← Product Code [GEX](/productcode/GEX) · K980389

# MEGABEAM REUSABLE FIBER OPTIC HANDPIECE AND NEEDLES (K980389)

_Ceram Optec, Inc. · GEX · Jul 29, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K980389

## Device Facts

- **Applicant:** Ceram Optec, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jul 29, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The MegaBeam Reusable Fiber Optic Handpiece and Needles are intended to be used with the company's disposable MegaBeam Fiber Optic Delivery Systems, which have already been cleared by FDA (K923953, K943445 and K946336). The MegaBeam Fiber Optic Delivery Systems, which contain disposable contact and non-contact fibers, are intended to vaporize, coagulate, incise and excise tissue and are cleared for any indication for which compatible Nd:YAG, 1.44 YAG, Ho:YAG, KTP and diode laser systems have been cleared by FDA.

## Device Story

MegaBeam Reusable Fiber Optic Handpiece and Needles serve as delivery accessories for laser surgical systems; compatible with Nd:YAG, 1.44 YAG, Ho:YAG, KTP, and diode lasers. Device accepts disposable optical fibers (400-1000 micron) threaded through a reusable handpiece and needle assembly. Needle available in 0°, 15°, 30°, and 70° angles. Used by clinicians in surgical settings to deliver laser radiation to target tissue for vaporization, coagulation, incision, or excision. Fiber optic component is discarded after each procedure; handpiece and needle are sterilized via EtO or steam for reuse. Facilitates precise laser energy delivery to surgical sites.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Reusable handpiece and needle assembly; threaded connector for needle attachment; compatible with 400, 600, 800, and 1000 micron optical fibers. Needle angles: 0°, 15°, 30°, 70°. Sterilization: EtO or steam. Energy source: external laser system (Nd:YAG, 1.44 YAG, Ho:YAG, KTP, diode).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- CeramOptec's original disposable handpieces ([K923953](/device/K923953.md))
- CeramOptec's disposable arthroscopy handpiece ([K946336](/device/K946336.md))
- CeramOptec's disposable general surgery straight handpiece and adjustable handpiece ([K943445](/device/K943445.md))
- TTI Medical's Optic Handpiece
- Coherent Medical's VersaLink reusable handpiece ([K960032](/device/K960032.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K980389

## 510(k) SUMMARY

# CeramOptec's MegaBeam Reusable Fiber Optic Handpiece and Needles

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Submitted by Regulatory Counsel for:

CeramOptec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 Phone: 413-525-0600 413-525-0611 Facsimile:

Carol Morello, V.M.D. Contact Person:

February 2, 1998 Date Prepared:

## Name of Device and Name/Address of Sponsor

MegaBeam Reusable Fiber Optic Handpiece and Needles

## Common or Usual Name

Nd:YAG Laser Fiber Optic Delivery System Handpiece; 1.44 YAG Laser Fiber Optic Delivery System Handpiece; Ho:YAG Laser Fiber Optic Delivery System Handpiece; KTP Laser Fiber Optic Delivery System Handpiece; and Diode Laser Fiber Optic Delivery System Handpiece;

#### Classification Name

Accessory to Laser Surgical Instruments

#### Predicate Devices

CeramOptec's original disposable handpieces (K923953), CeramOptec's disposable arthroscopy handpiece (K946336); CeramOptec's disposable general surgery straight handpiece and adjustable handpiece (K943445); TTI Medical's Optic Handpiece; and Coherent Medical's VersaLink reusable handpiece (K960032).

{1}------------------------------------------------

#### Intended Use

The MegaBeam Reusable Fiber Optic Handpiece and Needles are intended to be used with the company's disposable MegaBeam Fiber Optic Delivery Systems, which have already been cleared by FDA (K923953, K943445 and K946336). The MegaBeam Fiber Optic Delivery Systems, which contain disposable contact and non-contact fibers, are intended to vaporize, coagulate, incise and excise tissue and are cleared for any indication for which compatible Nd:YAG, 1.44 YAG, Ho:YAG, KTP and diode laser systems have been cleared by FDA.

## Technological Characteristics and Substantial Equivalence

The company's MegaBeam Reusable Fiber Optic Handpiece and Needles have the same general intended use as the previously cleared predicate devices. The MegaBeam Reusable Fiber Optic Handpiece and Needles are designed to be used with the company's cleared disposable MegaBeam Fiber Optic Delivery Systems which are intended to vaporize, coagulate, incise and excise tissue and which are designed for any indication for which compatible laser systems have been cleared by FDA. Similarly, the predicate devices also are components of delivery systems designed to deliver laser radiation to a specified point.

The MegaBeam Reusable Fiber Optic Handpiece and Needles have similar technological characteristics as their predicate devices. The needle is available with 0°, 15°, 30° and 70° angles. They are designed to fit 400 micron, 600 micron, 800 micron and 1000 micron disposable optical fibers. The labeling includes instructions for sterilizing the devices using EtO or steam. The needle is attached to the handpiece by a threaded connector. The fiber optic is then threaded through the handpiece and needle. At the end of the procedure, the fiber optic is discarded and the handpiece and needle are reusable.

CeramOptec's original, disposable handpieces that were cleared for use with the MegaBeam Delivery System (K923953) are compatible with Nd: YAG lasers and with 400, 600, 800 and 1000 micron fibers. In addition, the company's cleared disposable arthroscopy handpiece (K946336) is compatible with 1:44 YAG, Ho:YAG and Nd:YAG lasers and with 400 and 600 micron fibers. CeramOptec's cleared, disposable general surgery straight handpiece and adjustable handpiece (K943445) are compatible with Nd:YAG, KTP and diode lasers and with fibers ranging in size from 200-1200 microns.

Coherent's VersaLink Laser Delivery System consists of a reusable VersaLink handpiece assembly and VersaTip™ laser probe. The delivery system is compatible for use with Ho:YAG lasers (2.1 um wavelength). The reusable handpiece assembly, consisting of a handpiece, a fiber optic cable, a laser connector, and a protective cap, is shipped non-sterile. The handpiece assembly may be sterilized using either steam or EtO gas. The reusable VersaTip laser probe, which is

{2}------------------------------------------------

provided sterile, is available in the following configurations: 0° straight probe; 15° angled probe; 30° angled probe; and a 70° side fire probe. It must be sterilized using steam or EtO prior to reuse. The system is designed to be used with 600um fibers.

TTI Medical's Fiber Optic Handpiece is compatible with Nd:YAG and Ho:YAG lasers and is designed to be used with 600um, 800um and 1000um fibers. The reusable, malleable handpiece is supplied non-sterile and may be sterilized using EtO or steam in accordance with the manufacturer's instructions.

In sum, although there are some differences between the MegaBeam Reusable Fiber Optic Handpiece and Needles and their predicate devices, these differences are minor and raise no new questions of safety and effectiveness.

#### Performance Data

None provided.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of what appears to be an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# Jul 29 1998

Re:

CeramOptec, Inc. c/o Mr. Jonathan S. Kahan Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, D.C. 20004

> K980389 Trade Name: MegaBeam Reusable Fiber Optic Handpiece and Needles Regulatory Class: II Product Code: GEX Dated: May 1, 1998 Received: May 1, 1998

Dear Mr. Kahan:

We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Mr. Kahan

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# 510(k) Number (if known): K 9803 89

MegaBeam Reusable Handpiece ಾevice Name: _

## Indications For Use:

Intended to be used as a fiber optic delivery system handpiece with the capability to be reused with a disposable optical fiber.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|

| (Division Staff Off)            |         |
|---------------------------------|---------|
| Division of Restorative Devices |         |
| 510(k) Number                   | K980389 |

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K980389](https://fda.innolitics.com/device/K980389)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
