← Product Code [GCJ](/productcode/GCJ) · K974767

# SMITH & NEPHEW DYONICS ARTHOSCOPES/ENDOSCOPES, SMITH & NEPHEW DYONICS LAPAROSCOPES, SNITH & NEPHEW DYONICS HYSTEROSCOPES (K974767)

_Endoscopy Division · GCJ · Jan 23, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K974767

## Device Facts

- **Applicant:** Endoscopy Division
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 23, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Dyonics® line of rigid Arthroscopes/ENT Endoscopes is indicated to provide illumination and visualization in diagnostic and operative arthroscopic surgical procedures and in endoscopic examination and treatment of the nasal cavities and nasal pharynx. In addition, the Dyonics 4mm diameter Arthroscopes/ENT Endoscopes are indicated for use in the removal of loose bodies and soft tissue within the hip joint as size/length appropriate. Dyonics Laparoscopes are indicated for use in laparoscopic surgical procedures to provide illumination and visualization of the surgical site. The 5mm x 300mm Dyonics Laparoscope is also indicated for use in endoscopic surgery of superficial vessels and fascia of the lower extremities to provide illumination and visualization of the surgical site during decission, ligation and harvesting of vessels. Dyonics Hysteroscopes are used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. The abrasive cleaning agent is provided as an accessory to the Dyonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes and is intended to be used, when required, to clean the distal surfaces thus maintaining optical integrity and allowing the scopes to perform as per their intended uses.

## Device Story

Rigid endoscopes (arthroscopes, ENT endoscopes, laparoscopes, hysteroscopes) used for illumination and visualization of surgical sites during minimally invasive procedures. Device includes an abrasive cleaning agent accessory to maintain optical integrity of distal surfaces. Operated by physicians in clinical/surgical settings. Modification pertains solely to expanded cleaning instructions. No changes to device design, function, or technology.

## Clinical Evidence

Bench testing only; biocompatibility and residual chemical analyses performed to validate the safety of the modified cleaning instructions.

## Technological Characteristics

Rigid endoscopes; optical illumination and visualization system. Modification involves addition of an abrasive cleaning agent accessory. No changes to materials or core sensing/actuation principles.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Dyonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### K974767 510(k) Summary Smith & Nephew, Inc., Endoscopy Division Modification - Cleaning Instructions for Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes

# Substantial Equivalence:

The modified cleaning instructions defined in this premarket notification submission have no impact on the designed safety and efficacy of the Dvonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes. This conclusion is supported by the following testing: biocompatibility and residual chemical analyses.

## Predicate Device:

The predicate devices for this submission are the currently marketed Dyonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes.

## Summary of Device Function:

The Dyonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes are used during various minimally invasive endoscopic procedures to illuminate and visualize the surgical site.

## Intended Use of Device:

The Dyonics® Iine of rigid Arthroscopes/ENT Endoscopes is indicated to provide illumination and visualization in diagnostic and operative atthroscopic surgical procedures and in endoscopic examination and treatment of the nasal cavities and nasal pharynx. In addition, the Dyonics 4mm diameter Arthroscopes/ENT Endoscopes are indicated for use in the removal of loose bodies and soft tissue within the hip joint as size/length appropriate.

Dyonics Laparoscopes are indicated for use in laparoscopic surgical procedures to provide illumination and visualization of the surgical site.

The 5mm x 300mm Dyonics Laparoscope is also indicated for use in endoscopic surgery of superficial vessels and fascia of the lower extremities to provide illumination and visualization of the surgical site during decission, ligation and harvesting of vessels.

Dyonics Hysteroscopes are used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.

## Comparison of Technological Characteristics of Predicate Device:

The basic technologies, design and function of the Smith & Nephew Dyonics Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes are not changed by the expanded cleaning instructions described in this Premarket Notification Submission. The expanded cleaning instructions defined in this submission raise no new issues of safety and effectiveness.

Debrah J. Connor

Deborah J. Connors Regulatory Affairs Specialist

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol consisting of three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 3 1998

Ms. Deborah J. Connors "Senior Regulatory Affairs Specialist Smith & Nephew, Incorporated Endoscopy Division 160 Dascomb Road Andover, Massachusetts 01810

Re: K974767

> Trade Name: Modification - Cleaning Instructions for Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes Regulatory Class: II Product Code: GCJ Dated: December 19, 1997 Received: December 22, 1997

Dear Ms. Connors:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, . Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for

{2}------------------------------------------------

Page 2 - Ms. Connors

devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

tocoelfo

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

510(k) Number :

Device Name : Modification - Cleaning Instructions for Arthroscopes/ENT Endoscopes, Laparoscopes and Hysteroscopes

Indications for Use :

The Dyonics® line of rigid Arthroscopes/ENT Endoscopes is indicated to provide illumination and visualization in diagnostic and operative arthroscopic surgical procedures and in endoscopic examination and treatment of the nasal cavities and nasal pharynx. In addition, the Dyonics 4mm diameter Arthroscopes/ENT Endoscopes are indicated for use in the removal of loose bodies and soft tissue within the hip joint as size/length appropriate.

Dyonics Laparoscopes are indicated for use in laparoscopic surgical procedures to provide illumination and visualization of the surgical site.

The 5mm x 300mm Dyonics Laparoscope is also indicated for use in endoscopic surgery of superficial vessels and fascia of the lower extremities to provide illumination and visualization of the surgical site during decission, ligation and harvesting of vessels.

Dyonics Hysterosocpes are used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.

(PLEASE DO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|                                                                                                | <div> <img alt="Signature" src="signature.png"/> <div>(Division Sign-Off)</div> <div>Division of General Restorative Devices</div> <div>510(k) Number: K474767</div> </div> |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Use (Per 21 CFR 801.109) <img alt="Prescription Use" src="prescription_use.png"/> | OR Over-the-Counter ____                                                                                                                                                    |

(Optional Format 1-2-96)

{4}------------------------------------------------

J.C.H.L.D. 23

(R) 749 1443

NO.631 ь:3

p974767 ನ್ನ OR Us e Inclications

Page 2 of 2

The abrasive cleaning agent is provided as an accessory to the Dyonics Arthroscopes/ENT I he abrastoscopes and Hysteroscopes and is intended to be used, when required, to Ellooscopes, Laparoscopes and Arraces thus maintaining optical integrity and allowing the scopes to perform as per their intended uses.

**Prescription Use**
**(Per 21 CFR 801.109)**

(Divisio Sign-Off) Division of General Restorative 510(k) Number ________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K974767](https://fda.innolitics.com/device/K974767)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).