K980287 · Siemens Medical Solutions USA, Inc. · KLK · Aug 12, 1998 · Anesthesiology
Device Facts
Record ID
K980287
Device Name
SIEMENS SC9000/SC9015 TCP02/C02 MODULE
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KLK · Anesthesiology
Decision Date
Aug 12, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2500
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Siemens Transcutaneous tpO2/CO2 Gas Module is indicated for use to continuously monitor the noninvasive trending of transcutaneous oxygen in the neonatal population when the patient is not under gas anesthesia. Siemens Transcutaneous tp02/CO2 Gas Module is indicated for use to continuously monitor the noninvasive trending of transcutaneous carbon dioxide partial pressure in any patient population. The device is indicated for use in an environment where patient care is provided by healthcare professionals, i.e. Physicians, Nurses, Technicians, when the professional determines that the device is required. MRI Compatibility Statement: The Siemens tpO2/CO2 gas module is not intended for use in an MRI magnetic field.
Device Story
Transcutaneous Gas Module (tpO2/CO2 pod) functions as an add-on to Siemens SC9000 patient monitoring series; provides continuous, noninvasive trend monitoring of peripheral tissue oxygen and carbon dioxide partial pressures. Device utilizes solid-state sensors applied to patient skin; sensors feature user-selectable temperature settings (37-45°C) and heating power control. System performs single-point calibration; provides user-selectable alarms for temperature deviations. Output displayed on SC9000/SC9015 monitor; integrated with heart rate and respiratory values to generate oxycardiorespirograms. Operated by physicians, nurses, or technicians in clinical environments. Benefits include real-time physiological trend data to assist clinical decision-making for neonatal and general patient populations. Not MRI compatible.
Clinical Evidence
The device was deemed safe and effective based on clinical performance data provided in Section V of the submission. No specific study metrics (e.g., sensitivity, specificity) are detailed in the provided text.
Technological Characteristics
Solid-state transcutaneous O2/CO2 sensors; heating power 0-600 mW ± 10 mW; temperature range 37-45°C; single-point calibration; IEC 601-1 Type CF applied part; module form factor for integration with Siemens SC9000/SC9015 patient monitors.
Indications for Use
Indicated for continuous noninvasive trend monitoring of transcutaneous oxygen (neonates, not under gas anesthesia) and transcutaneous carbon dioxide (all patient populations). Used by healthcare professionals in clinical settings.
Regulatory Classification
Identification
A cutaneous oxygen (PcO2) monitor is a noninvasive, heated sensor (e.g., a Clark-type polargraphic electrode) placed on the patient's skin that is intended to monitor relative changes in the cutaneous oxygen tension.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9. The special control is FDA's “Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA.” See § 868.1(e) for the availability of this guidance document.
Predicate Devices
Siemens SIRECUST Transcutaneous Carbon Dioxide Monitoring System (K891481)
{1}------------------------------------------------
# AUG 1 2 1998
K980287
510(k) Notification
Siemens SC9000/SC9015 Transcutaneous Gas Module (tpO2/CO2)
# 510(k) SUMMARY
as required per 807.92(c)
#### Submitters Name, Address: 2.
Siemens Medical Systems, Inc. Electromedical Systems Group, PCS Danvers, MA 01923 Tel: (508) 750-7500 Fax: (508) 777-3398 Official Correspondent: David Simard, Director Quality Assurance & Regulatory Affairs Contact person for this submission: Jacqueline Emery Date submission was prepared: January 23, 1998
#### Trade Name, Common Name and Classification Name: 3.
### A. Trade Name:
Siemens Transcutaneous tpO2/CO2 Gas Module
- B. Common Name, Classification Name, Class and Regulation Number:
| Common Name | Classification<br>Number | Class | Regulation Number |
|--------------------------------------------------------------------|--------------------------|-------|-------------------|
| Monitor, Carbon-Dioxide, Cutaneous | 73LKD | II | 868.2480 |
| Monitor, Oxygen, Cutaneous, for infant not<br>under gas anesthesia | 73KLK | II | 868.2500 |
- Predicate Device Identification: 4.
Siemens SIRECUST Transcutaneous Carbon Diexide Monitoring System 510(k) Number: K891481 (02 sensor, Radiometer K900333)
#### ર. Device Description:
The Transcutaneous Gas Module (tpO/CO2 pod) is an addition to Siemens SC9000 patient monitoring series which provides transcutaneous measurement of partial pressures of the 02 and CO2 in the peripheral tissue of a patient.
- 6. Intended Use:
The intended use of Siemens Transcutaneous tpO2/CO2 Gas Module is for the continuous, noninvasive trend monitoring of transcutaneous oxygen and transcutaneous carbon dioxide partial pressure, in the neonatal patient population when the patient is not under gas anesthesia, by healthcare professionals, i.e. Physicians, Nurses, Technicians.
### 1
## COMPANY CONFIDENTIAL
Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
Image /page/0/Picture/25 description: The image shows a handwritten number nine inside of an oval shape. The number nine is written in a simple, cursive style. The oval shape surrounds the number nine, creating a border around it. The image is in black and white.
{2}------------------------------------------------
# 510(k) Notification Siemens SC9000/SC9015 Transcutaneous Gas Module (tpO2/CO2)
| | Substantial Equivalent Device | Substantial Equivalent Device | Applicant |
|----------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------|
| | Siemens Medical Systemst<br>SIRECUST 404/404N | Radiometer America, Inc.<br>Combined Transcutaneous<br>TcPO2/TCPCO2 Elect. | Siemens Medical Systems<br>SC9000/SC9015 tpO2/CO2 module |
| Manufacturer | Siemens Medical Systems | Radiometer | Siemens Medical Systems |
| 510(k) Number | K891481 | K900333 | To be assigned |
| Intended Use | Continuous, noninvasive trend monitoring of<br>transcutaneous carbon dioxide partial pressure | | Same |
| Intended Population | Adult, Neonatal | | Neonatal |
| Intended Environment | In an environment where healthcare is provided by<br>healthcare professionals, i.e.<br>doctors, nurses, technicians. | | Same |
| O2 only sensor | | Solid State | Same |
| O2 + CO2 sensor | Solid State | Solid State | Same |
| tpO2 | | | Same |
| Measuring Range | | 0 to 800 mmHg | |
| Correction Factor | | No | Same |
| Temperature setting | | 37 to 45°C | Same |
| Heating Power | | 0 to 600 mW ± 10 mW | Same |
| Calibration | | Single point | Same |
| Alarm | | Alarms >.2°C from the selected<br>electrode temperature | User Selectable |
| tpCO2 | | | Same |
| Measuring Range | 0 to 200 mmHg | | |
| Correction Factor | Yes | | Same |
| Temperature setting | 37 to 45°C | | Same |
| Heating Power | 0 to 600 mW ± 10 mW | | Same |
| Calibration | Single point | | Same |
| Alarm | Alarms >.2°C from the<br>selected electrode<br>temperature | | User Selectable |
| Oxycardiorespirogram | In conjunction with HR +<br>RESP values | | Same |
- 7 Table of Device Similarities and differences to predicate device
8. Assessment of non-clinical performance data for equivalence: Currently there are no FDA standards for this device.
- 9. Assessment of clinical performance data for equivalence: The transcutaneous gas module (tpO2/CO2) was deemed to be safe and effective. See Section V of this submission.
- 10. Biocompatability: Not applicable
- 11. Sterilization: Not applicable
... ..
- 12. Standards and Guidances: IEC 601-1 Type CF Applied Part
2
# COMPANY CONFIDENTIAL
# Siemens Medical Systems, Inc.
Electromedical Systems Group, PCS
16 Electronics Avenue Danvers, MA 01923 USA
10
Tel: (978) 907-7500 Fax: (978) 750-6879 Telex: 511958 (Siemensm SD)
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, with three abstract human figures facing right, representing health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 2 1998
Mr. David Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923
Re: K980287 SC9000/SC9015 Transcutaneous Gas Module (tpO2/tpCO2 pod) Regulatory Class: II (two) Product Code: 73 KLK Dated: May 20, 1998 Received: May 22, 1998
Dear Mr. Simard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
### Page 2 - Mr. David Simard
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K980287
Device Name: Siemens Transcutaneous TpO2/CO2 Gas Module
# Indications for Use:
Siemens Transcutaneous tpO2/CO2 Gas Module is indicated for use to continuously monitor the noninvasive trending of transcutaneous oxygen in the neonatal population when the patient is not under gas anesthesia.
Siemens Transcutaneous tp02/CO2 Gas Module is indicated for use to continuously monitor the noninvasive trending of transcutaneous carbon dioxide partial pressure in any patient population.
The device is indicated for use in an environment where patient care is provided by healthcare professionals, i.e. Physicians, Nurses, Technicians, when the professional determines that the device is required.
MRI Compatibility Statement: The Siemens tpO2/CO2 gas module is not intended for use in an MRI magnetic field.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | OR Over-The-Cou (Option |
|---------------------------------------|-------------------------|
|---------------------------------------|-------------------------|
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
| 510(k) Number. | |
|----------------|--|
|----------------|--|
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.