K980056 · Exigent Diagnostics, Inc. · JJF · Mar 6, 1998 · Clinical Chemistry
Device Facts
Record ID
K980056
Device Name
CARESIDE ANALYZER
Applicant
Exigent Diagnostics, Inc.
Product Code
JJF · Clinical Chemistry
Decision Date
Mar 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2170
Device Class
Class 1
Indications for Use
This product is intended for use with Exigent Diagnostics Cartridges when used by laboratory professionals for the in vitro measurement of various clinical chemistry analytes in human blood.
Device Story
CareSide™ Analyzer is a compact, semi-automated tabletop chemistry analyzer for clinical use. It processes whole blood, serum, or plasma specimens introduced via analyte-specific, single-use test cartridges. The analyzer performs automated centrifugation for sample separation, metering, dispensing, incubation, and colorimetric detection. It accepts up to 6 cartridges simultaneously. The system uses LED/photodiode pairs to measure reflectance on reagent film at defined intervals. Results are calculated using lot-specific polynomial coefficients stored in the cartridge barcode and a master dose-response curve. The device is operated by laboratory professionals via a touchscreen interface. Results are displayed on-screen and can be exported via floppy disk or RS-232 port. By automating sample preparation and analysis, the device provides rapid quantitative clinical chemistry results, facilitating timely clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated through precision studies (total CV 3.1% to 8.5% across 4 tests), linearity (correlation coefficient >= 0.95), and method comparison against the predicate device. Mean differences between CareSide™ and Vitros were approximately 3% for evaluated tests. No clinical prospective or retrospective studies on patient outcomes were required.
Technological Characteristics
Tabletop chemistry analyzer using dry film reflectance photometry. Sensing via 6 LED/photodiode pairs. Connectivity includes RS-232 port and 3.5" floppy drive for data export. Calibration is factory-set with lot-specific coefficients provided via cartridge barcodes and automated black/white reflectance standard checks. Software-controlled via Real-Time Processor (RTP) and User Interface Processor (UIP).
Indications for Use
Indicated for in vitro diagnostic use by laboratory professionals for the measurement of various clinical chemistry analytes in human whole blood, serum, or plasma.
Regulatory Classification
Identification
A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.
Predicate Devices
Johnson and Johnson's Vitros DT 60 II/DTSC II/I (formerly Kodak Ektachem DT 60 II System) (K912844/A)
Reference Devices
Abaxis Picollo
Submission Summary (Full Text)
{0}------------------------------------------------
# V. 510(k) Summary:
# CareSide™ Analyzer Safety and Effectiveness
MAR - 6 1998
# Applicant Information I.
- Applicant Name A.
- Applicant/Manufacturer Address B.
- Telephone Number C.
- D. Contact Person
- E. FAX Number
- e-Mail Address F.
- Date 510(k) Summary prepared G.
# Device Information II.
- A. Device Name (Trade)
- B. Device Name (Classification)
- C. Device Classification
CareSide™ Analyzer
Exigent Diagnostics, Inc.
Kenneth B. Asarch, Pharm.D., Ph.D.
6100 Bristol Parkway
310-338-6767
310-338-6789
February 23, 1998
Culver City, CA 90230
asarchk@worldnet.att.net
Micro chemistry analyzer for clinical use Clinical chemistry panel Micro chemistry analyzer for clinical use Regulation Number: 21 CFR 862.2170 Regulatory Class I Classification Number: 75JJF Tier I None applicable
- D. Device Tier
- Special controls and E. performance standards
# III. Substantial Equivalence Claim
- A. General equivalency claim
The ability to monitor analyte-specific biochemical reactions in dry film and other formats by reflectance and other techniques is widely recognized and has gained widespread acceptance for use in chemistry assays. Micro chemistry analyzers are already on the U.S. market. These products utilize dry film and other formats, and use reflectance and other types of signals. For example,
Johnson & Johnson Clinical Diagnostics, formerly Kodak Vitros DT 60 II Ektachem DT 60 II (analyzer using reflectance photometry and electrochemical measurement for signal detection)
Miles/Ames (analyzer using Seralyzer Reflectance Photometer formerly reflectance photometry for signal detection)
- B. Specific equivalency claim
The CareSide™ Analyzer is substantially equivalent in principle, intended use, and clinical performance to the currently marketed Vitros DT 60 II/DTSC II Module.
| Name of Predicate Device: | Johnson and Johnson's Vitros DT 60 II/DTSC II/I<br>(formerly Kodak Ektachem DT 60 II System) |
|-------------------------------|----------------------------------------------------------------------------------------------|
| Predicate Device 510K number: | K 912844/A |
| Product Code: | 75JJE |
{1}------------------------------------------------
#### IV. Device Description
#### CareSide™ System Overview A.
The Exigent Diagnostics CareSide™ System utilizes individual (analyte specific) test cartridges to perform a variety of clinical blood tests. The user introduces the whole blood, serum, or plasma specimen into the cartridge sample deposition well, closes the lid and inserts the cartridge into the Exigent Diagnostics CareSide™ Analyzer. The Exigent Diagnostics CareSide™ Analyzer automatically performs up to 6 quantitative test results in approximately ten minutes using approximately 85 microliters of human blood per test. Results are provided on-screen, and may be transferred to a floppy disk for subsequent uploading to a laboratory information system. The Exigent Diagnostics CareSide™ System components include:
- CareSide™ Analyzer t
- . Analyte-specific Cartridge(s)
# CareSide Cartridge™ Description B.
The CareSide™ Test Cartridge, a single use analyte-specific cartridge, performs the following functions: reagent and specimen temperature equilibration, sample separation, sample metering, sample dispensing, test incubation, and colorimetric detection. Each cartridge (patents pending) contains all reagents necessary to perform an analyte measurement.
The CareSide™ test cartridges are individually bar coded and packaged, containing both human readable and barcode labeling. The test cartridge package is hermetically sealed to assure stability over the shelf-life of the cartridge. Test specific requirements are provided in test specific package inserts. These cartridges consist of a pre-analytical element and an analytical element.
- Pre-analytical Element The pre-analytical element consists of a plastic 1. cartridge consisting of a series of channel and wells. On top of the cartridge is a hinged lid and a plastic membrane covered opening. The assembled cartridge contains various channels and chambers which, upon centrifugation within the Analyzer, separates the blood cells from the plasma or serum. The plasma or serum flow from the metering channel is initiated when the specimen deposition well is pressurized by deformation of its plastic cover within the analyzer. The pressure forces the plasma or serum out of the sample delivery passage onto the analytical element.
- 2. The analytical elements consist of a multi-layer film. Analytical Element Each different chemistry cartridge differs only with respect to the multi-layer film which is described in the analyte-specific test cartridge package insert.
# C. CareSide™ Analyzer Description and explanation of Device Function
The CareSide™ Analyzer is a compact tabletop chemistry analyzer that performs multiple discrete analyses on human whole blood, plasma, or serum specimens. The CareSide™ analyzer is semi-automated: the only operator intervention is the addition of the specimen to the test cartridge and the insertion of the dosed cartridge into the analyzer. The CareSide™ Analyzer automatically warms, separates, meters, dispenses, The and incubates the specimen before reading the signal and calculating results. CareSide™ Analyzer is intended only for use with Exigent Diagnostics CareSide™ test cartridges.
The user enters the patient identification and test(s) to be performed via the touch screen by following a series of menus and prompts. The CareSide™ Analyzer accepts up to 6 test cartridges from a single patient at the same time. Next, the user obtains the required test cartridge(s) from refrigerated storage and removes the cartridge from its individual hermetically sealed package.
{2}------------------------------------------------
Once loaded, the analyzer spins the cartridge(s) to heat the cartridge(s) and the contained specimen(s) to the reaction temperature and to to move the sample from the sample deposition well into the cartridge channels and chambers. As the cartridge spins, the cells accumulate in the separation well. A defined volume of plasma (or serum, as applicable) remain in the metering passage. Any excess sample flows into an overflow well. To dispense the metered volume of sample onto the film, a plunger displaces a flexible seal that covers the sample deposition well. As the flexible seal is displaced, air is forced through the metering passage, forcing the sample out and onto the reagent film. As the cartridges spin, the film passes by one of 6 LED/photodiode pairs. The LED/photodiodes perform reflectance measurements at a specified time(s). The analyzer uses the reflectance measurements and a chemistry-specific standard curve to calculate the concentration.
The analyte concentration is proportional to the concentration of dye on the film. The amount of dye is quantitated by the measurement of the reflected light from the reagent film using a light emitting diode/photodiode set under the reagent film. Reflectance is read at defined times during the reaction period.
- D. Analyzer Software
The CareSide™ Analyzer is software controlled. The software system utilizes a Real-Time Processor (RTP) and a User Interface Processor (UIP). New software will be provided and installed in the field to add additional tests as they are developed and receive U.S. FDA marketing clearance. The Analyzer is controlled through a touchscreen interface. Results are displayed on the interface screen. Results can also be down-loaded on to a 3-1/2 " diskette or on to a computer via a RS-232 port.
#### E. System Calibration
Similar to other automated analyzers in commercial distribution, such as the Abaxis Picollo, the CareSide™ Analyzer is factory calibrated. The user is not expected to perform calibration. The Analyzer is calibrated during each test by automatically reading a black and white reflectance standard and making adjustments if necessary.
Each CareSide™ Cartridge bar-code label contains lot-specific coefficients for a polynomial and is scanned by the CareSide™ Analyzer. The observed reflectance is adjusted by inputting it into the polynomial. The patient result is then calculated from the adjusted reflectance using the polynomial describing the master dose - response curve.
The user performs periodic calibration verification and quality control to confirm that the analyzer maintains its calibration over time.
#### V. Intended Use
#### Intended Use A.
This product is intended for use with Exigent Diagnostics Cartridges when used by laboratory professionals for the in vitro measurement of various clinical chemistry analytes in human blood.
- B. Indications for Use
This product is indicated for in vitro measurement of various clinical chemistry analytes in human blood.
{3}------------------------------------------------
# VI. Technological Characteristics
- A. Similarities
| | CareSide™ Analyzer | Vitros DT 60 II System |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Intended Use | Intended for the measurement<br>of various analytes using<br>reflectance photometry<br>measurements | Same |
| Indications | For <i>in vitro</i> diagnostic use by<br>professional laboratory<br>personnel | Same |
| Type | Quantitative | Same |
| Principle of<br>method | Dry film based: chromogen<br>quantitated by reflectance<br>measurement | Same |
| Compatible<br>reagents | CareSide™ Cartridges | Vitros DT Slides |
| Menu | 7 direct and calculated tests<br>(submitted) | 37direct and calculated tests<br>(cleared) |
| Specimen dilution | Not required | Not required |
| Detector | Reflectance | Reflectance |
| Test time | 4 minutes warm-up (on-<br>board) plus up to 6 minutes<br>test time. | 15 minutes warm-up (off-line)<br>plus 5 minutes test time. |
| Sample Type | serum, plasma, whole blood<br>(whole blood. applied sample,<br>plasma test sample)<br>Tests for urine not currently<br>available | serum, plasma<br>whole blood not acceptable<br>except for hemoglobin<br>Urine for some tests |
| Specimen volume | 85 µl (applied volume) | 10 µl |
| Calibration | Calibration information bar<br>coded on each cartridge.<br>Calibration information may<br>change with each lot | Run Kodak Ektachem DT II<br>calibrators whenever a new<br>slide lot is used or when<br>necessary |
{4}------------------------------------------------
# B. Differences
| | CareSide™ Analyzer | DT 60 II |
|--------------------------|--------------------------------|--------------------------------------------------------------------------|
| Access | Single patient per platter run | Single or multiple patient |
| Queuing | 6 cartridges | Multiple slides |
| Module | Single | Multiple |
| Software updating | 3-1/2 floppy | EPROM (CDM, CLM) |
| Direct blood<br>specimen | Yes, whole blood | No, requires separation of<br>whole blood prior to sample<br>application |
| Pipetting | Not required | Required |
| Accurate<br>pipetting | Not required | Required |
| Reagent<br>pre-warming | Not required | Required |
# C. Comparative Performance Characteristics
| | CareSide™ | DT 60 II |
|-------------------|---------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Precision | Typical total CV 3.1% to<br>8.5% (4 tests evaluated) | Typical total CV 1.5% to<br>5.2% (4 tests evaluated) |
| Relative Accuracy | Mean difference [Vitros - CareSide™] from method comparison means is test specific (3% for 4 tests evaluated) | |
| Linearity | Correlation coefficient >=<br>0.95 | Not provided |
| Interference | Susceptibility to interference is test specific. | Not provided |
| Detection limit | Test specific. | Not provided |
# D. Conclusion
The nonclinical and clinical data provided demonstrate that the CareSide™ Analyzer is as safe, effective, and performs as well as or better than the legally marketed predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration MAR - 6 1998 2098 Gaither Road Rockville MD 20850
Kenneth B. Asarch, Ph.D. · VP Quality Systems and Requlatory Affairs Exigent Diagnostics Inc. 6100 Bristol Parkway Culver City, California 90230
Re : K980056 CareSide™ Analyzer Regulatory Class: II Product Code: JJF, CDQ, CEK, CIX Dated: December 30, 1997 Received: January 6, 1998
Dear Dr. Asarch:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You mav, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਰ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the marret.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitxo diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact. the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" a
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Exigent Diagnostics, Inc. Page 16
CareSide Analyzer Premarket Notification December 30, 1997
# Indications for Use
510(k) Number: Device Name: Indications for use:
rc980056 To be assigned CareSide Analyzer For in vitro diagnostic use by professionals with Exigent Diagnostics Cartridges for the in vitro measurement of various clinical chemistry analytes in human blood.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div align="center">(Division Sign-Off)</div> Division of Clinical Laboratory Devices |
|---------------|---------------------------------------------------------------------------------------|
| 510(k) Number | |
| Prescription Use | OR | Over-The-Counter Use |
|----------------------|----|--------------------------|
| (Per 21 CFR 801.109) | | (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.