K974819 · Select Medical Systems, Inc. · HHK · Mar 23, 1998 · Obstetrics/Gynecology
Device Facts
Record ID
K974819
Device Name
SELECTCELLS(TM) MINI
Applicant
Select Medical Systems, Inc.
Product Code
HHK · Obstetrics/Gynecology
Decision Date
Mar 23, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1175
Device Class
Class 2
Indications for Use
Same as the Pipelle® (predicate device), the SelectCells™Mini (new device) is a single use, sterile, disposable endometrial sampling device designed to be used for obtaining a histologic biopsy of the uterine mucosal lining or specimen of the uterine menstrual content. The specimen obtained is then used for the following: (i) evaluation of infertility conditions, menstrual disorders, postmenopausal bleeding, abnormal cytology suspected of being of endometrial origin, hormonal replacement therapy; (ii) detection of endometrial carcinoma; (iii) diagnosis of luteal defect; (iv) endometrial dating and,(v) microscopic examination.
Device Story
Single-use, sterile, disposable endometrial suction curette. Comprises flexible polypropylene sheath and acetal copolymer rod with molded piston. Operated by physician in clinical setting. Device inserted into uterus; piston retracted to create negative pressure, aspirating endometrial tissue into sheath lumen through side opening. Device rotated and moved to-and-fro to collect sample. Sample expelled into receptacle for histologic/microscopic analysis. Smaller distal diameter compared to predicate improves patient comfort, reducing cramping and pain during insertion. Provides diagnostic tissue samples for evaluation of uterine conditions.
Clinical Evidence
Prospective, randomized, blind, controlled clinical study on 10 patients (20 samplings). Each patient served as own control, comparing SelectCells™ Mini to Pipelle®. Primary endpoint: histologic biopsy quality. Results showed 100% concordance in pathology diagnoses between devices. Sample preservation was uniform. Predicate yielded slightly larger samples, but diagnostic efficacy was equivalent. New device demonstrated improved patient comfort (less cramping/pain).
Technological Characteristics
Materials: Polypropylene sheath, acetal copolymer rod/piston. Principle: Manual suction via piston displacement. Dimensions: 29.8 cm overall length; distal 13 cm has 1.5 mm ID/1.9 mm OD, proximal section has 2.6 mm ID/3.1 mm OD. Features: 7 graduation marks, ergonomic finger grip. Single-use, sterile. Mechanical device; no software or energy source.
Indications for Use
Indicated for endometrial biopsy or menstrual content extraction in female patients. Applicable for infertility, menstrual disorders, postmenopausal bleeding, abnormal cytology, HRT monitoring, endometrial carcinoma detection, luteal defect diagnosis, and endometrial dating. Useful for patients with narrow cervix or cervical canal, including postmenopausal women. Note: sample quantity is reduced due to smaller device diameter.
Regulatory Classification
Identification
An endometrial suction curette is a device used to remove material from the uterus and from the mucosal lining of the uterus by scraping and vacuum suction. This device is used to obtain tissue for biopsy or for menstrual extraction. This generic type of device may include catheters, syringes, and tissue filters or traps.
Predicate Devices
Pipelle®
Submission Summary (Full Text)
{0}------------------------------------------------
K974819
P195
```
SELECT MEDICAL SYSTEMS INC.
Section N
Page 1 of 5
Food and Drug Administration Center For Devices and Radiological Health (HFZ-401) 9200 Corporate Boulevard Rockville MD 20850
MAR 2 3 1998 510(k)Summary
| Submitter | Select Medical Systems, Inc<br>2 Winter Sport Lane<br>P.O. Box 966<br>Williston VT 05495-0966<br>Ph (802) 862-1017 Fx (802) 862-3767 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Monique Girard |
| Date of Summary | 12-22-97<br>MM DD YY |
| Device proprietary name | SelectCells™ Mini |
| Device common name | Endometrial sampling device |
| Device classification name | Endometrial suction curette (per 21 CFR section 884.1175) |
| Predicate device name | Pipelle® |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).
## Intended use
Same as the Pipelle® (predicate device), the SelectCells™Mini (new device) is a single use, sterile, disposable endometrial sampling device designed to be used for obtaining a histologic biopsy of the uterine mucosal lining or specimen of the uterine menstrual content. The specimen obtained is then used for the following: (i) evaluation of infertility conditions, menstrual disorders, postmenopausal bleeding, abnormal cytology suspected of being of endometrial origin, hormonal replacement therapy; (ii) detection of endometrial carcinoma; (iii) diagnosis of luteal defect; (iv) endometrial dating and,(v) microscopic examination.
## Description of new device and comparison with predicate device
The SelectCells Mini, like the Pipelle, comprises a long clear, flexible, polypropylene sheath and a rod made of acetal copolymer onto which a piston is molded.
The sheath of the SelectCells Mini is a one part tube whose distal half has a smaller diameter than its proximal half. The sheath of the Pipelle is a one part tube with the same diameter throughout.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a white, curved shape against a black background. The shape resembles a stylized 'S' or a winding path. The contrast between the white shape and the black background makes the shape stand out. The image is simple and abstract.
K974819
p295
Section N Page 2 of 5
The device opening, by which the specimen is to be aspirated, is located at 2.5 mm distance (3 mm for Pipelle®) from the extreme distal tip of the sheath. The extreme distal end of the sheath is rounded and closed.
A piston is molded on the rod at almost its half length point (piston located at the distal end of the rod for Pipelle). This rod can be moved forward and backward within the proximal half length of the sheath lumen. The piston and rod are prevented from being totally withdrawn from the sheath during manipulation by the tapered proximal lumen end of the sheath (for Pipelle, by an identation in the sheath located at its proximal end).
The new device has an ergonomic 2.5 cm long finger grip which is molded onto the rod that protrudes the proximal end of the sheath. The Pipelle has a cylindrical 2.4 cm long knob which is glued to the rod that protrudes the proximal end of the sheath. This finger grip (Knob, for Pipelle) is used to move the rod within the sheath.
The SelectCells™Mini bears seven (7) coloured graduation marks with figures that are placed at 4. 5. 6. 7. 8. 9. and 10 cm distance from the extreme distal tip of the sheath. The predicate device bears four (4) coloured graduation marks with figures and three (3) marks in the form of a line that are placed at 4, 5, 6, 7, 8, 9 and 10 cm distance from the extreme distal tip of the sheath. The new and predicate devices bear an arrow on their sheaths indicating the location of the device opening. The graduation marks on the sheath serve as a guide for insertion depth of the device into the uterine cavity.
### Function of the device
As the Pipelle, the SelectCells Mini is to be used to remove material from the endometrium and from the mucosal lining of the uterine cavity by suctioning. The specimen obtained is then used for histologic biopsy or microscopic examination.
With the patient in position: the uterus is probed for depth and direction with a uterine sound. The uterine cervix lip is grasped with the proper instrument and according to its direction. The cervical curvature is straightened.
The device, with the piston rod fully depressed at the extreme distal end of the sheath, is inserted into the uterus up to the fundus (according to its depth as previously determined).
The piston is moved toward the proximal end of the sheath by pulling the piston rod until it stops. The device is then rotated through 360 degrees while making a gentle "to and fro" movement a few times. This procedure creates a negative pressure inside the sheath aspirating the sample into the sheath lumen through the device opening.
The device is then slowly withdrawn from the uterus. The sample is expelled in a specimen receptacle and sent to the laboratory for analysis.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a white, curved shape against a black background. The shape resembles a stylized flame or a winding path. The contrast between the white shape and the black background creates a stark and simple visual.
K974819
P395
Section N Page 3 of 5
# Technical characteristics - Comparison summary between new and predicate devices
| | SelectCells™ Mini (new device) | Pipelle® (predicate device) |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Length | 29.8 cm (overall)<br>25.6 cm (effective) | 26.5 cm (overall)<br>23.5 cm (effective) |
| I.D. (inner diameter)<br>O.D. (outer diameter)<br>I.D.<br>O.D. | 1.5 mm (distal 13 cm)<br>1.9 mm (distal 13 cm)<br>2.6 mm (proximal 13.7 cm)<br>3.1 mm (proximal 13.7 cm) | 2.6 mm (throughout)<br>3.1 mm (throughout) |
The minor difference in length and smaller distal 13 cm I.D. and O.D. do not affect the function, safety and efficacy of the new device as demonstrated in the Clinical Performance Data below. The new device has been designed with a smaller diameter on its distal portion to ease insertion and increase patient comfort during the procedure.
| | SelectCells™ Mini (new device) | Pipelle® (predicate device) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Graduation<br>marks | At 4, 5, 6, 7, 8, 9, and 10 cm distance<br>from the extreme distal tip of the<br>sheath.<br>An arrow indicating location of device<br>opening. | At 4, 5, 6, 7, 8, 9 and 10 cm<br>distance from the extreme<br>distal tip of the sheath.<br>An arrow indicating location<br>of device opening. |
The new device bears seven (7) graduation marks with figures as compared to the predicate device which has four (4) marks with figures and three (3) in the form of a line making the graduation marks clearer on the new device than the predicate device.
| | SelectCells ™ Mini (new device) | Pipelle® (predicate device) |
|-----------------------|--------------------------------------------------------------------------------|--------------------------------------------|
| Rod stop<br>mechanism | Proximal sheath end is tapered | Proximal sheath end bears<br>an identation |
| Rod grip | Molded finger grip | Glued knob |
| | These differences are minor and reflect only a preference in<br>manufacturing. | |
| Piston location | At half length on the rod | At distal end of the rod |
Because the new device has a smaller diameter on almost half of the sheath distal portion than its proximal portion, the piston had to be molded at half length point on the rod. This difference is minor and safety/efficacy performance of the new device is not affected as demonstrated in the Clinical Performance Data below.
| Design | Similar to Pipelle | Similar to SelectCells Min |
|-----------|--------------------|----------------------------|
| Materials | Similar to Pipelle | Similar to SelectCells Min |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a white, winding shape against a black background. The shape resembles a river or a winding road, curving from the bottom towards the top of the frame. The contrast between the white shape and the black background creates a stark and simple visual.
1974819
Section N Page 4 of 5
Non-clinical performance data
SelectCells™ Mini (new device)
Negative pressure
Average of 20,1 mm of water
Pipelle® (predicate device)
Average of 19.3 mm of water
The amount of negative pressure generated by the new and predicate devices was evaluated by an independent testing laboratory. On five sterile specimens of each device, negative pressure data were taken in triplicate using an open-tube water manometer. The sampling end (distal end) of the device was positioned to a designated position ensuring that each specimen was tested with the same amount of head space and displaced volume. A pressure relief valve was then opened to relieve any pressure built up in the system due to insertion of the device. Each measurement was obtained upon device piston rod displacements of ten centimeters. With the relief valve closed, the piston rod was pulled back continuously, until the maximum of ten centimeters on the devices' graduation marking was reached. The difference in the water levels in the manometer was then documented.
The reproducibility of the measurements was found to be very acceptable. Based on the data, it is apparent the negative pressure amount produced by the SelectCells Mini is greater than the negative pressure amount produced by the Pipelle.
These results indicate that the new device generates slightly more negative pressure than the predicate device thus, is able to suction an endometrial specimen as well as the predicate device. The Clinical Performance Data section supports this statement as well.
In conclusion, the results indicate that the new device is substantially equivalent to the predicate device.
| | SelectCells™ Mini (new device) | Pipelle® (predicate device) |
|---------------------|--------------------------------|-----------------------------|
| Tensile force | Average of 13.7 lbs | Average of 13.9 lbs |
| Tensile strength | Average of 3940 PSI | Average of 4010 PSI |
| Elongation at break | Average of 364% | Average of 605% |
The tensile properties of the new and predicate devices were evaluated by an independent laboratory per ASTM D 638-96. Five sterile specimens of each device were tested. The initial test used a 7" and 6" grip separation for the new device and the predicate device respectively. Due to the elongation properties of polypropylene beyond the limits of the universal tester. testing was performed under modified ASTM D 638-96 by choosing a smaller grip separation of 3.5".
Tensile force results are equivalent because both devices are made of the same material. There is no question regarding safety and efficacy of the new device compared to the predicate device as the results are equivalent. Tensile strength results, which measures a finite point on a cross section of the products, vary by less than 2% and are statistically insignificant. As expected, the new device has one half elongation of the predicate device as its diameter stepped down at its half way point. All breaks in the new device occurred at the region proximal to the change in diameter.
Neither the SelectCells Mini nor the Pipelle are subjected to any tensile force, strength, or elongation while using the product for endometrial sampling.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a black square with a white, curved shape inside. The white shape resembles a stylized 'S' or a winding path. The contrast between the black background and the white shape makes the shape stand out.
K974819
p595
Section N Page 5 of 5
## Clinical performance data
The safety and efficacy of the new device have been evaluated in a clinical study. To demonstrate that the new device provides an adequate endometrial specimen for a histological biopsy of the endometrium, a blind, controlled, prospective, randomized clinical study has been conducted on 10 patients.
Twelve (12) female patients aged 18 years or older having a clinical indication for endometrial sampling were recruited for the study. Results from two (2) patients were excluded as one patient could not be sampled with either device due to a cervical obstruction while the other patient had inadequate tissue obtained with both devices to evaluate.
Twenty (20) endometrial samplings were performed. Each of the ten (10) patients had her endometrium sampled twice. Five (5) patients were sampled with the new device followed by sampling with the predicate device. The other five (5) patients were sampled with the predicate device followed by sampling with the new device. Each patient served as her own control in order to eliminate any variations from patient to patient as to the sample produced.
No Protocol deviations, adverse events or complications occurred which demonstrate the safety of the new device.
In all ten (10) patients who had successful endometrial samplings, the pathology reports were concordant as to the diagnosis obtained with either sampling device. Results on preservation of histologic detail are comparable from both devices. It appears that the sequence of sampling had an effect but in none of the cases was the difference greater than one grade. As expected, the predicate device produced a larger sample size than the new device. The histologic grade results favor the predicate device but not more than by one and one half grades. These differences in the results do not afficacy of the new device as all the ten patient pathology reports demonstrate concordant diagnoses of paired specimens.
The specimen preservation is uniform throughout the specimens.
The new device provided greater patient comfort than the predicate device. The new device produced less cramping, cervical bleeding, cervical and uterine pain than the predicate device.
Based on the results of the Clinical Study, it can be concluded that the new device is substantially equivalent to the predicate device. The new device performed better than the predicate device in terms of patient comfort and diminished intensity of symptoms produced by the procedure itself.
### Substantial equivalence
The intended use, function and technical characteristics along with results obtained from the non-clinical and clinical tests indicate that the new device. the SelectCells™ Mini is substantially equivalent in safety, effectiveness and performance to the predicate device, the Pipelle®.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for the Department of Health. The logo features the words "DEPARTMENT OF HEALTH" in a circular arrangement on the left side of the image. To the right of the text is a symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines. The overall design is simple and conveys a sense of health and well-being.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# MAR 2 3 1998
Ms. Monique Girard Regulatory Affairs Manager Select Medical Systems, Inc. 2 Winter Sport Lane P.O. Box 966 Williston, VT 05495-0966
Re: K974819
SelectCells™ Mini (Endometrial Suction Curette and Accessories) Dated: December 22, 1997 Received: December 23, 1997 Regulatory Class: II 21 CFR 884.1175/Procode: 85 HHK
Dear Ms. Girard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (2) CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1
510 (k) NUMBER (IF KNOWN) : K974819
DEVICE NAME: SelectCells™ Mini
INDICATIONS FOR USE:
Biopsy to obtain a sample for histologic evaluation of the endometrium (uterine mucosal lining) or sample extraction of the uterine menstrual content for:
- Evaluation of infertility conditions, menstrual disorders, postmenopausal bleeding, abnormal cytology suggesting endometrial origin, hormonal replacement therapy (HRT);
- Detection of endometrial carcinoma (Please note: The most reliable method for the detection of endometrial carcinoma is dilation of the cervix and uterine curettage performed under general anesthesia);
- Evaluation of luteal defect (Evaluation of luteal defect is visually determined histologically from endometrial tissue obtained only during the secretory [progestational] phase of the menstrual cycle);
- Endometrial dating;
- Microscopic examination;
- and, this device is useful for patients presenting a narrow cervix or cervical canal and in postmenopausal women whose cervix might be narrower However, because of its smaller with any of the indications listed above. diameter, it should be recognized that the amount (quantity) of sample obtained with this device will be reduced.
Rev. 1 March 23, 1998
t
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use ﻟﺴﺎ (Per 21 CFR 801.109) Over-The-Counter-Use (Optional Format 1-2-96)
Riker D. Satkin/
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.