← Product Code [LLZ](/productcode/LLZ) · K974474

# PEGASYS INTOUCH (K974474)

_Adac Laboratories · LLZ · Feb 24, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K974474

## Device Facts

- **Applicant:** Adac Laboratories
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 24, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.

## Device Story

Pegasys InTouch™ is a PACS software application for remote viewing of medical images. It accepts Pegasys and DICOM image data transmitted via hospital LAN/WAN. The system enables physicians to display, store, and interact with images on standard WinTel PCs. Features include multi-zone display (up to 9 zones), tomographic comparative displays, and cardiac stress/rest comparative analysis (including cine motion). The device facilitates clinical interpretation by allowing simultaneous viewing of datasets at different locations. It does not alter image data. Benefits include improved physician access to diagnostic information outside the original study area, supporting clinical decision-making through remote image review.

## Clinical Evidence

Bench testing only. A prototype of the display application was used to download and display DICOM-format images in four display regions to verify image integrity and display functionality.

## Technological Characteristics

PC-compatible hardware (microprocessor, RAM, monitor, keyboard, mouse/trackball, hard drive). Connectivity via Ethernet or modem. Software runs on WinTel systems using Microsoft Internet Explorer 3.0 or Netscape Navigator 3.0. Supports DICOM standard. No specialized materials or energy sources; standard digital computing environment.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- AVP PACSPRO 640
- HyperPACS
- AMICAS Web/Intranet Image Server

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Pegasus InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 1 of 3
K974474

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA

## I. General Information

FEB 24 1998

A. Submitted By:
ADAC Laboratories
540 Alder Drive
Milpitas, CA 95035
Tel: (408) 321-9100
Fax: (408) 321-9686

Contact Person: Dennis Henkelman at address above

B. Device Trade Name: Pegasus InTouch™
Common Name: Picture Archive and Communication System (PACS)
Classification Name: Digital Image Communication System

C. Predicate Device: AVP PACSPRO 640
HyperPACS
AMICAS Web/Intranet Image Server

D. Device Description:

Pegasus InTouch™ has been developed to provide a physician with the ability to review Pegasus and DICOM images from locations outside the original study area. Pegasus InTouch™ has been designed to operate on standard, off-the-shelf hardware and software components. Using standard WinTel PC systems and the hospital proprietary LAN/WAN (either local or wide area), physicians will have the ability to review and interact with patient data from their desks.

Pegasus and other digital imaging communications systems consist of a computer with a high speed data modem connected to a standard telephone line or computer network that can transmit or receive digitized images. The system is programmed to send, receive, store, copy and display these images in a manner that assures that images are not altered. Operator interface features are also computer controlled so as to enhance the system's capabilities and make the system more user friendly. Images can be displayed with different resolutions and colors; the same image set can be viewed simultaneously at different locations on designated physician workstations.

{1}

Pegasys InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 2 of 3

The physician has the capability to select a variety of display formats for viewing. For example, the physician may select up to 9 separate datasets to be displayed simultaneously, including whole body and snapshot displays; comparative (tomographic) displays of oblique or transverse files; comparative cardiac SPECT data; and SPECT and gated SPECT data in a 10 frame-per-second cine mode. A brief description of the applications available to the physician workstation follows.

1. The physician may select from a number of display formats incorporating from 1 to 9 zones as described previously.
2. A Tomographic Display is used to display multiple orientations of the same image set in a comparative manner.
3. The Cardiac Stress/Rest Comparative application facilitates the visual comparison of the same cardiac locations during rest and during stress.
4. Cardiac Comparatives is similar to the Stress/Rest application above, but displays the stress and rest projections in cine motion.

E. Indications for Use:

Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.

F. Technological Comparison:

Pegasys InTouch™ is similar to a wide variety of Picture Archive and Communication Systems (PACS) currently on the market. These systems are designed to send, receive, store, copy and display high resolution digitized images. The images can be digitally scanned into the system or captured from various imaging modalities such as CT and MRI. Each system consists of either a single unit or multiple units networked. Image data is sent from a sending unit through digital network lines or through standard telephone lines by a high speed modem to a receiving station. Once the image is received, it can be stored, copied and displayed. Displayed images may be adjusted for user preference to assist in interpretation.

{2}

Pegasus InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 3 of 3

Pegasus InTouch™ and other currently marketed systems consist of a PC compatible computer with a microprocessor with RAM, monitor, keyboard, mouse/trackball, internal hard drive, modem or Ethernet connection, and Microsoft Internet Explorer 3.0 or Netscape Navigator 3.0.

## II. Testing

An image was generated using a prototype of the display applications. The DICOM-format image was downloaded using standard web-based data transfer and displayed into the image viewer. The same image was displayed in each of the four display regions.

{3}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
FEB 24 1998

Dennis W. Henkelman
Director, Regulatory Affairs
and Quality Assurance
ADAC Laboratories
540 Alder Drive
Milpitas, CA 95035

Re: K974474
Pegasys InTouch™ (PACS)
Dated: November 25, 1997
Received: November 26, 1997
Regulatory class: Unclassified
Procode: 90 LLZ

Dear Mr. Henkelman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{4}

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K974474

Device Name: Pegasys InTouch™

Sponsor Name: ADAC Laboratories

## Indications for Use

Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.

Do Not Write Below This Line - Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑
Over-The-Counter Use ☐

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K974474

---

**Source:** [https://fda.innolitics.com/device/K974474](https://fda.innolitics.com/device/K974474)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
