Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.
Device Story
Pegasys InTouch™ is a PACS software application for remote viewing of medical images. It accepts Pegasys and DICOM image data transmitted via hospital LAN/WAN. The system enables physicians to display, store, and interact with images on standard WinTel PCs. Features include multi-zone display (up to 9 zones), tomographic comparative displays, and cardiac stress/rest comparative analysis (including cine motion). The device facilitates clinical interpretation by allowing simultaneous viewing of datasets at different locations. It does not alter image data. Benefits include improved physician access to diagnostic information outside the original study area, supporting clinical decision-making through remote image review.
Clinical Evidence
Bench testing only. A prototype of the display application was used to download and display DICOM-format images in four display regions to verify image integrity and display functionality.
Technological Characteristics
PC-compatible hardware (microprocessor, RAM, monitor, keyboard, mouse/trackball, hard drive). Connectivity via Ethernet or modem. Software runs on WinTel systems using Microsoft Internet Explorer 3.0 or Netscape Navigator 3.0. Supports DICOM standard. No specialized materials or energy sources; standard digital computing environment.
Indications for Use
Indicated for physicians to remotely view and interpret Pegasys and DICOM medical images (e.g., SPECT, cardiac stress/rest) via hospital LAN/WAN networks.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
AVP PACSPRO 640
HyperPACS
AMICAS Web/Intranet Image Server
Submission Summary (Full Text)
{1}
Pegasus InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 1 of 3
K974474
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. General Information
FEB 24 1998
A. Submitted By:
ADAC Laboratories
540 Alder Drive
Milpitas, CA 95035
Tel: (408) 321-9100
Fax: (408) 321-9686
Contact Person: Dennis Henkelman at address above
B. Device Trade Name: Pegasus InTouch™
Common Name: Picture Archive and Communication System (PACS)
Classification Name: Digital Image Communication System
C. Predicate Device: AVP PACSPRO 640
HyperPACS
AMICAS Web/Intranet Image Server
D. Device Description:
Pegasus InTouch™ has been developed to provide a physician with the ability to review Pegasus and DICOM images from locations outside the original study area. Pegasus InTouch™ has been designed to operate on standard, off-the-shelf hardware and software components. Using standard WinTel PC systems and the hospital proprietary LAN/WAN (either local or wide area), physicians will have the ability to review and interact with patient data from their desks.
Pegasus and other digital imaging communications systems consist of a computer with a high speed data modem connected to a standard telephone line or computer network that can transmit or receive digitized images. The system is programmed to send, receive, store, copy and display these images in a manner that assures that images are not altered. Operator interface features are also computer controlled so as to enhance the system's capabilities and make the system more user friendly. Images can be displayed with different resolutions and colors; the same image set can be viewed simultaneously at different locations on designated physician workstations.
{2}
Pegasys InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 2 of 3
The physician has the capability to select a variety of display formats for viewing. For example, the physician may select up to 9 separate datasets to be displayed simultaneously, including whole body and snapshot displays; comparative (tomographic) displays of oblique or transverse files; comparative cardiac SPECT data; and SPECT and gated SPECT data in a 10 frame-per-second cine mode. A brief description of the applications available to the physician workstation follows.
1. The physician may select from a number of display formats incorporating from 1 to 9 zones as described previously.
2. A Tomographic Display is used to display multiple orientations of the same image set in a comparative manner.
3. The Cardiac Stress/Rest Comparative application facilitates the visual comparison of the same cardiac locations during rest and during stress.
4. Cardiac Comparatives is similar to the Stress/Rest application above, but displays the stress and rest projections in cine motion.
E. Indications for Use:
Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.
F. Technological Comparison:
Pegasys InTouch™ is similar to a wide variety of Picture Archive and Communication Systems (PACS) currently on the market. These systems are designed to send, receive, store, copy and display high resolution digitized images. The images can be digitally scanned into the system or captured from various imaging modalities such as CT and MRI. Each system consists of either a single unit or multiple units networked. Image data is sent from a sending unit through digital network lines or through standard telephone lines by a high speed modem to a receiving station. Once the image is received, it can be stored, copied and displayed. Displayed images may be adjusted for user preference to assist in interpretation.
{3}
Pegasus InTouch™
ADAC Laboratories
510(k) Premarket Notification
Appendix VIII, 510(k) Summary of Safety and Effectiveness Data
Page 3 of 3
Pegasus InTouch™ and other currently marketed systems consist of a PC compatible computer with a microprocessor with RAM, monitor, keyboard, mouse/trackball, internal hard drive, modem or Ethernet connection, and Microsoft Internet Explorer 3.0 or Netscape Navigator 3.0.
## II. Testing
An image was generated using a prototype of the display applications. The DICOM-format image was downloaded using standard web-based data transfer and displayed into the image viewer. The same image was displayed in each of the four display regions.
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
FEB 24 1998
Dennis W. Henkelman
Director, Regulatory Affairs
and Quality Assurance
ADAC Laboratories
540 Alder Drive
Milpitas, CA 95035
Re: K974474
Pegasys InTouch™ (PACS)
Dated: November 25, 1997
Received: November 26, 1997
Regulatory class: Unclassified
Procode: 90 LLZ
Dear Mr. Henkelman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K974474
Device Name: Pegasys InTouch™
Sponsor Name: ADAC Laboratories
## Indications for Use
Pegasys InTouch™ software is intended to transfer and provide interpretive displays of Pegasys and DICOM image data either compressed or uncompressed to physician workstations via a local area network (LAN) interconnected with Ethernet or via a wide area network (WAN) comprised of two or more LANs in different locations, using a hospital intranet, not the Internet.
Do Not Write Below This Line - Continue on Another Page if Needed
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑
Over-The-Counter Use ☐
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K974474
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.