← Product Code [KNT](/productcode/KNT) · K980305

# AMT BUTTON REPLACEMENT GASTROSTOMY DEVICE (K980305)

_Applied Medical Technology, Inc. · KNT · Apr 16, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K980305

## Device Facts

- **Applicant:** Applied Medical Technology, Inc.
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Apr 16, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The AMT Button is intended to be used as a replacement gastrostomy feeding device. Other uses are not recommended.

## Device Story

The AMT Button is a replacement gastrostomy feeding device used to provide nutrition, medication, or gastric decompression in patients with an established stoma. The device is an injection-molded silicone tube featuring a dome and an anti-reflux valve; the dome and valve contain barium for radiopacity. It is provided in various sizes, with the appropriate size determined by a physician or healthcare professional based on the length of the existing stoma tract. The device is supplied in a kit containing an obturator rod, a stoma measuring device, and a patient care kit. It is intended for single-user, disposable use.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on identical design, materials, and manufacturing processes to the predicate device.

## Technological Characteristics

Materials: Silicone; Barium (for radiopacity). Form factor: Injection-molded gastrostomy tube with dome and anti-reflux valve. Sterilization: Sterile (device), non-sterile (kit components). Connectivity: None.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- AMT Button ([K843392](/device/K843392.md))
- Microvasive/Boston Scientific Corp. Button and Replacement Button
- C.R.Bard Button and Replacement Button

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

16980305 1- 68 1

APR 16 1998

| 1. Submitter Name:                                                    | Applied Medical Technology, Inc.                       |
|-----------------------------------------------------------------------|--------------------------------------------------------|
| 2. Address:                                                           | 15653 Neo Parkway<br>Cleveland, OH 44128 USA           |
| 3. Telephone:                                                         | 216-475-5577                                           |
| 4. FAX:                                                               | 216-475-5575                                           |
| 5. Contact Person:                                                    | Tom Parkinson - Quality Assurance / Regulatory Affairs |
| 6. Date of preparation:                                               | January 22, 1998                                       |
| 7. Device Trade or Proprietary Name:                                  | AMT Button                                             |
| 8. Device Common Name:                                                | AMT Button Gastrostomy Feeding Device                  |
| 9. Device Classification Name:                                        | Gastrointestinal Tube and Accessories (78KGC)          |
| 10. Substantial Equivalency is claimed against the following devices: |                                                        |

510(k) Summary

| 1. Applied Medical Technology, Inc. -    | AMT Button - K843392          |
|------------------------------------------|-------------------------------|
| 2. Microvasive/Boston Scientific Corp. - | Button and Replacement Button |
| 3. C.R.Bard -                            | Button and Replacement Button |

### 11. Device of Description:

AMT Button is injection molded of Silicone. The Dome is injected molded from Silicone and Barium mixed for radiopaic detection if desired and is glued (with silicone) to the body of the Button. The Anti-Reflux Valve is injected molded from Silicone and Barium mixed for radiopaic detection if desired and is glued (with silicone) to the body of the Button. The device is sold in various sizes. Size is determined by the Physician or Professional Health Giver (dependant on length of existing stoma tract). The AMT Button is packaged in a kit which includes Directions For Use, Obturator Rod, Patient Care Kit, Stoma Measuring Device. The AMT Button Replacement Gastrostomy Device is sterile, single user, and disposable. The Patient Care Kit and Stoma Measuring Device are Non-Sterile and packaged for single user, disposable.

## 12. Intended Use:

The AMT Button is intended to be used as a replacement gastrostomy feeding device. Other uses are not recommended.

#### 13. Safety and Effectiveness of the device:

The AMT Button is as safe and effective as other predicate devices cited above.

# 14. Summary comparing technological characteristics with other predicate devices:

The AMT Button identical in design and manufacture, using the same materials and processes used in 510(k) K843392 of Applied Medical Technology, Inc.

AMT Button U.S. Patent No .: 4,863,438

Applied Medical Technology, Inc. - 510(k) Submission AMT Button Replacement Gastrostomy Device Section H - Page H.2.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or head in profile, with three overlapping shapes suggesting multiple faces or layers. The symbol is rendered in black, contrasting with the white background of the seal.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 6 1998

Mr. Thomas W. Parkinson · Leader - Quality Assurance / Regulatory Affairs Applied Medical Technology, Inc. 15653 Neo Parkway Cleveland, Ohio 44128

Re: K980305 AMT Button Replacement Gastrostomy Device Regulatory Class: II 21 CFR 876.5980/Procode: 78 KNT Dated: January 22, 1998 Received: January 27, 1998

Dear Mr. Parkinson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510 (k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Thomas W. Parkinson

Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Robert R Ratting/

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# SECTION - I

11480 XV3

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications For Use:

.. . . . . . .

The AMT Button Replacement Gastrostomy Device is to be used as a replacement gastrostomy tube for a patient with a well established gastrostomy tract. This device will assist in providing nutrition directly into the stomach through an established stoma in a human patient who is unable to consume nutrition by conventional means. The AMT Button Replacement Gastrostomy Device can also deliver medication and allow for decompression of the stomach.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

Concurrence of CDRH, Office of Device Evaluation (ODE)

|               | <div> <p><i>Division Sign-Off</i></p> </div>                       |
|---------------|--------------------------------------------------------------------|
|               | Division of Reproductive, Abdominal, ENT, and Radiological Devices |
| 510(k) Number | K980305                                                            |

| Prescription Use | <div> <p>(Per 21 CFR 801.109)</p> </div> | OR | Over-The-Counter Use |
|------------------|------------------------------------------|----|----------------------|
|------------------|------------------------------------------|----|----------------------|

(Optional Format 1-2-96)

Applied Medical Technology, Inc. - 510(k) Submission

AMT Button Replacement Gastrostomy Device Section 1 - Page 1.10.

---

**Source:** [https://fda.innolitics.com/device/K980305](https://fda.innolitics.com/device/K980305)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
