INFRARED GROUP AMPLIFICATION SYSTEM, INFRARED EMITTER, INFRARED BODY WORN RECEIVER, INFRARED HEADSET RECEIVER-STEREO, IT
K974658 · Phonic Ear, Inc. · EPF · Feb 25, 1998 · Ear, Nose, Throat
Device Facts
Record ID
K974658
Device Name
INFRARED GROUP AMPLIFICATION SYSTEM, INFRARED EMITTER, INFRARED BODY WORN RECEIVER, INFRARED HEADSET RECEIVER-STEREO, IT
Applicant
Phonic Ear, Inc.
Product Code
EPF · Ear, Nose, Throat
Decision Date
Feb 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3320
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Model PE 600 Series devices form the Impaired Group Amplification System which is used by adults and children in public halls, theaters, churches, courtrooms and other public gatherings and provide auditory assistance to hearing impaired persons.
Device Story
Infrared group amplification system; transmits audio signals via infrared light to assist hearing-impaired users. System comprises PE600E infrared emitter (power amplifier + LED array) and various receivers (PE600R body-worn, PE601R/PE602R headsets). Emitter processes audio inputs, broadcasting via 2.3 MHz and 2.8 MHz frequencies to eliminate electrical/lighting interference. Receivers detect infrared signals, converting them to audio for the user. Used in auditoriums, classrooms, and courtrooms; operated by facility staff or users. Benefits include improved speech understanding, word discrimination, and reduced impact of distance/reverberation. Receivers can power tele-coils in personal hearing aids or provide direct audio via headsets. Supports dual-channel transmission for language translation or stereo programs.
Clinical Evidence
No clinical data generated for this specific device. Effectiveness claims are substantiated by clinical studies of FM Auditory Training Systems, including reports by Dr. David Hawkins, Dr. Mark Ross, Robert Benoit, M.A., and Dr. Thomas Giolas.
Technological Characteristics
System uses infrared light transmission at 2.3 MHz and 2.8 MHz. Components include solid-state surface mount integrated circuits. Emitter (PE600E) combines power/amplifier circuits with infrared LED array. Receivers feature light-gathering lenses. Housed in cycolac plastic. Powered by low-voltage batteries with fused circuits. Supports multiple listening accessories and tele-coil integration.
Indications for Use
Indicated for hearing impaired adults and children requiring auditory assistance in public environments such as halls, theaters, churches, and courtrooms.
Regulatory Classification
Identification
A group hearing aid or group auditory trainer is a hearing aid that is intended for use in communicating simultaneously with one or more listeners having hearing impairment. The device is used with an associated transmitter microphone. It may be either monaural or binaural, and it provides coupling to the ear through either earphones or earmolds. The generic type of device includes three types of applications: hardwire systems, inductance loop systems, and wireless systems.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
Submission Summary (Full Text)
{1}------------------------------------------------
K97 4658
## )))))))))Phonic Ear
Innovative Communication Technologies
FEB 2 5 1998
## 510 (K) SUMMARY
## of the
# SAFETY AND EFFECTIVENESS OF THE DEVICE
### 510 (K) NOTIFICATION
## INFRARED GROUP AMPLIFICATION SYSTEM
### PHONIC EAR INC.
MODEL PE600 SERIES
Submitted by:
PHONIC EAR INCORPORATED 3880 Cypress Drive Petaluma, CA 94954-7600
Establishment Registration No. 2918633
Telephone Number: (707) 769-1110 Fax Number: (707) 769-9624 Official Correspondent:: Robert L. Reinke, Ext. 298
December 1, 1997
Phonic Ear Inc., 3880 Cypress Drive · Petaluma, CA 94954-7600 · 707/769-1110 · 707/769-9624 FAX
{2}------------------------------------------------
))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))))
Innovative Communication Technologies
## 510(k) NOTIFICATION Phonic Ear Model PE600 Series Infrared Group Amplification System December 1, 1997
#### SUMMARY OF THE SAFETY OF THE SYSTEM
- 1. The safety of this Infrared Amplification System is equivalent to approved systems of the same type.
Page 1
- 2. While moderate sound pressure levels are produced by the PE600 Series receivers, this output can be limited to the output required by the user. Limiting can be accomplished by the setting of the volume controls in accordance with the tests made by a clinician, audiologist or hearing professional prior to use by the hearing impaired user.
- 3. Instructions for the user to follow, should the system not function as expected, are contained in the user information.
- 4. Complete technical specifications are contained in the user information.
- 5. The systems are distributed directly to qualified dispensers or others who are trained in the proper fitting and use of the system.
- 6. The system elements are housed in cases made of cycolac plastic which has few known side effects caused by the material.
- 7. A complete line of accessories to be used with the system elements are available from the manufacturer and a list of these accessories and a description of their use is contained in the user information.
- 8. Addresses and telephone numbers of authorized service centers are available to the user should any questions arise about the function of the system.
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{3}------------------------------------------------
#### 510(k) NOTIFICATION
Phonic Ear Model PE600 Series Infrared Group Amplification System Page 2 December 1, 1997
- 9. Detailed descriptions on the use, maintenance, adjustment and function of each element of the system are contained in the user information.
- 10. The system elements are powered by low voltage batteries which are contained within battery compartments that are fitted with tight snap fitting covers to prevent inadvertent opening. The battery circuits are protected by fusing devices to prevent over heating should a short circuit occur.
- 11. The user worn receivers are fitted with removable covers over the earphones that contact the skin and ears of the users, which can be removed for cleaning or exchange to prevent the spread or transmission of possible contamination.
{4}------------------------------------------------
1755
510(k) NOTIFICATION Phonic Ear Model PE600 Series Infrared Group Amplification System December 1, 1997
Page 1
### SUMMARY OF THE EFFECTIVENESS OF THE SYSTEM
- 1. The effectiveness of this Infrared Amplification System is equivalent to approved systems of the same type.
- 2. The infrared light gathering lenses on the receivers are designed to detect the sound signals from any angle providing for clear reception at any location within the installation.
- 3. The use of solid state surface mount components and integrated circuits allow the size and weight of the system elements to be reduced without the loss of any of the beneficial features provided in the design. Units are less obtrusive to the user because of the smaller size and easier to use because of the reduction in weight.
- 4. The use of several optional listening accessories provide a more natural sound to the user in different environments because the sounds are processed more like the sounds received by normal hearing persons.
- 5. The ability of the system elements to utilize a variety of listening accessories allow the user a wider choice of fitting possibilities to better meet their personal needs.
- 6. The inclusion of a battery charging LED indicator on the PE600R receiver insures that the device's rechargeable batteries will receive a full charge by confirming to the user that a positive charging connection has been made.
- 7. The ability of the receivers to operate on either rechargeable or throw-away batteries allow the user to select the power source most beneficial to their use.
- 8. The provision of optional receiver types provide the user with a selection which will best suit their own needs.
{5}------------------------------------------------
#### 510(k) NOTIFICATION
14
Phonic Ear Model PE600 Series Infrared Group Amplification System Page 2 December 1. 1997
- 9. The combining of the power and amplifier circuits with the infrared LED circuits into one unit , provide easier installation and adjustment of the infrared emitter PE600E.
- 10. The ability of the PE600E to operate on both 2.3 Mhz and 2.8 Mhz simultaneously provides greater flexibility for the use of the system. Either stereo or 2-channel programs can be transmitted by the same system.
- 11. The use of the higher 2.3 Mhz and 2.8 Mhz transmission frequencies eliminates the interference which may be caused by electrical circuits and lighting, thus providing a clearer, interference free sound to the listener.
- 12. The ability of the PE600E Infrared Emitter to be mounted in several ways provide more flexibility for use as either a small portable system or in permanently installed systems for large auditoriums.
- 13. The ability of the systems to be enlarged by the addition of more infrared emitters provides full sound coverage even in very large installations.
- 14. The ability of the infrared receivers to power the tele-coil found in most hearing aids enables the users to utilize their own hearing aids which provide more convenience of use as well as allowing the user to hear the same sounds they would normally hear using only their hearing aid.
- 15. The design of the infrared detecting elements in the receivers allow the user to sit anywhere within the infrared environment without experiencing the dead spots or sound fade-outs usually found in most installations.
- 16. The use of an Infrared Group Amplification System provides increased speech understanding and word discrimination for the hearing impaired in listening environments such as school classrooms, concert halls, theaters or other similar situations where noises, reverberation or poor acoustical treatment may be found. The use of these systems also reduce the detrimental effects caused by the distance between the speaker or sound source and the hearing impaired listener. (See note.)
{6}------------------------------------------------
#### 510(k) NOTIFICATION
Phonic Ear Model PE600 Series Infrared Group Amplification System Page 3 December 1, 1997
- 17. The ability of the infrared system to operate on two channels simultaneously allows the system to be used for language translation thus integrating users of different languages into the same environment.
- 18. The PE600 Infrared Systems are especially adaptable for use in installations where security of the sound is desired , such as in court rooms or jury deliberation rooms.
Effectiveness claims outlined in paragraph 16 have been NOTE: substantiated by the results of many clinical studies made over the years since the introduction of FM Auditory Training Systems. Three such study reports by Dr. David Hawkins, Dr. Mark Ross, Robert Benoit, M.A. and Dr. Thomas Giolas form a part of this 510(k) submission.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 5 1998
Robert L. Reinke Phonic Ear, Inc. 3880 Cypress Drive Petaluma, CA 94954-7600 Re: K974658
Phonic Ear Model: PE600E Infrared Emitter, PE600R Infrared Receiver, PE601R Infrared Headset Receiver and PE602R Infrared Headset Receiver Dated: December 1, 1997 Received: December 15, 1997 Regulatory class: II 21 CFR 874.3320/Procode: 77 EPF
#### Dear Mr. Reinke:
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
h.T.Liau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "Phonic Ear". The logo consists of a series of curved lines on the left, followed by the words "Phonic Ear" in a bold, sans-serif font. A registered trademark symbol is located to the right of the word "Ear".
ovative Communication Technologies
December 1. 1997
INDICATIONS FOR USE
Page 1 of 1
510(k) Number: K97465 Q
Device Name: INFRARED GROUP AMPLIFICATION SYSTEM - MODELS PE 600 SERIES
Indications For Use:
The Model PE 600 Series devices form the Impaired Group Amplification System which is used by adults and children in public halls, theaters, churches, courtrooms and other public gatherings and provide auditory assistance to hearing impaired persons.
- 1. The Model PE 600E is a power amplifier which is combined with an infrared LED array. Audio signals from a sound system are processed and broadcast into the listening area by means of infrared light.
- 2. The Model PE 600R is a body wom infrared receiver. It receives the infrared light signals from the PE 600E and converts these signals to high level audio signals required by the hearing impaired listener. The audio signals can be adjusted to meet the user.
- The PE 601R is an Infrared Headset Receiver. It receives the infrared light signals from the 3. PE 600E and converts these signals to audio signals required by the hearing impaired listener. This devise is capable of stereo reception so that different audio signals can be received simultaneously and fed to each ear separately.
- The PE 602R is an Infrared Headset Receiver. It receives the infrared light signals from the 4. PE 600E and converts these signals to audio signals required by the hearing impaired listener. The device is capable of receiving infrared light signals transmitted on two separate frequencies. A switch on the headset allows the user to select the program frequency desired. The selected program is heard in each ear.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
> Restricted Device Per 874.420 and 421
OR
Over-The-Counter Use
(Optional Format 1-2-96)
David A. Lynn
---
**(Division Sign-Off)**
Division of Reproductive, Abdominal, E and Radiological De 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.