1. Iniection and administration of anesthetic regional blocks (e.g., paracervical, uterosacral, and pudendal) that may be used during labor, delivery, or both. 2. Injection and administration of anesthetic regional blocks (e.q., paracervical, uterosacral, and pudendal) that may be used during minor surgical procedures.
Device Story
Gynex Extended Reach Needle is a disposable, 27-gauge, 90mm stainless steel needle used for administering regional anesthetic blocks (paracervical, uterosacral, pudendal) during labor, delivery, or minor surgical procedures. The device features a dual-diameter shaft; the section from the hub to 5mm from the tip has a larger diameter to assist in estimating insertion depth and preventing over-insertion into deep tissue. Available with standard or threaded hubs, the needle is supplied sterile and non-pyrogenic. It is intended for use by clinicians in clinical or operating room settings. The device facilitates precise anesthetic delivery, potentially improving patient comfort and safety during obstetric or surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, material specifications, and design comparison to legally marketed predicate devices.
Technological Characteristics
302 instrument-grade stainless steel construction; 27-gauge, 90mm length; dual-diameter shaft for depth control; Luer lock or threaded hub options; radiation sterilization (SAL 10^-6); non-pyrogenic. Complies with ISO 9000 series and various ASTM standards.
Indications for Use
Indicated for patients requiring regional anesthetic blocks (paracervical, uterosacral, or pudendal) during labor, delivery, or minor surgical procedures.
Regulatory Classification
Identification
An obstetric anesthesia set is an assembly of antiseptic solution, needles, needle guides, syringes, and other accessories, intended for use with an anesthetic drug. This device is used to administer regional blocks (e.g., paracervical, uterosacral, and pudendal) that may be used during labor, delivery, or both.
Pudendal and paracervical block anesthesia instruments (V. Mueller, 1973)
Submission Summary (Full Text)
{0}------------------------------------------------
K980238
1103
# ed Reach Needle arv of Safety and s Information Notification. Section 510(k)
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
## 1. Device Name:
Trade Name: Gynex Extended Reach Needle . Disposable Needle, disposable anesthetic needle Common Name: Classification Name: Obstetric anesthesia set & accessories
## 2. Establishment Name & Registration Number:
GYNEX Corporation Name: . Number: 3032109
## 3. Classification:
\$ 884.5100 Obstetric anesthesia set. (a) Identification. An obstetric anesthesia set is an assembly of antiseptic solution, needles, needle guides, syringes, and other accessories, intended for use with an anesthetic drug. This device is used to administer regional blocks (e.g., paracervical, uterosacral, and pudendal) that may be used during labor, delivery, or both. (b) Classification. Class II (performance standards).
## 4. Establishment Name & Registration Number:
Name: GYNEX Corporation Number: 3032109
## 5. Classification:
Device Class: Class II Classification Panel: Ob/Gyn Devices Panel Product Code: 85HEE
## 6. Preamendments Device:
- 1. Pudendal and paracervical block anesthesia instruments, Gynecology, Obstetrics, V. Mueller, The Surgical Armamentarium, 1973., pp. 219
## 7. Equivalent Legally Marketed Device(s):
- 1. Potocky Needle Disposable Injection Needle, K910252, Cooper Surgical.
..
- - - -
{1}------------------------------------------------
K980238
P293
## 8. Device Description:
The Gynex Extended Reach Needles are made of 302 instrument grade stainless steel. There are two styles each being 27 gauge and 90mm in length. One is equipped with a standard I hore are the other basing thas a threaded hub. Both are disposable. As is typical for this type of anesthetic needle, the shaft of the needle is dual diameter. That is, the shaft of the needle from the hub to within 5mm of the tip a is larger diameter. This allows for easy estimation of insertion depth during anesthetic administration and prevents over-insertion of needle into deeper tissue levels or spaces.
## Applicant Name & Address: 9.
GYNEX Corporation 16700 NE 79th St., Suite 204 Redmond, WA 98052
## Company Contact: 10.
Mr. Steve Angelo, President GYNEX Corporation 16700 NE 79th St., Suite 204 Redmond, WA 98052 425.882.1179 * 425.895.0115 - fax
## 11. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 1000 Burnett Av.. Suite 450 Concord, CA 94520 510.356.2640 - 510.356.2654 - fax
#### 12. Comparison Table:
| FEATURE | Gynex Extended Reach Needle | Pudendal/Paracervical Needle &<br>Potocky Needle | SE? |
|----------------------------|------------------------------------------------|-------------------------------------------------------------------------|------------|
| Indications for<br>Use(s): | Administer Paracervical pudendal<br>anesthetic | Administer Paracervical pudendal<br>anesthetic | YES |
| Design: | Leuer lock & threaded hub | Leuer lock | YES |
| Sterilization<br>Method: | Radiation - Supplied sterile -<br>disposable | Steam Autoclave - reusable<br>Potocky -Supplied sterile -<br>disposable | NO&<br>YES |
| Sizes: | 90mm w/ 5mm 27 gauge tip | 6-8 inches - 20-25 gauge<br>Potocky - 90mm, 27 gauge | YES |
| Material: | 302 Stainless steel | 302 Stainless Steel | YES |
| Country of<br>Origin :: | Asia - Germany | Germany<br>Potocky - Asia - Germany | YES |
| Manufacturer: | Gynex Corporation | V. Mueller, Cooper Surgical | YES |
| Product Code: | 84HEE | 84HEE | YES |
| K - Number: | Pending | Preamendment & K910252<br>respectively | YES |
{2}------------------------------------------------
K910238
P3093
## 13. Voluntary Standards:
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include ASTM, GYNEX Corporation. Standard Operating Procedures (SOP), vendor certification and qualification procedures, Quality Systems Regulations, ISO materials standards and ISO 9000 series quality regulations.
#### 14. Storage, Packaging & Sterilization Information:
The Gynex Iris/Angle Extended Reach Needles are supplied sterile and non-pyrogenic.
All needles are individually packaged & labeled and supplied in multi-unit packages and boxes for ease in shipping and storage. All packaging should be inspected on arrival for evidence of shipping damage. Damaged packaging may indicate the presence of unsafe product and it should not be used until carefully inspected. Any shipping damaged product may not be sterile and should be returned promotiv. Product for use in the operating room must be processed, opened, handled and placed into use following accepted operating room sterile technique.
Sterility Assurance Level (SAL) is at least 10°. A minimum of 2.5 Mrrad radiation exposure is The product is dosemetry released. The product is also non-pyrogenic. Vendor utilized. certification and secondary sampling is performed and documented.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized symbol with three curved lines resembling a bird or abstract human figure. The text "DEPARTMENT OF HEALTH & HUM" is partially visible, arranged vertically along the left side of the symbol. The image is in black and white.
AUG 21 1998
Gynex Corporation c/o Mr. David Schlerf Buckman Company, Inc. 200 Gregorv Lane, Suite C-100 Pleasant Hill, CA 94523-3389
9200 Corporate Boulevard Rockville MD 20850
Food and Drug Administration
Re: K980238
GYNEX Extended Reach Needle Dated: June 20, 1998 Received: August 4, 1998 Regulatory Class: II 21 CFR 884.5100/Procode: 85 HEE
Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in intersiate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 80 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1
510(k) Number (if known): __ K 980236
Device Name(s):
GYNEX Extended Reach Needles
Intended Use:
- 1. Iniection and administration of anesthetic regional blocks (e.g., paracervical, uterosacral, and pudendal) that may be used during labor, delivery, or both.
- 2. Injection and administration of anesthetic regional blocks (e.q., paracervical, uterosacral, and pudendal) that may be used during minor surgical procedures.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert R. Sothin
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K980238
OR
Prescription Use X
Over-The-Counter Use
(Per 21 CFR 801.109)
(Optional format 1-2-96)
GYNEX-EXTEND.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.