K973390 · Ventana Medical Systems, Inc. · DEM · Oct 3, 1997 · Immunology
Device Facts
Record ID
K973390
Device Name
CHEMMATE L26 (CD20)
Applicant
Ventana Medical Systems, Inc.
Product Code
DEM · Immunology
Decision Date
Oct 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5130
Device Class
Class 2
Indications for Use
The ChemMate™ L26 (CD20) is intended for laboratory use to qualitatively identify by light microscopy human lymphocytes of B-cell lineage in normal and pathological paraffin embedded tissues processed in zinc formalin, neutral buffered formalin, Bouin's or B5 fixative. Positive results aid in the classification of lymphomas as B-cell in origin and must be interpreted by a pathologist within the context of clinical data, gross and microscopic morphological criteria and multiple chemical and immunohistochemical stains.
Device Story
ChemMate™ L26 (CD20) is a mouse monoclonal antibody (clone L26, IgG2a/Kappa) used for immunohistochemical staining of paraffin-embedded tissue sections. It targets the CD20 antigen (33kD and 30kD components) on the plasma membrane of B-cells. The device is intended for laboratory use; it is optimized for use with the TechMate™ automated staining system or manual secondary detection kits (Peroxidase/DAB). Pathologists examine stained tissue slides via light microscopy. Positive staining identifies B-cell lineage, aiding in the differential diagnosis of lymphomas and other lymphoproliferative disorders. The antibody is designed to survive standard tissue fixation processes, allowing for reliable detection in formalin-fixed specimens. Clinical utility includes distinguishing B-cell lymphomas from T-cell lymphomas and non-lymphoid tumors.
Clinical Evidence
Evidence includes literature-based studies and internal performance testing on 281 tissue specimens. Studies demonstrated L26 reactivity in 397/430 (92%) of B-cell non-Hodgkin's lymphomas and 1/89 (1.1%) of T-cell lymphomas. Specificity was confirmed across various normal tissues and 139 non-lymphoid pathological tissues (0/139 reactive). Reproducibility was verified across three separate daily runs.
Technological Characteristics
Mouse monoclonal antibody (clone L26, IgG2a/Kappa). Prediluted for use with secondary detection kits (Peroxidase/DAB). Optimized for automated immunohistochemical staining (TechMate™). Applicable to paraffin-embedded tissues fixed in zinc formalin, neutral buffered formalin, Bouin's, or B5.
Indications for Use
Indicated for qualitative identification of human B-cell lineage cells by recognizing CD20 antigen in normal and pathologic paraffin-embedded tissues (fixed in neutral buffered formalin, B5, or Bouin's). Used to aid in classification of lymphomas of B-cell origin. Must be interpreted by a pathologist in context of clinical data and other stains.
Regulatory Classification
Identification
An alpha-1-antitrypsin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the alpha-1-antitrypsin (a plasma protein) in serum, other body fluids, and tissues. The measurements aid in the diagnosis of several conditions including juvenile and adult cirrhosis of the liver. In addition, alpha-1-antitrypsin deficiency has been associated with pulmonary emphysema.
Submission Summary (Full Text)
{1}------------------------------------------------
KU 13340
### 510(k) Summary
OCT - 3 1997
Date: September 29, 1997
VENTANA MEDICAL SYSTEMS, INC. 3865 North Business Center Drive Tucson, Arizona 85705
(520) 887-2155 Telephone: (520) 887-2558 Facsimile:
Stephen A. Tillson, Ph.D. Contact: Vice President Scientific Affairs/ Quality Assurance
Registration #: 2028492
ChemMate™ L26 (CD20) Trade Name:
Class II
FOR IN VITRO DIAGNOSTIC USE. Intended Use:
The ChemMate™ L26 (CD20) is intended for laboratory use to qualitatively identify by light microscopy human lymphocytes of B-cell lineage in normal and pathological paraffin embedded tissues processed in zinc formalin, neutral buffered formalin, Bouin's or B5 fixative. 8. 12, 19 Positive results aid in the classification of lymphomas as B-cell in origin and must be interpreted by a pathologist within the context of clinical data, gross and microscopic morphological criteria and multiple chemical and immunohistochemical stains.
This mouse monoclonal antibody has been optimally prediluted for use with the ChemMate™ SDK605 or SDK305 Secondary Detection - Peroxidase/DAB kit. Additionally, the prediluted ChemMate™ L26 (CD20) antibody as well as the ChemMate™ SDK605 or SDK305 Secondary Detection -Peroxidase/DAB kit has been optimized for use with the TechMate™ for automated immunohistochemical staining.
{2}------------------------------------------------
# 510K SUMMARY OF SAFETY AND EFFECTIVENESS
# Summary And Explanation
The ChemMate™ L26 (CD20) is comprised of a mouse monoclonal antibody, clone L26. and is of the IgG2a / Kappa light chain class of immunoglobulins. 123 The antibody reacts with two non-covalently associated components (33kD and 30kD) present in the majority of B-cells. 12 Positive staining occurs on the cells' plasma membrane. Recent studies demonstrated that the L26 antibody positively stained COS-1 cells transfected with CDNA encoding for the CD20 antigen (a pan B-cell marker).4 L26 has been classified as a CD20 antibody by the Fifth International Workshop on Human Leukocyte Differentiation Antigens.5
A unique feature of the ChemMate™ L26 (CD20) is the antibody's ability to recognize epitopes on Ivmphocvtes that survive the rigors of processing and fixation. 23.6 The resultant advantage in the use of this antibody is its applicability with formalin-fixed specimens.
The CD20 antigen is expressed in most B-cells present in peripheral blood and lymphoid tissue including germinal center blasts, mantle zone lymphocytes and scattered interfollicular lymphocytes. 13 Staining also occurs in most non-Hodgkin's lymphomas of B-cell lineage.3,10
In a study encompassing benign lymphoid tissues. Cartun et. al. demonstrated that L26 was reactive in known B-cell rich regions; i.e. germinal centers, mantle zone lymphocytes of lymph node and tonsil and white pulp of spleen. Thymus and bone marrow were unreactive for L26 with only scattered positive cells in the medulla regions of thymus. L26 was unreactive in all normal non-lymphoid tissues testedand negative in 124 non-lymphoid tumors with L26 reactivity confined to immunoreactive lymphocytes infiltrated within the tumors. Among non-Hodgkin's lymphoma cases, L26 was shown to be reactive in 73 out of 74 B-cell lymphomas, including 44 out of 44 cases of diffuse large cell and immunoblastic B-cell lymphoma. Only 37 of 44 of these B cell Lymphomas were positive with leukocyte Common Antigen (LCA). This finding suggests that a panel of antibodies including L26 should be used to exclude B-cell pathogenesis. No T-cell tumors (0/8) were reactive. Four of four hairy cell leukemias were positive for L26.7
The majority of non-Hodgkin's lymphomas will stain positively for L26. Elghetany, et al. reported on 44 cases of paraffin-embedded non-Hodgkin's lymphomas stained with a panel of monoclonal antibodies to T and B cells. Results revealed that immunophenotyping was the same as genotypic analysis in 37 of the 39 cases. Five could not be assigned a specific B or T-
{3}------------------------------------------------
cell lineage. L26 labeled 32 out of 33 B-cell Ivmphomas and one of the six Tcell lymphomas.8 In a study on 13 diffuse mixed cell non-Hodgkin's lymphomas. Katzin et al. emploved immunohistochemical and gene rearrangement analyses to determine whether these mixed cell lymphomas were of monoclonal, oligoclonal or dual lineage. Large lymphoid cells were L26 positive in 12 of 13 cases and in no cases were the small Ivmphoid cells positive for L26.11
And in studies conducted by Wolf et al. in which thyroid neoplasms were reevaluated, immunophenotyping results showed that 65 of the 68 neoplasms assessed were non-Hodgkin's lymphomas (60 out of 65 were immunoreactive with L26). Of these lymphomas, 63 out of 65 were determined to be of B-cell origin and two were not able to be determined. The remaining three cases were characterized as epithelial in origin (carcinomas) and negative for L26. Immunohistochemical results converted a preliminary diagnosis of small cell carcinoma to a non-Hodgkin's lymphoma in many cases. Small cell carcinomas are rare in the thyroid with most thyroid neoplasms being Ivmphomas of B-cell origin. 12
Additionally, Hodgkin's disease that has been classified as lymphocyte predominance stained positively for L26. Nicholas, et al re-evaluated archival cases previously diagnosed as lymphocyte predominance Hodgkin's disease (LPHD) using a panel of antibodies, including L26. Immunostaining revealed that of the 30 cases all were positive for L26. These cases were further subdivided as LPHD with nodular architecture (21 cases), diffuse LPHD (four cases) and LPHD with inconspicuous nodularity. Mixed cellularity Hodakin's (13 cases), nodular sclerosing Hodgkin's (two cases), and interfollicular Hodgkin's (one case) were all negative for L26. Additionally, the LPHD biopsies all contained polylobated Reed-Sternberg cell variants that were positive for L26.13
L26 reactivity has been reported in several neoplastic and lymphoproliferative disorders. In a study encompassing 110 B-5 fixed, paraffin-embedded neoplastic and lymphoid proliferative disorders, Segal et. al. confirmed L26 reactivity in 73 of the 76 B-cell lymphomas and found L26 was unreactive in 5/5 T-cell lymphomas. Additionally, 11 of 12 cases positive for the T-cell marker CD43 were also positive for L26, indicating a differentiation of malignant B-cell lymphoma.14
Reactivity of L26 with hairy cell leukemias and monocytoid B-cell lymphomas was further assessed by Stroup, et al. by using a large panel of antibodies in 42 cases of hairy cell leukemia (HCL) and 24 cases of monocytoid B-cell lymphoma (MBCL). L26 was immunoreactive in 85% of the hairy cell leukemias tested. Similarly, L26 was positive in 84% of the monocytoid B-cell lymphomas tested. 15
{4}------------------------------------------------
A panel of antibodies, including L26, were used by Wolf et al. to stain 34 paraffin-embedded cases of gastrointestinal tract lymphoma. Previous phenotyping studies through extensive frozen section immunohistochemistry revealed that 31 of the 34 cases were B-cell in origin and the remaining three T-cell in origin. 24 of the 31 paraffin sections were immunoreactive with L26 and L26 was negative in all three of the T-cell Ivmphomas. 16
Further, in studies conducted by Macon, et al. L26 labeled 33 out of 38 B-cell lymphomas tested and 0 out of 76 peripheral T-cell Ivmphomas.17
No L26 cross reactivity has been observed with normal or malignant nonlymphoid cells with the following exception7.20. L26 positive dendritic cells have been noted in the medulla of normal thymi and thymomas along with scattered B-Cells 19, 23, 24. The L26 positive dendritic cells coexpress keratin and are frequently associated with Hasall's corpuscles. It is recommended that an epithelial marker, such as keratin, be used in conjunction with the L26 antibody to aid in the differential diagnosis of mediastinal non-Hodqkin's lymphomas.
A summary of immunoreactivity in lymphoid and hematopoietic cells/tissues and neoplasms may be found in tables I-V of Performance Characteristics in this insert.
# Product Specific Limitations:
- 1. A few cases of diffuse large (2/26. 7.7%) and diffuse mixed (1/12, 8.3%) non-Hodgkin's T-Cell lymphomas coexpress the CD20 antigen10,20
- 2. In poorly fixed tissue specimens, nonspecific staining of non-lymphoid tissues may be observed, particularly epithelium and smooth muscle.
- 3. In some cases of Hodgkin's disease, particularly the lymphocyte predominant type, membrane, cytoplasmic, or paranuclear staining of some Reed-Sternberg cells has been observed 5,7.
- 4. The differential diagnosis of mediastinal tumors with the L26 antibody may be compromised by the presence of L26 and keratin positive dendritic cells found in the medulla of normal thymi and some thymomas 19, 23, 24. Use a panel of antibodies to distinguish mediastinal tumors.
- 5. No L26 cross-reactivity has been observed with normal or malignant nonlymphoid cells with the following exception7.20. L26 positive dendritic cells have been noted in the medulla of normal thymi and thymomas along with
{5}------------------------------------------------
scattered B-Cells 19, 23, 24. The L26 positive dendritic cells coexpress keratin and are frequently associated with Hassall's corpuscles. It is recommended that an epithelial marker, such as keratin, be used in conjunction with L26 antibody to aid in the differential diagnosis of mediastinal non-Hodgkin's lymphomas.
- 6. Dako® CD20, L26 positive B-Cells have been observed in granulomatous lesions of unknown significance, granulomas associated with toxoplasmosis; and sarcoid reaction affiliated with seminoma, Hodgkin's Disease. non-Hodgkin's lymphoma, and cancer of the breast, colon, and lung25.
- 7. Occasional cases of T-cell lymphoma have been reported to co-express the CD20 antigen.
- 8. In Hodgkin's disease, Reed-Sternberg cells have been reported as staining positively for L26.13,14
- 9. L26-positive B-cells may be present in tissues other than those of Ivmphoid origin. Though these reactions are positive for L26, interpretation should always be considered within the context of the predominant cell type of the tissue in question26.
# Performance Characteristics:
Reproducibility: ChemMate™ L26 (CD20), and Negative Control Reagent have been tested on serial sections of 281 tissue specimens (both normal and tumor specimens were included in the study). Runs were performed a total of three times, with each run being performed on a different day. Consistent staining results have been obtained with run-to-run and withinrun antibody testing.
Immunoreactivity: The following immunoreactivities have been demonstrated in paraffin-embedded tissues.. The list provided below is not exhaustive but characterizes the types of immunoreactivity reported in the literature for the L26 clone contained in ChemMate™ L26 (CD20).
{6}------------------------------------------------
# TABLE I REACTIVE NORMAL TISSUES/CELLS '188
| Tissues |
|------------------------------------------------------------------------------|
| Lymph Node:<br>germinal center cells<br>mantle zone lymphocytes (most cells) |
| Spleen:<br>white pulp (B-cell) areas |
| Tonsil:<br>germinal center cells<br>mantle zone lymphocytes (most cells) |
| Thymus:<br>medulla (some scattered areas)* |
| Whole Blood (Peripheral):<br>circulating B lymphocytes |
, 24 4 L26 positive dendritic cells have also been identified
{7}------------------------------------------------
#### nonreactive normal tissues/cells17 TABLE II
| Lymphoid Tissues | Non-Lymphoid Tissues |
|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Lymph Node:<br>Interfollicular & Medullary Areas*<br>Histiocytes (Interdigitating Reticulum Cells)<br>Dendritic Reticulum Cells, Sinus Histiocytes | Skin:<br>Epidermis<br>Sebacious Glands<br>Eccrine Glands<br>Hair Follicles |
| Spleen:<br>Red pulp areas* | Thyroid:<br>Follicular Epithelium |
| Tonsil:<br>Interfollicular & Medullary Areas*<br>Histiocytes (Interdigitating Reticulum Cells)<br>Dendritic Reticulum Cells, Sinus Histiocytes | Lung:<br>Pneumocytes<br>Bronchial Epithelium |
| Thymus:<br>Cortex* | Brain:<br>Neurons<br>Glial Cells |
| Bone Marrow:<br>Erythroid Cells<br>Myeloid Cells<br>Megakaryocytes | Pancreas:<br>Acini<br>Intercalated & Interlobular Ducts<br>Islets |
| | Liver:<br>Hepatocytes<br>Bile Duct Epithelium |
| | Prostate:<br>Epithelial Glands<br>Stroma |
| | Kidney:<br>Proximal Convoluted Tubules<br>Distal Convoluted Tubules<br>Glomeruli<br>Collecting Ducts |
| | Uterus:<br>Endometrium<br>Myometrium<br>Ectocervix<br>Endocervix |
| | Muscle:<br>Cardiac<br>Smooth<br>Skeletal |
| | Placenta |
| | Other:<br>Fibroblasts<br>Mesothelium |
| | Peripheral Blood:<br>Erythrocytes<br>Granulocytes<br>Monocytes<br>Platelets |
*Scatted positive B-cells identified
{8}------------------------------------------------
# TABLE IIINON-LYMPHOID PATHOLOGICAL NEGATIVE TISSUES/CELLST,9,12,20
| Epithelial (0/90): | Endo |
|--------------------------------------------|---------------|
| Adenoma, Parathyroid (0/1) | Neuro |
| | Caro |
| Adenocarcinoma (0/49): | Lung |
| Breast (0/11) | Merk |
| Colon(0/7) | Panc |
| Liver, hepatocellular (0/5) | Thyro |
| Lung (0/10) | Othe |
| Metastatic (0/4) | Pitu |
| | Pro |
| Ovary (0/1) | |
| Ovary, cystadenocarcinoma (0/2) | |
| Pancreas (0/1) | Mixe |
| Stomach (0/3) | Car |
| Thyroid (0/3) | Fibr |
| Thyroid, papillary (0/2) | Uter |
| Ameloblastoma (0/1) | Neuro |
| | Astro |
| Carcinoma (0/20): | Glio |
| Bladder, transitional cell (0/5) | Gan |
| Nasopharyngeal (0/1) | Neuro |
| Squamous cell (0/10) | |
| Small cell (0/3) | Small |
| Thyroid, anaplastic (0/1) | Ewir<br>Neuro |
| Mesothelioma (0/5) | Wilr |
| Melanoma (0/14): | |
| Amelanotic (0/10) | |
| Unspecified (0/4) | |
| Mesenchymal (0/25): | |
| Angiosarcoma (0/3) | |
| Chordoma (0/1) | |
| Chondrosarcoma, extraskeletal myxoid (0/1) | |
| Endometrial Stromal Sarcoma (0/3) | |
| Fibrosarcoma (0/1) | |
| Leiomyosarcoma (0/3) | |
| Malignant Fibrous Histiocytoma (0/6) | |
| Rhabdomyosarcoma (0/7) | |
| Embryonal (0/2) | |
| Unspecified (0/5) | |
### ocrine (0/24):
roendocrine (0/21) inoid (0/5) g, small cell undifferentiated (0/5) el cell tumor (0/4) creas (0/3) oid, medullary carcinoma (0/4) r (0/3) uitary adenoma (0/1) olactinoma (0/2)
d (0/5): cinoma, Salivary Gland (0/1) roadenoma, breast (0/3) rus (0/1)
- o/Glial (0/9): rocytoma (0/5) ma, malignant (0/2) glioneuroma (0/1) rofibroma (0/1)
Cell Tumors (0/9): ng's Sarcoma (0/4) roblastoma (0/4) n's (0/1)
Note: None of the 139 non-lymphoid pathological tissues listed above were reactive for L26. The number of reactive tissues over the total number of tissues is recorded by the specimen type. Confidence Intervals have been assigned for sample values greater than 1: 0/2 = 0-84.2%, 0/3 = 0-70.8%, 0/5 = 0-52.2%, 0/7 = 0-41.0%, 0/9 = 0-33.6%, 0/10 = 0-30.8%, 0/14 = 0-23.2%, 0/17 = 0-19.5%, 0/21 = 0-16.1%, 0/24 = 0-14.2% 0/25 = 0-13.7%, 0/61 = 0-6 %, 0/99 = 0-3.5%
{9}------------------------------------------------
#### TABLE IV NON-HODGKIN'S LYMPHOMAS and OTHER LYPMPHOPROLIFERATIVE DISORDERS
| Tissue | B-Cell | % | 95% CI | T-Cell | % | 95% CI |
|----------------------------------------------------|---------|------|---------------|--------|------|----------|
| Non-Hodgkin's Lymphoma | 397/430 | 92% | | 1/89 | 1.1% | 0.2-4.5% |
| Low Grade: 7,10,11,17,21 | 176/184 | 96% | 93.5<br>98.5% | -- | -- | |
| -Small Lymphocytic | 26/32 | 81% | 63-93% | -- | -- | |
| -Small cleaved, follicular &<br>diffuse | 47/47 | 100% | 92-100% | 0/5 | -- | 0-52% |
| -Mixed small and large | 7/7 | 100% | 59-100% | 1/22 | 4.5% | 0.7-27% |
| -Large cleaved | 2/2 | 100% | 15.8-100% | -- | -- | |
| -Large diffuse | 94/96 | 98% | 93-99.8% | 0/13 | -- | 0-24.7% |
| Intermediate: 7.21 | 25/25 | 100% | 86.3-100% | -- | -- | |
| -Larger cell | 25/25 | 100% | 86.3-100% | -- | -- | |
| High Grade: 7,10,12,11,14,16 | 102/106 | 96% | | -- | -- | |
| -Small non-cleaved | 4/4 | 100% | 39.8-100% | -- | -- | |
| -Large immunoblastic | 26/27 | 96% | 81-99% | 0/2 | -- | 0-84.7% |
| -Large lymphoblastic (pre B) | 1/1 | 100% | -- | 0/2 | -- | 0-84.7% |
| -Malignant | 71/74 | 96% | 87-99% | -- | -- | |
| Miscellaneous: 7,10,14,15,16,17,21,22 | 21/21 | 95% | | -- | -- | |
| -Immunoblastic Sarcoma | 2/2 | 100% | 15.8-100% | -- | -- | |
| -Marginal Zone Lymphoma | 1/1 | 100% | -- | -- | -- | |
| -Sinonasal Large-Cell<br>Lymphoma | 0/1 | -- | -- | -- | -- | |
| -Monocytoid Lymphoma | 17/17 | 100% | | -- | -- | |
| -T-cell Lymphoma | | | | 0/44 | -- | 0-11.5% |
| -High Grade Lymphoblastic<br>Lymphoma | | | | 0/1 | -- | |
| Other Lymphoproliferative<br>Disorders: 7,14,16,17 | 74/94 | 79% | 78-95% | -- | -- | |
| -Hairy Cell Leukemia | 40/46 | 87% | 73-95% | -- | -- | |
| -Plasmacytoma/Plasma Cell<br>Myeloma | 1/11 | 10% | 0.5-42% | -- | -- | |
| -Lymph nodules/Bone marrow<br>biopsies 27 | | | | | | |
| ---Polycythemia Vera | 4/4 | 100% | 39.8-100% | -- | -- | |
| ---Idiopathic Thrombocythaemia | 7/7 | 100% | 59-100% | -- | -- | |
| ---Chronic Myeloid Leukemia | 0/4 | -- | 0-60.2% | -- | -- | |
| Myelofibrosis/<br>Osteomyelosclerosis | 17/17 | 100% | 80.5-100% | -- | -- | |
| Borderline | 5/5 | 100% | 48-100% | -- | -- | |
Note: The number of reactive tissues over the total number of tissues is recorded by the specimen type. Confidence Intervals have been assigned for sample values greater than 1.
{10}------------------------------------------------
| TISSUE | NO. REACTIVE /<br>NO. TESTED | % | 95% CONFIDENCE<br>INTERVAL |
|-----------------------------|------------------------------|------|----------------------------|
| Lymphocyte Depleted | 1/9 | 11% | 0.3-48.2% |
| Lymphocyte<br>Predominant | 49/51 | 96% | 86.5-98.9% |
| Mixed Cellularity | 8/44 | 18% | 8-30% |
| Nodular Sclerosing | 2/33 | 6% | 1-20.5% |
| Inconspicuous<br>Nodularity | 5/5 | 100% | 48-100% |
| Diffuse | 4/4 | 100% | 39.8-100% |
| Interfollicular | 0/1 | --- | --- |
| Unclassified | 0/3 | --- | 0-70.8% |
| Total: | 69/150 | 46% | --- |
#### HODGKIN'S LYMPHOMA 9,10,13,19,20 TABLE V
Note: The number of reactive tissues over the total number of tissues is recorded by the specimen type. Confidence Intervals have been assigned for sample values greater than 1.
# References
- 1. Ishii Y, et al. Six distinct antigen systems of human B-cells as defined by monoclonal antibodies. In: Reinherz EL, et al, eds. Leukocyte Typing II. New York. Springer Verlag 1986
- 2. Ishii Y. et al. Two distinct antigen systems in human B-lymphocytes; identification of cell surface and intercellular antigens using monoclonal antibodies. Clin Exp Immun 2.
- 3. Takami T, et al. B20.3 Reactivity and specificity of L26 (pan-B-cell mAb) on 322 cases of fresh and paraffin-embedded lymphoproliferative diseases. In Knapp W et al, eds. Leukocyte Typing IV. White Cell Differentiation Antigens. Oxford, Oxford University Press, 1989: 134
- 4. Mason DY, et al. Antibody L26 recognizes an intercellular epitope on the B-cell associated CD20 antigen. Am J Path. 1990, 136:1215
- 5. Fifth International Workshop on Leukocyte Differentiation Antigens. Boston, 1993.
- 6. Takami, et al. Three distinct antigen systems on human B cell subpopulations as defined by monoclonal antibodies. J Immun. 1985, 134:828
- 7. Cartun RW, et al. Utilization of monoclonal antibody L26 in the identification and confirmation of B-cell lymphomas: a sensitive and
{11}------------------------------------------------
specific marker applicable to formalin- and B5 fixed, paraffin embedded tissues. Am J Path 1987, 129: 415
- 8. Takami T, et al. Reactivity and specificity of L26 (pan-B-cell mAb) on 322 cases of freshly and paraffin-embedded lymphoproliferative diseases. In Knapp w. et al., eds. Leukocvte Typing IV. White cell differentiation antigens. Oxford University Press., 1989: 134
- 9. Coles, FB et al. Hodgkin's disease lymphocyte predominant type: immunoreactivity with B-cell antibodies. Mod Path. 1988, 1:274
- 10. Elghteany MT et al. Immunophenotyping of non-Hodgkin's lymphomas in paraffin-embedded tissue sections. Am J Clin Path. 1991, 95:517
- 11. Katzin WE et al. Immunophenotypic and genotypic characterization of diffuse mixed non-Hodgkin's lymphomas. Am J Pathol. 1989, 135:615
- 12. Wolf, BC et al. Immunohistochemical analysis of small cell tumors of the thyroid gland. Hum Pathol. 1992, 23:1252
- 13. Nicholas DS, et al. Lymphocyte predominance Hodgkin's disease. Histopath. 1990, 16:157
- 14. Segal GH, et al. Reliable and cost-effective paraffin section immunohistology of lymphoproliferative disorders. Am J Surg Path. 1991, 15(11):1034
- 15. Stroup, R and Sheibani, K. Antigenic phenotypes of hairy cell leukemia and monocytoid B-cell lymphoma. Hum Path. 1992, 23:172
- 16. Wolf BC, et al. Non-Hodgkin's lymphomas of the gastrointestinal tract. Am J Clin Path. 1990, 93:233
- 17. Macon WR. et al. Leu-22(L60), a more sensitive marker than UCHL 1 for peripheral T-cell lymphomas particularly large-cell types. Am J Clin Path. 1991, 95:696
- 18. Taylor CR and Cote RJ. Immunomicroscopy: A diagnostic tool for the surgical pathologist. WB Saunders Company, Philadelphia. 1992, second edition. pg 42-70
- 19. Chilosi M., et al. Neoplastic epithelial cells in a subset of human thymomas express the B-Cell associated CD20 antigen. Am J Surg Path 1992; 16(10): 988
- 20. Linder J. et al. Monoclonal antibodies marking B-cell non-Hodgkin's lymphoma in paraffin-embedded tissue. Mod Path. 1988, 1:29
- 21. Norton AJ and Isaacson PG. Monoclonal antibody L26: an antibody that is reactive with normal and neoplastic B-lymphocytes in routinely fixed and paraffin wax embedded tissue. J Clin Path. 1897, 40:1405.
- 22. Franco V, et al. Immunohistochemical evaluation of bone marrow lymphoid nodules in chronic myeloproliferative disorders. Virchows Archiv A Path Anat. 1991, 419:261
- 23. Masanuga A, et al. Strong expression of L26 antigen in epithelial cells of human neonatal thymus. Thymus 1993, 21:65
- 24. Taubenberger JK, et al. Thymoma with abundant L26 positive 'asteroid' cells. Arch Path Lab Med 1991, 115:1254
{12}------------------------------------------------
- 25. Brincker H, and Pederson NT. Immunohistologic separation of B-cell positive granulomas from B-cell negative granulomas in paraffin embedded tissues with special reference to tumor related sarcoid reaction. APMIS 1991, 99:282
- 26. Ernst PB, et al. Immunity in mucosal tissues. Basic and Clinical Immunology, 6th edition, chapter 12. Stites, DP, et al. Editors. Norwalk, CT, Appleton & Lang Press, 1987:159
- 27. Battifora H, Alsabeh R, Jenkins KA, Gown A. Epitope Retrieval (unmasking) in immunohistochemistry. Advances in Pathology and Laboratory Medicine 1995; 8:117.
{13}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the word "DEPARTMENT" in bold, black font. The word is in all capital letters and is positioned horizontally. To the left of the word is a graphic that is partially visible.
OCT - 3 1997
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Stephen A. Tillson, Ph.D. Vice President Scientific Affair/ Quality Assurance VENTANA MEDICAL SYSTEMS, INC. 3865 North Business Center Drive Tucson, Arizona 85705
Re: K973390 ChemMate™ L26 (CD20) Trade Name: Requlatory Class : II Product Code: DEM Dated: July 7, 1997 Received: July 10, 1997
Dear Dr. Tillson:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject-to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{14}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CUIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K973390
ChemMate L26(CD20) Device Name:
Antibody Reagent
# Indications For Use:
ﺎﻧﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ
To qualitative aid in the identification by light microscopy of human cells of B-cell lineage, by recognizing CD20 antigen in normal and pathologic paraffin embedded tissues processed in neutral buffered formalin, B5, or Bouin's fixative. Positive results aid in the classification of lymphomas of B-cell origin and must be interpreted by a pathologist within the context of clinical data, gross and microscopic morphologic criteria and multiple chemical and immunohistochemical stains.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Titia E. Malins
(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number .
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter. Use_
(Optional Format 1-2-96)
O
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.