K973419 · Nellcor Puritan Bennett, Inc. · BTM · Mar 25, 1998 · Anesthesiology
Device Facts
Record ID
K973419
Device Name
DMR PLUS DISPOSABLE MANUAL RESUSCITATOR
Applicant
Nellcor Puritan Bennett, Inc.
Product Code
BTM · Anesthesiology
Decision Date
Mar 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5915
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison. The DMR2 Plus™ is available in three sizes and is for use on the following patient populations: Adult is indicated for use on adults >40 kg. Child is indicated for use on children 10 40 kg. Infant is indicated for use on infants 5 10 kg The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as hospital, transport, mobile and home.
Device Story
Manual resuscitator bag (natural rubber or PVC) with integrated CO2 detector; provides manual ventilatory support using ambient air or supplemental oxygen. Operated by qualified healthcare professionals in hospital, transport, or home settings. Integrated CO2 detector uses color comparison to verify endotracheal/nasotracheal tube placement. Output is visual color change indicating CO2 presence; assists clinicians in confirming correct airway placement during resuscitation.
Clinical Evidence
Bench testing only. Compliance with ISO 8382 (resuscitators), ISO 5356-1 (conical connectors), and ASTM F 920-93 (performance/safety). Environmental testing conducted per FDA guidance.
Technological Characteristics
Materials: processed natural rubber or PVC (bag), K-resin (housings). Components: compressible bag, nonrebreathing valve, oxygen accumulator, integrated colorimetric CO2 detector. Standards: ISO 8382, ISO 5356-1, ASTM F 920-93. Non-sterile, single-patient use.
Indications for Use
Indicated for patients requiring manual ventilatory support. Sizes: Adult (>40 kg), Child (10-40 kg), Infant (5-10 kg). Used for verification of endotracheal or nasotracheal tube placement.
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
Predicate Devices
CAPNO-FLO™ End-Tidal CO2 Detector/Monitor (K933626)
{0}------------------------------------------------
K973419
| Nellcor Puritan Bennett DMR <sup>2</sup> Plus <sup>™</sup> Disposable Manual Resuscitator | |
|-------------------------------------------------------------------------------------------|-------------|
| with integrated CO2 detection | |
| 510(k) Summary | |
| December 18, 1997 | |
| Page 1 of 3 | MAR 25 1998 |
#### 1. Company Information
Establishment:
Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588 (510) 463-4177
DMR2 Plus™ Disposable Manual
Resuscitator with integrated CO2
Disposable Manual Resuscitator
with integrated CO2 detection
- Shervil A. Mathews Official Correspondent: Regulatory Affairs Assistant
#### 2. Device Name
Proprietary:
Common/Usual:
Classification:
73BTM
detection
#### 3. Equivalent Device
- 1. CAPNO-FLO™ End-Tidal CO2 Detector/Monitor Kirk Specialty Systems - K933626
- 2. Nellcor Puritan Bennett DMR/CO2 Combo Kit K960468
## Device Description 4.
The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support.
The DMR2 Plus™ is available in three sizes: adult, child and infant. It may be used by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated; such as, hospital, transport, mobile and home.
{1}------------------------------------------------
Nellcor Puritan Bennett DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection 510(k) Summary December 18, 1997 Page 2 of 3
The DMR2 Plus™ consists of a compressible resuscitator bag that is available in either processed natural rubber or PVC. The swivel, patient connection and CO2 detector cartridge are connected to the nonrebreathing Valve (NRV) housing - all housings are K-resin. The oxygen accumulator assembly enables the use of three different styles of oxygen accumulators: large bore, small bore or flat bag.
During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison.
## 5. Intended Use
The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison.
The DMR2 Plus™ is available in three sizes and are intended for use on the following patient populations:
- · Adult is indicated for use on adults >40 kg.
- · Child is indicated for use on children 10 40 kg.
- · Infant is indicated for use on infants 5 10 kg
The DMR2 Plus™ is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as hospital, transport, mobile and home.
{2}------------------------------------------------
Nellcor Puritan Bennett DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection 510(k) Summary December 18, 1997 Page 3 of 3
## 6. Technological Characteristics
The method of operation and the CO2 detector technology of the DMR2 Plus™ are equivalent to the CAPNO-FLO and the Nellcor Puritan Bennett DMR/CO2 Combo Kit. A comparison of the devices' design features, methods of operation and labeling support technical equivalency.
The safety and effectiveness of the DMR2 Plus™ has been demonstrated by design and testing. Testing was conducted in accordance with the requirements of ISO 8382: Resuscitators Intended for Use With Humans; ISO 5356-1: Anaesthetic and respiratory equipment - Conical connectors; and ASTM F 920-93: Standard Specification for Minimum Performance and Safety Requirements for Resuscitators Intended for Use With Humans.
In addition, environmental testing was conducted as recommended by applicable sections of the Draft Reviewer's Guidance for Premarket Notification Submissions, November 1993.
Test results demonstrate and/or labeling indicates how all applicable requirements have been addressed.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a bird or eagle, with three curved lines forming the body and wings.
MAR 25 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Sherry Mathews Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588-2719
Re: K973419 DMR2 Plus Disposable Manual Resuscitator Regulatory Class: II (two) Product Code: 73 BTM January 7, 1998 Dated: Received: January 9, 1998
Dear Ms. Mathews:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
# Page 2 - Ms. Sherry Mathews
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________ of
510(k) Number (if known):
Unknown Ka73419
Device Name: DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection
Indications For Use:
The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison.
The DMR2 Plus™ is available in three sizes and is for use on the following patient populations:
- Adult is indicated for use on adults >40 kg. .
- Child is indicated for use on children 10 40 kg. .
- · Infant is indicated for use on infants 5 10 kg
The DMR2 Plus™ Disposable Manual Resuscitator with integrated CO2 detection is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated such as hospital, transport, mobile and home.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
3/24/98 C. Avery for Superstar, Branch, teling Chief
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number.
OR
Over-The-Counter Use **__**
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.