← Product Code [LHI](/productcode/LHI) · K973654

# VIAL-MATE RECONSTITUTION DEVICE (2B8071) (K973654)

_Baxter Healthcare Corp · LHI · Oct 24, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K973654

## Device Facts

- **Applicant:** Baxter Healthcare Corp
- **Product Code:** [LHI](/productcode/LHI.md)
- **Decision Date:** Oct 24, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Vial-Mate Reconstitution Device is intended to provide the pharmacist or health practitioner a means of connecting a standard 20 mm single dose drug vial to an I.V. solution container without reconstituting and mixing the vial contents with the diluent until immediately before administration of the drug to the patient.

## Device Story

Vial-Mate Reconstitution Device serves as an accessory for connecting standard 20 mm single dose drug vials to I.V. solution containers. Operated by pharmacists or health practitioners in clinical settings; device facilitates drug reconstitution and transfer. Unlike predicate devices using a one-step frangible cannula, Vial-Mate utilizes a two-step rotating action for activation. Device allows pre-attachment to the vial without immediate penetration of the stopper; penetration occurs only upon activation. This mechanism ensures vial contents remain separated from the diluent until immediately before administration. By enabling controlled, on-demand mixing, the device supports safe drug delivery and minimizes premature reconstitution.

## Clinical Evidence

Bench testing only. Biological and chemical reactivity of the new elastomeric solution contact material assessed via ISO 10993-1 and USP Physicochemical tests. Functional performance data confirmed the device meets or exceeds all requirements.

## Technological Characteristics

Accessory device for 20 mm vials. Features a two-step rotating activation mechanism. Includes a new elastomeric material in the septum, tested per ISO 10993-1 and USP standards. Other materials previously cleared for IV administration. Non-electronic, mechanical device.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- ReconPlus™ Reconstitution Device ([K910793](/device/K910793.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K173654

510(k) Premarket Notification Vial-Mate Reconstitution Device

### 510(k) SUMMARY

### Vial-Mate Reconstitution Device

# Submitted by:

OCT 2 4 1997

Tamima Itani, Ph.D. Associate Director. Regulatory Affairs Baxter Healthcare Corporation I.V. Systems Division Rte. 120 and Wilson Road Round Lake, IL 60073

#### Date Prepared:

September 24, 1997

#### Proposed Device:

Vial-Mate Reconstitution Device

Predicate Devices: ReconPlus™ Reconstitution Device

#### Proposed Device Description:

The Vial-Mate Reconstitution Device is an accessory intended for use by the pharmacist or health practitioner in the reconstitution and transfer of drugs into Baxter solution containers. This product is similar to the currently marketed Baxter ReConPlus™ Reconstitution Device, cleared under K910793 on March 21, 1991. Like the ReConPlus™ Reconstitution Device, it provides a means of connecting a standard 20 mm single dose drug vial to an I.V. solution container without reconstituting and mixing the vial contents with the diluent until immediately before administration of the drug to the patient.

There is one solution contact material in the proposed Vial-Mate Reconstitution Device which is new to Baxter. It is an elastomeric material which is used in the septum of the device. The other solution materials used in the proposed device were previously tested and used in other Baxter devices for similar IV solution administration applications.

#### Statement of Intended Use:

The proposed Vial-Mate Reconstitution Device has the same intended use as previously cleared devices of its type. The Vial-Mate Reconstitution Device is intended to provide the pharmacist or health practitioner a means of connecting a standard 20 mm single dose drug vial to an I.V. solution container without reconstituting and mixing the vial contents with the diluent until immediately before administration of the drug to the patient.

s:\510k\vialmate\vialmate

{1}------------------------------------------------

# Summary of Technological Characteristics of New Device to Predicate Devices

The existing ReConPlus™ Reconstitution Device uses a one step motion for reconstitution where a frangible plastic cannula (breakaway seal) is broken. The proposed Vial-Mate Reconstitution Device uses a two-step rotating action for drug reconstitution. In addition, while the ReConPlus™ device penetrates the drug vial stopper when attached, the Vial-Mate device can be pre-attached without penetration into the stopper until ready for activation and use through the two-step rotating action.

# Discussion of Nonclinical Tests and Referenced Studies Reported in Published Literature

The biological and chemical reactivity of the new solution contact material have been assessed using biological methods specified in ISO Standard 10993-1 and USP Physicochemical tests. The material was found to be acceptable for its intended use.

Data regarding the functional performance of the proposed Vial-Mate have been generated and submitted. The data indicate that the proposed reconstitution device meets or exceeds all functional requirements.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. The text is in all capital letters and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856

Tamima Itani, Ph.D. Associate Director, Regulatory Affairs Baxter Healthcare Corporation I.V. Systems Division Route 120 and Wilson Road Round Lake, Illinois 60073

OCT 2 4 1997

Re: K973654 Vial-Mate Reconstitution Device (2B8071) Trade Name: Requlatory Class: II Product Code: LHI September 24, 1997 Dated: Received: September 25, 1997

Dear Dr. Itani:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Dr. Itani

through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

.Enclosure

{4}------------------------------------------------

# Indication for Use

510(k) Number: Not Available

Vial-Mate Reconstitution Device Device Name:

## Indication for Use:

The proposed Vial-Mate Reconstitution Device has the same intended use as previously cleared devices of its type. The Vial-Mate Reconstitution Device is intended to provide the pharmacist or health practitioner a means of connecting a standard 20 mm single dose drug vial to an I.V. solution container without mixing the vial contents with the diluent until immediately before administration of the reconstituted drug to the patient.

Patrica Crescenti
(Division Sign-On

"sion of Dental, In Pheral Hospita ilia Number

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

s:\510k\vialmate\vialmate

---

**Source:** [https://fda.innolitics.com/device/K973654](https://fda.innolitics.com/device/K973654)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
