in order to block tear drainage through the canaliculus. Each Punctal Plug is molded from medical grade silicone and comes premounted on an inserter. The plugs are available in the following sizes:

|        | GAUGE SIZE | HEAD DIA. | NOSE DIA. | LENGTH |
|--------|------------|-----------|-----------|--------|
| Micro  | 0.4        | 0.6 mm    | 0.6 mm    | 1.1 mm |
| Mini   | 0.5        | 0.7 mm    | 0.7 mm    | 1.3 mm |
| Petite | 0.6        | 0.9 mm    | 0.9 mm    | 1.4 mm |
| Small  | 0.7        | 0.9 mm    | 1.0 mm    | 1.8 mm |
| Medium | 0.8        | 0.9 mm    | 1.4 mm    | 2.0 mm |

## Substantial Equivalence Comparison:

The OASIS Medical Silicone Punctal Plug is substantially equivalent to the OASIS Medical Silicone Punctal Plug (K945906). The OASIS Medical Silicone Punctal Plug is identical to the predicate device except for its physical design and a modifieed inserter.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized design that resembles an eagle or a similar bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 3 0 1998

Mr. James Christensen Oasis Medical, Inc. 514 S. Vermont Avenue Glendora, CA 91741

Re: K980437 Trade Name: Soft Plug® Silicone Punctal Plug Regulatory Class: Unclassified Product Code: 86 LZU Dated: February 2, 1998 Received: February 4, 1998

Dear Mr. Christensen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. James Christensen

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): k 980437

Device Name: -----------------------------------Silicone Punctal Plug

Indications for Use:

The OASIS® Medical Silicone Punctal Plug is intended for use in patients experiencing dry eye symptoms such as redness, burning, reflex tearing, itching or foreign body sensations which can be relieved by blockage of the punctum

Silicone Punctal Plug may be used in the treatment of dry eye syndrome and the dry eye components of various ocular surface diseases such as corneal ulcers, conjunctivitis, pyterygium, blepharitis, keratitis, red lid margins, recurrent chalazions, recurrent corneal erosion, filamentry keratitis and other external eye diseases.

Other patients that may benefit are: cataract patients with arthritis, patients medicated for hypertension, contact wearers of any age, seasonal allergy sufferers, patients who live in dry climates, or spend extended periods in air conditioning, and any others who suffer from dry eye irritation. It is also reasonable that eye drops of many kinds would be more effective if retained on the surface of the eye, rather than drained into the sinus.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Claudine V. Knoegyh for DC
(Division Sign-Off)

Division of Ophthalmic Devices
510(k) Number K980437

Prescription Use X (Per 21 CFR 801.109)

Over-The-Counter Use

OR

---

**Source:** [https://fda.innolitics.com/device/K980437](https://fda.innolitics.com/device/K980437)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
