The ActiveLock Modular Femoral Hip System is intended for primary or revision reconstruction, with or without bone cement, of the femoral portion of severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present.
Device Story
ActiveLock Modular Femoral Hip System is a modular hip prosthesis for primary or revision reconstruction. System comprises interchangeable components: stem, body, actuator, and femoral head. Components are locked together to form a single femoral hip prosthesis. Stem is cylindrical, fluted, and straight/curved; body is wedge-shaped with porous CpTi plasma spray or HA coating; actuator is a NiTi shape memory alloy locking component; femoral head is CoCr alloy. Device is implanted by orthopedic surgeons in a clinical/surgical setting. Modular design allows customization of fit to patient anatomy. System provides mechanical fixation to the proximal metaphyseal region of the femur. Output is a reconstructed hip joint, intended to restore function and reduce pain in patients with severe hip joint disability.
Clinical Evidence
Bench testing only. Static and fatigue testing performed to evaluate mechanical integrity. Corrosion testing performed to evaluate material resistance. No clinical data provided.
Technological Characteristics
Modular femoral hip prosthesis. Materials: Ti-6Al-4V ELI (ASTM F136), Co-Cr-Mo (ASTM F1537), NiTi shape memory alloy. Surface treatments: porous CpTi plasma spray or HA coating. Modular assembly via modified Morse taper and NiTi actuator locking mechanism. No software or electronic components.
Indications for Use
Indicated for patients with noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, revision procedures after treatment failure, and nonunion/fractures of the proximal femur with head involvement unmanageable by other techniques.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
510(k) Summary prepared December 30, 1997
APR - 3 1998
# Manufacturer Identification
Metagen, LLC 428 Technology Drive East Menomonie, Wisconsin 54751
# Official Contact Person
Wesley D. Johnson President Telephone (715) 232-4880 Fax (715) 235-9570
# DEVICE IDENTIFICATION
# Proprietary Name
ActiveLock™ Modular Femoral Hip System
# Common Name
Modular Hip Replacement System.
# Classification Name and Reference
Hip joint metal/ceramic/polymer semi-constrained or nonporous uncemented prosthesis. 21 CFR 888.3353 [HA version]
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis. 21 CFR 888.3358 [TPS version]
# Proposed Regulatory Class
The Orthopedic Device Section of the Surgical and Rehabilitation Devices Panel of FDA has recommended Class II, Performance Standards, for this device.
### Device Product Codes
87 JDO - Device, Fixation, Proximal Femoral, Implant
# SUBSTANTIAL EQUIVALENCE INFORMATION
The ActiveLock M Modular Femoral Hip System demonstrates substantial equivalence, relative to the intended use, materials and design, to several pre-amendment and post-amendment
{1}------------------------------------------------
devices, including implants marketed by Biomet (Impact, K942027; Mallory-Head Modular Porous Series, K921181), Joint Medical Products Corporation/Johnson & Johnson (S-ROM, K851422, K912713, K913231), Wright Medical Technology (Infinity, K942115), and Smith & Nephew Richards (RMHS, K912593).
### DEVICE DESCRIPTION
### Intended Use
The ActiveLock Modular Femoral Hip System is intended for primary or revision reconstruction, with or without bone cement, of the femoral portion of severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present.
#### Design
The ActiveLock M Modular Femoral Hip System is a matrix of interchangeable components (STEM, BODY, ACTUATOR, FEMORAL HEAD) that can be locked together to form a single femoral hip prosthesis. A description of the components is given below.
### Stem
The Stem component is cylindrical, straight, and fluted with a slot in the coronal plane that extends along the distal third of its length. The proximal aspect of the Stem is a smooth tubular cvlinder designed to mate with the Body component. The Stem component is manufactured from wrought Ti 6Al-4V alloy and is offered in a variety of lengths, straight or curved, to accommodate primary and revision surgery.
### Body
The Body component is wedge-shaped and symmetrical in the coronal plane. Design variations of the Body include primary, revision, and calcar replacement components. All Body designs incorporate identical cylindrical bores for modular attachment to the Stem and the same modified Morse taper for attachment of Metagen Femoral Heads. All Body designs are available with either porous CpTi plasma spray or plasma-sprayed smooth HA (hydroxyapatite) coating. The Body component is manufactured from wrought Ti 6A1-4V alloy. When attached to the Stem component, the Body component provides fixation to the proximal metaphyseal region of the femur.
### Actuator
The Actuator is manufactured of Nickel-Titanium alloy (NiT). It is a cylindrical, tubular locking component.
{2}------------------------------------------------
## Femoral Head
The Femoral Head is a CoCr alloy articulating component with a modified female Morse taper. The components are spherically ground and polished to 22 mm, 26 mm, and 32 mm diameters, and are available in four offsets -- 5 mm, 0 mm, +5 mm, and +10 mm.
Components are matched according to four distinct "FAMILY" designations. The four families are the "A" family, "B" family, "C" family, and the "D" family. The "A" family corresponds to the smallest sizes, the "B" and "C" family to the mid-range sizes, and the "D" family to the largest sizes. The Stem and Body components are only interchangeable within their respective families, and can only be locked by the Actuator in that family.
# Composition/Materials
The ActiveLock "" Modular Femoral Hip System is composed of titanium alloy (Ti-6AI-4V ELI) conforming to ASTM Specification F136. The Femoral Head is made from wrought Co-Cr-Mo alloy conforming to ASTM Specification F1537.
The Actuator of the ActiveLock™ Modular Femoral Hip System is made from NiTi alloy, a nickel-titanium shape memory alloy. NiTi alloy is currently being used in a number of implant devices and has a long clinical history as an implant material. In addition, NiTi alloy is being used in orthopedic applications for bone anchors.
# Mechanical Characterization
The ActiveLock " Modular Hip System has been evaluated by use of the static and fatigue tests. It has been shown to be substantially equivalent to the predicate devices.
### Corrosion Testing
The components and materials of the ActiveLock™ Modular Hip System have been tested in corrosion. It has been shown that the materials of the system are resistant to corrosion.
{3}------------------------------------------------
Canada Petersi a
# Summary: Table of Substantial Equivalence
| | Subject<br>Device | Predicate Devices | | | | | |
|-------------------------------------------|-------------------------------|---------------------|-------------------------------|--------------------------------------------|-----------------------|-------------------------------------------------|--|
| MANUFACTURER | Metagen | Biomet | Joint<br>Medical/<br>J&J | Wright<br>Medical<br>Technology | S&N<br>Richards | | |
| SYSTEM NAME | Modular Femoral<br>Hip System | Impact<br>(K942027) | Mallory-<br>Head<br>(K921181) | S-ROM<br>(K851422,<br>K912713,<br>K913231) | Infinity<br>(K942115) | Modular<br>Hip<br>System<br>(RMHS)<br>(K912593) | |
| DESIGN FEATURES | | | | | | | |
| Cementless fixation | YES | YES | YES | YES | YES | YES | |
| Modular proximal<br>body/stem | YES | YES | YES | YES | YES | YES | |
| Straight cylindrical distal<br>stem | YES | | | YES | YES | YES | |
| Flexible distal stem | YES | YES | YES | YES | YES | | |
| Maximized projected<br>area | YES | YES | YES | YES | YES | YES | |
| Collarless | YES | YES | YES | YES | | YES | |
| Modular head with<br>modified Morse taper | YES | YES | YES | YES | YES | YES | |
| MATERIALS | | | | | | | |
| Ti-6Al-4V | YES | YES | YES | YES | YES | YES | |
| NiTi alloy Actuator | YES | | | | | | |
| HA on proximal surfaces | YES | | | YES | | | |
| TPS on proximal<br>surfaces | YES | YES | YES | YES | YES | | |
| Co-Cr-Mo Modular Head | YES | YES | YES | YES | YES | YES | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized lines or strokes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 3 1998
Mr. Wesley D. Johnson ·President Metagen, LLC 428 Technology Drive East Menomonie, Wisconsin 54751
K980020 Re : ActiveLock™ Modular Hip System Requlatory Class: II Product Codes: MEH, LPH, and JDI Dated: December 30, 1997 Received: January 5, 1998
Dear Mr. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on consideration of the specific design of stem and coating composition detailed in this application. You may, therefore, market the device, subject to the general controls provisions of the Act and the following limitation:
You may not label or in any way promote the HA coated versions of these devices for "biological attachment, enhanced clinical or radiographic performance, enhanced fixation and/or long-term stable fixation." The data presented support equivalence with no additional claims over a conventional press-fit hip prosthesis (i.e., mechanical interlock, only) --------------------------------------------------------------------------------------------------------------------------------------------------. . . . . . . . . . .................
Additional limitations for more specific claims of safety and effectiveness may be forthcoming. Should additional limitations be applied you will be contacted in writing to inform you of the additional labeling limitations.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
Page 2 - Mr. Wesley D. Johnson
You may market your device under the above limitations as class II devices. These devices would be considered not substantially equivalent to a legally marketed predicate device if labeled with other intended uses and/or claims of safety or effectiveness. Any other intended uses or claims may cause the device to be classified into Class III under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing.
Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to one regaration onerozon" (21 CFR 807.97) - Other general information on your responsibilities under the Act may be
{6}------------------------------------------------
Page 3 - Mr. Wesley D. Johnson
obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
a M. Witten, Ph.D., M.D. Celi Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Device Name: ACTIVELOCK™ MODULAR FEMORAL HIP SYSTEM
The ActiveLock™ Modular Femoral Hip System is intended for primary or revision reconstruction, with or without bone cement, of the femoral portion of severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present. Clinical Indications for Use are:
- noninflammatory degenerative joint disease including osteoarthritis and avascular 1. necrosis;
- 2. rheumatoid arthritis;
- 3. correction of functional deformity;
- 4. revision procedures where other treatments or devices have failed; and,
- 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------------------------|---------|
| (Division Sign-Off)<br>Division of General Restorative Devices<br>510(k) Number | K980020 |
Prescription Use X OR Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.