FRIEDL GLIDING NAIL SYSTEM

K974409 · Plus Orthopedics · KTT · Feb 19, 1998 · Orthopedic

Device Facts

Record IDK974409
Device NameFRIEDL GLIDING NAIL SYSTEM
ApplicantPlus Orthopedics
Product CodeKTT · Orthopedic
Decision DateFeb 19, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Friedl Gliding Nail System is an all-purpose locking nail system for ensuring primary load stability in: - pertrochanteric femoral fractures - subtrochanteric femoral fractures and - lateral femoral neck fractures. Internal fixation with the long Friedl Gliding Nail is indicated in all combination injuries involving the lateral femoral neck or trochanter region and femoral shaft fractures. Thanks to its biomechanical characteristics, the Friedl Gliding Nail system is also suitable for medial femoral neck fractures with retention of the head and simple femoral shaft fractures down to the distal metaphysis. Gliding nail fixation can also be used to secure pathological fractures or to provide weight-bearing stability after varus and valgus revision osteotomies of the proximal femur.

Device Story

The FRIEDL Gliding Nail System is an orthopedic implant for internal fixation of femoral fractures. The system comprises intramedullary nails, femoral neck blades, and distal locking pins. It provides primary load stability for various femoral fractures and osteotomies. Used by orthopedic surgeons in a clinical/surgical setting, the device is implanted to stabilize bone segments, facilitating healing and weight-bearing. It functions as a mechanical fixation system.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Wrought High Nitrogen Stainless Steel (ISO 5832-9). Components: Intramedullary nails, femoral neck blades, distal locking pins. Mechanical fixation system.

Indications for Use

Indicated for patients with pertrochanteric, subtrochanteric, and lateral femoral neck fractures; combination injuries of lateral femoral neck/trochanter and femoral shaft; medial femoral neck fractures with head retention; simple femoral shaft fractures; pathological fractures; and post-osteotomy (varus/valgus) stabilization of the proximal femur.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} HHL CONSULTING FEB 19 1998 K974409 ^{}[] 510(k) Summary of Safety and Effectiveness Information (as required by 807.92c), prepared by Hartmut Loch, President of HHL Consulting for PLUS Orthopedics in San Diego, California November 10, 1997 Trade name: FRIEDL Gliding Nail System Common name: Compression Hip Nail Classification name: Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, 87KTT, Regulation Number 888.3030. Description: The FRIEDL Gliding Nail System is an all-purpose locking nail system for primary load stability in pertrochanteric femoral fractures, subtrochanteric femoral fractures, and lateral femoral neck fractures. This system consists of several sizes of intramedullary nails, femoral neck blades, and distal locking pins. The Friedl Gliding Nail System is made of Wrought High Nitrogen Stainless Steel according to ISO 5832-9. Indications: The Friedl Gliding Nail System is an all-purpose locking nail system for ensuring primary load stability in: - pertrochanteric femoral fractures - subtrochanteric femoral fractures and - lateral femoral neck fractures. Internal fixation with the long Friedl Gliding Nail is indicated in all combination injuries involving the lateral femoral neck or trochanter region and femoral shaft fractures. Thanks to its biomechanical characteristics, the Friedl Gliding Nail system is also suitable for medial femoral neck fractures with retention of the head and simple femoral shaft fractures down to the distal metaphysis. Gliding nail fixation can also be used to secure pathological fractures or to provide weight-bearing stability after varus and valgus revision osteotomies of the proximal femur. Comparable Features to Predicate Device: The FRIEDL Gliding Nail System is comparable to the Gliding Nail from Encore Orthopedics, Inc., which was cleared for marketing by FDA on July 13, 1995 (K-951809). HNNN NNNN - SNN 1060-2LS-059-1 61:11 8661/21/20 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Hartmut Loch President HHL Consulting Representing PLUS Orthopedics 835 Cortez Lane Foster City, California 94404 FEB 19 1998 Re: K974409 Trade Name: FRIEDL Gliding Nail System Regulatory Class: II Product Code: KTT Dated: November 10, 1997 Received: November 24, 1997 Dear Mr. Loch: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 - Mr. Hartmut Loch This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number (if known): K974409 Device Name: FRIEDL GLIDING NAIL SYSTEM Indications for Use: The Friedl Gliding Nail System is an all-purpose locking nail system for ensuring primary load stability in: - pertrochanteric femoral fractures - subtrochanteric femoral fractures and - lateral femoral neck fractures. Internal fixation with the long Friedl Gliding Nail is indicated in all combination injuries involving the lateral femoral neck or trochanter region and femoral shaft fractures. Thanks to its biomechanical characteristics, the Friedl Gliding Nail system is also suitable for medial femoral neck fractures with retention of the head and simple femoral shaft fractures down to the distal metaphysis. Gliding nail fixation can also be used to secure pathological fractures or to provide weight-bearing stability after varus and valgus revision osteotomies of the proximal femur. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-1.jpeg](img-1.jpeg) Prescription Use ☑ (Per 21 CFR 80.1.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%