K980065 · C/T Med System, Ltd. · KCT · May 11, 1998 · General Hospital
Device Facts
Record ID
K980065
Device Name
C/T MED-SYSTEMS CASSETTE SYSTEM
Applicant
C/T Med System, Ltd.
Product Code
KCT · General Hospital
Decision Date
May 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
General dental / medical instrument cassettes intended to hold instruments and accessories in place throughout the entire instrument use, cleaning and sterilization cycle. These cassettes are suitable for Gravity Steam, Pre-vacuum steam and EO sterilization.
Device Story
The C/T Med-Systems Instrument Cassette System is a reusable metal container made of aluminum and stainless steel, designed to organize and secure dental and medical instruments. The system is used in hospitals, operating rooms, and dental offices to hold instruments throughout the entire cycle of use, cleaning, and sterilization. It supports multiple sterilization methods, including Gravity Steam, Pulsing High Vacuum steam, and Ethylene Oxide (EO). Instruments can also be cleaned via chemical sterilant while inside the cassette. The device provides a structured environment for instrument management, reducing handling and potential injury during the sterilization process. Healthcare providers use the cassettes to maintain instrument organization and sterility from the point of use through the cleaning and sterilization workflow.
Clinical Evidence
No clinical data. Performance was established through bench testing in accordance with AAMI TIR No. 12-1994. Testing validated the cassette system's performance under three sterilization methods: Pulsing High Vacuum steam, Gravity steam, and Ethylene Oxide (EO). Ultrasonic cleaning efficacy was also verified by Indiana University.
Technological Characteristics
Materials: Aluminum and stainless steel. Form factor: Metal containers in various sizes (typically 1.5" x 3.5" x 1" to 8" x 15.5" x 2.5"). Sterilization compatibility: Gravity Steam, Pulsing High Vacuum steam, and Ethylene Oxide (EO). Design features: Internal instrument holding mechanisms, optional latch system.
Indications for Use
Indicated for holding dental and medical instruments and accessories in place during use, cleaning, and sterilization cycles in hospital, operating room, physician, and dental office settings.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for CT Med-Systems LTD., INC. The logo features the letters "CT" in a bold, sans-serif font, with a caduceus symbol placed between them. Below the logo, the text "MED-SYSTEMS LTD., INC." is printed. The date "MAY 11 1998" is printed at the bottom of the image.
K980065
## Non-Confidential Summary of Safety and Effectiveness
page 1 of 3 January 6, 1998
C/T Med- Systems Ltd., Inc. 3755 N. Arlington Ave. Indianapolis, IN 46218
Tel - (317) 543-4250 Fax - (317) 543-4203
Alan Booker - Operations Manager Official Contact: C/T Med-Systems Cassette System Proprietary or Trade Name: Sterilization Cassettes Common/Usual Name: Sterilization Wrapper Pack, Bag and Accessories Classification Name: C/T Med-Systems Cassette system Device: Sterilization Cassettes - K962545 Predicate Devices: Hu-Friedy - IMS - K844002
## Device Description:
A cassette system and accessories designed to hold various dental and medical instrument to be cleaned and sterilized. The design is a metal, aluminum and stainless steel container which has various methods of holding the instruments in place. Available in various sizes ranging which have a typical size of 1.5" x 3.5" x 1" to 8" x 15.5" x 2.5".
Indicated Use --General dental / medical instrument cassettes intended to hold instruments and accessories in place throughout the entire instrument use, cleaning and sterilization cycle. These cassettes are suitable for Gravity Steam, Pulsing High Vacuum steam and EO sterilization. Instruments may be cleaned by chemical sterilant.
Environment of Use -- Hospital, Operating Room (OR), physician and dental office or places where instruments are sterilized. Page 3 of 75
> Industry Member of Safety and Effectiveness Non-Confidential Sur
> > Office Sterilization & Asepsis Procedures Research Foundation
{1}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness
page 2 of 3 January 6, 1998
# Summary of Performance testing --
The C/T Med-System Cassette system was independently tested according to AAMI TIR No. 12-1994 for its performance under three (3) sterilization methods - Pulsing High Vacuum and Gravity steam and Ethylene Oxide (EO). It was also tested for ultrasonic cleaning by Indiana University.
Comparison to Predicate Devices: ﺭﻳﺎﺿﻴﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮ
| | CALL CARDER STATE | Sterilization . Din Teixilen . De<br>ల్లాలు |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| | | |
| | | |
| Hributo | | THZAUOTE<br>1.15 2 2 |
| | . | |
| Attribus | | 2. 25.5 |
| 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 | 62222<br>156 2011 | 18.0 |
| | ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ | . 8/4 . |
| S | | |
| | Seems<br>Carl Ass | ा (1) - |
| | | |
| | --------------------------------------- | 20.00000000<br>11 2018 2 |
| 1 2011 2013<br><br> | | |
| | | 19993 |
| | 20022 | PRODUCT |
| Call . 11 . | | |
| | 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - | 10 0 20 20 |
| -1.9 | 170200 | ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |
| 11-51-2017<br>.<br>112 | 中心数据 | |
| Carlos Comments<br>11652 | | |
| Use | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-----------|---------------|
| Indicated for holding dental and<br>medical instruments and accessories<br>in place throughout entire instrument<br>use, cleaning and sterilization cycle | Yes | Yes | Assumed |
| Intended to be reused | Yes | Yes | Yes |
| Sterilization by Gravity Steam,<br>Pulsing High Vacuum Steam and<br>Ethylene oxide | Yes | Autoclave | Not specified |
| Design | | | |
| Various sizes offered ranging typically<br>ranging from 2.5" x 8" x 1.5" to<br>8" x 15.5" x 2.5" | Yes | Yes | Yes |
| Utilizes various methods of holding<br>instruments in place | Yes | Yes | Yes |
| May incorporate latch system to hold | | | |
{2}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness
. . .
## page 3 of 3
# January 6, 1998
| Performance Standards / Specifications | | | |
|------------------------------------------------------------------------------------------|-----|-----------------------------|-----------|
| Tested in accordance to AAMI | | | |
| TIR No. 12-1994 | Yes | No but other<br>method used | Not known |
| Validation study performed with half<br>cycles to challenge sterilization<br>method used | Yes | No | Not known |
| AAMI ST 33 -1992 requirements | | | |
| 3.4-3.5 Instructions provided for<br>different cleaning methods of cassette | Yes | Assumed | Assumed |
| 3.6-4.2 Instructions for inspections | Yes | Assumed | Assumed |
| 6.2.2 Sterilization manufacturer<br>documentation available | Yes | Yes | Not known |
| 6.2.3 Drying time in labeling | Yes | Not known | Not known |
| 6.2.4 EO residual removal in labeling | Yes | Not known | Not known |
| 6.3 User (basic) responsibilities listed<br>in labeling | Yes | Not known | Not known |
| 7.3.1 Routine inspection in labeling | Yes | Not known | Not known |
| Materials | | | |
| Aluminum and stainless steel | Yes | Yes | Yes |
# Differences between Other Legally Marketed Predicate Devices
There are no significant differences between the intended device and the predicates - Sterilization Cassette - K962545 and Hu-Friedy - IMS - K844002.
{3}------------------------------------------------
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one above the other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY | | 1998
Mr. Alan Booker ·Operations Manager C/t Med Systems, Ltd., Incorporated 3755 North Arlington Avenue Indianapolis, Indiana 46218
Re : K980065 C/T Med-Systems - Sterilization Cassette Trade Name: System Regulatory Class: II Product Code: KCT Dated: April 22, 1998 Received: April 23, 1998
Dear Mr. Booker:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{4}------------------------------------------------
Page 2 - Mr. Booker
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda/gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timot Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### TIONSTORUSET INDICA
| 510(k) Number: | (to be assigned) |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | C/T Med-Systems Instrument Cassette System |
| Intended Use : | General dental / medical instrument cassettes intended to hold<br>instruments and accessories in place throughout the entire<br>instrument use, cleaning and sterilization cycle. These cassettes are<br>suitable for Gravity Steam, Pre-vacuum steam and EO sterilization. |
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per CFR 801.109)
*****************************************************************************************************************************************************************************
or
Over-the-counter use _________________________________________________________________________________________________________________________________________________________
(Division Sign-Off) Stearce ) Division of Dental, Infection Control, PhD and General Hospital Devices
510(k) Number _
Page 10 of 75
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.