IGEL(R) 56 (HEFILCON C) SOFT (HYDROPHLIC) CONTACT LENS
Applicant
Igel Vision Care Pte , Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Jul 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The Igel®56 (hefilcon C) Soft (hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism, for the spherical lens up to 1.50 diopters that does not interfere with visual acuity, and for the toric lens up to 7.00 diopters) in not-aphakic persons with non-diseased eyes. The lens may be disinfected using chemical disinfecting systems only.
Device Story
Igel® 56 is a hemispherical, flexible, transparent soft contact lens made of hefilcon C (copolymer of 2-hydroxyethylmethacrylate and N-vinyl pyrrolidinone). Used for daily wear to correct refractive ametropia (myopia, hyperopia, astigmatism) by focusing light rays on the retina. Lens is 56% water by weight when hydrated. Available in spherical and toric configurations with variable base curves and powers. Operated by patient for daily vision correction. Disinfection requires chemical systems only. Benefits include improved visual acuity for patients with non-diseased eyes.
Clinical Evidence
Clinical trial of 37 subjects over 4 weeks of daily wear. Primary endpoints: safety (adverse events, slit lamp findings) and efficacy (visual acuity, lens cleanliness). Results: 0 adverse events; 91% of lenses clinically clean; visual acuity maintained within 1 line of initial best-corrected acuity. Slit lamp findings showed no significant ocular irritation or edema. Study concluded device is safe and effective.
Technological Characteristics
Material: hefilcon C (copolymer of 2-hydroxyethylmethacrylate and N-vinyl pyrrolidinone). Water content: 56%. Refractive index: 1.41 (hydrated). Oxygen permeability (Dk): 21x10^-11 (cm/sec)(ml O2/ml x mmHg) at 35°C. Specific gravity: 1.16. Tensile strength: 2.85 kg/cm2. Modulus of Elasticity: 5.20 kg/cm2. Form factor: hemispherical, flexible shell. Sterilization: validated manufacturing process.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, and astigmatism in non-aphakic persons with non-diseased eyes. Spherical lens correction up to 1.50 diopters; toric lens correction up to 7.00 diopters.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Gold Medalist Toric lenses
Submission Summary (Full Text)
{1}------------------------------------------------
K974837, Amendment
510(k) SUMMARY FOR FREEDOM OF INFORMATION IGEL® 56 (hefilcon C) SOFT (hydrophilic) CONTACT LENS FOR DAILY WEAR Igel Vision Care PTE, Ltd 1. Submitted by: 139 Joo Seng Road, #06-01 ATD Centre Singapore 368362 John M. Szabocsik, Ph.D. Contact : official agent Szabocsik and Associates 203 N. Wabash, Ste 1200 Chicago, IL 60601 (312) 553-0828 June 5, 1998 2. Date prepared: 3. Device: Igel® 56 (hefilcon C) Soft (hydrophilic) Common Name Contact Lens Igel® 56 (hefilcon C) Soft (hydrophilic) Trade Name Contact Lens Class II (Performance Standards) 4. Classification 21 CFR 886.5925 Soft (hydrophilic) contact lens This product is substantially 5. Substantial equivalent to other currently marketed equivalence hefilcon lenses, such as Gold Medalist Toric lenses The Igel® 56 (hefilcon C) soft 6. Device (hydrophilic) contact lens is a description hemispherical, flexible transparent shell of the following dimensions: 14 - 2mm Chord diameter:
Center thickness: Sphere: 0.09mm (at-3.00 D) variable optic zone Optic Zone: with power Base Curve: 8.60mm -0.50D to -8.00D (in Power : 0.25D steps): -8.50D -12.00D (in to 0.50D steps); +0.50D to +6.00D (in 0.25D
steps)
{2}------------------------------------------------
Chord diameter: 14 - 6mm Toric: 0.15mm Center thickness: 8.45mm Optic Zone: 8.70mm Base Curve: Sphere: Plano to -8.00 D (in 0.25 D steps) Plano to +6.00 D (in 0.25 D steps) up to 7.00 D Cvlinder: 80, 90, 100, 160, 170, 180, 10 Axis: & 20°
The lens material (hefilcon C) is a copolymer of 2 hydroxethylmethacrylate and N-vinyl pyrrolidinone. The blue tinted lens also contains D&C Green #6. The apparent color of the visibility tint may decrease slowly following repeated disinfection, but this will not affect the safety or performance of the lenses. When fully hydrated, the lens is 56% water by weight.
The Igel®56 (hefilcon C) Soft 7. Intended use: (hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism, for the spherical lens up to 1.50 diopters that does not interfere with visual acuity, and for the toric lens up to 7.00 diopters) in not-aphakic persons with non-diseased eyes. The lens may be disinfected using chemical disinfecting systems only.
8. Comparison to predicate devices: see following table
{3}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE
| Material | Igel® 56<br>hefilcon C | Gold Medalist™ Toric<br>hefilcon C |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | random copolymer of<br>2-hydroxyethyl<br>methacrylate and N-<br>vinyl-2-pyrrolidone | random copolymer of<br>2-hydroxyethyl<br>methacrylate and N-<br>vinyl-pyrrolidone |
| Water Content | 56% in buffered<br>saline | 57% in normal saline |
| Specific<br>Gravity | 1.16 | 1.1 |
| Refractive<br>Index | 1.41 (hydrated) | 1.41 |
| Light<br>Transmittance<br>(visible) | 94.5% (clear)<br>90.3% (blue handling<br>tint) | at least 90% |
| Oxygen<br>permeability | $21x10^{-11}$ (cm/sec)(ml<br>O₂/mlxmm Hg)<br>Measured at 35°C<br>(Fatt method). | $17x10^{-11}$ (cm/sec) (ml<br>O₂/mlxmm Hg) Measured<br>at 35°C (Fatt<br>method). |
| Actions | When placed on the<br>human cornea, the<br>hydrated Igel® 56<br>(hefilcon C) Soft<br>(hydrophilic)<br>Contact Lens acts as<br>a corrective<br>refracting medium to<br>focus light rays on<br>the retina.<br> | In its hydrated<br>state, the Bausch &<br>Lomb Gold Medalist™<br>Toric (hefilcon C)<br>Visibility Tinted<br>Contact Lens when<br>placed on the cornea<br>acts as a corrective<br>refracting medium to<br>focus light rays on<br>the retina. |
| Chord Diameter* | 14.2mm | 12.0 to 18.0mm |
| Center<br>thickness* | 0.09mm (at -3.00D) | 0.02mm to 1.0mm |
| Base Curve* | 8.6mm | 8.3mm to 9.2mm |
| Powers<br>(Spherical)<br>(Cylinder) | -12.00 to +6.00D<br>up to 7.00D | +20.00 to -20.00D<br>up to 5.00D |
| Optical Zone | vary with power | |
| Chord diameter: | 14.2mm | |
| Center thickness: | 0.09mm (at-3.00 D) | |
| Optic Zone: | variable optic zone with<br>power | |
| Base Curve: | 8.60mm | |
| Power: | -0.50D to -8.00D (in 0.25D<br>steps) -8.50D to -12.00D (in<br>0.50D steps) +0.50D to +6.00D<br>(in 0.25D steps) | |
{4}------------------------------------------------
Introduction
The Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lens is a spherical lens for the correction of nearsightedness (myopía) and farsightedness (hyperopia). A toric version of the same lens is able to correct for astigmatism from 0.50D to 7.00D.
Contained in the submission are comparisons of the product to the predicate device, information on the chemistry and manufacturing, results of toxicological and microbiological tests, and the report of a clinical trial of 37 subjects, who have used the product over a period of four weeks .
I. Chemistry and Manufacturing
The lens material (hefilcon C) is a copolymer of 2 hydroxethylmethacrylate and N-Vinyl pyrrolidinone. The blue tinted lens also contains D&C Green #6. The apparent color of the visibility tint may decrease slowly following repeated disinfection, but this will not affect the safety or performance of the lenses. When fully hydrated, the lens is 56% water by weight.
The Igel® 56 (hefilcon C) Soft (hydrophilic) Spherical Contact Lens is a hemispherical, flexible transparent shell of the following dimensions:
The Igel® 56 (hefilcon C) Soft (hydrophilic) Toric Contact Lens has the following characteristics: Chord diameter: 14.6mm 0.15mm Center thickness: 8.45mm Optic Zone: 8.70mm Base Curve: Plano to ~8.00 D (in 0.25 D steps) Sphere: Plano to +6.00 D (in 0.25 D steps) up to 7.00 D Cylinder: 80, 90, 100, 160, 170, 180, 10 & 20 Axis: Dynamic Stabilizing Stabilizing Mechanism: Sectors
{5}------------------------------------------------
| | The physical properties of the lens (sphere or toric) |
|------------------------|----------------------------------------------------------------------------|
| are as follows: | |
| Specific Gravity: | 1.16 |
| Refractive index: | 1.41 (hydrated) |
| Light transmission: | 94.5% (clear)<br>90.3% (blue handling<br>tint) |
| Surface character: | hydrophilic |
| Water content: | 56% weight in normal<br>buffered saline |
| Tensile strength: | 2.85 kg/cm2 |
| Elongation: | 72% |
| Modulus of Elasticity: | 5.20 kg/cm2 |
| Oxygen permeability: | Dk: 21x10-11(cm/sec)(ml<br>O2/mlxmm Hg) Measured at<br>35°C (Fatt method). |
A. Extractables
Samples of lenses were extracted in saline, and the extracts analyzed by high performance liquid chromatography to determine if any monomers were extracted from the material. There were no extractables detected.
B. Process validation
The manufacturing process for this cast-moulded lens was shown to be valid both for lens quality and sterility.
# II. Toxicology
The toxicological testing is summarized below. The lens material was shown to be non-toxic in all tests .
Agar Overlay Cytotoxicity: A.
Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses were tested in a direct contact cytotoxicity The lenses were noncytotoxic. assay.
Systemic toxicity: B.
Saline and cottonseed oil extracts of Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses were evaluated for systemic toxicity by intraperitoneal (ip) injection in healthy mice, 50ml/kg body weight. The animals were observed over a 72 hour period, and showed no difference from control animals. The lenses passed the test requirements, that there be no difference between the response of test and control animals.
{6}------------------------------------------------
Acute Ocular irritation: C.
Saline and cottonseed oil extracts of Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses were evaluated for ocular irritation by instillation into the inferior ocular cul-de-sac of rabbits. The eyes were examined over a 72 hour period and showed no irritation.
# III. Microbiology
A -Sterility
The Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses passed the requirements of sterility testing.
B. Stability
Accelerated shelf-life testing has demonstrated a shelf-life in excess of 16 months. Sterility stability testing, accelerated and real-time, continue.
# IV. Clinical Studies
A clinical trial of 4 weeks usage of the Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses by 37 subjects, wearing lenses on a daily wear schedule, showed that the product is substantially equivalent to other lenses available on the market. The clinical summary follows.
The study was conducted over 1 month of wear, subjects being seen initially, and after 1, 2 and 4 The study was initiated July 11, 1997 and weeks. completed August 19, 1997.
Two (2) investigators enrolled a total of 37 test subjects. The age range of the test population was from 17 to 61, with 26 (70.3%) females and 11 (29.7%) males.
Of the 37 subjects, 35 (94.6%) completed 1 month of wear, and 2 (5.4%) discontinued.
FINDINGS
অ SAFETY :
(1) Adverse Reactions
The FDA regulations for medical devices (21 CFR 812.3) define an unanticipated adverse device effect as:
"any serious adverse effect on health or safety or any life-threatening problem or death caused by, or associated with, a device, if that effect, problem or death was not previously
{7}------------------------------------------------
identified in nature, severity or degree of incidence in the investigational plan or application (including a supplementary plan or application), or any other unanticipated serious problem associated with a device that related to the rights, safety or welfare of the subject"
There were no adverse events during this study.
Slit Lamp Findings: (2)
A positive slit lamp finding is a routinely occurring complication that can be expected with or without the presence of contact lenses. The degree of severity may range from very slight, representing no medical concern, to serious, requiring medical treatment.
Table B shows the incidence of slit lamp findings in the investigational group at the initial and final visits.
The slit lamp finding listed as "Other" at the initial visit was an old healed neovascularization.
### TABLE B
# INCIDENCE OF SLIT LAMP
### FINDINGS
b
| FINDING | INITIAL | FINAL |
|--------------|---------|-------|
| NO FINDINGSa | 51.4 | 58.6 |
| EDEMAb | 0.0 | 0.0 |
| NEOVASCULAR | 18.9 | 11.4 |
| STAINING | 5.4 | 1.4 |
| INJECTION | 0.0 | 0.0 |
| TARSAL | 48.6 | 40.0 |
| OTHER | 1.4 | 0.0 |
» Percent of eyes examined with no findings,
regularly scheduled visits only
Percent of eyes with finding, regularly scheduled visits only
# (3) Symptoms, Problems and Complaints:
Symptoms, problems and complaints were reported by the investigators at each visit. Lens awareness, handling problems and reading problems were the most frequently reported symptoms among all subjects. Table C shows the overall incidence of these selected symptoms .
{8}------------------------------------------------
The "other" symptoms were dryness (reported by 2 subjects); torn lenses reported by 1 subject.
## TABLE C
# SELECTED SYMPTOMS, PROBLEMS AND COMPLAINTS
| SYMPTOM | Incidence |
|-------------------|-----------|
| NONE* | 62.6% |
| LENS AWARENESSb | 14.5% |
| HANDLING PROBLEMS | 8.4% |
| READING PROBLEMS | 7.9% |
- 다 Percent of eyes examined with no findings, regularly scheduled visits only
- D Percent of eyes with finding, regularly scheduled visits only
### Discontinuations: ( 4 )
Throughout the study, 2 subjects (5.4%) were discontinued.
### b. EFFICACY :
#### (1) Visual Acuity:
All but 2 eyes had a final visual acuity within 1 line of the initial best corrected acuity. The appropriate acuity was achieved by all eyes, since those two eyes were in subjects fit for monovision.
(2) Wear Time:
Wear time remained essentially unchanged over the one month of the study, indicating continuing comfort and cleanliness with the investigational solution.
#### (3) Lens Cleanliness:
Lenses were evaluated at each visit, according to a modified Rudko classification. Overall, the reports of the Rudko evaluation showed that 91% of test lenses were clinically clean during the study. (Rudko grades I and II are considered clinically clean. }
{9}------------------------------------------------
These results confirm that the Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses are effective in the correction of myopia and hyperopia.
# Gender Comparisons:
The overall test population was 70.3% female, 29.7% male, and the visit distribution over the study was 69% female, 31% male. Overall, there were no slit lamp findings at 58% of the visits; among females, at 63%, among males at 47% of the visits. There were no symptoms reported at 61% of the visits; among females, no symptoms were reported at 64%, among males, at 56%. There is no significant difference in the findings, and no further analysis was warranted. Because of the small sample size, the gender analysis may be statistically irrelevant.
# OVERALL CONCLUSION OF THE CLINICAL STUDY:
The data of the clinical trial confirm that the Igel® 56 (hefilcon C) Soft (hydrophilic) Contact Lenses are substantially equivalent to currently marketed lenses in safety and efficacy.
{10}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
6 1998 JUL
Igel Vision Care PTE, Ltd. c/o John M. Szabocsik, Ph.D. Szabocsik and Associates 203 N. Wabash, St. 1200 Chicago, IL 60601
Re: K974837
Trade Name: IGEL ® 56 (hefilcon C) soft (hydrophilic) Contact Lens for Daily Wear (clear or visitint.cast molded)
Regulatory Class: II Product Code: 86 LPL Dated: June 8, 1998 Received: June 9, 1998
Dear Dr. Szabocsik:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{11}------------------------------------------------
Page 2 - Dr. John Szabocsik
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
June 29, 1998
Page 1 of 1
510(k) NUMBER (IF KNOWN) K974837
IGEL® 56 (hefilcon C) SOFT DEVICE NAME (HYDROPHILIC) CONTACT LENS
INDICATIONS FOR USE
The Igel®56 (hefilcon C) Soft (hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism, for the spherical lens up to 1.50 diopters that does not interfere with visual acuity, and for the toric lens up to 7.00 diopters) in not-aphakic persons with non-diseased eyes. The lens may be disinfected using chemical disinfecting systems only.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Mig-chen Star
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR
Over-The-Counter-Use _________________________________________________________________________________________________________________________________________________________
21 CFR 801.109)
(Per
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.