← Product Code [FSY](/productcode/FSY) · K974433

# CHROMOPHARE D 500, CHROMOPHARE D 530 (K974433)

_Berchtold Holding GmbH · FSY · Feb 20, 1998 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K974433

## Device Facts

- **Applicant:** Berchtold Holding GmbH
- **Product Code:** [FSY](/productcode/FSY.md)
- **Decision Date:** Feb 20, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4580
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The surgical lights, CHROMOPHARE® D 500 and D 530, are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.

## Device Story

CHROMOPHARE® D 500 and D 530 are ceiling-mounted surgical lights; provide high-intensity, shadow-free, "cold" illumination for surgical sites. Utilize optical double filter technique to produce whiter, cooler light. D 530 provides up to 110,000 lux; D 500 provides up to 85,000 lux. Features include swivel arms, auto-switching to secondary lamp upon primary failure, and exchangeable lamp cartridges. D 530 includes optional CCD-video camera and "EndoLite" version for endoscopic procedures. Operated by surgical staff in clinical/OR settings. Enhances visualization of surgical field; improves procedural outcomes through superior lighting.

## Clinical Evidence

Bench testing only; compliance with IEC 601 and UL 2601 safety regulations.

## Technological Characteristics

Ceiling-mounted surgical light system; optical double filter technique for light quality; intensity up to 110,000 lux (D 530) or 85,000 lux (D 500); mechanical swivel arm assembly; auto-switching lamp redundancy; optional CCD-video camera integration; meets IEC 601 and UL 2601 safety standards.

## Regulatory Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- CHROMOPHARE® D 650 ([K965130](/device/K965130.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

BERCHTOLD
K 974433
715
FÜNFUNGSLEBZIG JAHRE

K. 510(k) SUMMARY (As Required By 21 CFR 807.3):
FEB 20 1998

Date: November 20, 1997

Submitter:
Berchtold GmbH &amp; Co
Ludwigstaler Str. 25
D-78532 Tuttlingen
Germany

Phone number: 0049 7461 181-0
Fax number: 0049 7461 181 100
Contact person: Wolfram K. Hill
Manager R &amp; D

Device Name:

Trade Name: CHROMOPHARE® D 500
CHROMOPHARE® D 530

Classification name: Light, Surgical, Ceiling Mounted

Device Description:

The new Berchtold CHROMOPHARE® D 500 and D 530 surgical lights are suitable for all types of surgical procedures and offer the physician a natural (whiter) and "colder" light. The light quality is based on a optical double filter technique.

The CHROMOPHARE® D 530 provides a light intensity of up to 110000 lux and the CHROMOPHARE® D 500 up to 85000 lux.

The lights have features like easy-to-operate swivel arms, auto-switching on the second lamp in case of failure of the main lamp and an easy to exchange lamp cartridge. For the CHROMOPHARE® D 530 there is also an optional CCD-video camera available and a special version with „EndoLite“ for endoscopical procedures. The lights could be combined with other Berchtold lights. The lights have been designed to and meet the requirements of the IEC 601 and UL 2601 regulation for safety.

Intended Use:

The surgical lights, CHROMOPHARE® D 500 and D 530, are intended to illuminate locally the operating site on the patient's body with a high intensity, shadow free, "cold" light.

{1}

4
BERCHTOLD
FÜNFUNGSFÜHRUNG
715
FÜNFUNGSFÜHRUNG JAHRE

## Substantial Equivalence:

The CHROMOPHARE® D 500 and D 530 surgical lights are substantially equivalent to the surgical light CHROMOPHARE® D 650 which is approved by FDA under K965130.

Any difference that exists between the CHROMOPHARE® D 500/530 and the CHROMOPHARE® D 650 has no negative effect on safety or efficacy and actually enhances the usefulness in the chosen application. Safety of the lights, as indicated above, has been established by meeting the requirements of IEC 601 and UL 2601.

{2}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850

Mr. Wolfram K. Hill
Manager R &amp; D
Berchtold GmbH &amp; Company
Ludwigstaler Street 25
D-78532 Tuttlingen
Germany

Fe: K974433
Trade Name: CHROMOPHARE® D 500 and D 530 Surgical Lights
Regulatory Class: II
Product Code: FSY
Dated: November 21, 1997
Received: November 24, 1997

Dear Mr. Hill:

We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}

Page 2 - Mr. Wolfram K. Hill

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your device and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

![img-0.jpeg](img-0.jpeg)

Enclosure

{4}

6
BERCHTOLD
FÜNFUNGSFÜRZIG JAHRE

## M. INDICATION FOR USE:

The surgical lights, CHROMOPHARE® D 500 and D 530, are intended to illuminate locally the operating site on the patient’s body with a high intensity, shadow free, "cold" light.

Wolfram Hill

Nov. 21, 98
Date

Prescription Use ☑ (Per 21 CFR 801.109)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number E974433

---

**Source:** [https://fda.innolitics.com/device/K974433](https://fda.innolitics.com/device/K974433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
