ALCON CONTACT LENS CASE

K974281 · Alcon Laboratories · LRX · Jan 20, 1998 · Ophthalmic

Device Facts

Record IDK974281
Device NameALCON CONTACT LENS CASE
ApplicantAlcon Laboratories
Product CodeLRX · Ophthalmic
Decision DateJan 20, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.5928
Device ClassClass 2

Intended Use

For storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection.

Device Story

Alcon Contact Lens Case is a plastic storage container for contact lenses. Device consists of a side-by-side double-well design with two screw-on plastic closures. Used by patients for lens storage during chemical disinfection. Each well has an overflow capacity of 5.0 mL to ensure complete lens immersion. Device is a passive storage accessory; no electronic or mechanical components.

Clinical Evidence

Bench testing only. Device performance verified through capacity testing (5.0 mL per well) and physico-chemical/toxicological testing of materials.

Technological Characteristics

Plastic, side-by-side double-well case with screw-on closures. 5.0 mL overflow capacity per well. Passive device; no energy source, software, or connectivity.

Indications for Use

Indicated for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection.

Regulatory Classification

Identification

A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.

Special Controls

*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 13, 1997 # 510(k) SUMMARY Submitted by: Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, Texas 76132 (817) 551-4702 (Phone) (817) 551-4630 (Fax) JAN 20 1998 ALCON LABORATORIES, INC. 6201 SOUTH FREEWAY FORT WORTH, TEXAS 76134-2099 (817) 293-0450 K974281 #### Device Name: Common Name: Contact Lens Case Proprietary Name: Alcon Contact Lens Case #### Device Classification: Contact lens have not been officially classified, but have been recommended by the Ophthalmic Devices Panel to be classified in Class II and have historically been regulated through the 510(k) process. ## Description: The Alcon Contact Lens Case is intended for use for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. It consists of a side by side double well plastic case with two screw-on plastic closures. ## Performance: The Alcon Contact Lens Case is designed to have an overflow capacity of 5.0 mL in each lens well. This will be a sufficient volume to assure that the contact lens will remain immersed under use conditions. ## Substantial Equivalence Comparison: The Alcon Contact Lens Case is substantially equivalent to the Alcon Contact Lens Case (K971618). It is intended for storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. All the materials currently being used in the lens cases or proposed in this submission have passed the relevant physico-chemical, and toxicological tests. {1}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 20 1998 Mr. Ralph H. Larsen Manager, Regulatory Affairs Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, TX 76134-2099 Re: K974281 Trade Name: Alcon Contact Lens Case Regulatory Class: unclassified Product Code: LRX (0) Dated: November 13, 1997 Received: November 14, 1997 Dear Mr. Larsen: " We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: -General regulation (21 CFR-Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Mr. Ralph H. Larsen This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, A. Roerl. Rosenthal A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________ The Alcon Contact Lens Case Indications for Use: For storage of soft (hydrophilic), rigid gas permeable, and hard contact lenses during chemical disinfection. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic Devices 510(k) Number K974281 Prescription Use_ (Per 21 CFR 801.109) OR Over-The-Counter Use_X
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...