CMVGEN

K974456 · Instrumentation Laboratory CO · LJO · May 1, 1998 · Microbiology

Device Facts

Record IDK974456
Device NameCMVGEN
ApplicantInstrumentation Laboratory CO
Product CodeLJO · Microbiology
Decision DateMay 1, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3175
Device ClassClass 2

Indications for Use

CMVgen is an in vitro diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total antibodies (IgG and IgM) to cytomegalovirus (CMV) in human serum or plasma to determine prior exposure to cytomegalovirus. This assay has not been FDA cleared for use in screening blood or plasma donors.

Device Story

CMVgen is an in vitro diagnostic rapid latex particle agglutination test. It uses human serum or plasma samples to detect total antibodies (IgG and IgM) against cytomegalovirus. The device operates via passive latex agglutination, where antigen-coated latex particles agglutinate in the presence of specific CMV antibodies. The test is performed in a clinical laboratory setting by trained personnel. Results are interpreted qualitatively or semiquantitatively to determine prior CMV exposure. The output assists clinicians in assessing patient immune status regarding CMV. It is not intended for blood or plasma donor screening.

Clinical Evidence

Method comparison study (n=165 serum samples) showed 97.5% sensitivity compared to a commercial total antibody test. A second study (n=131 serum samples, including 31 organ transplant recipients) showed 91.1% sensitivity compared to a commercial IgG EIA test. Reproducibility study (n=10 samples tested over 3 days) demonstrated 100% agreement within one 2-fold dilution.

Technological Characteristics

In vitro diagnostic latex particle agglutination assay. Uses passive latex agglutination principle. Manual test format. No electronic components, software, or energy sources required.

Indications for Use

Indicated for qualitative and semiquantitative detection of total CMV antibodies (IgG and IgM) in human serum or plasma to assess prior CMV exposure. Not for use in screening blood or plasma donors.

Regulatory Classification

Identification

Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.

Predicate Devices

Submission Summary (Full Text)

{0} K974456 MAY 1 1998 # SECTION 3 ## CMVgen - 510(k) SUMMARY ### (Summary of Safety and Effectiveness) **Submitted by:** Carol Marble Regulatory Affairs Engineer Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173 Phone: (781) 861-4467 Fax: (781) 861-4464 **Contact Persons:** Carol Marble Phone: (781) 861-4467 **Summary Prepared:** November 24, 1997 (Revised on February 23, 1998) **Name of the device:** CMVgen **Classification name(s):** 866.3175 Cytomegalovirus serological reagents Class II 83LJO Antigen, IHA, Cytomegalovirus **Identification of predicate device(s):** CMVscan K841520 **Description of the device/intended use(s):** CMVgen is an *in vitro* diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total antibodies (IgG and IgM) to cytomegalovirus (CMV) in human serum or plasma (EDTA) to determine prior exposure to cytomegalovirus. This product is not FDA cleared for use in screening blood or plasma (EDTA) donors. **Statement of How the Technological Characteristics of the Device compare to the Predicate Device:** CMVgen uses the same test principle (passive latex agglutination) as the predicate CMVscan and is substantially equivalent in performance, intended use, and safety and effectiveness with the exception that CMVscan is labeled for use to screen donor specimens. **Summary of Performance Data:** In a method comparison study performed at UMMC Clinical Microbiology Laboratories (Massachusetts) evaluating 165 serum samples, the sensitivity of CMVgen as compared to a commercially available total antibody test was 97.5%. In a separate study performed at Cambridge University (U.K.) evaluating 131 serum samples (ninety of which were collected from 31 organ transplant recipients who had experienced either a primary CMV infection or a reactivation of CMV post-transplant), the sensitivity of CMVgen as compared to a commercially available IgG EIA test was 91.1%. In a reproducibility study, panels of 10 serum and plasma samples were tested on 3 consecutive days using the semiquantitative procedure for CMVgen. The results, as defined by the ability to give agreement to within one 2-fold dilution on the replicates, indicated 100% reproducibility. Section 3 CMVgen 510(k) Page 1 of 1 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service MAY 1 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Carol Marble Senior Regulatory Affairs Specialist Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02173-3190 Re: K974456 Trade Name: CMVgen Regulatory Class: II Product Code: LJO Dated: February 23, 1998 Received: February 24, 1998 Dear Ms. Marble: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} # Indications for Use Statement 510(k) Number (if known): _________________________ Device Name: CMVgen ## Indications for Use: CMVgen is an *in vitro* diagnostic, rapid latex particle agglutination test for the qualitative and semiquantitative determination of total antibodies (IgG and IgM) to cytomegalovirus (CMV) in human serum or plasma to determine prior exposure to cytomegalovirus. This assay has not been FDA cleared for use in screening blood or plasma donors. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-0.jpeg](img-0.jpeg) Prescription Use ☑ (Per 21 CFR 801.019) OR Over-The-Counter Use _________ Section 2 CMVgen 510(k) Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...