The Lasershare Print Spooler and/or Software is a distributable filming management system that receives digital images and format control data from various image sources (including, but not limited to, CT scanners, MR scanners, Ultrasound systems, R/F units, Computed and Direct Radiography devices, secondary capture devices, imaging gateways, removable storage media or other imaging sources) connected point to point or via networks. The incoming data formats are standard medical imaging formats or proprietary to the modality source vendor. The Lasershare Print Spooler and/or Software allows the operator to manage print job requests, create filming sessions and route image and patient data to one or more laser film printers, thermal dye sublimation and dry printers and other medical imaging print devices.
Device Story
Lasershare Print Spooler is a software-based filming management system; accepts digital images and format control data from various medical imaging modalities (CT, MR, Ultrasound, etc.) via point-to-point or network connections. System provides spooling, queue management, and routing of image/patient data to laser film, thermal dye sublimation, or dry printers. Operates on off-the-shelf general-purpose computing hardware; modular architecture allows components to reside on local or remote machines. Used by clinical staff to manage print job requests and create filming sessions. Facilitates efficient hard-copy production of medical images for diagnostic review.
Clinical Evidence
No clinical data; bench testing only. The device is a software port of a previously marketed system, with performance verified through functional assessment of the spooling and routing capabilities across hardware platforms.
Technological Characteristics
Modular software system; runs on off-the-shelf general-purpose computing hardware. Uses network interfaces for communication between modules. Supports standard and proprietary medical imaging data formats. Designed for portability across operating systems.
Indications for Use
Indicated for use as a filming management system to receive, manage, and route digital medical images from various imaging modalities (CT, MR, Ultrasound, R/F, CR/DR) to medical imaging print devices for clinical printing.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Lasershare Print Spooler and/or Software (DeJarnette Research Systems, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
# 973422
## 1. 510(k) Summary
1 0
## 510(k) Summary
[As required by 21 CFR 807.92(c)]
OCT 28 1997
| 1.0 | Submitter Information | | | | | |
|-----|-----------------------------------------------------------------------------------|------------------|-----|---------------------------------------|----------------------------------------------------------------------------------------|--|
| 1.1 | DeJarnette Research Systems<br>401 Washington Avenue Suite 700<br>Towson MD 21204 | | 1.2 | Contact:<br>Voice:<br>Fax:<br>E-mail: | Eric John Finegan<br>+1 (410) 583 - 0680<br>+1 (410) 583 - 0696<br>info@dejarnette.com | |
| 1.3 | Date Prepared: | 8 September 1997 | | | | |
#### Device Identification 2.0
- Lasershare Print Spooler and/or Software 2.1 Trade Name:
- Digital Imaging Network Spooler 2.2 Common Name:
- Classification Name: System, Digital Image Communication, accessory 2.3
#### 3.0 Predicate Devices
Lasershare Print Spooler and/or Software [DeJarnette Research Systems, Inc.]
#### Device Description 4.0
- Function: 4.1
The Lasershare Print Spooler is designed to accept images from various modalities or digital image networks and direct images to one or more medical imaging print devices, providing spooling and queue management facilities.
The Lasershare Print Spooler and/or Software is a different model of the current Lasershare Print Spooler and/or Software [predicate device], ported to another operating system. The Lasershare is a modular system that uses network interfaces for communication between the modules. Network interfaces enable the various modules to reside on local or remote machines.
### Physical and Performance Characteristics: 4.2
The Lasershare Print Spooler and/or Software is designed to run on off-the-shelf, general purpose computing equipment. The application software is designed for maximum portability across operating systems and hardware platforms. Performance of the application software is primarily a function of network load; secondarily a function of the hardware platform's computational speed and finally the performance of the printer(s). Intrinsic performance of the application does not change significantly as it is ported from one operating system to another.
{1}------------------------------------------------
## EJARNETTE
#### Intended Use: 5.0
The Lasershare Print Spooler and/or Software is a distributable filming management system that receives digital images and format control data from various image sources (including, but not limited to, CT scanners, MR scanners, Ultrasound systems, R/F units, Computed and Direct Radiography devices, secondary capture devices, imaging gateways, removable storage media or other imaging sources) connected point to point or via networks. The incoming data formats are standard medical imaging formats or proprietary to the modality source vendor. The Lasershare Print Spooler and/or Software allows the operator to manage print job requests, create filming sessions and route image and patient data to one or more laser film printers, thermal dye sublimation and dry printers and other medical imaging print devices.
#### 6.0 Statement of Substantial Equivalence:
The Lasershare Print Spooler and/or Software is substantially equivalent to the previously marketed device (as listed above in Part 3) in design, composition, function, intended use, safety and efficacy.
Any differences between the Lasershare Print Spooler and/or Software and the predicate devices have no significant influence on safety or efficacy.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features the department's name in a circular arrangement on the left side. To the right of the text is a stylized symbol consisting of three overlapping human figures or silhouettes. The figures are depicted in a simple, abstract manner.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 28 1997
Wayne T. DeJarnette, Ph.D. President, DeJarnette Research Systems, Inc. 401 Washington Avenue Suite 700 Towson, MD 21204
Re: K973422
Lasershare Print Spooler Dated: September 8, 1997 Received: September 10, 1997 Unclassified Procode: 90 LMC
Dear Dr. DeJarnette:
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the inclosure) to devices marketed in interstate in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W. Niau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## 2. Statement of Indications for Use
Device: Lasershare Print Spooler and/or Software
## Statement of Indications for Use
[As required by 21 CFR 801.109]
The Lasershare Print Spooler and/or Software is a distributable filming management system that receives digital images and format control data from various image sources (including, but not limited to, CT scanners, MR scanners, Ultrasound systems, R/F units, Computed and Direct Radiography devices, secondary capture devices, imaging gateways, removable storage media or other imaging sources) connected point to point or via networks. The incoming data formats are standard medical imaging formats or proprietary to the modality source vendor. The Lasershare Print Spooler and/or Software allows the operator to manage print job requests, create filming sessions and route image and patient data to one or more laser film printers, thermal dye sublimation and dry printers and other medical imaging print devices.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Garold A. Heggen
(Division Sign-Off)
Division of Reproductive, Abde ... ENT, and Radiological Devic 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)
✓
OR
Over-The-Counter Use
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.