BRANEMARK SYSTEM DEC 600 DRILLING UNIT

K974123 · Nobel Biocare Uas, Inc. · DZE · May 26, 1998 · Dental

Device Facts

Record IDK974123
Device NameBRANEMARK SYSTEM DEC 600 DRILLING UNIT
ApplicantNobel Biocare Uas, Inc.
Product CodeDZE · Dental
Decision DateMay 26, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2

Intended Use

The Nobel Biocare's Brånemark System® DEC 600 Drilling Unit is intended to be used in oral and maxillo facial surgery including placement of the Branemark System endosseous implants. It is intended to be used to drill into the upper or lower jaw to prepare placement of an endosseous fixture, in the treading and mounting of fixtures, while drilling measure and record torque resistance for bone quality, as well as to tighten cover and abutment screws.

Device Story

The Nobel Biocare Branemark System® DEC 600 Drilling Unit is an accessory to a bone cutting instrument used in oral and maxillofacial surgery. The device functions as a motorized drilling unit for preparing jaw bone sites for endosseous implant fixtures. During the drilling process, the unit measures and records torque resistance to assess bone quality. It also facilitates the threading and mounting of fixtures and the tightening of cover and abutment screws. The device is operated by a clinician in a surgical setting. By providing controlled drilling and real-time torque feedback, the device assists the surgeon in achieving proper implant placement and assessing bone density, potentially improving surgical outcomes and implant stability.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Motorized drilling unit for oral and maxillofacial surgery. Features include torque measurement and recording capabilities for bone quality assessment. Designed for drilling, threading, and mounting endosseous fixtures and tightening screws. Operates as an accessory to bone cutting instruments.

Indications for Use

Indicated for patients undergoing oral and maxillofacial surgery requiring placement of endosseous implants. Used for drilling, threading, and mounting fixtures in the upper or lower jaw, and tightening cover and abutment screws.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAY 2 6 1998 Center for Devices and Radiological Health September 24, 1997 Page -7- K974123 ### Summary of Safety and Effectiveness #### Name and Address A. This Summary of Safety and Effectiveness is being submitted by Nobel Biocare USA, 777 Oakmont Lane, Suite 100, Westmont, IL 60559. Their telephone number is: (630) 654-9100 and the contact person will be Betsy A. Brown, V.P., Regulatory Affairs. #### B. Name of the Device This device is generally known as an accessory to a bone cutting instrument with the trade name "Nobel Biocare Branemark System® DEC 600 Drilling Unit." #### C. The Predicate Product The predicate product used in this Premarket Notification is the Branemark System® Drill Equipment DEC 500 handpiece subject to premarket Notification K905436. #### Description of the Device D. The Nobel Biocare Branemark System® DEC 600 Drilling Unit is intended for use in oral and maxillofacial surgery including preparing sites for implantation of an endosseous implant fixture. #### E. Intended Use of the Device The Nobel Biocare Branemark System® DEC 600 Drilling Unit is intended to be used in oral and maxillo-facial surgery including placement of the Branemark System® endosseous implant. It is intended to be used to drill into the upper or lower jaw to prepare placement of an endosseous fixture, in the threading and mounting of fixtures, while drilling measure and record torque resistance for bone quality as well as to tighten cover and abutment screws. #### F. Comparison of Technological Characteristics The technological characteristics between the DEC 600 Drilling Unit and the predicate product are similar and the intended use is the same. Both products are used for oral and maxillofacial surgery and are used for cutting, drilling and shaping jaw bone. {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 6 1998 Ms. Betsy A. Brown · Vice President, Regulatory Affairs Nobel Biocare USA, Incorporated 777 Oakmont Lane, Suite 100 ୧୦୧୮୨ Westmont, Illinois Re : K974123 Branemark System® Dec 600 Drilling Unit Trade Name: Regulatory Class: III Product Code: DZE Dated: February 25, 1998 February 25, 1998 Received: Dear Ms. Brown: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਸੀ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 {2}------------------------------------------------ Page 2 - Ms. Brown through 542 of the Act for devices under the Electronic enroduct Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA asboirsof in four and equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure _.... . {3}------------------------------------------------ 510(k) Number (if known): Brånemark System® DEC 600 Drilling Unit Device Name: Indications For Use: The Nobel Biocare's Brånemark System® DEC 600 Drilling Unit is intended to be used in oral and maxillo facial surgery including placement of the Branemark System endosseous implants. It is intended to be used to drill into the upper or lower jaw to prepare placement of an endosseous fixture, in the treading and mounting of fixtures, while drilling measure and record torque resistance for bone quality, as well as to tighten cover and abutment screws. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEED ============================================================================================================================================================================== Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use v OR Over-The-Counter Use (Per 21 CFR 801.109) On Alles IN MDA (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices (Optional Format 1-2-96) 510(k) Number i
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...