K980308 · Echotech 3D Imaging Systems GmbH · IYO · May 6, 1998 · Radiology
Device Facts
Record ID
K980308
Device Name
ECHOTECH 3D FREESCAN
Applicant
Echotech 3D Imaging Systems GmbH
Product Code
IYO · Radiology
Decision Date
May 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The ECHOTECH 3D FreeScan system is indicated for acquisition of related sets of 2D ultrasound images and 3 dimensional reconstruction of diagnostic ultrasound images. It is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing. It is an add-on accessory for existing ultrasound imaging systems, and is intended to record position and movement of ultrasound transducers for the systematic acquisition of 2 dimensional image slices throughout a volume of interest. It is intended as a general purpose digital 3D ultrasound image processing tool for cardiology, radiology, neurology, gastroenterology, urology, surgery, orthopedics, oncology, obstetrics and gynecology.
Device Story
EchoTech 3D FreeScan is an add-on computer system for existing diagnostic ultrasound machines; facilitates 3D image reconstruction. Input: sequential 2D ultrasound images and spatial transducer position data. Operation: magnetic field sensor (attached to probe via plastic holding plate) tracks transducer in six degrees of freedom; system digitizes 2D slices and converts them into 3D data volumes. Output: 3D ultrasound image volumes for analysis, storage, and retrieval. Used in clinical settings by healthcare providers to assist in diagnostic imaging. Benefits: enables 3D visualization of anatomy from standard 2D ultrasound equipment.
Clinical Evidence
Bench testing only. Software validation performed at module and system levels per internal protocols. Additional system testing conducted by a third-party standards test house. Results confirmed device performance conforms to design intent and system specifications.
Technological Characteristics
Computer system based on Intel motherboard; Microsoft DOS/Windows OS. Includes image digitizer circuit board. Spatial tracking via electromagnetic sensor (pulsed magnetic field). Connectivity: add-on accessory for ultrasound systems. Software: proprietary image processing.
Indications for Use
Indicated for patients requiring 3D reconstruction of diagnostic ultrasound images across various clinical specialties including cardiology, radiology, neurology, gastroenterology, urology, surgery, orthopedics, oncology, obstetrics, and gynecology.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
6 1998 MAY
1 11
Image /page/0/Picture/1 description: The image contains a handwritten string of characters. The string appears to be "K980308". The characters are written in a dark ink on a white background.
January 23, 1998
# 510(k) Summary EchoTech 3D FreeScan
# Submitter (Consultant) Name and Address
Morningstar Consulting Group, Inc. P. O. Box 218 Indian Hills, CO 80454
# Submitter (Consultant) Contact Person
Kevin Morningstar, Senior Consultant phone (303) 697-8198 fax: call first
## Manufacturer Name and Address
EchoTech 3D Imaging Systems GmbH Zeppelinstr. 4 D-85399 Hallbergmoos Germany
## Manufacturer Contact Person
Peter Wiczek, Product Line Manager Phone 49 811 5556 0 fax 49 811 5556 20
## Common, Classification & Proprietary Names
Common Name: Classification Name: Proprietary Name:
Digital Ultrasound Image Analysis System Ultrasonic Pulsed Echo Imaging System EchoTech 3D FreeScan
P-2
{2}------------------------------------------------
#### Predicate Device
TomTec Echo-Scan K963807 dated December 18, 1996.
#### Device Description
The EchoTech 3D FreeScan is a high performance computer system based on Intel motherboard and Microsoft DOS/Windows standards. It incorporates a commercially available image digitizer circuit board and proprietary software for the acquisition, analysis, storage and retrieval of digital 3D ultrasound image data sets. The device is an add-on accessory for any existing diagnostic imaging ultrasound system.
The device records ultrasound transducer spatial position in six degrees of freedom during use. Coordinate tracking is achieved with a miniature magnetic field sensor within a transmitted pulsed magnetic field. This is done by attaching a plastic holding plate to the probe of the host ultrasound system, to which the receiver of an electromagnetic sensor device is attached.
2D ultrasound images are acquired sequentially in a series of steps as the ultrasound transducer is swept across the patient scan site. The resulting set of digitized 2D images is then converted into a 3D data volume.
#### Intended Use
The ECHOTECH 3D FreeScan system is indicated for acquisition of related sets of 2D ultrasound images and 3 dimensional reconstruction of diagnostic ultrasound images. It is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 3-dimensional image processing. It is an add-on accessory for existing ultrasound imaging systems, and is intended to record position and movement of ultrasound transducers for the systematic acquisition of 2 dimensional image slices throughout a volume of interest. It is intended as a general purpose digital 3D ultrasound image processing tool for cardiology, radiology, neurology, gastroenterology, urology, surgery, orthopedics, oncology, obstetrics and gynecology.
#### Technological Characteristics Comparison
The EchoTech 3D FreeScan is nearly identical in performance and intended use to the TomTec EchoScan, except that the EchoScan features an assortment of computer controlled probe carriage devices and incorporates respiration gating.
{3}------------------------------------------------
## Test Discussion
Testing was performed according to internal company procedures. Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted. Test results were reviewed by designated technical professionals before software proceeded to release. Additional system testing was done by a third party standards test house.
### Test Conclusions
Test results support the conclusion that actual device performance satisfies the design intent. Actual device performance as tested internally and by a third party conforms to the system performance specifications.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle, with its wings forming three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
C 1398 MAY
EchoTech 3D Imaging Systems GmbH c/o Kevin Morningstar, Senior Consultant Morningstar Consulting Group, Inc. P.O. Box 218 Indian Hills, CO 80454
Re:
K980308 EchoTech 3D FreeScan Dated: April 4, 1998 Received: April 6, 1998 Regulatory class: II 21 CFR 892.1560/Procode: 90 IYO
Dear Mr. Morningstar:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odrh/dsmaldsmamain.html".
Sincerely yours,
Kilian Yin
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K 980308
510(k) Number (if known):
ECHOTECH 3D FreeScan Device Name:
Indications for Use:
The ECHOTECH 3D FreeScan system is indicated for acquisition of related sets of 2D ritle LOTOTEON of 11000an of econstruction of diagnostic ultrasound images. It is ultrasound images and 3 differial retrieve digital ultrasound images for computerized 3-Intended to acquire, analyze, other and rother accessory for existing ultrasound imaging dimensional intego prococeing to to arritorn and movement of ultrasound transducers for the systems, and is michaod to room poal image slices throughout a volume of interest. It is intended as a general purpose digital 3D ultrasound image processing tool for radiology, intendou as a general parpos angery, orthopedics, oncology, obstetrics and gynecology.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jamil A. Segerson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.