The QUIDEL CH50 Eq EIA measures the total classical complement pathway activity in human serum, thereby allowing detection of a deficiency of one or more of the complement components C1 through C9. The CH50 Eq EIA is for laboratory and professional use.
Device Story
The QUIDEL CH50 Eq EIA is an in vitro diagnostic test for quantifying total classical complement pathway activity in human serum. The device uses a three-step process: (1) complement activation by adding serum to microassay wells containing a complement activator; (2) sample dilution and binding of generated terminal complement complexes (TCC) to monoclonal antibody-coated microassay wells; (3) detection via HRP-conjugate and chromogenic enzyme substrate. Absorbance (A405) is measured spectrophotometrically; color intensity is proportional to TCC concentration and CH50 units. Results are expressed in CH50 unit equivalents per milliliter (CH50 U Eq/mL). The test is performed in a laboratory setting by professional staff. By providing a direct measure of TCC, the device aids clinicians in diagnosing immunologic disorders associated with complement component deficiencies, offering an alternative to traditional lytic CH50 hemolytic assays.
Clinical Evidence
Clinical evaluation involved 226 patient samples tested at a major U.S. clinical reference laboratory. Accuracy was assessed via Receiver Operator Curve (ROC) analysis, comparing the device to the combined results of three hemolytic assays and one EIA-based assay. The device demonstrated an overall accuracy greater than 97%. Additional bench testing confirmed intra- and inter-assay precision, lot-to-lot consistency, and lack of interference from potentially interfering substances.
Technological Characteristics
Enzyme-linked immunosorbent assay (EIA) using monoclonal antibodies to capture terminal complement complexes (TCC). Employs HRP-conjugate and chromogenic enzyme substrate for detection. Readout via spectrophotometer (A405). Standardized to CH50 unit equivalents per milliliter. Designed for laboratory use.
Indications for Use
Indicated for patients requiring measurement of total classical complement pathway activity in human serum to detect deficiencies in complement components C1 through C9. For laboratory and professional use.
Regulatory Classification
Identification
A complement components immunological test system is a device that consists of the reagents used to measure by immunochemical techniques complement components C1q , C1r , C1s , C2 , C3 , C4 , C5 , C6 , C7 , C8 , and C9 , in serum, other body fluids, and tissues. Complement is a group of serum proteins which destroy infectious agents. Measurements of these proteins aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.
Submission Summary (Full Text)
{0}------------------------------------------------
Jul 2 9 1998
# 1974111
#### SUMMARY OF SAFETY AND EFFECTIVENESS XI.
Product: OUIDEL CH50 Eq EIA
### Manufacturer:
QUIDEL Corporation 10165 McKellar Court San Diego, CA 92121
#### Device Classification:
The device, QUIDEL CH50 Eq EIA, is similar to other FDA-cleared devices used to measure by immunochemical techniques complement components Clo, C1, C1, C1, C2, C4, C5, C6, C7, C8 and C9 in serum, other body fluids and tissues. Measurements of these proteins aid in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components (21 CFR 866.5240).
The Food and Drug Administration (FDA) published a final regulation in the Federal Register classifying Complement components immunological test system as Class II.
#### Intended Use:
The QUIDEL CH50 Eq EIA measures the total classical complement pathway activity in human serum, thereby allowing detection of a deficiency of one or more of the complement components C1 through C9. The CH50 Eq EIA is for laboratory and professional use.
### Physiologic Basis for the Test:
The classical complement pathway is triggered by the binding of C1q component of C1 to immune complexes. This activation results in a cascade of enzymatic and non-enzymatic reactions, culminating in the formation of terminal complement complexes (TCC). Under standard conditions the level of TCC which can be generated in serum is a quantitative expression of the serum's total classical complement activity.
{1}------------------------------------------------
The traditional method for measuring the total classical complement activity in serum is the CH50 test. This test is a lytic assay which uses antibodysensitized sheep erythrocytes (EA) as the activator of the classical complement pathway and various dilutions of the test serum to determine the amount required to give 50% lysis. The percent hemolysis is determined spectrophotometrically. The CH50 test is an indirect measure of TCC, since the TCC themselves are directly responsible for the hemolysis which is measured.
The QUIDEL CH50 Eq EIA provides a direct measure of the total classical complement activity in serum by quantifying the amount of TCC generated under standard conditions. The test uses a monoclonal antibody to a unique neoantigen to capture the TCC analyte. Since both the QUIDEL CH50 Eq EIA and the CH50 test rely on the generation of TCC and correlate, the QUIDEL CH50 Eq EIA's results are expressed in CH50 unit equivalents per milliliter.
#### Principle of the Test:
The OUIDEL CH50 Eq EIA for quantifying the total classical complement activity in human serum involves three basic procedures: (1) complement activation; (2) sample dilution; and, (3) assay for terminal complement complexes (TCC).
To activate the classical complement pathway, undiluted human serum samples and the Controls are added to microassay wells containing the complement Activator. During an incubation the classical pathway of complement is triggered and TCC are generated.
In the second step, the activated sera are diluted in microassay wells or test tubes and dispensed, together with the Standards, directly into a precoated microassay plate. The TCC present in the activated samples bind to the monoclonal antibodies coating the surface of the microassay wells.
In the third step, the TCC microassay plate is washed and loaded with an HRP-conjugate which will bind to the bound TCC. After washing, the TCC microplate is loaded with a chromogenic enzyme substrate. After incubation a reagent is added to stop color development. The absorbancies (A405 values) generated with the Controls, Standards and test specimens are measured spectrophotometrically. The color intensity of the reaction mixture is proportional to the concentration of TCC present and to CH50 units. Using the kit standard curve, assay results are expressed in CH50 unit equivalents per milliliter (CH50 U Eq/mL).
{2}------------------------------------------------
## Safety and Effectiveness:
Numerous studies were undertaken to document the performance characteristics and the substantial equivalence of the QUIDEL CH50 Eq EIA to other commercially available products. These studies included the following:
- The test was shown to be similar to other commercially available tests in terms of . features and intended use.
- The test was shown to have good intra- and inter-assay precision. ●
- Lot-to-lot consistency analyses showed the test to be reproducibly manufacturable. ●
- Potentially interfering substances were shown not to interfere with the test's . performance.
- Two hundred and twenty six (226) patient samples were tested at a major clinical . reference laboratory in the United States and evaluated by Receiver Operator Curve (ROC) Analysis. By this method the overall accuracy of the QUIDEL CH50 Eq EIA, when compared to the combined results from three hemolytic and one EIA-based assay, was greater than 97%.
- . Stability studies are underway to establish the shelf-life of the product as its optimal storage and shipping conditions.
#### Conclusion:
These studies demonstrated the substantial equivalence of the QUIDEL CH50 Eq EIA to existing products already marketed. They further demonstrated the suitability of the product for laboratory and professional use. Such studies are a critical element in establishing the fundamental safety and effectiveness of the product and its appropriateness for commercial distribution.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing or supporting another person, represented by abstract shapes.
Jul 2 9 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Edward J. McMullen Manager, Regulatory Compliance Ouidel Corporation 10165 McKellar Court San Diego, California 92121
Re : K974111/S1 Trade Name: QUIDEL CH50 Eq EIA Regulatory Class: II Product Code: DAE Dated: May 29, 1998 Received: June 2, 1998
Dear Mr. McMullen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure --------------------------------------------------------------------------------------------------------------------------------------------------------------------
{5}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________
| 510(k) Number (if known): | K974111 |
|---------------------------|---------|
|---------------------------|---------|
Device. Name:
QUIDEL CH50 Eq EIA
Indications for Use:
. ·
The QUIDEL CH50 Eq EIA intended to measure the total classical complement pathway activity in human serum, thereby allowing detection of a deficiency of one or more of the complement components C1 through C9. The QUIDEL CH50 Eq EIA is intended for laboratory and professional use.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Device
510(k) Number K974111
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The Counter Use |
|---------------------------------------|----|----------------------|
|---------------------------------------|----|----------------------|
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.